7 SOW for RFQ.pdf
PDF 393 KB Posted
- Attached to
- NOAA: Relative Humidity Calibration Standard built to Specifications Federal contract opportunity
- Solicitation number
- 1305M226Q0197
About this file
This is a Statement of Work for the procurement of a Relative Humidity/Dewpoint Primary Standard Replacement (Contract Number 142082-26-0391) for the National Weather Service Sterling Field Support Center (SFSC) Temperature and Humidity Laboratory (HTL).
The procurement requires the purchase of a maximum of two pieces of relative humidity/dewpoint measurement equipment to replace an aging primary calibration standard that is experiencing frequent failures and requiring costly vendor repairs. The equipment will be used to provide internal calibration for 16 humidity probes and accuracy testing services for programs testing new humidity sensing equipment. Key sensor measurement requirements include: dew/frost point measurement range equal to or better than -62°C to 54°C with accuracy of ±0.1°C maximum and reproducibility of ±0.05°C maximum; relative humidity measurement range of 1–100% at 20°C; and display resolution of 0.01°C or better. The equipment must include a heated hose for gas sampling with a minimum length of 55 inches that fits within a 1.874-inch diameter port hole, feature a front-facing digital display, PT100 Class A temperature probe, and 316L stainless steel electropolished sample gas circuit. Additional requirements specify maximum dimensions of 20 inches width x 12 inches height x 24 inches length, maximum weight of 75 pounds, 100–120 VAC 50/60HZ power source, serial communications capability with third-party terminal programs, 1-year minimum calibration interval, and ISO-17025 accredited calibration with certification. The equipment must operate in standard laboratory environments with temperatures of 15°C to 35°C and ambient humidity of 20%–70% RH. Delivery shall be completed at Sterling Field Support Center, 43741 Weather Service Road, Sterling, VA 20166. The contractor must provide a full successful demonstration in accordance with a vendor-supplied test plan with federal representative signature, and acceptance test results must be submitted with the invoice for payment approval. The government point of contact is Kathryn Franc (kathryn.franc@noaa.gov, 301-388-8565).
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 2 SF-30 1305M226Q0197_Amendment 0001_v.1.pdf | ||
| 5 SOL 1305M226Q0197_V.1.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work 142082-26-0391
Relative Humidity/Dewpoint Primary Standard Replacement
The National Weather Service Sterling Field Support Center (SFSC) Temperature and Humidity Laboratory (HTL) provides internal calibration for 16 humidity probes as well as accuracy testing services for programs testing new humidity sensing equipment. In order to provide accurate calibrations, the use of an accurate relative humidity/dewpoint standard is required. The primary calibration and testing standard currently used by SFSC is reaching the end of its life cycle. Staff have noted a dramatic increase in the number of failures that the system has experienced over the past few years. These failures often require that the unit be returned to the vendor for repair. This can be costly and often results in long periods of time that the unit is not available for use, increasing laboratory downtime, and decreasing testing availability for NWS programs. A replacement dewpoint refence must be purchased to reduce laboratory downtime and equipment repair costs while allowing for laboratory redundancy for a short period of time while the original unit is still functional to ensure installation and integration time for new equipment.
Performance Criteria
The equipment requirements listed below must all be met for the final acceptance of the system.
Number of Equipment to meet Performance Criteria: Maximum of two pieces of equipment o 1 to 2 pieces of equipment to meet full Sensor Measurement Requirements for Dew/Frost Point Range and RH Measurement Range. All other Sensor Measurement requirements must apply to both units if more than one unit is used.
Sensor Measurement Requirements:
o Dew/Frost Point Measurement Range: Equal to or better than -62°C to 54°C o Dew/Frost Point Accuracy (Max): +/- 0.1°C o Reproducibility (Max): +/- 0.05°C o RH Measurement Range at 20°C: 1 – 100% o Display Resolution: Less than or equal to 0.01°C o ppmv, minimum: 2.5
Heated Hose Length: Must include a heated hose for gas sampling with minimum length of 55 inches
Heated Hose Diameter: Must fit inside of a port hole with a 1.874 inch diameter
Display: Front facing Digital Display
Temperature Probe: PT100 Class A or equivalent
Sample gas circuit: 316L stainless steel, electropolished or high quality PTGE
Dimensions (Max): Width 20in x Height 12in x Length 24 in
Weight (Max): 75 lbs
Power Source: 100 to 120 VAC, 50/60HZ
Serial Communications Requirements: Must possess the ability to receive text strings sent from a third-party terminal program such as Hyper Terminal or PuTTy. Must have the ability to send text strings to a third-party terminal program containing information about the current measurement readings.
Calibration Interval: 1-year minimum calibration interval
Environmental Operational Requirements:
o Temperature Range: Must be capable of operating in a standard laboratory environment with a temperature range of 15°C to 35°C o Ambient Humidity Range: Must be capable of operating in a standard laboratory environment with a RH range of 20% - 70% RH
Calibration Requirements: Each unit must come with an ISO-17025 Accredited Calibration with a Calibration Certificate and Sticker for every unit
Maintenance: Maintenance costs minimized
Shipping: Shipping Cost included
Inspection and Acceptance
Upon completion, contractor shall notify the Government representative for approval and acceptance. The acceptance period will include a full successful demonstration in accordance with vendor supplied accepted test plan with the federal representative signature. Delivery of acceptance test results from the test plan must be submitted with invoice for processing and approval of payment.
Delivery/Point of Service to be Completed, if applicable:
Sterling Field Support Center 43741 Weather Service Rd Sterling VA, 20166
National Weather Service POC:
Kathryn Franc kathryn.franc@noaa.gov
301-388-8565
File details come from the government source that posted it. Updated .