The solicitation requires a comprehensive analytical laboratory system consisting of seventeen line items totaling one unit each. The primary equipment is the Agilent 8900 ICP-MS bundle package (G3665AA) that includes a personal computer and MassHunter software for data analysis, along with the MassHunter Chromatographic data analysis option (G7205C). Supporting equipment includes a recirculating chiller (G8414A) with non-CFC water circulation and temperature control, and an SPS 4 compact autosampler (G8415A) designed for high-throughput laboratory operations. The specification also includes consumables and calibration materials: checkout solutions, PA tuning solution set, and stock tuning solution containing 10 mg/L of lithium, yttrium, cerium, thallium, and cobalt in 2 percent nitric acid. Gas handling components comprise a carrier gas filter kit, tube cutter, stainless steel ORS cell gas tubing (two units at 6 meters each), wall mounting bracket, and five gas regulators for helium, hydrogen, oxygen, argon, and ammonia/helium reaction gases. Each regulator features two-stage construction with brass or stainless-steel bodies, PTFE seals, and stainless-steel diaphragms with specified inlet pressures of 3000 psig and outlet pressures ranging from 15 to 100 psig. The procurement concludes with on-site method and application consulting (H2149A) for a maximum of four participants.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SPECIFICATIONS:
Agilent ICP-MS System - Salient characteristics
Description Qty
1 G3665AA
Bundle package, Agilent 8900 (G3665A) ICP-MS, PC and MassHunter software
2 G7205C ICP-MS MassHunter Chromatographic data analysis option. 1
3 G8414A Recirculating Chiller for ICP-MS. Non-CFC water recirculator with temperature control.
4 G8415A SPS 4 is a compact 4 rack autosampler, ideal for high throughput labs requiring the ultimate in performance without compromising bench
5 5185-5850 ICP-MS Checkout Solutions
6 5188-6524 PA tuning solution set
7 5188-6564 ICP-MS stock tuning solution (100 mL). Contains 10 mg/L: Li, Y, Ce, Tl and Co in a matrix of
2 percent HNO3
8 CP17976
Gas Clean Carrier Gas filter kit (includes 1 connecting unit 1/8 and 2 Carrier Gas filters)
9 8710-1709 Cutter, tube
10 G3270-65035 ORS cell gas tubing, stainless steel, 1/8 in od, 6 m. For supply connection of Helium, Hydrogen or reaction gas from regulator to instrument.
11 CP7981
Wall Mounting Bracket, Gas Clean Wall Mounting Bracket for 1-position connecting unit only
12 0101-1536 Helium gas regulator, two-stage, brass body, PTFE seals, stainless steel diaphragms, CGA580 3000 psig inlet, 15 psig with 30 psig gauge, 1/8 inch Swagelok outlet fitting, suits ICP-MS. Available in US only. Recommended for ORS cell gas for routine ICP-MS x-lens applications.
13 0101-1535 Hydrogen gas regulator, two-stage, brass body, PTFE seals, stainless-steel diaphragms, CGA350 3000 psig inlet, 15 psig with 30 psig gauge, 1/8 inch Swagelok outlet fitting, suits
ICP-MS
14 0101-1537 Oxygen gas regulator, two-stage, brass body, PTFE seals, stainless-steel diaphragms, CGA540 3000 psig inlet,15 psig with 30 psig gauge, 1/8 inch Swagelok outlet fitting, suits
ICP-MS.
15 0101-1400
Argon high purity gas regulator two-stage, stainless-steel body and diaphragms, CGA580 3000 psig inlet, 100 psig outlet with 200 psig gauge, 1/4 inch FPT fitting, suits ICP-MS.
16 5188-5374 Ammonia/helium reaction gas regulator, two-stage, high purity stainlesssteel body and diaphragms, CGA705 3000 psig inlet, 15 psig outlet with 30 psig gauge, 1/4 inch FPT fitting, for ICP-MS.
17 H2149A Method and Application Consulting On-site consulting for a maximum of 4 participants.