Statement of Work Aberdeen Test Center Sindie Gen 3 Analyzer
Description of Purchase:
Sindie Gen 3 m-series analyzer uses Monochromatic Wavelength Dispersive X-ray Fluorescence (MWDXRF) with state-of-the-art focusing and monochromating optics to increase excitation intensity and dramatically improve signal-to-background ratio compared to traditional WDXRF instruments, testing to ASTM D7039. The analyzer uses hassle free Accucells for samples and only 1.0 mL of sample. The detection limit of the analyzer should be 0.15 ppm at 300 s with a dynamic detection range of 0.15 ppm - 3000 ppm sulfur. The measurement time of the analyzer should be between 30- 900 seconds with a 75W air-cooled excitation tube. The analyzer should have an intuitive touch screen to use and be ready to plug-in and measure right away.
System must include complete delivery, installation, and on-site familiarization training for up to 4 people. Installation and training must be provided by a certified service engineer.
System must be installed turn-key ready for analysts to performed analysis in accordance with internal operating procedures and ISO 17025.
Period of performance (POP):
Instrument should be delivered 30 days after contract is awarded along with the training.
Instrument must be delivered to: 358 Michealsville Rd B363 APG MD 21050.
The POC for this instrument is Mark Michie. mark.w.michie.civ@army.mil
Service on the instrument should be performed yearly based on the awarded contract date and delivery date. Service should also be performed at Instrument must be delivered to: 358 Michealsville Rd B363 APG MD 21050.
Technical Requirements:
Part number 400905-01m, Sindie, Gen 3 Analyzer, m-series, quantity 1 Setup/Train, 1day, flat fee per analyzer, quantity 1 Part number 490002-01, Kitset, Accucell, Mylar, 100 Ea (10 x 10 tubes), quantity 3 BT 2nd & 3rd Year Warranty, quantity 1 Preventive Maintenance - BT CONTRACT, quantity 2 Setup/Train Travel Expenses, quantity 1 mailto:mark.w.michie.civ@army.mil
Security:
Service engineers will need to be granted access to base during training and service.
Instrument is non-network capable and will not interface with the ATC network in any way.
Warranty:
The Sindie will be covered with a 100% coverage on all parts, labor, and travel for 2 years after your original factory warranty expires. ATC will also receive priority remote technical and application assistance.
The 2nd and 3rd Year Warranty also comes with 2 PM service visits by our field engineers (1 each year) which includes: inspection, cleaning, calibration, software updates, radiation survey, primary window replacement, fan filter replacement, and a vacuum pump rebuild.
The Sindie will also contain an extension of 2 years for Preventative Maintenance (PM) visits by XOS field engineers and 15% off any parts during the visits. ATC also receives priority remote technical and application assistance. The PM service includes:
inspection, cleaning, calibration, software updates, radiation survey, primary window replacement, fan filter replacement, and a vacuum pump rebuild.