A05 - SOW MEDRAD Injector.pdf
PDF 99 KB Posted
- Solicitation number
- 75H70926BAO-007-RFQ
About this file
This Statement of Work (SOW) outlines preventive maintenance services for Bayer/MEDRAD Power Injector equipment at the Crow/Northern Cheyenne Indian Health Service Radiology Department. The contractor will perform three physical maintenance checks annually on two specific medical devices: a Medrad Portegra2 Overhead Counterpoise System (Serial No. 0418/22951) and a Medrad Stellant D CT Injection System (Serial No. 204358).
Specific maintenance tasks include calibration per OEM specifications, operational inspections, flow rate and volume checks, software and hardware updates, proactive part replacements, and a complete functional check of the head mounting system. The contract period runs from 12/01/2025 to 11/30/2028, with the contractor required to submit a service report to Program Manager James Reed within two weeks of each maintenance check. The work will be performed at the Radiology Department located at 1 Hospital Way in Crow Agency, Montana, and the contractor may be subject to COVID-19 screening prior to facility entry.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| A05 - SOW MEDRAD Injector amended.pdf | ||
| B01 -75H70926BAO-007-RFQ MEDRAD.pdf | ||
| B01 - Price Schedule.xlsx | XLSX spreadsheet |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work (SOW)
1. PROJECT TITLE: Preventive maintenance for Bayer/MEDRAD Power Injector unit
2. BACKGROUND/PURPOSE: Bayer/MEDRAD Power Injector is used to inject calculated amount of contrast for CT scans.
3. SCOPE: The Contractor will come to the Crow Service Unit Radiology department and preform preventative maintanence services on the Bayer/MEDRAD Power Inject Unit
4. APPLICABLE DOCUMENTS: Joint Commision Equipment Maintenance Standards
5. SPECIFIC TASKS: The Contractor will conduct the following Prevenative
Maintenance checks on the Medrad Portegra2 Overhead Counterpoise System SERIAL NO. 0418/22951 and Medrad Stellant D CT Injection System SERIAL NO.
204358.
a. Calibration per OEM specifications and Procedures.
b. operational inspection, flow rate and volume check, plunger position check and pressure limit check
c. Proprietary software and hardware updates
d. EPM proactive certified part replacement
e. Complete inspection and functional check of head mounting system(OCS, Table Brackets, and Pedestals)
f. Warranty for travel, parts and labor are covered under the contract.
6. DELIVERABLES AND DELIVERY SCHEDULE:
a. The Contractor will at a minimum complete a physical preventative maintenance check (3) three times a year and monitor the equipment centrally.
b. The Contractor will provide a service report to the Program Manager James Reed within 2 weeks of completing preventative maintenance check.
7. GOVERNMENT-FURNISHED EQUIPMENT AND INFORMATION: N/A
8. PLACE OF PERFORMANCE:
a. Crow/Northern Cheyenne Indian Health Service
Radiology Department 1 Hospital Way Crow Agency, Montana 59022
9. PERIOD OF PERFORMANCE: 12/01/2025-11/30/2028
10. SECURITY: The Contractor could be screened for Covid-19 prior to entry into facility.