The National Weather Service requires a new stable temperature calibration bath to replace a non-functional unit used for internal calibration of 50-60 platinum resistance thermometers and accuracy comparison services. The equipment must meet specific performance criteria including internal bath dimensions between 6in x 6in x 8in (minimum) and 9in x 14.5in x 16in (maximum), with an immersion depth of 8in to 16in and volume capacity of 1 to 11 gallons. The bath must operate with Xiameter PMX-200 Silicone Fluid 1cSt for temperatures between -80°C to 10°C and distilled water for temperatures between 5°C to 60°C. Critical performance specifications include a low temperature range of -80°C or lower, high temperature range of 60°C or higher, full range stability and uniformity of ±0.02°C or better, combined stability and uniformity of ±0.025°C or better, heating rate of ≥0.625°C/minute, cooling rate of ≥0.4375°C/minute, display resolution of ≤0.01°C, set-point accuracy of ≤0.5°C, and set-point repeatability of ≤0.01°C. The equipment must operate on 115 or 230 VAC power, fit within a 31in W x 31in L x 60in H footprint, and include serial communications capability to receive and send commands via third-party terminal programs such as Hyper Terminal or PuTTy to control temperature setpoints. Shipping costs and minimized maintenance costs must be included. Delivery is required to the Sterling Field Support Center in Sterling, VA 20166, with acceptance contingent upon a full successful demonstration in accordance with a vendor-supplied test plan and submission of acceptance test results with the invoice. The National Weather Service point of contact is Kathryn Franc (kathryn.franc@noaa.gov; 301-388-8565).
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 6 SOL for SAM_1305M226Q0240_V.1.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work 142082-26-0415
SFSC Temperature Calibration Bath Replacement
The National Weather Service Sterling Field Support Center (SFSC) Temperature and Humidity laboratory provides internal calibration for 50-60 platinum resistance thermometers and accuracy comparison services for NWS programs testing new temperature sensors and equipment. In order to provide accurate calibrations, the use of an extremely stable and accurate temperature source is required. Staff utilize two stable temperature calibration bath sources, one for low temperature ranges and another for high temperature ranges. Currently one of our temperature calibration baths required for temperature accuracy testing and sensor calibrations is no longer working correctly.
The unit can no longer cool properly and based on in house assessments the unit will likely stop functioning completely. The lack of system cooling also causes a major safety risk along with the inability for the temperature bath to function in a correct manner. The unit must be replaced with a new stable temperature source to continue to provide accurate calibration and temperature accuracy testing for NWS programs and sensors.
Performance Criteria
The equipment requirements listed below must all be met for the final acceptance of the system.
Internal Bath Dimensions (Minimum): 6in x 6in x 8in
Internal Bath Dimensions (Maximum): 9in x 14.5in x 16in
Immersion Depths: 8in to 16in
Access Opening (Minimum): 6in x 6in
Access Opening (Maximum): 14in x 8.5in
Volume: 1 to 11 gallons
Bath Fluid Type: Must operate with Xiameter PMX-200 Silicone Fluid 1cSt at temperatures between -80°C to 10°C and Distilled Water at temperatures between 5°C to 60°C
Low Temperature Range: Less than or equal to -80°C
High Temperature Range: Greater than or equal to 60°C
Power Source: 115 or 230 VAC
Environmental Operating Range:
o Relative Humidity: 80% at 31°C to 50% at 40°C o Temperature : 5 to 50°C
Full Range Stability: Less than or equal to 0.02°C
Full Range Uniformity: Less than or equal to 0.02°C
Combined Stability and Uniformity: Less than or equal to 0.025°C
Heating Rate: Greater than or equal to 0.625°C / minute
Cooling Rate: Greater than or equal to 0.4375°C / minute
Display Resolution: Less than or equal to 0.01°C
Set-Point Accuracy: Less than or equal to 0.5°C
Set-Point Repeatability: Less than or equal to 0.01°C
Serial communications:
o Must possess the ability to receive commands sent from a third-party terminal program such as Hyper Terminal or PuTTy. These commands must have the ability to control the operation of the temperature bath specifically the ability to set and maintain a temperature setpoint. Must also have the ability to send commands that can be received by a third-party terminal program such as Hyper Terminal or PuTTy.
Overall Equipment Dimensions: All equipment must fit within 31in W x 31in L x 60in H area
Maintenance: Maintenance costs minimized
Shipping: Shipping Cost included
Inspection and Acceptance
Upon completion, contractor shall notify the Government representative for approval and acceptance. The acceptance period will include a full successful demonstration in accordance with vendor supplied accepted test plan with the federal representative signature. Delivery of acceptance test results from the test plan must be submitted with invoice for processing and approval of payment.