This document is a detailed STERIS Harmony Custom Integration System installation guideline for Operating Rooms (OR) #3-#10 at the VA Bronx Medical Center in New York. The comprehensive set of technical drawings and specifications outlines the installation requirements for an advanced surgical room integration system, including precise mounting instructions for surgical lights, utility hubs, booms, displays, speakers, and control systems. The documentation provides explicit responsibilities for both STERIS and the customer's electrical contractor, covering conduit specifications, electrical requirements, equipment placement, and site preparation guidelines.
The integration system involves installing sophisticated medical technology across multiple operating rooms, with specific technical details for each room's layout. Key components include HarmonyAIR utility hubs, A-Series surgical lights, wall-mounted displays (LG 55" 4K IPS LED Commercial Grade), integrated wall controls, and IP status room cameras. The documents emphasize strict installation protocols, including conduit specifications, power requirements, junction box placements, and specific mounting heights and clearances. The most recent revision (Revision 5) was made on 16APR2024, updating the system to current standards and changing OR6-8 to the HV1000 system.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C24225B0012 0001.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
00 COVER
01 PROPOSED EQUIPMENT LAYOUT
DRAWING INDEX
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
SIGN OFF ACCEPTANCES
__________________________________ (OPTIONAL)
__________________________________ (OPTIONAL)
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
DATE
SCALE
CKDR
CHECK PRINT
PAGE
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
LJD 22SEP2005
COVER SHEET
NOT TO SCALE 00
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
VESTIBULE
2E-18B
STERILIZA.
2E-20A
STERILIZA.
2E-18A
O.R. #10 2E-19A
F.A .C
STORAGE
2E-20B
O.R. #8 2E-19
O.R. #9 2E-20
O.R.#7CARD
CARD.EQUP.
2E-18
BLOOD RM.
2E-01E
2E-17
X-RAY&EQUP
2E-15
SPLT.&PLS.
2E-14
O.R. #2 2E-05
O.R. #3 2E-06
CONTROL RM
2E-10A
O.R. #4 2E-07
O.R. #5 2E-08
ORH.O.R.#6
2E-09
01E 01E 01E
01E
01E01E
63"
60" 48" 53" 46" 60" 48"
60" 48"411 2"
60" 48" 761 8"
53 8"
48" 60" 641 2"
5 8"
48" 60" 671 2"
5 8"C7
B1
C1
B3
C5
C6
C2 C2
C5
B3 C6
01E 01E
4"
1 8"
3 8"
48" 843 4"
1 2"
K1
K1
K1
K1 K1
K1K1
MOUNT C2 LIMITED
UPPER ARM
ROTATION AT WALL
CEILING LIGHTS MODIFIED TO
ACCOMMODATE STERIS
EQUIPMENT
C1
E1
E1
E2E2
K1F
C1F
B1F
C2F
B4B4
B1
E2
E1 C6
C5
B3
EQUIPMENT LIST
TAG QTY DESCRIPTION
01E 8 Existing Amsco 3085 SP Surgical Table
B1 2 A Series-Surgical Light(s)
B1F 1 Future A Series-Surgical Light(s)
B3 3 A Series-Surgical Light(s)
B4 2 A Series-Surgical Light(s)
C1 2 HarmonyAIR Utility Hub-Surgical
C1F 1 Future HarmonyAIR Utility Hub-Surgical
C2 2 HarmonyAIR Adj Arm Spring-Anesthesia
C2F 1 Future HarmonyAIR Adj Arm Spring-Anesthesia
C5 3 Tandem HarmonyAIR Adj Arm Spring - A Series-Anesthesia, Accessory Arm(s)
C6 3 Tandem HarmonyAIR Utility Hub - A Series-Power Boom Support, Accessory Arm(s)
C7 1 HarmonyAIR-Straight-Drop-Anesthesia
E1 3 LG 55" 4K IPS LED Commercial Grade Display - Wall Mounted
E2 3 LG 55" 4K IPS LED Commercial Grade Display - Wall Mounted
K1 7 HarmonyAIR A-Series Integrated Wall Control
K1F 1 Future HarmonyAIR A-Series Integrated Wall Control
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
DISCLAIMER: THE SERVICE CLEARANCES INDICATED DO NOT MEET THE MINIMUM SAFETY CLEARANCES RECOMMENDED IN THE EQUIPMENT DRAWINGS PROVIDED BY STERIS CORPORATION DISCLAIMER: THE WALLS AND OR EQUIPMENT SHOWN IN THIS DRAWING HAVE BEEN MOVED/MODIFIED TO ACCOMMODATE MINIMUM RECOMMENDED CLEARANCES FOR STERIS EQUIPMENT
DATE
SCALE
CKDR
CHECK PRINT
PAGE
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
LJD 22SEP2005
OPERATING ROOMS #3 - #10
PROPOSED EQUIPMENT LAYOUT
1/4" = 1'-0" 01
HV1000 INSTALLATION GUIDELINES
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
STERIS
GENERAL NOTES
1. SEE INFRASTRUCTURE RESPONSIBILITIES PAGE (1A) FOR ADDITIONAL DETAILS & DESCRIPTIONS FOR SPECIFIC STAKEHOLDER RESPONSIBILITIES
2. SUPPLY AND INSTALLATION OF ALL CONDUIT, POWER CIRCUITS, BACK BOXES, ELECTRICAL DEVICES, ELECTRICAL WIRING IS TO BE PROVIDED BY THE
CLIENTS ELECTRICAL CONTRACTOR UNLESS NOTED OTHERWISE
3. ONE WIRE HANGER FOR CABLE COILS REQUIRED FOR EACH MOUNT
4. CLIENT CONTRACTOR TO PROVIDE BLANK COVER PLATES FOR ALL JUNCTIONS BOXES WHERE NECESSARY
5. ALL CABLE RUNS OUTSIDE INTEGRATED ROOMS ARE THE RESPONSIBILITY OF HOSPITAL/CONTRACTOR; ALL CABLE RUNS SHOULD BE CONTINUOUS
EXCEPT WHERE NOTED
6. 24x24" ACCESS MUST BE PROVIDED WITHIN 36" OF ALL JUNCTION BOXES
7. CABLE RUNS THRU PLENUM SPACE MUST BE PLENUM RATED OR SHALL BE INSTALLED WITH CONDUIT
8. CABLES NOT RUN IN CONDUIT MUST HAVE J-HOOKS EVERY 4' FOR SUPPORT (NOT BY STERIS)
9. WHERE J-HOOKS ARE USED IN CEILING, CONDUIT IS REQUIRED IN ALL CONCEALED LOCATIONS (SIZED PER SCHEDULES)
10. EACH RACK REQUIRES (4) NETWORK PORTS PER O.R.
11. ALL POWER FOR INTEGRATION EQUIPMENT SHOULD COME FROM A SINGLE ELECTRICAL PANEL IN THE OR IT SERVES
12. CONFORMANCE TO NATIONAL, STATE AND LOCAL CODES FOR ELECTRICAL CIRCUITS AND CONDUIT IS THE RESPONSIBILITY OF THE HOSPITAL OR
THEIR CONTRACTED ELECTRICIAN
13. THE EQUIPMENT AND ROUGH-IN REQUIREMENTS SHOWN ARE FOR BUDGETARY PLANNING PURPOSES ONLY; EQUIPMENT IS SUBJECT TO CHANGE AND DIFFERENT OPTIONS ARE AVAILABLE TO MEET SPECIFIC CUSTOMER DESIGN CRITERIA; POST-ORDER DRAWINGS WILL REFLECT EXACT EQUIPMENT
SPECIFICATIONS, AND REQUIRES CUSTOMER SIGNED APPROVAL
14. USE OF ISOLATED GROUND POWER TO BE DETERMINED BY ELECTRICAL CONTRACTOR
ALL STERIS TERMS AND CONDITIONS RELATIVE TO STERIS AND CUSTOMER,
DESIGNATED CONTRACTOR RESPONSIBILITIES, SITE READINESS, AND CHANGE
ORDERS APPLY. THESE WILL BE PROVIDED TO ALL PARTIES ABOVE AT THE
PROJECT KICKOFF MEETING BY YOUR STERIS REPRESENTATIVEABBREVIATIONS:
A.F.F. = ABOVE FINISHED FLOOR
B.C.P. = BLANK COVER PLATE
C. = CONDUIT
C.P. = COVER PLATE
C.K.T. = CIRCUIT
E.C. = ELECTRICAL CONTRACTOR
N.S.E. = NURSE STATION EQUIPMENT
INDEX
PAGE # SHEET TITLE ROOM #
01 TITLE PAGE, INDEX -
01A INFRASTRUCTURE RESPONSIBILITIES TERMS & CONDITIONS -
02 PROPOSED OPERATING ROOM #6 LAYOUT 2E-09
02A PROPOSED OPERATING ROOM #7 LAYOUT 2E-17
02B PROPOSED OPERATING ROOM #8 LAYOUT 2E-19
03 PROPOSED OPERATING ROOM CONDUIT RISER DIAGRAM #6-8 -
04 PROPOSED EQUIPMENT INSTALLATION DETAILS -
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HV1000 INTEGRATION SYSTEM
TITLE PAGE, INDEX
NTS 01
STERIS Responsibilities:
· Provide cable conduit drawings, electrical requirements, and mounting requirements for the room.
· Provide space, structural, and heat load requirements for the base unit.
· Provide site preparation specifications.
· Deliver all custom back boxes, mounting brackets, and construction-related items to the hospital in accordance with the mutually agreed upon construction or renovation schedule.
· Participate in project meetings in person, via conference call or via e-mail.
· Ship equipment in protective and labeled packaging.
· Ship cables and equipment on the scheduled ship date
· Pull and terminate non-plenum cabling inside the OR through specified conduits and junction boxes as specified in the Product
Description and Final Project Plan.
· Provide external cables and detailed specifications for the interconnections only if external cabling is specified in the Customer
Proposal and Final Project Plan.
· Provide a pre-installation checklist to verify site preparation is complete.
· Mount, install, and connect integration equipment - including video, audio, and control cables, cable terminations, wall cameras, wall monitors (does not include wall mounting brackets), base unit, touch screen control, and surgical monitors - as applicable to project.
· Test and validate integration system to complete customer satisfaction.
· Provide the hospital with a certificate of completion signed by the install team and the hospital.
· Adjust STERIS surgical light brakes as necessary after installation of STERIS monitors.
· In-service hospital staff at the time of the install. Further in-service training may be scheduled as mutually agreed.
· Leave a tidy work area each evening and remove packing material to a customer provided on-site container.
Customer or Customer's Designated Contractor Responsibilities:
· Ensure that site preparations and construction are complete and ready for final integration installation - including electrical and network connections, finished floor and paint.
· Provide complete room schematics including electrical, mechanical, and equipment location details.
· Install cable conduit, junction boxes, electrical outlets, and mounting brackets according to project specifications.
· Ensure that all electrical requirements comply with local code.
· Ensure space, structural, and heat load requirements are met according to project specs and plan.
· Receive shipment and safely, securely store equipment if shipment arrives prior to requested installation date.
· Provide adequate access from delivery/storage area to installation site.
· Relocate equipment from on-site storage/delivery area to installation site.
· Run all cables outside the OR- Including conduit, J-hooks, penetrations through firewall, video cables, radio antennae cables, or fiber optics to networking hubs and video conferencing hubs, communications closets, conference rooms, centralized nurse desks, rooftop, or other departments.
· Ensure that site preparations and construction are complete and ready for final integration installation - including electrical and network connections, finished floor and paint.
· Proactively schedule the installation in conjunction with the STERIS project manager.
· Provide a minimum of two (2) weeks advance notice for requested changes to the installation schedule.
· Remove old equipment or furniture from the room.
· Provide a clean, dry, dust free environment for equipment storage and installation.
· Provide a secure work site for storing ladders, toolboxes etc. during install.
· Provide easy access to all rooms involved in the installation during normal business hours and upon request after hours.
· Ensure that all non-STERIS or third-party equipment and systems that are to be part of the integrated solution are on-site and functional at the time of installation - otherwise full testing and connections cannot be completed.
· Participate in a final walk-through evaluation of the system followed by a sign-off of the certificate of completion.
· Provide an on-site container for trash, including shipping and packing material. Empty container as needed and on a regular basis.
Cabling and Fiber Responsibilities:
· All cables and fiber optic runs to locations outside the OR are the responsibility of the hospital.
Minimum Site Readiness Successful and efficient installation of a complete integration system requires that the site be carefully prepared according to the project documentation and schematics. A site that is not ready when the installation team arrives will cause delays and additional expenses to both parties. If the installation site is not ready per the mutually agreed specifications, STERIS reserves the right to charge a change order fee and extend the timeline needed to finish project past the planned date given at the project's inception.
To be ready a site should have -
1. All floor work must be complete
2. All painting must be complete
3. All electrical requirements must be installed and powered
4. Room must be clean and free from dust
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HV1000 INTEGRATION SYSTEM
INFRASTRUCTURE RESPONSIBILITIES
TERMS & CONDITIONS
NTS 01A
CONTROL RM
2E-10A ORH.O.R.#6
2E-09
K1
E2
SPEAKER PLACEMENT (G1-G4) FOR
REFERENCE ONLY; SITE VERIFY
G2G3
E1
Q1
G4
G1 D1 E3
C6
C5
B3
V2
T1
V3
TP1
INTEGRATION EQUIPMENT SCHEDULE
LEGEND DESCRIPTION RESPONSIBILITY
B3 SURGICAL LIGHT MOUNT STERIS
C5 TANDEM ANESTHESIA BOOM STERIS
C6 TANDEM UTILITY HUB MOUNT STERIS
D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION BOX
FOR HD UNIVERSAL INPUT ELECTRICAL CONTRACTOR
E1
WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR
E2
WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR
E3 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR OUTPUT TO DAVINCI ELECTRICAL CONTRACTOR
G1 CEILING MOUNTED SPEAKER STERIS
G3 CEILING MOUNTED SPEAKER STERIS
G2 CEILING MOUNTED SPEAKER STERIS
G4 CEILING MOUNTED SPEAKER STERIS
K1 SURGICAL LIGHT CONTROLLER STERIS
Q1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR
T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE CEILING ELECTRICAL CONTRACTOR
TP1
TOUCH PANEL STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR
V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER
DISTRIBUTION ELECTRICAL CONTRACTOR
V3 WALL MOUNTED LEVITON BOX FOR STERIS AUDIO
DISTRIBUTION ELECTRICAL CONTRACTOR
A SERIES LIGHT CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
K1 - B3 2 .75" ELECTRICAL CONTRACTOR
LIGHT CONDUIT SCHEDULE NOTES
1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT
AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz
2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET
3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO
REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE
SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM
4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN
DOUBLE RUNS ARE REQUIRED (LONG RUNS)
5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE
THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS
LIGHT SYSTEM IS UTILIZED
SHEET NOTES
1. TOUCH PANEL MUST BE LOCATED WITHIN 75' OF HV1000 BASE UNIT
2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS
3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE
RIGHT HALF OF THE BOX
4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE
AROUND ALL SIDES OF THE BOX
5. EMERGENCY POWER REQUIRED FOR LEVITON BOX
6. A 24" X 24" HINGED ACCESS PANEL IS REQUIRED AT ALL MOUNTS. COORDINATE LOCATION
& PARAMETERS OF ACCESS PANELS FOR MOUNTS AND ELECTRICAL BOXES WITH STERIS
REPRESENTATIVE
INTEGRATION CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
T1 - B3 1 1.5" ELECTRICAL CONTRACTOR
T1 - C5 2 1.5" ELECTRICAL CONTRACTOR
T1 - C6 2 1.5" ELECTRICAL CONTRACTOR
T1 - D1 1 1.5" ELECTRICAL CONTRACTOR
T1 - E1 1 1" ELECTRICAL CONTRACTOR
T1 - E2 1 1" ELECTRICAL CONTRACTOR
T1 - E3 1 1.5" ELECTRICAL CONTRACTOR
T1 - TP1 1 1" ELECTRICAL CONTRACTOR
T1 - V2 2 1.5" ELECTRICAL CONTRACTOR
V3 - Q1 1 1" ELECTRICAL CONTRACTOR
V3 - G1 1 .75" ELECTRICAL CONTRACTOR
V3 - G2 1 .75" ELECTRICAL CONTRACTOR
V3 - G3 1 .75" ELECTRICAL CONTRACTOR
V3 - G4 1 .75" ELECTRICAL CONTRACTOR
ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT
SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HV1000 INTEGRATION SYSTEM
PROPOSED OPERATING ROOM 6 LAYOUT
[2E-09]
NTS 02
O.R.#7CARD
CARD.EQUP.
2E-18
BLOOD RM.
2E-01E
2E-17
C5
B3 C6
K1 G1 G4
SPEAKER PLACEMENT (G1-G4) FOR
REFERENCE ONLY; SITE VERIFY
G2 G3
D1
E3
E1
E2
SERVES OR 7
Q1
TP1
V2
T1
V3
INTEGRATION EQUIPMENT SCHEDULE
LEGEND DESCRIPTION RESPONSIBILITY
B3 SURGICAL LIGHT MOUNT STERIS
C5 TANDEM ANESTHESIA BOOM STERIS
C6 TANDEM UTILITY HUB MOUNT STERIS
D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION BOX
FOR HD UNIVERSAL INPUT ELECTRICAL CONTRACTOR
E1
WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR
E2
WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR
E3 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR OUTPUT TO DAVINCI ELECTRICAL CONTRACTOR
G1 CEILING MOUNTED SPEAKER STERIS
G3 CEILING MOUNTED SPEAKER STERIS
G2 CEILING MOUNTED SPEAKER STERIS
G4 CEILING MOUNTED SPEAKER STERIS
K1 SURGICAL LIGHT CONTROLLER STERIS
Q1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR
T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE CEILING ELECTRICAL CONTRACTOR
TP1
TOUCH PANEL STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR
V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER
DISTRIBUTION ELECTRICAL CONTRACTOR
V3 WALL MOUNTED LEVITON BOX FOR STERIS AUDIO
DISTRIBUTION ELECTRICAL CONTRACTOR
A SERIES LIGHT CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
K1 - B3 2 .75" ELECTRICAL CONTRACTOR
LIGHT CONDUIT SCHEDULE NOTES
1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT
AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz
2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET
3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO
REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE
SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM
4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN
DOUBLE RUNS ARE REQUIRED (LONG RUNS)
5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE
THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS
LIGHT SYSTEM IS UTILIZED
SHEET NOTES
1. TOUCH PANEL MUST BE LOCATED WITHIN 75' OF HV1000 BASE UNIT
2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS
3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE
RIGHT HALF OF THE BOX
4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE
AROUND ALL SIDES OF THE BOX
5. EMERGENCY POWER REQUIRED FOR LEVITON BOX
6. A 24" X 24" HINGED ACCESS PANEL IS REQUIRED AT ALL MOUNTS. COORDINATE LOCATION
& PARAMETERS OF ACCESS PANELS FOR MOUNTS AND ELECTRICAL BOXES WITH STERIS
REPRESENTATIVE
INTEGRATION CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
T1 - B3 1 1.5" ELECTRICAL CONTRACTOR
T1 - C5 2 1.5" ELECTRICAL CONTRACTOR
T1 - C6 2 1.5" ELECTRICAL CONTRACTOR
T1 - D1 1 1.5" ELECTRICAL CONTRACTOR
T1 - E1 1 1" ELECTRICAL CONTRACTOR
T1 - E2 1 1" ELECTRICAL CONTRACTOR
T1 - E3 1 1.5" ELECTRICAL CONTRACTOR
T1 - TP1 1 1" ELECTRICAL CONTRACTOR
T1 - V2 2 1.5" ELECTRICAL CONTRACTOR
V3 - Q1 1 1" ELECTRICAL CONTRACTOR
V3 - G1 1 .75" ELECTRICAL CONTRACTOR
V3 - G2 1 .75" ELECTRICAL CONTRACTOR
V3 - G3 1 .75" ELECTRICAL CONTRACTOR
V3 - G4 1 .75" ELECTRICAL CONTRACTOR
ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT
SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HV1000 INTEGRATION SYSTEM
PROPOSED OPERATING ROOM 7 LAYOUT
[2E-17]
NTS 02A
O.R. #8 2E-19
CARD.EQUP.
2E-18
BLOOD RM.
2E-01E
B3
C5
C6
K1
G1 G4
SPEAKER PLACEMENT (G1-G4) FOR
REFERENCE ONLY; SITE VERIFY
G2 G3
D1
E3
E1
E2
N3
SERVES
OR 8
Q1
TP1
V2
T1
V3
INTEGRATION EQUIPMENT SCHEDULE
LEGEND DESCRIPTION RESPONSIBILITY
B3 SURGICAL LIGHT MOUNT STERIS
C5 TANDEM ANESTHESIA BOOM STERIS
C6 TANDEM UTILITY HUB MOUNT STERIS
D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION BOX
FOR HD UNIVERSAL INPUT ELECTRICAL CONTRACTOR
E1
WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR
E2
WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR
E3 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR OUTPUT TO DAVINCI ELECTRICAL CONTRACTOR
G1 CEILING MOUNTED SPEAKER STERIS
G3 CEILING MOUNTED SPEAKER STERIS
G2 CEILING MOUNTED SPEAKER STERIS
G4 CEILING MOUNTED SPEAKER STERIS
K1 SURGICAL LIGHT CONTROLLER STERIS
Q1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR
T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE CEILING ELECTRICAL CONTRACTOR
TP1
TOUCH PANEL STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR
V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER
DISTRIBUTION ELECTRICAL CONTRACTOR
V3 WALL MOUNTED LEVITON BOX FOR STERIS AUDIO
DISTRIBUTION ELECTRICAL CONTRACTOR
A SERIES LIGHT CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
K1 - B3 2 .75" ELECTRICAL CONTRACTOR
LIGHT CONDUIT SCHEDULE NOTES
1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT
AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz
2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET
3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO
REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE
SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM
4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN
DOUBLE RUNS ARE REQUIRED (LONG RUNS)
5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE
THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS
LIGHT SYSTEM IS UTILIZED
SHEET NOTES
1. TOUCH PANEL MUST BE LOCATED WITHIN 75' OF HV1000 BASE UNIT
2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS
3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE
RIGHT HALF OF THE BOX
4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE
AROUND ALL SIDES OF THE BOX
5. EMERGENCY POWER REQUIRED FOR LEVITON BOX
6. A 24" X 24" HINGED ACCESS PANEL IS REQUIRED AT ALL MOUNTS. COORDINATE LOCATION
& PARAMETERS OF ACCESS PANELS FOR MOUNTS AND ELECTRICAL BOXES WITH STERIS
REPRESENTATIVE
INTEGRATION CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
T1 - B3 1 1.5" ELECTRICAL CONTRACTOR
T1 - C5 2 1.5" ELECTRICAL CONTRACTOR
T1 - C6 2 1.5" ELECTRICAL CONTRACTOR
T1 - D1 1 1.5" ELECTRICAL CONTRACTOR
T1 - E1 1 1" ELECTRICAL CONTRACTOR
T1 - E2 1 1" ELECTRICAL CONTRACTOR
T1 - E3 1 1.5" ELECTRICAL CONTRACTOR
T1 - TP1 1 1" ELECTRICAL CONTRACTOR
T1 - V2 2 1.5" ELECTRICAL CONTRACTOR
V3 - Q1 1 1" ELECTRICAL CONTRACTOR
V3 - G1 1 .75" ELECTRICAL CONTRACTOR
V3 - G2 1 .75" ELECTRICAL CONTRACTOR
V3 - G3 1 .75" ELECTRICAL CONTRACTOR
V3 - G4 1 .75" ELECTRICAL CONTRACTOR
ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT
SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HV1000 INTEGRATION SYSTEM
PROPOSED OPERATING ROOM 8 LAYOUT
[2E-19]
NTS 02B
B3 C6C5
SURGICAL
LIGHT MOUNT
SURGICAL
LIGHT HEAD
(3)
SURGICAL
IN-LIGHT
CAMERA
TANDEM
UTILITY HUB
MOUNT
HD/4K
UNIVERSAL
INPUT
1.5"C.
(2)1.5"C.
(2)1.5"C.
PW
R
CK
T
#1
T1
HD
UNIVERSAL
INPUT
1.5"C.
(E1)
WALL
DISPLAY
1"C.
D1
PW
R
CK
T
#1
(E2)
WALL
DISPLAY
1"C.
(V2)
LEVITON
BOX
PW
R
CK
T
#2
(2)1.5"C.
PW
R
CK
T
#3NURSE
DESK
TOUCH PANEL
TP1
1"C.
EACH LEVITON BOX REQUIRES
DEDICATED EMERGENCY
BACK-UP 120V ON (2) 20 AMP
POWER CIRCUIT (NOT BY STERIS)
G1 G3
.75"C.
.75"C.
.75"C. STUB
WITHIN 18" OF
SPEAKER
G2 G4
.75"C.
.75"C.
.75"C. STUB
WITHIN 18" OF
SPEAKER
(V3)
LEVITON
BOX
PW
R
CK
T
#4
1"C.
NURSE
DESK
Q1
MUSIC
AUDIO
INPUT
JUNCTION BOX SCHEDULE
J-BOX # SIZE TYPE MOUNTING
T1 12"H X 12"W X 6"D NEMA 1, SCREW COVER ABOVE CEILING
JUNCTION BOX SCHEDULE NOTES:
1. ALL BOXES ARE PROVIDED BY THE CLIENTS CONTRACTOR UNLESS NOTED OTHERWISE
2. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS OTHERWISE NOTED/DETAILED
3. ALL BOXES AND COVER PLATES ARE PROVIDED BY THE CLIENTS CONTRACTOR UNLESS NOTED OTHERWISE
4. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS OTHERWISE NOTED/DETAILED
5. BOXES/OUTLETS/RECEPTACLES LOCATED ABOVE THE CEILING MUST BE WITHIN 18" OF THE CEILING AND BE ACCESSIBLE FROM THE SPACE BELOW; CLIENT/G.C. TO PROVIDE SUITABLE CEILING ACCESS PANELS AT DEVICE
LOCATIONS
6. T1 JUNCTION BOX - 12"H x 12"W x 6"D (MOUNTED VERTICALLY; NOT BY STERIS); 24" X 24" CEILING ACCESS PANEL REQUIRED FOR SERVICE ACCESS; ALL CONDUIT TO ENTER T1 BOX ANYWHERE EXCEPT FOR THE BACK SIDE;
T1 BOX SHOULD NOT BE HIGHER THAN 12' ABOVE FINISHED FLOOR
*NOT TO BE LOCATED DIRECTLY ABOVE NURSE DESKS OR ANY OTHER FIXED EQUIPMENT*
CONDUIT NOTES:
1. IF A DIRECT CONNECTION IS NEEDED BETWEEN TWO PIECES OF EQUIPMENT, CONDUIT MUST BE RUN
2. EXACT CONDUIT PATH DETERMINED BY ELECTRICAL CONTRACTOR
3. CONDUIT SHOULD NOT EXCEED 50'
4. MUST BE A PULL BOX FOR EVERY 270° OF CONDUIT BEND
5. ALL CONDUITS SHOULD BE PROVIDED WITH PULL STRINGS AND MUST HAVE INSULATED BUSHINGS ON ALL
OPEN ENDS
6. USE OF FLEXIBLE CONDUIT IS NOT ACCEPTABLE
7. STERIS WILL NOT BE RESPONSIBLE FOR FIRE STOPPING IN ANY FORM
STERIS IS NOT RESPONSIBLE FOR SPEC-ING POWER OR DATA REQUIREMENTS OF NON-STERIS
PROVIDED EQUIPMENT
ABBREVIATIONS:
A.F.F. = ABOVE FINISHED FLOOR
B.C.P. = BLANK COVER PLATE
C. = CONDUIT
C.P. = COVER PLATE
C.K.T. = CIRCUIT
E.C. = ELECTRICAL CONTRACTOR
N.S.E. = NURSE STATION EQUIPMENT
RISER LEGEND
SYMBOL DESCRIPTION MOUNTING/COMMENT
AP CEILING ACCESS PANEL FOR REQUIRED ACCESS TO BOXES
B1 EQUIPMENT LABEL DENOTED CEILING MOUNT
N1 CONDUIT OR CABLE RUN JUNCTION BOX REFERENCE
CONDUIT STUB OUT CONDUIT STUB OUT WITH BUSHING
G1 CEILING SPEAKER CEILING, FLUSH, RECEPTACLE HEIGHT
DUPLEX POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT
QUAD POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT
DT DATA/TELEPHONE BOX WALL, FLUSH, RECEPTACLE HEIGHT
TANDEM
ANESTHESIA
MOUNT
SURGICAL
DISPLAY
SURGICAL
DISPLAY
(E3)
OUTPUT TO
DAVINCI
1.5"C.
HD
UNIVERSAL
INPUT
INTEGRATION EQUIPMENT SCHEDULE
LEGEND DESCRIPTION RESPONSIBILITY
B3 SURGICAL LIGHT MOUNT STERIS
C5 TANDEM ANESTHESIA BOOM STERIS
C6 TANDEM UTILITY HUB MOUNT STERIS
D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION BOX
FOR HD UNIVERSAL INPUT ELECTRICAL CONTRACTOR
E1
WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR
E2
WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR
E3 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR OUTPUT TO DAVINCI ELECTRICAL CONTRACTOR
G1 CEILING MOUNTED SPEAKER STERIS
G3 CEILING MOUNTED SPEAKER STERIS
G2 CEILING MOUNTED SPEAKER STERIS
G4 CEILING MOUNTED SPEAKER STERIS
K1 SURGICAL LIGHT CONTROLLER STERIS
Q1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR
T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE CEILING ELECTRICAL CONTRACTOR
TP1
TOUCH PANEL STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG
MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR
V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER
DISTRIBUTION ELECTRICAL CONTRACTOR
V3 WALL MOUNTED LEVITON BOX FOR STERIS AUDIO
DISTRIBUTION ELECTRICAL CONTRACTOR
SHEET NOTES
1. TOUCH PANEL MUST BE LOCATED WITHIN 75' OF HV1000 BASE UNIT
2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS
3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE
RIGHT HALF OF THE BOX
4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE
AROUND ALL SIDES OF THE BOX
5. EMERGENCY POWER REQUIRED FOR LEVITON BOX
6. A 24" X 24" HINGED ACCESS PANEL IS REQUIRED AT ALL MOUNTS. COORDINATE LOCATION
& PARAMETERS OF ACCESS PANELS FOR MOUNTS AND ELECTRICAL BOXES WITH STERIS
REPRESENTATIVE
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HV1000 INTEGRATION SYSTEM
PROPOSED OPERATING ROOM
CONDUIT RISER DIAGRAM
NTS 03
E1,E2
TYPICAL WALL DISPLAY INSTALLATION
(OPTIONAL)
*REFERENCE SHEET NOTES #4,#5,#6
TYPICAL LEVITON BOX FOR STERIS
POWER DISTRIBUTION INSTALLATION
*REFERENCE SHEET NOTES #1,#2,#3
CUSTOMER
VERIFY
3"3"
" A
BO
VE
FIN
IS
HE
D
FL
O
O R
(6 8"
IF
U
SE
D
FO
R
X-
RA
Y
VI
EW
IN
G
ST
AT
IO
N
CU
ST
O M
ER
V
ER
IFY
A
LL
" A
BO
VE
FIN
IS
HE
D
FL
O
O R
4"
1 4"
POWER (SEE
SHEET NOTES
FOR DETAILS)
3"
V2
TP1
TYPICAL WALL MOUNTED
TOUCH PANEL INSTALLATION
(OPTIONAL)
" A
BO
VE
FIN
IS
HE
D
FL
O
O R V3
A SERIES SURGICAL LIGHT CONTROLLER
K1 13 1/
2"
16 1/2"2"
3/
4"
" M
IN
. A
BO
VE
F
IN
IS
HE
D
FL
O O
R
6"
2" 6"
SURGICAL LIGHT
CONTROL
ROUGH OPENING
9 1/2"
EQUIPMENT SCHEDULE
LEGEND DESCRIPTION RESPONSIBILITY
E1,E2 WALL MOUNTED DISPLAY STERIS
G1,G3,G2,G4 CEILING SPEAKER, 8.7" DIAMETER x 8.8" HEIGHT; 7.4 LB STERIS
K1 SURGICAL LIGHT CONTROLLER STERIS
TP1 WALL MOUNTED TOUCH PANEL STERIS
V2 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR
V3 LEVITON BOX FOR STERIS AUDIO DISTRIBUTION ELECTRICAL CONTRACTOR
SHEET NOTES
1. EACH LEVITON BOX REQUIRES DEDICATED EMERGENCY BACK-UP 120V ON (2)20 AMP POWER CIRCUITS FROM THE O.R. IT SERVES (NOT
BY STERIS)
2. LEVITON BOX (NOT BY STERIS) WITH VENTED DOOR FOR STERIS POWER & AUDIO DISTRIBUTION; BOX PRODUCT NUMBER: 47605-42N
(REQUIRES MINIMUM 2" SERVICE CLEARANCE); DOOR (NOT BY STERIS) PRODUCT NUMBER: 47605-42S; LEVITON AC DUAL POWER MODULE (NOT BY STERIS - BOLTED TO BOTTOM OF LEVITON BOX - LOCATED INSIDE OF WALL) PRODUCT NUMBER: 47605-DP
3. LEVITON BOX REQUIRES CONDUIT TO ENTER FROM THE TOP RIGHT OF THE BOX
4. WALL DISPLAY BACKING TO BE PROVIDED AND INSTALLED BY CONTRACTOR; WALL DISPLAY BACKING (NOT BY STERIS) MUST SUPPORT
A 250LBS LOAD; WALL BRACKET NOT INSTALLED BY STERIS
5. WALL DISPLAY REQUIRES 4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING
6. WALL DISPLAY REQUIRES DUPLEX OUTLET 120 VAC ON 20 AMP SERVICE (STANDARD OR EMERGENCY)
7. FINAL SPEAKER LOCATIONS TO BE DECIDED BY ARCHITECT/CONTRACTOR; CONTRACTOR TO STUB OUT ~ 18" ABOVE FINISHED CEILING
FOR SPEAKER CONNECTION; CONTRACTOR TO PROVIDE AND INSTALL SEISMIC SUPPORT WIRE FOR SPEAKER ANCHORING TO THE BUILDING STRUCTURE; CONTRACTOR TO CUT ROUND HOLE IN CEILING AT THE SPEAKER LOCATION ACCORDING TO THE SPEAKER INSTALLATION TEMPLATE REQUIRES FLEXIBLE CONDUIT WITH CONNECTOR ON SPEAKER END (NOT BY STERIS; TYPICAL AT EACH
SPEAKER LOCATION)
G1,G3,G2,G4
TYPICAL JBL 16C/T SPEAKER INSTALLATION (OPTIONAL)
*REFERENCE SHEET NOTE #7
CEILING
8.7"
9.
4"
8.
8"
8.
2"
"M
AX
FLEX
CONDUIT
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HV1000 INTEGRATION SYSTEM
PROPOSED EQUIPMENT
INSTALLATION DETAILS
NTS 04
HARMONY CUSTOM INTEGRATION SYSTEM INSTALLATION
GUIDELINES
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
STERIS
INDEX
PAGE # SHEET TITLE ROOM #
01 TITLE PAGE, INDEX -
01A INFRASTRUCTURE RESPONSIBILITIES TERMS & CONDITIONS -
02 PROPOSED OPERATING ROOM 4 LAYOUT 2E-07
02A PROPOSED OPERATING ROOM 5 LAYOUT 2E-08
02B PROPOSED OPERATING ROOM 6 LAYOUT 2E-09
02C PROPOSED OPERATING ROOM 9 LAYOUT 2E-20
02D PROPOSED OPERATING ROOM 10 LAYOUT 2E-19A
03 PROPOSED EQUIPMENT INSTALLATION DETAILS -
04 CEILING MOUNT INSTALLATION REQUIREMENTS -
GENERAL NOTES
1. SEE INFRASTRUCTURE RESPONSIBILITIES PAGE (1A) FOR ADDITIONAL DETAILS & DESCRIPTIONS FOR SPECIFIC STAKEHOLDER RESPONSIBILITIES
2. SUPPLY AND INSTALLATION OF ALL CONDUIT, POWER CIRCUITS, BACK BOXES, ELECTRICAL DEVICES, ELECTRICAL WIRING IS TO BE PROVIDED BY THE
CLIENTS ELECTRICAL CONTRACTOR UNLESS NOTED OTHERWISE
3. ONE WIRE HANGER FOR CABLE COILS REQUIRED FOR EACH MOUNT
4. CLIENT CONTRACTOR TO PROVIDE BLANK COVER PLATES FOR ALL JUNCTIONS BOXES WHERE NECESSARY
5. ALL CABLE RUNS OUTSIDE INTEGRATED ROOMS ARE THE RESPONSIBILITY OF HOSPITAL/CONTRACTOR; ALL CABLE RUNS SHOULD BE CONTINUOUS
EXCEPT WHERE NOTED
6. 24x24" ACCESS MUST BE PROVIDED WITHIN 36" OF ALL JUNCTION BOXES
7. CABLE RUNS THRU PLENUM SPACE MUST BE PLENUM RATED OR SHALL BE INSTALLED WITH CONDUIT
8. CABLES NOT RUN IN CONDUIT MUST HAVE J-HOOKS EVERY 4' FOR SUPPORT (NOT BY STERIS)
9. WHERE J-HOOKS ARE USED IN CEILING, CONDUIT IS REQUIRED IN ALL CONCEALED LOCATIONS (SIZED PER SCHEDULES)
10. EACH RACK REQUIRES (4) NETWORK PORTS PER O.R.
11. ALL POWER FOR INTEGRATION EQUIPMENT SHOULD COME FROM A SINGLE ELECTRICAL PANEL IN THE OR IT SERVES
12. CONFORMANCE TO NATIONAL, STATE AND LOCAL CODES FOR ELECTRICAL CIRCUITS AND CONDUIT IS THE RESPONSIBILITY OF THE HOSPITAL OR
THEIR CONTRACTED ELECTRICIAN
13. THE EQUIPMENT AND ROUGH-IN REQUIREMENTS SHOWN ARE FOR BUDGETARY PLANNING PURPOSES ONLY; EQUIPMENT IS SUBJECT TO CHANGE AND DIFFERENT OPTIONS ARE AVAILABLE TO MEET SPECIFIC CUSTOMER DESIGN CRITERIA; POST-ORDER DRAWINGS WILL REFLECT EXACT EQUIPMENT
SPECIFICATIONS, AND REQUIRES CUSTOMER SIGNED APPROVAL
14. USE OF ISOLATED GROUND POWER TO BE DETERMINED BY ELECTRICAL CONTRACTOR
ABBREVIATIONS:
A.F.F. = ABOVE FINISHED FLOOR
B.C.P. = BLANK COVER PLATE
C. = CONDUIT
C.P. = COVER PLATE
C.K.T. = CIRCUIT
E.C. = ELECTRICAL CONTRACTOR
N.S.E. = NURSE STATION EQUIPMENT
ALL STERIS TERMS AND CONDITIONS RELATIVE TO STERIS AND CUSTOMER,
DESIGNATED CONTRACTOR RESPONSIBILITIES, SITE READINESS, AND CHANGE
ORDERS APPLY. THESE WILL BE PROVIDED TO ALL PARTIES ABOVE AT THE
PROJECT KICKOFF MEETING BY YOUR STERIS REPRESENTATIVE
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HARMONY CUSTOM INTEGRATION SYSTEM
TITLE PAGE, INDEX
NTS 01
STERIS Responsibilities:
· Provide cable conduit drawings, electrical requirements, and mounting requirements for the room.
· Provide space, structural, and heat load requirements for the base unit.
· Provide site preparation specifications.
· Deliver all custom back boxes, mounting brackets, and construction-related items to the hospital in accordance with the mutually agreed upon construction or renovation schedule.
· Participate in project meetings in person, via conference call or via e-mail.
· Ship equipment in protective and labeled packaging.
· Ship cables and equipment on the scheduled ship date
· Pull and terminate non-plenum cabling inside the OR through specified conduits and junction boxes as specified in the Product
Description and Final Project Plan.
· Provide external cables and detailed specifications for the interconnections only if external cabling is specified in the Customer
Proposal and Final Project Plan.
· Provide a pre-installation checklist to verify site preparation is complete.
· Mount, install, and connect integration equipment - including video, audio, and control cables, cable terminations, wall cameras, wall monitors (does not include wall mounting brackets), base unit, touch screen control, and surgical monitors - as applicable to project.
· Test and validate integration system to complete customer satisfaction.
· Provide the hospital with a certificate of completion signed by the install team and the hospital.
· Adjust STERIS surgical light brakes as necessary after installation of STERIS monitors.
· In-service hospital staff at the time of the install. Further in-service training may be scheduled as mutually agreed.
· Leave a tidy work area each evening and remove packing material to a customer provided on-site container.
Customer or Customer's Designated Contractor Responsibilities:
· Ensure that site preparations and construction are complete and ready for final integration installation - including electrical and network connections, finished floor and paint.
· Provide complete room schematics including electrical, mechanical, and equipment location details.
· Install cable conduit, junction boxes, electrical outlets, and mounting brackets according to project specifications.
· Ensure that all electrical requirements comply with local code.
· Ensure space, structural, and heat load requirements are met according to project specs and plan.
· Receive shipment and safely, securely store equipment if shipment arrives prior to requested installation date.
· Provide adequate access from delivery/storage area to installation site.
· Relocate equipment from on-site storage/delivery area to installation site.
· Run all cables outside the OR- Including conduit, J-hooks, penetrations through firewall, video cables, radio antennae cables, or fiber optics to networking hubs and video conferencing hubs, communications closets, conference rooms, centralized nurse desks, rooftop, or other departments.
· Ensure that site preparations and construction are complete and ready for final integration installation - including electrical and network connections, finished floor and paint.
· Proactively schedule the installation in conjunction with the STERIS project manager.
· Provide a minimum of two (2) weeks advance notice for requested changes to the installation schedule.
· Remove old equipment or furniture from the room.
· Provide a clean, dry, dust free environment for equipment storage and installation.
· Provide a secure work site for storing ladders, toolboxes etc. during install.
· Provide easy access to all rooms involved in the installation during normal business hours and upon request after hours.
· Ensure that all non-STERIS or third-party equipment and systems that are to be part of the integrated solution are on-site and functional at the time of installation - otherwise full testing and connections cannot be completed.
· Participate in a final walk-through evaluation of the system followed by a sign-off of the certificate of completion.
· Provide an on-site container for trash, including shipping and packing material. Empty container as needed and on a regular basis.
Cabling and Fiber Responsibilities:
· All cables and fiber optic runs to locations outside the OR are the responsibility of the hospital.
Minimum Site Readiness Successful and efficient installation of a complete integration system requires that the site be carefully prepared according to the project documentation and schematics. A site that is not ready when the installation team arrives will cause delays and additional expenses to both parties. If the installation site is not ready per the mutually agreed specifications, STERIS reserves the right to charge a change order fee and extend the timeline needed to finish project past the planned date given at the project's inception.
To be ready a site should have -
1. All floor work must be complete
2. All painting must be complete
3. All electrical requirements must be installed and powered
4. Room must be clean and free from dust
Customer or Customer's Designated Contractor Rack Clearance & Ventilation Guidelines:
· Reference site specific rack layout document provided by STERIS for total rack BTU estimation information.
· Customer or Customer's Designated Contractor responsible for ensuring proper ventilation of STERIS Integration Equipment.
· Contact your STERIS representative for any questions regarding STERIS Integration Equipment ventilation and/or BTU ratings.
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HARMONY CUSTOM INTEGRATION SYSTEM
INFRASTRUCTURE RESPONSIBILITIES
TERMS & CONDITIONS
NTS 01A
O.R. #3 2E-06M1
V1
K1
C1
B1
C2
A SERIES LIGHT CONDUIT SCHEDULE - OR 5&6
FROM - TO QUANTITY SIZE RESPONSIBILITY
K1 - B1 2 .75" ELECTRICAL CONTRACTOR
LIGHT CONDUIT SCHEDULE NOTES
1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT
AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz
2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET
3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO
REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE
SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM
4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN
DOUBLE RUNS ARE REQUIRED (LONG RUNS)
5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE
THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS
LIGHT SYSTEM IS UTILIZED
*REFER TO TITLE PAGE FOR ADDITIONAL ELECTRICAL NOTES*
CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
M1 - HOSPITAL
NETWORK 1 1" ELECTRICAL CONTRACTOR
*INTEGRATION EQUIPMENT
LEGEND DESCRIPTION RESPONSIBILITY
B1 FUTURE SURGICAL LIGHT MOUNT STERIS
C1 FUTURE UTILITY HUB STERIS
C2 FUTURE ANESTHESIA BOOM STERIS
K1 FUTURE SURGICAL LIGHT CONTROLLER STERIS
M1
IP STATUS ROOM CAMERA/NURSE OBSERVATION SYSTEM STERIS
4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING
REQUIRED FOR IP CAMERA ELECTRICAL CONTRACTOR
ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT
SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HARMONY CUSTOM INTEGRATION SYSTEM
PROPOSED OPERATING ROOM 3 LAYOUT
[2E-06]
NTS 02
O.R. #4 2E-07
B1
C1
C2
K1
M1
V1 V1
A SERIES LIGHT CONDUIT SCHEDULE - OR 5&6
FROM - TO QUANTITY SIZE RESPONSIBILITY
K1 - B1 2 .75" ELECTRICAL CONTRACTOR
LIGHT CONDUIT SCHEDULE NOTES
1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT
AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz
2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET
3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO
REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE
SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM
4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN
DOUBLE RUNS ARE REQUIRED (LONG RUNS)
5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE
THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS
LIGHT SYSTEM IS UTILIZED
*REFER TO TITLE PAGE FOR ADDITIONAL ELECTRICAL NOTES*
CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
B1 - V1 1 1.25" ELECTRICAL CONTRACTOR
C1 - V1 1 1.25" ELECTRICAL CONTRACTOR
M1 - HOSPITAL
NETWORK 1 1" ELECTRICAL CONTRACTOR
*INTEGRATION EQUIPMENT
LEGEND DESCRIPTION RESPONSIBILITY
B1 SURGICAL LIGHT MOUNT STERIS
C1 UTILITY HUB STERIS
C2 ANESTHESIA BOOM STERIS
K1 SURGICAL LIGHT CONTROLLER STERIS
M1
IP STATUS ROOM CAMERA/NURSE OBSERVATION SYSTEM STERIS
4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING
REQUIRED FOR IP CAMERA ELECTRICAL CONTRACTOR
V1 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR
ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT
SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HARMONY CUSTOM INTEGRATION SYSTEM
PROPOSED OPERATING ROOM 4 LAYOUT
[2E-07]
NTS 02A
X-RAY&EQUP
2E-15
CONTROL RM
2E-10A
O.R. #5 2E-08
C2
K1
C1
M1
V1
B1 V2
V3
A SERIES LIGHT CONDUIT SCHEDULE - OR 5&6
FROM - TO QUANTITY SIZE RESPONSIBILITY
K1 - B1 2 .75" ELECTRICAL CONTRACTOR
LIGHT CONDUIT SCHEDULE NOTES
1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT
AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz
2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET
3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO
REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE
SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM
4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN
DOUBLE RUNS ARE REQUIRED (LONG RUNS)
5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE
THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS
LIGHT SYSTEM IS UTILIZED
*REFER TO TITLE PAGE FOR ADDITIONAL ELECTRICAL NOTES*
CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
B1 - V1 1 1.25" ELECTRICAL CONTRACTOR
C1 - V1 1 1.25" ELECTRICAL CONTRACTOR
M1 - HOSPITAL
NETWORK 1 1" ELECTRICAL CONTRACTOR
*INTEGRATION EQUIPMENT
LEGEND DESCRIPTION RESPONSIBILITY
B1 SURGICAL LIGHT MOUNT STERIS
C1 UTILITY HUB STERIS
C2 ANESTHESIA BOOM STERIS
K1 SURGICAL LIGHT CONTROLLER STERIS
M1
IP STATUS ROOM CAMERA/NURSE OBSERVATION SYSTEM STERIS
4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING
REQUIRED FOR IP CAMERA ELECTRICAL CONTRACTOR
V1 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR
ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT
SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HARMONY CUSTOM INTEGRATION SYSTEM
PROPOSED OPERATING ROOM 5 LAYOUT
[2E-08]
NTS 02B
VESTIBULE
2E-18B
STERILIZA.
2E-20A
STERILIZA.
2E-18A
F.A .C
STORAGE
2E-20B
O.R. #9 2E-20
K1
MODIFIED TO
V1
D1 B4
M1
*REFER TO TITLE PAGE FOR ADDITIONAL ELECTRICAL NOTES*
CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
B4 - V1 1 1.25" ELECTRICAL CONTRACTOR
D1 - V1 1 1.25" ELECTRICAL CONTRACTOR
M1 - HOSPITAL
NETWORK 1 1" ELECTRICAL CONTRACTOR
*INTEGRATION EQUIPMENT
LEGEND DESCRIPTION RESPONSIBILITY
B4 SURGICAL LIGHT MOUNT STERIS
D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION
BOX FOR HDSDI INPUT ELECTRICAL CONTRACTOR
K1 SURGICAL LIGHT CONTROLLER STERIS
M1
IP STATUS ROOM CAMERA/NURSE OBSERVATION SYSTEM STERIS
4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING
REQUIRED FOR IP CAMERA ELECTRICAL CONTRACTOR
V1 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR A SERIES LIGHT CONDUIT SCHEDULE - OR 9&10
FROM - TO QUANTITY SIZE RESPONSIBILITY
K1 - B4 2 .75" ELECTRICAL CONTRACTOR
LIGHT CONDUIT SCHEDULE NOTES
1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT
AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz
2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET
3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO
REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE
SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM
4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN
DOUBLE RUNS ARE REQUIRED (LONG RUNS)
5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE
THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS
LIGHT SYSTEM IS UTILIZED
ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT
SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE
REVISION SUMMARY
REVISION # DESCRIPTION DESIGNER/DATE CHECKER
5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION
SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873
DATE
SCALE
CKDR
CHECK PRINT
ST
ER
IS
d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.
VA BRONX MEDICAL CENTER
BRONX, NEW YORK
6313D5R5-INT
JTP 27JUL2023
PAGE
HARMONY CUSTOM INTEGRATION SYSTEM
PROPOSED OPERATING ROOM 9 LAYOUT
[2E-20]
NTS 02C
STERILIZA.
2E-20A
STERILIZA.
2E-18A
O.R. #10 2E-19A
STORAGE
2E-20B
C7 K1
V1 V1
D1
D1B4
M1
*REFER TO TITLE PAGE FOR ADDITIONAL ELECTRICAL NOTES*
CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
B4 - V1 1 1.25" ELECTRICAL CONTRACTOR
D1 - V1 1 1.25" ELECTRICAL CONTRACTOR
M1 - HOSPITAL
NETWORK 1 1" ELECTRICAL CONTRACTOR
*INTEGRATION EQUIPMENT
LEGEND DESCRIPTION RESPONSIBILITY
B4 SURGICAL LIGHT MOUNT STERIS
C7 ANESTHESIA BOOM STERIS
D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION
BOX FOR HDSDI INPUT ELECTRICAL CONTRACTOR
K1 SURGICAL LIGHT CONTROLLER STERIS
M1
IP STATUS ROOM CAMERA/NURSE OBSERVATION SYSTEM STERIS
4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING
REQUIRED FOR IP CAMERA ELECTRICAL CONTRACTOR
V1 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR
A SERIES LIGHT CONDUIT SCHEDULE - OR 9&10
FROM - TO QUANTITY SIZE RESPONSIBILITY
K1 - B4 2 .75" ELECTRICAL CONTRACTOR
LIGHT CONDUIT SCHEDULE NOTES
1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT
AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz
2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET
3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO
REFERENCED WITHIN THE…