This document contains the technical specifications for an Altitude/Humidity/Temperature Chamber required by the Defense Logistics Agency (DLA) Land and Maritime Engineering & Technical Support Directorate Product Test Center Division, Electronics Lab in Columbus, Ohio. The chamber must have a minimum 16 cubic foot capacity and be capable of simulating altitudes up to 100,000 feet, with simultaneous temperature and humidity control. Key requirements include a temperature range of -68°C to +190°C, cooling and heating rates of 60 minutes or less, a humidity range of 10% to 95%, and other specifications for power, insulation, safety features, training, and installation. The chamber must be new, with a 1-year warranty, and have been on the market for a minimum of 3 years. The contractor must provide references, spare parts, and on-site installation, training, and acceptance testing. Pricing for shipping, NIST calibration, and an extended service contract are also required.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| SP4706-24-Q-0050 0001.pdf | ||
| Altitude Chamber SOW.pdf | ||
| SP470624Q0050.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Rev 0
Technical Specifications
Altitude/Humidity/Temperature Chamber DLA Land and Maritime Engineering &
Technical Support Directorate
Product Test Center Division, Electronics Lab, VTP
I. The contractor shall complete Altitude/Humidity/Temperature Chamber to the Defense
Supply Center Columbus (DSCC), DLA Land and Maritime, Product Test Center Division, Electronics Lab, Columbus, Ohio.
A. An Altitude/Temperature/Humidity Chamber with the following salient characteristics:
1. Shall be a new, not refurbished.
2. Shall be a minimum 16 cubic foot altitude/temperature/humidity chamber.
3. Chamber height simulation range: minimum of site level to 100,000 feet.
4. Simultaneous control of temperature and simulated altitude shall range from site level to
60,000 feet.
5. Stand-alone chamber temperature range: minimum of -68 degrees C to +190 degrees C.
6. Temperature control stability: +/-1 degree C.
7. Cooling capacity: Shall be able to pull-down from +24 degrees C to -68 degrees C in 60 minutes or less when operated in the stand-alone mode with an empty chamber at site level.
8. Heating capacity: Shall be able to heat up from +24 degrees C to +190 degrees C in 60 minutes or less when operated in the stand-alone mode with an empty chamber at site level.
9. Live load capacity of stand-alone chamber: minimum of 3,000 watts at -40 degrees C.
10. Relative humidity range: minimum of 10% to 95% relative humidity.
11. Relative humidity control stability: +/-2% relative humidity.
12. Cascade compressors shall have integral air cooling.
13. Shall have either liquid filled refrigeration gauges and/or sensors which can be read and monitored by controller.
14. The chamber shall have refrigeration service taps.
15. Electrical Power: 230V/3 Ph/60Hz or 460V/3 Ph/60Hz.
16. Must include vacuum pump and humidity water recirculation system with appropriately sized water holding tank to hold necessary amount of water for humidity operation included.
17. The maximum sound level must be less than 80 dBA at a distance of 1 meter from the chamber.
18. The chamber shall have an interior light and front door with fog free heated window.
19. The chamber shall have heavy duty casters.
20. Electrical access to DUT is required. Preferred method is an air-tight bank of environment resistant BNC connectors on a side of the chamber which are female coupler connections to connect a cable from the exterior to and another cable on the interior of the chamber. BNC connectors shall be able to withstand the environment of this machine for the life of the machine.
21. The chamber shall have over and under temperature protection.
22. The chamber shall include one (1) adjustable stainless steel shelf
Altitude/Humidity/Temperature Chamber DLA Land and Maritime Engineering &
23. Shall have a color touch screen controller with user friendly Windows like environment (a
Watlow F4 controller is not acceptable).
24. The controller shall have the capability to upload and download files/data/programs to a computer through a USB or Ethernet port.
25. Exterior: 48” wide x 48” deep x 84” high (approximate).
26. Interior Test Space shall be proportionate: 32” wide x 32” high x 32” deep (approximately
18 cubic feet).
27. Stainless steel interior.
28. Painted steel exterior.
29. Shall be double walled construction, and sufficiently insulated to prevent sweating and excessive heat transfer to or from the ambient atmosphere.
B. Other Requirements:
1. Minimum warranty period for parts, labor, and travel is 1 year.
2 Provide three references from labs/companies with installed quoted system.
3. Quoted temperature chamber shall have been on the market in the USA for a minimum of
3 years.
4. The contractor shall have a field service department capable to respond within 48 hours for system repair.
5. The contractor shall have an inventory of spare parts within the USA that can be shipped within 48 hours in the event of system repair.
6. Basic operator training (estimated ½ day) shall be on-site for not more than 5 students.
7. All training materials shall be supplied by the contractor.
8. The contractor will be responsible to remove the system from the shipper’s truck and place the components in their final position.
9. The contractor shall be responsible for installation and start up, which consists of 1) verifying proper wiring of the chamber to the power disconnect box, 2) turn system on, run diagnostics, and verify system operation, 3) demonstrate to PTC personnel the system meets specifications.
10. After installation and prior to acceptance, the system shall be burned-in for 48 hours (not necessarily continuous) and then pass all installation checkout procedures. In the event of a system fault, the system will be repaired and then another 48 hour burn-in is required.
11. Cost for shipping must be included.
12. Cost for initial NIST calibration documentation must be included.
13. The system shall be delivered to :
Defense Supply Center Columbus
Product Testing Center
3990 E. Broad Street
Building 11, Section 7
Columbus, Ohio 43218-3990
Altitude/Humidity/Temperature Chamber DLA Land and Maritime Engineering &
Other Requirements:
1. Extended warranty, service and calibration contract for all equipment must be offered.
2. ISO 17025 or NIST calibration certification and data must be included.
3. All software must include permanently licensed, and not be a subscription-based license including calibration verification features.
4. Shipping cost shall be quoted as a separate line item.
5. Contractor shall contact Larry Baas (614-692-9893, larry.baas@dla.mil) or Eric Kelly
(614-692-9862, eric.kelly@dla.mil) to schedule the delivery.
6. Payment of invoices will be accomplished by payment through the Defense Finance and Accounting Service. Invoices will be submitted to the Wide Area Workflow system in accordance with DFARS 252.232-7003 Electronic Submission of Payment
Requests and Receiving Reports and DFARS 252.232-7006 Wide Area Workflow
Payment Instructions within 10 workdays after the conclusion of work performed.