This is an Agency Specifications document for the procurement of a Troxler Model 5151 Automatic Drying Unit (ADU) with a separate vacuum pump assembly. The equipment is a moisture analysis device that must meet specific technical requirements, including a maximum overall size of 25" W x 10" D x 12" H, weighing between 20 and 23 pounds. The unit must operate effectively across a temperature range of 2 to 38 degrees Celsius and achieve drying times of 10 to 20 minutes depending on moisture content. The ADU must be powered by 12V DC, include a chamber capable of accommodating samples up to 150 millimeters in diameter, and comply with ASTM D7227 and AASHTO PP75 standards, with internal pressure checks performed daily as a minimum requirement. Delivery is required to 399 25th Street NE, Fayette, Alabama 35555, and the selected equipment must include a standard one-year manufacturer's warranty.

The procurement requires that bidders submit exact compliance with all stated specifications with no alternative brands or substitutions permitted. Storage temperature capabilities must range from -55 to 85 degrees Celsius. Bidders must include shipping and freight charges in their proposals if applicable, and the equipment must be tax-exempt. No information regarding pricing terms, set-asides, response deadlines, award dates, or contract terms beyond the one-year warranty is provided in this specifications document.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications for a Troxler Model 5151 Automatic Drying Unit (ADU), to include separate vacuum pump assembly

· Overall size: max. 25” W x 10” D x12” H (63.5x25.4x30.5 cm)

· Weight: min. 20 lbs max. 23 lbs

· Storage Temperature: min. -55 to max. 85 C (-67 to 185 F)

· Operating temperature: min. 2 to max. 38 C (55 to 101 F)

· Drying time: moisture content dependent, min. 10 to max. 20 minutes average

· Power: 12 V DC

· Calibration: internal pressure check performed daily, minimum

· Standards: comply with ASTM D7227 and AASHTO PP75

· Chamber shall be sized for samples up to 150 mm in diameter

To be shipped/delivered to:

399 25th Street NE Fayette, AL 35555

Include shipping/freight charge in bid if applicable Unit to include 1 year (12 months) standard manufacturer’s warranty

**Must meet all specifications/No alternative brands/No substitutions** Must be tax exempt