SATPC0040194 Tab 04 4 SOW.pdf

PDF 104 KB Posted

Attached to
Ardara quadrupole mass spectrometer with cross-beam ionizer Federal contract opportunity
Solicitation number
80NSSC26922480Q
Issued by
National Aeronautics and Space Administration Shared Services Center

About this file

This Statement of Work (SOW) details the procurement of a quadrupole mass spectrometer (QMS) with cross-beam ionizer for an unspecified federal agency. The required instrument must have a mass range of 1-50 AMU, a 12 mm quadrupole mass filter operating at a minimum 3.3 MHz frequency, and a 0.92" ionization volume with a 0.375" entrance aperture. Specific technical requirements include an analytical quadrupole section at least 8" long, yttrium oxide coated iridium filaments, and a complete system including ion detector, power supplies, cables, and data acquisition software.

The procurement specifies a delivery timeline of 60-90 days after order receipt. The vendor must supply a QMS capable of distinguishing atomic and molecular hydrogen, with no specified place of performance noted in the document. The SOW provides precise technical specifications without additional context about the procurement process, funding, or end-user application.

View the file

Other files for this federal contract opportunity

Other files attached to Ardara quadrupole mass spectrometer with cross-beam ionizer, newest first.
File Type Posted
SATPC0040194 Tab 07 Capability Statement SAM.gov.pdf PDF
SATPC0040194 Tab 06 RDSS..pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

1. Objective/Requirements Vendor shall supply quadrupole mass spectrometer with cross-beam ionizer.

2. Characteristics, Scope, and Specs

A quadrupole mass spectrometer (QMS) with cross-beam ionizer capable of distinguishing atomic and molecular hydrogen with the following features:

Mass range 1-50 AMU 12 mm quadrupole mass filter with frequency of at least 3.3 MHz ionization volume with ID 0.92” entrance aperture 0.375” analytical portion of the quadrupole of at least 8” length yttrium oxide coated iridium filaments

QMS shall include appropriate ion detector, power supplies, cables, and data acquisition software.

3. Place of Performance N/A

4. Period of Performance Delivery is 60-90 days after receipt of order.