6313D5R5-INT CHECKPRINT.pdf

PDF 2 MB Posted

Attached to
Y1AZ--NRM_CONST_526-21-521 REPLACE SURGICAL LIGHTS Federal contract opportunity
Solicitation number
Not on record
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 2

About this file

This document is a detailed STERIS Harmony Custom Integration System installation guideline for Operating Rooms (OR) #3-#10 at the VA Bronx Medical Center in New York. The comprehensive set of technical drawings and specifications outlines the installation requirements for an advanced surgical room integration system, including precise mounting instructions for surgical lights, utility hubs, booms, displays, speakers, and control systems. The documentation provides explicit responsibilities for both STERIS and the customer's electrical contractor, covering conduit specifications, electrical requirements, equipment placement, and site preparation guidelines.

The integration system involves installing sophisticated medical technology across multiple operating rooms, with specific technical details for each room's layout. Key components include HarmonyAIR utility hubs, A-Series surgical lights, wall-mounted displays (LG 55" 4K IPS LED Commercial Grade), integrated wall controls, and IP status room cameras. The documents emphasize strict installation protocols, including conduit specifications, power requirements, junction box placements, and specific mounting heights and clearances. The most recent revision (Revision 5) was made on 16APR2024, updating the system to current standards and changing OR6-8 to the HV1000 system.

View the file

Other files for this federal contract opportunity

Other files attached to Y1AZ--NRM_CONST_526-21-521 REPLACE SURGICAL LIGHTS, newest first.
File Type Posted
36C24225B0012 0001.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

00 COVER

01 PROPOSED EQUIPMENT LAYOUT

DRAWING INDEX

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

SIGN OFF ACCEPTANCES

__________________________________ (OPTIONAL)

__________________________________ (OPTIONAL)

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

DATE

SCALE

CKDR

CHECK PRINT

PAGE

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

LJD 22SEP2005

COVER SHEET

NOT TO SCALE 00

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

VESTIBULE

2E-18B

STERILIZA.

2E-20A

STERILIZA.

2E-18A

O.R. #10 2E-19A

F.A .C

STORAGE

2E-20B

O.R. #8 2E-19

O.R. #9 2E-20

O.R.#7CARD

CARD.EQUP.

2E-18

BLOOD RM.

2E-01E

2E-17

X-RAY&EQUP

2E-15

SPLT.&PLS.

2E-14

O.R. #2 2E-05

O.R. #3 2E-06

CONTROL RM

2E-10A

O.R. #4 2E-07

O.R. #5 2E-08

ORH.O.R.#6

2E-09

01E 01E 01E

01E

01E01E

63"

60" 48" 53" 46" 60" 48"

60" 48"411 2"

60" 48" 761 8"

53 8"

48" 60" 641 2"

5 8"

48" 60" 671 2"

5 8"C7

B1

C1

B3

C5

C6

C2 C2

C5

B3 C6

01E 01E

4"

1 8"

3 8"

48" 843 4"

1 2"

K1

K1

K1

K1 K1

K1K1

MOUNT C2 LIMITED

UPPER ARM

ROTATION AT WALL

CEILING LIGHTS MODIFIED TO

ACCOMMODATE STERIS

EQUIPMENT

C1

E1

E1

E2E2

K1F

C1F

B1F

C2F

B4B4

B1

E2

E1 C6

C5

B3

EQUIPMENT LIST

TAG QTY DESCRIPTION

01E 8 Existing Amsco 3085 SP Surgical Table

B1 2 A Series-Surgical Light(s)

B1F 1 Future A Series-Surgical Light(s)

B3 3 A Series-Surgical Light(s)

B4 2 A Series-Surgical Light(s)

C1 2 HarmonyAIR Utility Hub-Surgical

C1F 1 Future HarmonyAIR Utility Hub-Surgical

C2 2 HarmonyAIR Adj Arm Spring-Anesthesia

C2F 1 Future HarmonyAIR Adj Arm Spring-Anesthesia

C5 3 Tandem HarmonyAIR Adj Arm Spring - A Series-Anesthesia, Accessory Arm(s)

C6 3 Tandem HarmonyAIR Utility Hub - A Series-Power Boom Support, Accessory Arm(s)

C7 1 HarmonyAIR-Straight-Drop-Anesthesia

E1 3 LG 55" 4K IPS LED Commercial Grade Display - Wall Mounted

E2 3 LG 55" 4K IPS LED Commercial Grade Display - Wall Mounted

K1 7 HarmonyAIR A-Series Integrated Wall Control

K1F 1 Future HarmonyAIR A-Series Integrated Wall Control

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

DISCLAIMER: THE SERVICE CLEARANCES INDICATED DO NOT MEET THE MINIMUM SAFETY CLEARANCES RECOMMENDED IN THE EQUIPMENT DRAWINGS PROVIDED BY STERIS CORPORATION DISCLAIMER: THE WALLS AND OR EQUIPMENT SHOWN IN THIS DRAWING HAVE BEEN MOVED/MODIFIED TO ACCOMMODATE MINIMUM RECOMMENDED CLEARANCES FOR STERIS EQUIPMENT

DATE

SCALE

CKDR

CHECK PRINT

PAGE

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

LJD 22SEP2005

OPERATING ROOMS #3 - #10

PROPOSED EQUIPMENT LAYOUT

1/4" = 1'-0" 01

HV1000 INSTALLATION GUIDELINES

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

STERIS

GENERAL NOTES

1. SEE INFRASTRUCTURE RESPONSIBILITIES PAGE (1A) FOR ADDITIONAL DETAILS & DESCRIPTIONS FOR SPECIFIC STAKEHOLDER RESPONSIBILITIES

2. SUPPLY AND INSTALLATION OF ALL CONDUIT, POWER CIRCUITS, BACK BOXES, ELECTRICAL DEVICES, ELECTRICAL WIRING IS TO BE PROVIDED BY THE

CLIENTS ELECTRICAL CONTRACTOR UNLESS NOTED OTHERWISE

3. ONE WIRE HANGER FOR CABLE COILS REQUIRED FOR EACH MOUNT

4. CLIENT CONTRACTOR TO PROVIDE BLANK COVER PLATES FOR ALL JUNCTIONS BOXES WHERE NECESSARY

5. ALL CABLE RUNS OUTSIDE INTEGRATED ROOMS ARE THE RESPONSIBILITY OF HOSPITAL/CONTRACTOR; ALL CABLE RUNS SHOULD BE CONTINUOUS

EXCEPT WHERE NOTED

6. 24x24" ACCESS MUST BE PROVIDED WITHIN 36" OF ALL JUNCTION BOXES

7. CABLE RUNS THRU PLENUM SPACE MUST BE PLENUM RATED OR SHALL BE INSTALLED WITH CONDUIT

8. CABLES NOT RUN IN CONDUIT MUST HAVE J-HOOKS EVERY 4' FOR SUPPORT (NOT BY STERIS)

9. WHERE J-HOOKS ARE USED IN CEILING, CONDUIT IS REQUIRED IN ALL CONCEALED LOCATIONS (SIZED PER SCHEDULES)

10. EACH RACK REQUIRES (4) NETWORK PORTS PER O.R.

11. ALL POWER FOR INTEGRATION EQUIPMENT SHOULD COME FROM A SINGLE ELECTRICAL PANEL IN THE OR IT SERVES

12. CONFORMANCE TO NATIONAL, STATE AND LOCAL CODES FOR ELECTRICAL CIRCUITS AND CONDUIT IS THE RESPONSIBILITY OF THE HOSPITAL OR

THEIR CONTRACTED ELECTRICIAN

13. THE EQUIPMENT AND ROUGH-IN REQUIREMENTS SHOWN ARE FOR BUDGETARY PLANNING PURPOSES ONLY; EQUIPMENT IS SUBJECT TO CHANGE AND DIFFERENT OPTIONS ARE AVAILABLE TO MEET SPECIFIC CUSTOMER DESIGN CRITERIA; POST-ORDER DRAWINGS WILL REFLECT EXACT EQUIPMENT

SPECIFICATIONS, AND REQUIRES CUSTOMER SIGNED APPROVAL

14. USE OF ISOLATED GROUND POWER TO BE DETERMINED BY ELECTRICAL CONTRACTOR

ALL STERIS TERMS AND CONDITIONS RELATIVE TO STERIS AND CUSTOMER,

DESIGNATED CONTRACTOR RESPONSIBILITIES, SITE READINESS, AND CHANGE

ORDERS APPLY. THESE WILL BE PROVIDED TO ALL PARTIES ABOVE AT THE

PROJECT KICKOFF MEETING BY YOUR STERIS REPRESENTATIVEABBREVIATIONS:

A.F.F. = ABOVE FINISHED FLOOR

B.C.P. = BLANK COVER PLATE

C. = CONDUIT

C.P. = COVER PLATE

C.K.T. = CIRCUIT

E.C. = ELECTRICAL CONTRACTOR

N.S.E. = NURSE STATION EQUIPMENT

INDEX

PAGE # SHEET TITLE ROOM #

01 TITLE PAGE, INDEX -

01A INFRASTRUCTURE RESPONSIBILITIES TERMS & CONDITIONS -

02 PROPOSED OPERATING ROOM #6 LAYOUT 2E-09

02A PROPOSED OPERATING ROOM #7 LAYOUT 2E-17

02B PROPOSED OPERATING ROOM #8 LAYOUT 2E-19

03 PROPOSED OPERATING ROOM CONDUIT RISER DIAGRAM #6-8 -

04 PROPOSED EQUIPMENT INSTALLATION DETAILS -

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HV1000 INTEGRATION SYSTEM

TITLE PAGE, INDEX

NTS 01

STERIS Responsibilities:

· Provide cable conduit drawings, electrical requirements, and mounting requirements for the room.

· Provide space, structural, and heat load requirements for the base unit.

· Provide site preparation specifications.

· Deliver all custom back boxes, mounting brackets, and construction-related items to the hospital in accordance with the mutually agreed upon construction or renovation schedule.

· Participate in project meetings in person, via conference call or via e-mail.

· Ship equipment in protective and labeled packaging.

· Ship cables and equipment on the scheduled ship date

· Pull and terminate non-plenum cabling inside the OR through specified conduits and junction boxes as specified in the Product

Description and Final Project Plan.

· Provide external cables and detailed specifications for the interconnections only if external cabling is specified in the Customer

Proposal and Final Project Plan.

· Provide a pre-installation checklist to verify site preparation is complete.

· Mount, install, and connect integration equipment - including video, audio, and control cables, cable terminations, wall cameras, wall monitors (does not include wall mounting brackets), base unit, touch screen control, and surgical monitors - as applicable to project.

· Test and validate integration system to complete customer satisfaction.

· Provide the hospital with a certificate of completion signed by the install team and the hospital.

· Adjust STERIS surgical light brakes as necessary after installation of STERIS monitors.

· In-service hospital staff at the time of the install. Further in-service training may be scheduled as mutually agreed.

· Leave a tidy work area each evening and remove packing material to a customer provided on-site container.

Customer or Customer's Designated Contractor Responsibilities:

· Ensure that site preparations and construction are complete and ready for final integration installation - including electrical and network connections, finished floor and paint.

· Provide complete room schematics including electrical, mechanical, and equipment location details.

· Install cable conduit, junction boxes, electrical outlets, and mounting brackets according to project specifications.

· Ensure that all electrical requirements comply with local code.

· Ensure space, structural, and heat load requirements are met according to project specs and plan.

· Receive shipment and safely, securely store equipment if shipment arrives prior to requested installation date.

· Provide adequate access from delivery/storage area to installation site.

· Relocate equipment from on-site storage/delivery area to installation site.

· Run all cables outside the OR- Including conduit, J-hooks, penetrations through firewall, video cables, radio antennae cables, or fiber optics to networking hubs and video conferencing hubs, communications closets, conference rooms, centralized nurse desks, rooftop, or other departments.

· Ensure that site preparations and construction are complete and ready for final integration installation - including electrical and network connections, finished floor and paint.

· Proactively schedule the installation in conjunction with the STERIS project manager.

· Provide a minimum of two (2) weeks advance notice for requested changes to the installation schedule.

· Remove old equipment or furniture from the room.

· Provide a clean, dry, dust free environment for equipment storage and installation.

· Provide a secure work site for storing ladders, toolboxes etc. during install.

· Provide easy access to all rooms involved in the installation during normal business hours and upon request after hours.

· Ensure that all non-STERIS or third-party equipment and systems that are to be part of the integrated solution are on-site and functional at the time of installation - otherwise full testing and connections cannot be completed.

· Participate in a final walk-through evaluation of the system followed by a sign-off of the certificate of completion.

· Provide an on-site container for trash, including shipping and packing material. Empty container as needed and on a regular basis.

Cabling and Fiber Responsibilities:

· All cables and fiber optic runs to locations outside the OR are the responsibility of the hospital.

Minimum Site Readiness Successful and efficient installation of a complete integration system requires that the site be carefully prepared according to the project documentation and schematics. A site that is not ready when the installation team arrives will cause delays and additional expenses to both parties. If the installation site is not ready per the mutually agreed specifications, STERIS reserves the right to charge a change order fee and extend the timeline needed to finish project past the planned date given at the project's inception.

To be ready a site should have -

1. All floor work must be complete

2. All painting must be complete

3. All electrical requirements must be installed and powered

4. Room must be clean and free from dust

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HV1000 INTEGRATION SYSTEM

INFRASTRUCTURE RESPONSIBILITIES

TERMS & CONDITIONS

NTS 01A

CONTROL RM

2E-10A ORH.O.R.#6

2E-09

K1

E2

SPEAKER PLACEMENT (G1-G4) FOR

REFERENCE ONLY; SITE VERIFY

G2G3

E1

Q1

G4

G1 D1 E3

C6

C5

B3

V2

T1

V3

TP1

INTEGRATION EQUIPMENT SCHEDULE

LEGEND DESCRIPTION RESPONSIBILITY

B3 SURGICAL LIGHT MOUNT STERIS

C5 TANDEM ANESTHESIA BOOM STERIS

C6 TANDEM UTILITY HUB MOUNT STERIS

D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION BOX

FOR HD UNIVERSAL INPUT ELECTRICAL CONTRACTOR

E1

WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR

E2

WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR

E3 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR OUTPUT TO DAVINCI ELECTRICAL CONTRACTOR

G1 CEILING MOUNTED SPEAKER STERIS

G3 CEILING MOUNTED SPEAKER STERIS

G2 CEILING MOUNTED SPEAKER STERIS

G4 CEILING MOUNTED SPEAKER STERIS

K1 SURGICAL LIGHT CONTROLLER STERIS

Q1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR

T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE CEILING ELECTRICAL CONTRACTOR

TP1

TOUCH PANEL STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR

V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER

DISTRIBUTION ELECTRICAL CONTRACTOR

V3 WALL MOUNTED LEVITON BOX FOR STERIS AUDIO

DISTRIBUTION ELECTRICAL CONTRACTOR

A SERIES LIGHT CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

K1 - B3 2 .75" ELECTRICAL CONTRACTOR

LIGHT CONDUIT SCHEDULE NOTES

1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT

AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz

2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET

3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO

REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE

SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM

4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN

DOUBLE RUNS ARE REQUIRED (LONG RUNS)

5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE

THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS

LIGHT SYSTEM IS UTILIZED

SHEET NOTES

1. TOUCH PANEL MUST BE LOCATED WITHIN 75' OF HV1000 BASE UNIT

2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS

3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE

RIGHT HALF OF THE BOX

4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE

AROUND ALL SIDES OF THE BOX

5. EMERGENCY POWER REQUIRED FOR LEVITON BOX

6. A 24" X 24" HINGED ACCESS PANEL IS REQUIRED AT ALL MOUNTS. COORDINATE LOCATION

& PARAMETERS OF ACCESS PANELS FOR MOUNTS AND ELECTRICAL BOXES WITH STERIS

REPRESENTATIVE

INTEGRATION CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

T1 - B3 1 1.5" ELECTRICAL CONTRACTOR

T1 - C5 2 1.5" ELECTRICAL CONTRACTOR

T1 - C6 2 1.5" ELECTRICAL CONTRACTOR

T1 - D1 1 1.5" ELECTRICAL CONTRACTOR

T1 - E1 1 1" ELECTRICAL CONTRACTOR

T1 - E2 1 1" ELECTRICAL CONTRACTOR

T1 - E3 1 1.5" ELECTRICAL CONTRACTOR

T1 - TP1 1 1" ELECTRICAL CONTRACTOR

T1 - V2 2 1.5" ELECTRICAL CONTRACTOR

V3 - Q1 1 1" ELECTRICAL CONTRACTOR

V3 - G1 1 .75" ELECTRICAL CONTRACTOR

V3 - G2 1 .75" ELECTRICAL CONTRACTOR

V3 - G3 1 .75" ELECTRICAL CONTRACTOR

V3 - G4 1 .75" ELECTRICAL CONTRACTOR

ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT

SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HV1000 INTEGRATION SYSTEM

PROPOSED OPERATING ROOM 6 LAYOUT

[2E-09]

NTS 02

O.R.#7CARD

CARD.EQUP.

2E-18

BLOOD RM.

2E-01E

2E-17

C5

B3 C6

K1 G1 G4

SPEAKER PLACEMENT (G1-G4) FOR

REFERENCE ONLY; SITE VERIFY

G2 G3

D1

E3

E1

E2

SERVES OR 7

Q1

TP1

V2

T1

V3

INTEGRATION EQUIPMENT SCHEDULE

LEGEND DESCRIPTION RESPONSIBILITY

B3 SURGICAL LIGHT MOUNT STERIS

C5 TANDEM ANESTHESIA BOOM STERIS

C6 TANDEM UTILITY HUB MOUNT STERIS

D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION BOX

FOR HD UNIVERSAL INPUT ELECTRICAL CONTRACTOR

E1

WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR

E2

WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR

E3 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR OUTPUT TO DAVINCI ELECTRICAL CONTRACTOR

G1 CEILING MOUNTED SPEAKER STERIS

G3 CEILING MOUNTED SPEAKER STERIS

G2 CEILING MOUNTED SPEAKER STERIS

G4 CEILING MOUNTED SPEAKER STERIS

K1 SURGICAL LIGHT CONTROLLER STERIS

Q1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR

T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE CEILING ELECTRICAL CONTRACTOR

TP1

TOUCH PANEL STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR

V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER

DISTRIBUTION ELECTRICAL CONTRACTOR

V3 WALL MOUNTED LEVITON BOX FOR STERIS AUDIO

DISTRIBUTION ELECTRICAL CONTRACTOR

A SERIES LIGHT CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

K1 - B3 2 .75" ELECTRICAL CONTRACTOR

LIGHT CONDUIT SCHEDULE NOTES

1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT

AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz

2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET

3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO

REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE

SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM

4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN

DOUBLE RUNS ARE REQUIRED (LONG RUNS)

5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE

THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS

LIGHT SYSTEM IS UTILIZED

SHEET NOTES

1. TOUCH PANEL MUST BE LOCATED WITHIN 75' OF HV1000 BASE UNIT

2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS

3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE

RIGHT HALF OF THE BOX

4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE

AROUND ALL SIDES OF THE BOX

5. EMERGENCY POWER REQUIRED FOR LEVITON BOX

6. A 24" X 24" HINGED ACCESS PANEL IS REQUIRED AT ALL MOUNTS. COORDINATE LOCATION

& PARAMETERS OF ACCESS PANELS FOR MOUNTS AND ELECTRICAL BOXES WITH STERIS

REPRESENTATIVE

INTEGRATION CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

T1 - B3 1 1.5" ELECTRICAL CONTRACTOR

T1 - C5 2 1.5" ELECTRICAL CONTRACTOR

T1 - C6 2 1.5" ELECTRICAL CONTRACTOR

T1 - D1 1 1.5" ELECTRICAL CONTRACTOR

T1 - E1 1 1" ELECTRICAL CONTRACTOR

T1 - E2 1 1" ELECTRICAL CONTRACTOR

T1 - E3 1 1.5" ELECTRICAL CONTRACTOR

T1 - TP1 1 1" ELECTRICAL CONTRACTOR

T1 - V2 2 1.5" ELECTRICAL CONTRACTOR

V3 - Q1 1 1" ELECTRICAL CONTRACTOR

V3 - G1 1 .75" ELECTRICAL CONTRACTOR

V3 - G2 1 .75" ELECTRICAL CONTRACTOR

V3 - G3 1 .75" ELECTRICAL CONTRACTOR

V3 - G4 1 .75" ELECTRICAL CONTRACTOR

ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT

SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HV1000 INTEGRATION SYSTEM

PROPOSED OPERATING ROOM 7 LAYOUT

[2E-17]

NTS 02A

O.R. #8 2E-19

CARD.EQUP.

2E-18

BLOOD RM.

2E-01E

B3

C5

C6

K1

G1 G4

SPEAKER PLACEMENT (G1-G4) FOR

REFERENCE ONLY; SITE VERIFY

G2 G3

D1

E3

E1

E2

N3

SERVES

OR 8

Q1

TP1

V2

T1

V3

INTEGRATION EQUIPMENT SCHEDULE

LEGEND DESCRIPTION RESPONSIBILITY

B3 SURGICAL LIGHT MOUNT STERIS

C5 TANDEM ANESTHESIA BOOM STERIS

C6 TANDEM UTILITY HUB MOUNT STERIS

D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION BOX

FOR HD UNIVERSAL INPUT ELECTRICAL CONTRACTOR

E1

WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR

E2

WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR

E3 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR OUTPUT TO DAVINCI ELECTRICAL CONTRACTOR

G1 CEILING MOUNTED SPEAKER STERIS

G3 CEILING MOUNTED SPEAKER STERIS

G2 CEILING MOUNTED SPEAKER STERIS

G4 CEILING MOUNTED SPEAKER STERIS

K1 SURGICAL LIGHT CONTROLLER STERIS

Q1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR

T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE CEILING ELECTRICAL CONTRACTOR

TP1

TOUCH PANEL STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR

V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER

DISTRIBUTION ELECTRICAL CONTRACTOR

V3 WALL MOUNTED LEVITON BOX FOR STERIS AUDIO

DISTRIBUTION ELECTRICAL CONTRACTOR

A SERIES LIGHT CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

K1 - B3 2 .75" ELECTRICAL CONTRACTOR

LIGHT CONDUIT SCHEDULE NOTES

1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT

AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz

2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET

3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO

REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE

SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM

4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN

DOUBLE RUNS ARE REQUIRED (LONG RUNS)

5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE

THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS

LIGHT SYSTEM IS UTILIZED

SHEET NOTES

1. TOUCH PANEL MUST BE LOCATED WITHIN 75' OF HV1000 BASE UNIT

2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS

3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE

RIGHT HALF OF THE BOX

4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE

AROUND ALL SIDES OF THE BOX

5. EMERGENCY POWER REQUIRED FOR LEVITON BOX

6. A 24" X 24" HINGED ACCESS PANEL IS REQUIRED AT ALL MOUNTS. COORDINATE LOCATION

& PARAMETERS OF ACCESS PANELS FOR MOUNTS AND ELECTRICAL BOXES WITH STERIS

REPRESENTATIVE

INTEGRATION CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

T1 - B3 1 1.5" ELECTRICAL CONTRACTOR

T1 - C5 2 1.5" ELECTRICAL CONTRACTOR

T1 - C6 2 1.5" ELECTRICAL CONTRACTOR

T1 - D1 1 1.5" ELECTRICAL CONTRACTOR

T1 - E1 1 1" ELECTRICAL CONTRACTOR

T1 - E2 1 1" ELECTRICAL CONTRACTOR

T1 - E3 1 1.5" ELECTRICAL CONTRACTOR

T1 - TP1 1 1" ELECTRICAL CONTRACTOR

T1 - V2 2 1.5" ELECTRICAL CONTRACTOR

V3 - Q1 1 1" ELECTRICAL CONTRACTOR

V3 - G1 1 .75" ELECTRICAL CONTRACTOR

V3 - G2 1 .75" ELECTRICAL CONTRACTOR

V3 - G3 1 .75" ELECTRICAL CONTRACTOR

V3 - G4 1 .75" ELECTRICAL CONTRACTOR

ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT

SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HV1000 INTEGRATION SYSTEM

PROPOSED OPERATING ROOM 8 LAYOUT

[2E-19]

NTS 02B

B3 C6C5

SURGICAL

LIGHT MOUNT

SURGICAL

LIGHT HEAD

(3)

SURGICAL

IN-LIGHT

CAMERA

TANDEM

UTILITY HUB

MOUNT

HD/4K

UNIVERSAL

INPUT

1.5"C.

(2)1.5"C.

(2)1.5"C.

PW

R

CK

T

#1

T1

HD

UNIVERSAL

INPUT

1.5"C.

(E1)

WALL

DISPLAY

1"C.

D1

PW

R

CK

T

#1

(E2)

WALL

DISPLAY

1"C.

(V2)

LEVITON

BOX

PW

R

CK

T

#2

(2)1.5"C.

PW

R

CK

T

#3NURSE

DESK

TOUCH PANEL

TP1

1"C.

EACH LEVITON BOX REQUIRES

DEDICATED EMERGENCY

BACK-UP 120V ON (2) 20 AMP

POWER CIRCUIT (NOT BY STERIS)

G1 G3

.75"C.

.75"C.

.75"C. STUB

WITHIN 18" OF

SPEAKER

G2 G4

.75"C.

.75"C.

.75"C. STUB

WITHIN 18" OF

SPEAKER

(V3)

LEVITON

BOX

PW

R

CK

T

#4

1"C.

NURSE

DESK

Q1

MUSIC

AUDIO

INPUT

JUNCTION BOX SCHEDULE

J-BOX # SIZE TYPE MOUNTING

T1 12"H X 12"W X 6"D NEMA 1, SCREW COVER ABOVE CEILING

JUNCTION BOX SCHEDULE NOTES:

1. ALL BOXES ARE PROVIDED BY THE CLIENTS CONTRACTOR UNLESS NOTED OTHERWISE

2. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS OTHERWISE NOTED/DETAILED

3. ALL BOXES AND COVER PLATES ARE PROVIDED BY THE CLIENTS CONTRACTOR UNLESS NOTED OTHERWISE

4. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS OTHERWISE NOTED/DETAILED

5. BOXES/OUTLETS/RECEPTACLES LOCATED ABOVE THE CEILING MUST BE WITHIN 18" OF THE CEILING AND BE ACCESSIBLE FROM THE SPACE BELOW; CLIENT/G.C. TO PROVIDE SUITABLE CEILING ACCESS PANELS AT DEVICE

LOCATIONS

6. T1 JUNCTION BOX - 12"H x 12"W x 6"D (MOUNTED VERTICALLY; NOT BY STERIS); 24" X 24" CEILING ACCESS PANEL REQUIRED FOR SERVICE ACCESS; ALL CONDUIT TO ENTER T1 BOX ANYWHERE EXCEPT FOR THE BACK SIDE;

T1 BOX SHOULD NOT BE HIGHER THAN 12' ABOVE FINISHED FLOOR

*NOT TO BE LOCATED DIRECTLY ABOVE NURSE DESKS OR ANY OTHER FIXED EQUIPMENT*

CONDUIT NOTES:

1. IF A DIRECT CONNECTION IS NEEDED BETWEEN TWO PIECES OF EQUIPMENT, CONDUIT MUST BE RUN

2. EXACT CONDUIT PATH DETERMINED BY ELECTRICAL CONTRACTOR

3. CONDUIT SHOULD NOT EXCEED 50'

4. MUST BE A PULL BOX FOR EVERY 270° OF CONDUIT BEND

5. ALL CONDUITS SHOULD BE PROVIDED WITH PULL STRINGS AND MUST HAVE INSULATED BUSHINGS ON ALL

OPEN ENDS

6. USE OF FLEXIBLE CONDUIT IS NOT ACCEPTABLE

7. STERIS WILL NOT BE RESPONSIBLE FOR FIRE STOPPING IN ANY FORM

STERIS IS NOT RESPONSIBLE FOR SPEC-ING POWER OR DATA REQUIREMENTS OF NON-STERIS

PROVIDED EQUIPMENT

ABBREVIATIONS:

A.F.F. = ABOVE FINISHED FLOOR

B.C.P. = BLANK COVER PLATE

C. = CONDUIT

C.P. = COVER PLATE

C.K.T. = CIRCUIT

E.C. = ELECTRICAL CONTRACTOR

N.S.E. = NURSE STATION EQUIPMENT

RISER LEGEND

SYMBOL DESCRIPTION MOUNTING/COMMENT

AP CEILING ACCESS PANEL FOR REQUIRED ACCESS TO BOXES

B1 EQUIPMENT LABEL DENOTED CEILING MOUNT

N1 CONDUIT OR CABLE RUN JUNCTION BOX REFERENCE

CONDUIT STUB OUT CONDUIT STUB OUT WITH BUSHING

G1 CEILING SPEAKER CEILING, FLUSH, RECEPTACLE HEIGHT

DUPLEX POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT

QUAD POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT

DT DATA/TELEPHONE BOX WALL, FLUSH, RECEPTACLE HEIGHT

TANDEM

ANESTHESIA

MOUNT

SURGICAL

DISPLAY

SURGICAL

DISPLAY

(E3)

OUTPUT TO

DAVINCI

1.5"C.

HD

UNIVERSAL

INPUT

INTEGRATION EQUIPMENT SCHEDULE

LEGEND DESCRIPTION RESPONSIBILITY

B3 SURGICAL LIGHT MOUNT STERIS

C5 TANDEM ANESTHESIA BOOM STERIS

C6 TANDEM UTILITY HUB MOUNT STERIS

D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION BOX

FOR HD UNIVERSAL INPUT ELECTRICAL CONTRACTOR

E1

WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR

E2

WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR WALL MOUNTED DISPLAY ELECTRICAL CONTRACTOR

E3 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR OUTPUT TO DAVINCI ELECTRICAL CONTRACTOR

G1 CEILING MOUNTED SPEAKER STERIS

G3 CEILING MOUNTED SPEAKER STERIS

G2 CEILING MOUNTED SPEAKER STERIS

G4 CEILING MOUNTED SPEAKER STERIS

K1 SURGICAL LIGHT CONTROLLER STERIS

Q1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR

T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE CEILING ELECTRICAL CONTRACTOR

TP1

TOUCH PANEL STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH SINGLE GANG

MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR

V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER

DISTRIBUTION ELECTRICAL CONTRACTOR

V3 WALL MOUNTED LEVITON BOX FOR STERIS AUDIO

DISTRIBUTION ELECTRICAL CONTRACTOR

SHEET NOTES

1. TOUCH PANEL MUST BE LOCATED WITHIN 75' OF HV1000 BASE UNIT

2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS

3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE

RIGHT HALF OF THE BOX

4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE

AROUND ALL SIDES OF THE BOX

5. EMERGENCY POWER REQUIRED FOR LEVITON BOX

6. A 24" X 24" HINGED ACCESS PANEL IS REQUIRED AT ALL MOUNTS. COORDINATE LOCATION

& PARAMETERS OF ACCESS PANELS FOR MOUNTS AND ELECTRICAL BOXES WITH STERIS

REPRESENTATIVE

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HV1000 INTEGRATION SYSTEM

PROPOSED OPERATING ROOM

CONDUIT RISER DIAGRAM

NTS 03

E1,E2

TYPICAL WALL DISPLAY INSTALLATION

(OPTIONAL)

*REFERENCE SHEET NOTES #4,#5,#6

TYPICAL LEVITON BOX FOR STERIS

POWER DISTRIBUTION INSTALLATION

*REFERENCE SHEET NOTES #1,#2,#3

CUSTOMER

VERIFY

3"3"

" A

BO

VE

FIN

IS

HE

D

FL

O

O R

(6 8"

IF

U

SE

D

FO

R

X-

RA

Y

VI

EW

IN

G

ST

AT

IO

N

CU

ST

O M

ER

V

ER

IFY

A

LL

" A

BO

VE

FIN

IS

HE

D

FL

O

O R

4"

1 4"

POWER (SEE

SHEET NOTES

FOR DETAILS)

3"

V2

TP1

TYPICAL WALL MOUNTED

TOUCH PANEL INSTALLATION

(OPTIONAL)

" A

BO

VE

FIN

IS

HE

D

FL

O

O R V3

A SERIES SURGICAL LIGHT CONTROLLER

K1 13 1/

2"

16 1/2"2"

3/

4"

" M

IN

. A

BO

VE

F

IN

IS

HE

D

FL

O O

R

6"

2" 6"

SURGICAL LIGHT

CONTROL

ROUGH OPENING

9 1/2"

EQUIPMENT SCHEDULE

LEGEND DESCRIPTION RESPONSIBILITY

E1,E2 WALL MOUNTED DISPLAY STERIS

G1,G3,G2,G4 CEILING SPEAKER, 8.7" DIAMETER x 8.8" HEIGHT; 7.4 LB STERIS

K1 SURGICAL LIGHT CONTROLLER STERIS

TP1 WALL MOUNTED TOUCH PANEL STERIS

V2 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR

V3 LEVITON BOX FOR STERIS AUDIO DISTRIBUTION ELECTRICAL CONTRACTOR

SHEET NOTES

1. EACH LEVITON BOX REQUIRES DEDICATED EMERGENCY BACK-UP 120V ON (2)20 AMP POWER CIRCUITS FROM THE O.R. IT SERVES (NOT

BY STERIS)

2. LEVITON BOX (NOT BY STERIS) WITH VENTED DOOR FOR STERIS POWER & AUDIO DISTRIBUTION; BOX PRODUCT NUMBER: 47605-42N

(REQUIRES MINIMUM 2" SERVICE CLEARANCE); DOOR (NOT BY STERIS) PRODUCT NUMBER: 47605-42S; LEVITON AC DUAL POWER MODULE (NOT BY STERIS - BOLTED TO BOTTOM OF LEVITON BOX - LOCATED INSIDE OF WALL) PRODUCT NUMBER: 47605-DP

3. LEVITON BOX REQUIRES CONDUIT TO ENTER FROM THE TOP RIGHT OF THE BOX

4. WALL DISPLAY BACKING TO BE PROVIDED AND INSTALLED BY CONTRACTOR; WALL DISPLAY BACKING (NOT BY STERIS) MUST SUPPORT

A 250LBS LOAD; WALL BRACKET NOT INSTALLED BY STERIS

5. WALL DISPLAY REQUIRES 4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING

6. WALL DISPLAY REQUIRES DUPLEX OUTLET 120 VAC ON 20 AMP SERVICE (STANDARD OR EMERGENCY)

7. FINAL SPEAKER LOCATIONS TO BE DECIDED BY ARCHITECT/CONTRACTOR; CONTRACTOR TO STUB OUT ~ 18" ABOVE FINISHED CEILING

FOR SPEAKER CONNECTION; CONTRACTOR TO PROVIDE AND INSTALL SEISMIC SUPPORT WIRE FOR SPEAKER ANCHORING TO THE BUILDING STRUCTURE; CONTRACTOR TO CUT ROUND HOLE IN CEILING AT THE SPEAKER LOCATION ACCORDING TO THE SPEAKER INSTALLATION TEMPLATE REQUIRES FLEXIBLE CONDUIT WITH CONNECTOR ON SPEAKER END (NOT BY STERIS; TYPICAL AT EACH

SPEAKER LOCATION)

G1,G3,G2,G4

TYPICAL JBL 16C/T SPEAKER INSTALLATION (OPTIONAL)

*REFERENCE SHEET NOTE #7

CEILING

8.7"

9.

4"

8.

8"

8.

2"

"M

AX

FLEX

CONDUIT

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HV1000 INTEGRATION SYSTEM

PROPOSED EQUIPMENT

INSTALLATION DETAILS

NTS 04

HARMONY CUSTOM INTEGRATION SYSTEM INSTALLATION

GUIDELINES

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

STERIS

INDEX

PAGE # SHEET TITLE ROOM #

01 TITLE PAGE, INDEX -

01A INFRASTRUCTURE RESPONSIBILITIES TERMS & CONDITIONS -

02 PROPOSED OPERATING ROOM 4 LAYOUT 2E-07

02A PROPOSED OPERATING ROOM 5 LAYOUT 2E-08

02B PROPOSED OPERATING ROOM 6 LAYOUT 2E-09

02C PROPOSED OPERATING ROOM 9 LAYOUT 2E-20

02D PROPOSED OPERATING ROOM 10 LAYOUT 2E-19A

03 PROPOSED EQUIPMENT INSTALLATION DETAILS -

04 CEILING MOUNT INSTALLATION REQUIREMENTS -

GENERAL NOTES

1. SEE INFRASTRUCTURE RESPONSIBILITIES PAGE (1A) FOR ADDITIONAL DETAILS & DESCRIPTIONS FOR SPECIFIC STAKEHOLDER RESPONSIBILITIES

2. SUPPLY AND INSTALLATION OF ALL CONDUIT, POWER CIRCUITS, BACK BOXES, ELECTRICAL DEVICES, ELECTRICAL WIRING IS TO BE PROVIDED BY THE

CLIENTS ELECTRICAL CONTRACTOR UNLESS NOTED OTHERWISE

3. ONE WIRE HANGER FOR CABLE COILS REQUIRED FOR EACH MOUNT

4. CLIENT CONTRACTOR TO PROVIDE BLANK COVER PLATES FOR ALL JUNCTIONS BOXES WHERE NECESSARY

5. ALL CABLE RUNS OUTSIDE INTEGRATED ROOMS ARE THE RESPONSIBILITY OF HOSPITAL/CONTRACTOR; ALL CABLE RUNS SHOULD BE CONTINUOUS

EXCEPT WHERE NOTED

6. 24x24" ACCESS MUST BE PROVIDED WITHIN 36" OF ALL JUNCTION BOXES

7. CABLE RUNS THRU PLENUM SPACE MUST BE PLENUM RATED OR SHALL BE INSTALLED WITH CONDUIT

8. CABLES NOT RUN IN CONDUIT MUST HAVE J-HOOKS EVERY 4' FOR SUPPORT (NOT BY STERIS)

9. WHERE J-HOOKS ARE USED IN CEILING, CONDUIT IS REQUIRED IN ALL CONCEALED LOCATIONS (SIZED PER SCHEDULES)

10. EACH RACK REQUIRES (4) NETWORK PORTS PER O.R.

11. ALL POWER FOR INTEGRATION EQUIPMENT SHOULD COME FROM A SINGLE ELECTRICAL PANEL IN THE OR IT SERVES

12. CONFORMANCE TO NATIONAL, STATE AND LOCAL CODES FOR ELECTRICAL CIRCUITS AND CONDUIT IS THE RESPONSIBILITY OF THE HOSPITAL OR

THEIR CONTRACTED ELECTRICIAN

13. THE EQUIPMENT AND ROUGH-IN REQUIREMENTS SHOWN ARE FOR BUDGETARY PLANNING PURPOSES ONLY; EQUIPMENT IS SUBJECT TO CHANGE AND DIFFERENT OPTIONS ARE AVAILABLE TO MEET SPECIFIC CUSTOMER DESIGN CRITERIA; POST-ORDER DRAWINGS WILL REFLECT EXACT EQUIPMENT

SPECIFICATIONS, AND REQUIRES CUSTOMER SIGNED APPROVAL

14. USE OF ISOLATED GROUND POWER TO BE DETERMINED BY ELECTRICAL CONTRACTOR

ABBREVIATIONS:

A.F.F. = ABOVE FINISHED FLOOR

B.C.P. = BLANK COVER PLATE

C. = CONDUIT

C.P. = COVER PLATE

C.K.T. = CIRCUIT

E.C. = ELECTRICAL CONTRACTOR

N.S.E. = NURSE STATION EQUIPMENT

ALL STERIS TERMS AND CONDITIONS RELATIVE TO STERIS AND CUSTOMER,

DESIGNATED CONTRACTOR RESPONSIBILITIES, SITE READINESS, AND CHANGE

ORDERS APPLY. THESE WILL BE PROVIDED TO ALL PARTIES ABOVE AT THE

PROJECT KICKOFF MEETING BY YOUR STERIS REPRESENTATIVE

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HARMONY CUSTOM INTEGRATION SYSTEM

TITLE PAGE, INDEX

NTS 01

STERIS Responsibilities:

· Provide cable conduit drawings, electrical requirements, and mounting requirements for the room.

· Provide space, structural, and heat load requirements for the base unit.

· Provide site preparation specifications.

· Deliver all custom back boxes, mounting brackets, and construction-related items to the hospital in accordance with the mutually agreed upon construction or renovation schedule.

· Participate in project meetings in person, via conference call or via e-mail.

· Ship equipment in protective and labeled packaging.

· Ship cables and equipment on the scheduled ship date

· Pull and terminate non-plenum cabling inside the OR through specified conduits and junction boxes as specified in the Product

Description and Final Project Plan.

· Provide external cables and detailed specifications for the interconnections only if external cabling is specified in the Customer

Proposal and Final Project Plan.

· Provide a pre-installation checklist to verify site preparation is complete.

· Mount, install, and connect integration equipment - including video, audio, and control cables, cable terminations, wall cameras, wall monitors (does not include wall mounting brackets), base unit, touch screen control, and surgical monitors - as applicable to project.

· Test and validate integration system to complete customer satisfaction.

· Provide the hospital with a certificate of completion signed by the install team and the hospital.

· Adjust STERIS surgical light brakes as necessary after installation of STERIS monitors.

· In-service hospital staff at the time of the install. Further in-service training may be scheduled as mutually agreed.

· Leave a tidy work area each evening and remove packing material to a customer provided on-site container.

Customer or Customer's Designated Contractor Responsibilities:

· Ensure that site preparations and construction are complete and ready for final integration installation - including electrical and network connections, finished floor and paint.

· Provide complete room schematics including electrical, mechanical, and equipment location details.

· Install cable conduit, junction boxes, electrical outlets, and mounting brackets according to project specifications.

· Ensure that all electrical requirements comply with local code.

· Ensure space, structural, and heat load requirements are met according to project specs and plan.

· Receive shipment and safely, securely store equipment if shipment arrives prior to requested installation date.

· Provide adequate access from delivery/storage area to installation site.

· Relocate equipment from on-site storage/delivery area to installation site.

· Run all cables outside the OR- Including conduit, J-hooks, penetrations through firewall, video cables, radio antennae cables, or fiber optics to networking hubs and video conferencing hubs, communications closets, conference rooms, centralized nurse desks, rooftop, or other departments.

· Ensure that site preparations and construction are complete and ready for final integration installation - including electrical and network connections, finished floor and paint.

· Proactively schedule the installation in conjunction with the STERIS project manager.

· Provide a minimum of two (2) weeks advance notice for requested changes to the installation schedule.

· Remove old equipment or furniture from the room.

· Provide a clean, dry, dust free environment for equipment storage and installation.

· Provide a secure work site for storing ladders, toolboxes etc. during install.

· Provide easy access to all rooms involved in the installation during normal business hours and upon request after hours.

· Ensure that all non-STERIS or third-party equipment and systems that are to be part of the integrated solution are on-site and functional at the time of installation - otherwise full testing and connections cannot be completed.

· Participate in a final walk-through evaluation of the system followed by a sign-off of the certificate of completion.

· Provide an on-site container for trash, including shipping and packing material. Empty container as needed and on a regular basis.

Cabling and Fiber Responsibilities:

· All cables and fiber optic runs to locations outside the OR are the responsibility of the hospital.

Minimum Site Readiness Successful and efficient installation of a complete integration system requires that the site be carefully prepared according to the project documentation and schematics. A site that is not ready when the installation team arrives will cause delays and additional expenses to both parties. If the installation site is not ready per the mutually agreed specifications, STERIS reserves the right to charge a change order fee and extend the timeline needed to finish project past the planned date given at the project's inception.

To be ready a site should have -

1. All floor work must be complete

2. All painting must be complete

3. All electrical requirements must be installed and powered

4. Room must be clean and free from dust

Customer or Customer's Designated Contractor Rack Clearance & Ventilation Guidelines:

· Reference site specific rack layout document provided by STERIS for total rack BTU estimation information.

· Customer or Customer's Designated Contractor responsible for ensuring proper ventilation of STERIS Integration Equipment.

· Contact your STERIS representative for any questions regarding STERIS Integration Equipment ventilation and/or BTU ratings.

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HARMONY CUSTOM INTEGRATION SYSTEM

INFRASTRUCTURE RESPONSIBILITIES

TERMS & CONDITIONS

NTS 01A

O.R. #3 2E-06M1

V1

K1

C1

B1

C2

A SERIES LIGHT CONDUIT SCHEDULE - OR 5&6

FROM - TO QUANTITY SIZE RESPONSIBILITY

K1 - B1 2 .75" ELECTRICAL CONTRACTOR

LIGHT CONDUIT SCHEDULE NOTES

1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT

AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz

2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET

3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO

REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE

SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM

4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN

DOUBLE RUNS ARE REQUIRED (LONG RUNS)

5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE

THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS

LIGHT SYSTEM IS UTILIZED

*REFER TO TITLE PAGE FOR ADDITIONAL ELECTRICAL NOTES*

CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

M1 - HOSPITAL

NETWORK 1 1" ELECTRICAL CONTRACTOR

*INTEGRATION EQUIPMENT

LEGEND DESCRIPTION RESPONSIBILITY

B1 FUTURE SURGICAL LIGHT MOUNT STERIS

C1 FUTURE UTILITY HUB STERIS

C2 FUTURE ANESTHESIA BOOM STERIS

K1 FUTURE SURGICAL LIGHT CONTROLLER STERIS

M1

IP STATUS ROOM CAMERA/NURSE OBSERVATION SYSTEM STERIS

4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING

REQUIRED FOR IP CAMERA ELECTRICAL CONTRACTOR

ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT

SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HARMONY CUSTOM INTEGRATION SYSTEM

PROPOSED OPERATING ROOM 3 LAYOUT

[2E-06]

NTS 02

O.R. #4 2E-07

B1

C1

C2

K1

M1

V1 V1

A SERIES LIGHT CONDUIT SCHEDULE - OR 5&6

FROM - TO QUANTITY SIZE RESPONSIBILITY

K1 - B1 2 .75" ELECTRICAL CONTRACTOR

LIGHT CONDUIT SCHEDULE NOTES

1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT

AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz

2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET

3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO

REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE

SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM

4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN

DOUBLE RUNS ARE REQUIRED (LONG RUNS)

5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE

THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS

LIGHT SYSTEM IS UTILIZED

*REFER TO TITLE PAGE FOR ADDITIONAL ELECTRICAL NOTES*

CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

B1 - V1 1 1.25" ELECTRICAL CONTRACTOR

C1 - V1 1 1.25" ELECTRICAL CONTRACTOR

M1 - HOSPITAL

NETWORK 1 1" ELECTRICAL CONTRACTOR

*INTEGRATION EQUIPMENT

LEGEND DESCRIPTION RESPONSIBILITY

B1 SURGICAL LIGHT MOUNT STERIS

C1 UTILITY HUB STERIS

C2 ANESTHESIA BOOM STERIS

K1 SURGICAL LIGHT CONTROLLER STERIS

M1

IP STATUS ROOM CAMERA/NURSE OBSERVATION SYSTEM STERIS

4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING

REQUIRED FOR IP CAMERA ELECTRICAL CONTRACTOR

V1 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR

ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT

SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HARMONY CUSTOM INTEGRATION SYSTEM

PROPOSED OPERATING ROOM 4 LAYOUT

[2E-07]

NTS 02A

X-RAY&EQUP

2E-15

CONTROL RM

2E-10A

O.R. #5 2E-08

C2

K1

C1

M1

V1

B1 V2

V3

A SERIES LIGHT CONDUIT SCHEDULE - OR 5&6

FROM - TO QUANTITY SIZE RESPONSIBILITY

K1 - B1 2 .75" ELECTRICAL CONTRACTOR

LIGHT CONDUIT SCHEDULE NOTES

1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT

AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz

2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET

3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO

REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE

SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM

4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN

DOUBLE RUNS ARE REQUIRED (LONG RUNS)

5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE

THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS

LIGHT SYSTEM IS UTILIZED

*REFER TO TITLE PAGE FOR ADDITIONAL ELECTRICAL NOTES*

CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

B1 - V1 1 1.25" ELECTRICAL CONTRACTOR

C1 - V1 1 1.25" ELECTRICAL CONTRACTOR

M1 - HOSPITAL

NETWORK 1 1" ELECTRICAL CONTRACTOR

*INTEGRATION EQUIPMENT

LEGEND DESCRIPTION RESPONSIBILITY

B1 SURGICAL LIGHT MOUNT STERIS

C1 UTILITY HUB STERIS

C2 ANESTHESIA BOOM STERIS

K1 SURGICAL LIGHT CONTROLLER STERIS

M1

IP STATUS ROOM CAMERA/NURSE OBSERVATION SYSTEM STERIS

4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING

REQUIRED FOR IP CAMERA ELECTRICAL CONTRACTOR

V1 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR

ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT

SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HARMONY CUSTOM INTEGRATION SYSTEM

PROPOSED OPERATING ROOM 5 LAYOUT

[2E-08]

NTS 02B

VESTIBULE

2E-18B

STERILIZA.

2E-20A

STERILIZA.

2E-18A

F.A .C

STORAGE

2E-20B

O.R. #9 2E-20

K1

MODIFIED TO

V1

D1 B4

M1

*REFER TO TITLE PAGE FOR ADDITIONAL ELECTRICAL NOTES*

CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

B4 - V1 1 1.25" ELECTRICAL CONTRACTOR

D1 - V1 1 1.25" ELECTRICAL CONTRACTOR

M1 - HOSPITAL

NETWORK 1 1" ELECTRICAL CONTRACTOR

*INTEGRATION EQUIPMENT

LEGEND DESCRIPTION RESPONSIBILITY

B4 SURGICAL LIGHT MOUNT STERIS

D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION

BOX FOR HDSDI INPUT ELECTRICAL CONTRACTOR

K1 SURGICAL LIGHT CONTROLLER STERIS

M1

IP STATUS ROOM CAMERA/NURSE OBSERVATION SYSTEM STERIS

4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING

REQUIRED FOR IP CAMERA ELECTRICAL CONTRACTOR

V1 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR A SERIES LIGHT CONDUIT SCHEDULE - OR 9&10

FROM - TO QUANTITY SIZE RESPONSIBILITY

K1 - B4 2 .75" ELECTRICAL CONTRACTOR

LIGHT CONDUIT SCHEDULE NOTES

1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT

AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz

2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET

3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO

REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE

SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM

4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN

DOUBLE RUNS ARE REQUIRED (LONG RUNS)

5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO ENSURE

THAT THE MOST CURRENT & DETAILED INFORMATION OF THIS

LIGHT SYSTEM IS UTILIZED

ALL LIGHT MOUNTS REQUIRE 2 J-HOOKS EACH, 8" APART -OR- EQUIVALENT

SUSPENSION POINTS ABOVE THE STRUCTURAL MOUNTING PLATE

REVISION SUMMARY

REVISION # DESCRIPTION DESIGNER/DATE CHECKER

5 CHANGED OR6-8 TO HV1000 SYSTEM; UPDATED CUSTOM INTEGRATION

SYSTEM TO CURRENT STANDARDS TM/16APR2024 JTP

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

Healthcare Design Group 5960 Heisley Road Mentor, OH 44060 USA 800-548-4873

DATE

SCALE

CKDR

CHECK PRINT

ST

ER

IS

d ra wi ng s d o no t s up er se de th e dr aw in gs fr om th e ar ch ite ct o f r ec or d.

VA BRONX MEDICAL CENTER

BRONX, NEW YORK

6313D5R5-INT

JTP 27JUL2023

PAGE

HARMONY CUSTOM INTEGRATION SYSTEM

PROPOSED OPERATING ROOM 9 LAYOUT

[2E-20]

NTS 02C

STERILIZA.

2E-20A

STERILIZA.

2E-18A

O.R. #10 2E-19A

STORAGE

2E-20B

C7 K1

V1 V1

D1

D1B4

M1

*REFER TO TITLE PAGE FOR ADDITIONAL ELECTRICAL NOTES*

CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

B4 - V1 1 1.25" ELECTRICAL CONTRACTOR

D1 - V1 1 1.25" ELECTRICAL CONTRACTOR

M1 - HOSPITAL

NETWORK 1 1" ELECTRICAL CONTRACTOR

*INTEGRATION EQUIPMENT

LEGEND DESCRIPTION RESPONSIBILITY

B4 SURGICAL LIGHT MOUNT STERIS

C7 ANESTHESIA BOOM STERIS

D1 WALL MOUNTED HUBBELL - RACO 698 4 GANG JUNCTION

BOX FOR HDSDI INPUT ELECTRICAL CONTRACTOR

K1 SURGICAL LIGHT CONTROLLER STERIS

M1

IP STATUS ROOM CAMERA/NURSE OBSERVATION SYSTEM STERIS

4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING

REQUIRED FOR IP CAMERA ELECTRICAL CONTRACTOR

V1 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR

A SERIES LIGHT CONDUIT SCHEDULE - OR 9&10

FROM - TO QUANTITY SIZE RESPONSIBILITY

K1 - B4 2 .75" ELECTRICAL CONTRACTOR

LIGHT CONDUIT SCHEDULE NOTES

1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE INPUT

AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz

2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET

3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO

REFERENCED WITHIN THE…