This is an item description document from the Department of the Air Force's 76th Maintenance Support Group at Tinker Air Force Base, Oklahoma, specifying requirements for four Water Cooled X-ray Systems (A24-007).

The technical specifications detail a water-cooled X-ray system that must automatically adapt to input line voltages of 100-130VAC and 200-250VAC, with operating capabilities of 10-160kV and 0.5-5.0mA. The system features a beryllium tube window, radiation output of 14R/min at 50cm, and radiation leakage less than 2.0 Roentgens per hour at 1 meter. Physical requirements include tube head dimensions no greater than 7.25" diameter and 30" length, with component weights of 54 lbs for the cooling unit, 29 lbs for liquid cooling, and 26 lbs for the control unit. The system includes comprehensive safety devices such as pressure relief valves, thermal cut-outs, and flow sensors. Standard accessories include a 100' tubehead cable, 25' power cable, 50' coolant hose, operation manual, carrying case, and GUI control software for system operation and diagnostics.

View the file

Other files

File Type Posted
Solicitation - FA812525Q0027.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DEPARTMENT OF THE AIR FORCE

76th MAINTENANCE SUPPORT GROUP (76 MXSG)

TINKER AIR FORCE BASE, OKLAHOMA

Item Description

A24-007 Water Cooled Xray System

76 AMXG Request for the purchase of 4 ea.Water Cooled Xray System to include the following:

Input Line Requirements:

Automatically adapts to input line voltage100-130Vac, 50/60Hz, 20 Amperes maximum.

200-250Vac, 50/60Hz, 10 Amperes maximum 10 to 160kV, 0.5mA to 5.0mA (800 watts max.)

Constant potential, end-grounded anode, liquid cooled version X-Ray Tube Window: Beryllium 0.8mm (directional) 40° directional Radiation Output: 14R/min at 50cm filtered with 0.5 inches (12.7mm) aluminum at 160kV, 5mA Radiation Leakage: Less than 2.0 Roentgens per hour at 1 meter from the X-Ray tube target Effective Focal Spot: Standard: 0.7mm x 1.0mm (0.03in x 0.04in) EN12543

0.5mm x 0.5mm (0.02in x 0.02in) nominal IEC60336 Ambient Temp: 100% duty cycle @ 120°F (49°C) Duty Cycle: 100% Storage Temp: -30°F to 160°F (-35°C to 71°C) Anode Cooling: Liquid coolant solution closed loop between X-Ray tube anode and cooling unit.

Liquid Cooling Unit Weight: 54lbs. (15.4kg) approx.

Tube Head Dimensions: No greater than 7.25" Diameter, 30" Length Liquid Cooled: 29lbs. (13.15kg)

DEPARTMENT OF THE AIR FORCE

76th MAINTENANCE SUPPORT GROUP (76 MXSG)

TINKER AIR FORCE BASE, OKLAHOMA

X-Ray Control Unit: X-Ray Control Unit Weight: 26lbs. (11.8kg) approx.

Safety Devices:

Tubehead Pressure Relief Valve Tubehead Thermal Cut-Out Tubehead Pressure Gauge Tubehead Low Pressure Cut-Out @ 25psi (1.72 Bar) Coolant Flow Sensor (liquid cooled only) Control Unit Safety Keyswitch Microcomputer-based Self Diagnostics Continuous Exposure Parameter Display

Standard Accessories:

Operation manual Tubehead carrying case Tubehead Cable- 100’ with strain relief Extra key (1) for Control Unit Safety Lock Power cable - 25’ with strain relief Coolant hose - twin, 50’ with self-sealing terminations (liquid cooled only) Cooler power cable GUI Control Software for the GUI is specifically designed for controlling series systems. As an alternative to the front panel control, the GUI will allow the user to control all necessary functions of the system from a user friendly windows based menu. Additionally the GUI can be used as a diagnostic tool when the system is controlled via the front panel.