Amendment A00001_Questions and Answers.pdf
PDF 115 KB Posted
- Solicitation number
- 75F40124Q00378
About this file
This document is a set of Questions and Answers related to Solicitation 75F40124Q00378 for Annual Pipette Calibrations for the Office of Applied Research and Safety Assessment.
The key details are:
- ISO17025:2017 accreditation is not required for the calibration services, as it would be an additional expense beyond what is needed.
- There are no specific templates provided for Volumes I and II submissions - the contractor can use their own templates.
- The government will provide payment for additional parts needed for ad-hoc repairs outside the standard preventative maintenance.
- Calibration testing is required for each individual channel on multi-channel pipettes, and if one channel fails, the entire pipette will fail.
- The solicitation has a deadline of August 12, 2024 at 12:00 pm ET for questions, and a deadline of August 16, 2024 at 12:00 pm ET for quotes.
- The contracting agency is the Department of Health and Human Services Food and Drug Administration Office of Acquisition and Grant Services.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Questions and Answers
75F40124Q00378
Question 1: Please confirm that this is to be ISO17025:2017 accredited service with Certificates for each Pipette.
Response 1: NOT required, ISO17025:2017 accredited calibration services. This would be an additional expense beyond what is needed for our research purposes.
Question 2: Will you provide a standard format for Volume I and Volume II submissions? Where do we locate these templates?
Response 2: There are no specific templates. The Contractor can utilize their own templates for submission.
Question 3: Are all repair parts included in the unit price or will a government agent provide payment for additional parts needed for ad-hoc repairs outside the scope of a standard preventative maintenance?
Response 3: The government agent will provide payment for additional parts needed for ad-hoc repairs upon approval for such repairs.
Question 4: Can you clarify that we will be providing data for one channel on these pipettes rather than all channels?
Response 4: Yes, as long as it is understood that they will test each channel individually. If one channel fails on a multi-channel pipette, then the pipette as a whole should fail.