5 - DID_Calibration_Certification_Report.pdf

PDF 53 KB Posted

Attached to
B-1 Nose Radome Repair and New Buy - Final RFP
Solicitation number
FA811923R0004
Issued by

View the file

Other files

File Type Posted
FA811923R0004_______0005.pdf PDF
FA811923R0004_______0004.pdf PDF
26_PWS_Updated_20_March_23.pdf PDF
FA811923R0004_______0003.pdf PDF
FA811923R0004_______0002.pdf PDF
32 - Pricing Matrix Updated.xlsx XLSX spreadsheet
1 - AFMC 260 First Article Updated.pdf PDF
FA811923R0004_______0001.pdf PDF
18 - DID_Product_Engineering_Design_Data_and_Associated_Lists.pdf PDF
8 - DID_Certification_Data_for_Non_Nuclear_Level_I.pdf PDF
9 - DID_Computer_Software_Product_End_Items.pdf PDF
Final RFP FA811923R0004.pdf PDF
33 - Salient_Characteristics_Radome.pdf PDF
19 - DID_Production_Surge_Plan.pdf PDF
10 - DID_Conference_Agenda.pdf PDF
3 - DD_Form_1653_Transportation_Data_for_Solicitations.pdf PDF
28 - Item_Description_Antenna.pdf PDF
23 - DID_Test_Inspection_Report.pdf PDF
Exhibit A - CDRLs.pdf PDF
14 - DID_Counterfeit_Prevention_Plan.pdf PDF
6 - DID_CAV_Reporting.pdf PDF
29 - DD254_Security_Classification.pdf PDF
30 - Justification_for_Qualification_Requirements.pdf PDF
17 - DID_Presentation_Material.pdf PDF
7 - DID_Certificate_of_Compliance.pdf PDF
25 - IUID_Checklist.pdf PDF
32 - Pricing_Matrix.xlsx XLSX spreadsheet
20 - DID_Report_of_Shipping_Item_and_Packaging_Discrepancy.pdf PDF
4 - DD_Form_2169_Special_Packaging_Instructions.pdf PDF
16 - DID_Government_Property_Inventory_Report.pdf PDF
13 - DID_Contractors_Corrective_Action_Plan.pdf PDF
27 - Item_Description_Radome.pdf PDF
1 - AFMC_260_First_Article_Requirements.pdf PDF
22 - DID_Status_Report.pdf PDF
26 - PWS_Updated_26_Jan_23.pdf PDF
21 - DID_Request_for_Variance.pdf PDF
31 - Personnel_and_Skill_Requirement.pdf PDF
12 - DID_Contract_Depot_Maintenance_Production_Report.pdf PDF
11 - DID_Conference_Minutes.pdf PDF
24 - GFP.pdf PDF
2 - AFMC_158_Packaging_Requirements.pdf PDF
15 - DID_Engineering_Change_Proposal.pdf PDF
Show all 42

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DATA ITEM DESCRIPTION

Title: CALIBRATION CERTIFICATE/REPORT

Number: DI-QCIC-80798C Approved Date: 20170123 AMSC Number: F9766 Limitation: N/A DTIC Applicable: No GIDEP Applicable: No Preparing Activity: 36 Project Number: QCIC-2017-003 Applicable Forms: N/A Use/Relationship: The Calibration Certificate/Report documents calibration certificates or reports accompanying the calibrations performed on Department of Defense (DoD) test, measurement, and diagnostic equipment.

a. This Data Item Description (DID) contains the format and content preparation instructions for the data product generated by the specific and discrete task requirement as delineated in the contract.

b. This DID is applicable to contracts requiring calibration of test equipment prior to delivery.

This DID is consistent with ISO/IEC 17025, General requirements for the competence of testing and calibration laboratories, requirements for calibration reports or certificates.

(Copies of this document are available online at http://www.iso.org.)

c. This DID supersedes DI-QCIC-80798B.

Requirements:

1. Reference documents. The applicable issue of the documents cited herein, including their approval dates and dates of any applicable amendments, notices, and revisions, shall be as specified in the contract.

2. Format. Contractor format is acceptable.

3. Content. The calibration certificate or report shall include:

a. Title (e.g., calibration certificate or calibration report);

b. The name and address of the laboratory and the location where the calibrations were carried out, if different from the address of the laboratory;

c. Unique identification of the calibration certificate or test report (such as the serial number), and on each page an identification in order to ensure that the page is recognized as a part of the test report or calibration certificate, and a clear identification of the end of the calibration certificate or test report;

d. The name and address of the client or customer;

e. Identification of the method used;

f. A description of, the condition of, and the unambiguous identification of the item(s) tested or calibrated;

g. The date of receipt of the test or calibration item(s) where this is critical to the validity and application of the results, and the date(s) of performance of the test or calibration;

DISTRIBUTION STATEMENT A. Approved for public release. Distribution is unlimited.

Source: https://assist.dla.mil -- Downloaded: 2017-03-13T13:39Z Check the source to verify that this is the current version before use.

DI-QCIC-80798C

h. Reference to the sampling plan and procedures used by the laboratory or other bodies where these are relevant to the validity or application of the results;

i. The test or calibration results with, where appropriate, the units of measurement;

j. The name(s), function(s) and signature(s) or equivalent identification of person(s) authorizing the calibration certificate or test report;

k. Where relevant, a statement to the effect that the results relate only to the items tested or calibrated;

l. The conditions (e.g., environmental) under which the calibrations were made that have an influence on the measurement results;

m. The uncertainty of measurement or a statement of compliance with an identified metrological specification or clauses thereof;

n. Evidence that the measurements are traceable.

End of DI-QCIC-80798C.

Source: https://assist.dla.mil -- Downloaded: 2017-03-13T13:39Z Check the source to verify that this is the current version before use.