This sources sought notice requests information from commercial sources capable of providing clinical audiometers and related equipment to the Department of Veterans Affairs, Veterans Integrated Service Network 23 in support of the Iowa VA Healthcare System. The notice seeks two each of a Madsen Astera 2-Channel Clinical Audiometer, a Zodiac Diagnostic Stand Alone Tympanometer, Aurical Probe Microphone measurement system with HI-PRO, NOAH System 4 licenses, and other specified audiology equipment. Responses are due by 10:00 AM Central Time on December 18, 2023. The notice provides contact information, the NAICS code of 334510, questions for a sources sought questionnaire, and required specifications for the clinical audiometer and related equipment. It does not represent a commitment by the government to issue a solicitation or award a contract.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sources Sought Notice Sources Sought Notice

SUBJECT*
Clinical Audiometer (VA-24-00021267)

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
58102
SOLICITATION NUMBER*
36C26324Q0227
RESPONSE DATE/TIME/ZONE
12-18-2023 10:00 CENTRAL TIME, CHICAGO, USA
ARCHIVE
60 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N

SET-ASIDE

PRODUCT SERVICE CODE*
6515
NAICS CODE*
334510
CONTRACTING OFFICE ADDRESS
DEPARTMENT OF VETERANS AFFAIRS

NETWORK 23 CONTRACTING OFFICE

2101 North Elm Street Building 30 (04-S)

FARGO ND 58102

POINT OF CONTACT*

Contract Specialist Jahaun Stafford jahaun.stafford@va.gov 701-638-2745

PLACE OF PERFORMANCE

ADDRESS

Iowa City VA Health Care System

601 Highway 6 West

Iowa City IA

POSTAL CODE
52246
COUNTRY
USA

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS

EMAIL DESCRIPTION

DESCRIPTION

THIS IS A SOURCES SOUGHT NOTICE ONLY. THIS IS NOT A REQUEST FOR PROPOSAL OR QUOTE; THE VA IS NOT SEEKING PRICING OR OFFERS.

The Department of Veterans Affairs, VISN 23, in support of Iowa VA Healthcare System – IA, is currently requesting information to determine the existence of viable commercial sources capable of providing Clinical Audiometers (Sound and Medical Devices).

The standard delivery period is 60 Days ARO.

Description of Supplies: Madsen Astera Audiology Equipment (Brand Name or Equal)

NAICS Code: 334510 - Audiological equipment, electromedical, manufacturing and the small business size standard is 1,250 Employees.

The Government is not soliciting offers at this time, and it will be the Contracting Officer who will determine which suggested changes are accepted. This is a sources sought notice to gain knowledge of potential qualified sources. A solicitation is not currently available.

Responses to this request will be considered for the purpose of determining how the requirements will be stated and how the procurement will be conducted. This sources sought announcement is not to be construed as a commitment by the Government, implied or otherwise, to issue a solicitation or to award a contract. No reimbursement will be made for any costs associated with providing information in response to this synopsis or any follow-up information requests.

If you are interested and are capable of providing the required services, please complete the sources sought questionnaire below and email your response to jahaun.stafford@va.gov no later than 10 a.m. on 18 December 2023. Include any comments or questions you may have. Also provide your capability statements within your response.

A SOURCES SOUGHT QUESTIONNAIRE is below. Responses to the Questionnaire are encouraged from all interested parties. You are advised that providing responses to the questions will not automatically include you in the acquisition process for this solicitation. Please provide any additional feedback in your response that you feel is relevant.

Source Sought Questionnaire:

Company Name ___________________________________________

Phone Number ____________________________________________

Email Address ____________________________________________

DUNS ______________________

* Proof of Systems for Award Management Registration (SAM) is mandatory*

1. GSA Number if applicable_________________________________

2. Is your firm eligible for participation in one of the following small business programs? If so, please indicate the program. The North American Industry Classification System (NAICS) code is tentatively 334510 Electromedical and Electrotherapeutic Apparatus Manufacturing; the small business size standard for this NAICS code is 1,250 Employees.

A size standard, which is usually stated in number of employees or average annual receipts, represents the largest size that a business (including its subsidiaries and affiliates) may be to remain classified as a small business for Small Business Administration and federal contracting programs. The definition of "small" varies by industry.

[ ] yes [ ] no Small Business (SB) [ ] yes [ ] no HUBZone [ ] yes [ ] no Small Business 8(a) [ ] yes [ ] no Small Disadvantaged Business (SDB) [ ] yes [ ] no Economically-Disadvantaged Women-Owned Small Business (EDWOSB) [ ] yes [ ] no Women-Owned (WO) Small Business [ ] yes [ ] no Service Disabled Veteran Owned Small Business (SDVOSB) [ ] yes [ ] no Other (please specify)

3. If your company were selected, how many calendar days would your company need after the contract award date before completing delivery and services?

_____ 15 days _____ 30 days _____45 days _____other (please list)

4. In your opinion, what are the risks associated with this effort? Comment on any current or potential technical, cost, schedule, or performance risks that you would like to make the Government aware of that concerns this service.

5. Provide any additional feedback that you feel is relevant (i.e., problems or any other issues experienced with similar contracts).

Note: Do not include Proprietary, classified, confidential, or sensitive information in responses.

Functional/Performance/Specifications General Equipment Name: Audio Booth Equipment for Decorah & Galesburg CBOC Description & Salient Characteristics Audiometer (Astera 2, 2-Channel Clinical Audiometer / Brand Name or Equal) - All equipment must be compatible with Noah database approved by VBA/VHA and required for specifications for use in documenting clinic diagnostic exams. In the room monitor to provide counseling to patient required.

Audiometer, Tympanometer, and additional accessories to run equipment required.

If computer based, must have audiometer control panel as well as computer accessories for dual function.

Audiometer: Audiometer must be a two-channel clinical diagnostic manual audiometer.

• Pure Tone - Two Independent Channels

• Frequency Range:

Full Frequency Range: 125Hz to 20,000Hz Air Conduction: 125Hz to 12,000Hz High Frequency: 8,000Hz to 20,000Hz Bone Conduction: 250Hz to 8,000Hz Sound Field and Paired Inserts: 125Hz to 8,000Hz

• Intensity Range:

Air Conduction (TDH): -10dB HL to 120dB HL Bone Conduction: Mastoid: -10dB HL to 75dB HL, Forehead: -10dB HL to 65dB HL Sound Field: -10dB HL to 90dB HL Masking: Narrow Band Noise, White Noise, Speech Noise

• Format:

Steady, Pulsed or FM tone

• Communications and Monitoring:

Talk Forward, Talk Back, Monitor, Aux Intercom, On-Board VRA Control

• Special Tests (user defined):

Pure Tone Stenger, Speech Stenger, and Maryland CNC worklists. Ability to add speech tests to the existing list of pre-programmed tests such as the Speech Recognition in Noise Test (SPRINT) that is unique to the VBA.

• Special Tests:

ALT (ABLB), SISI, High Frequency, TEN, QuickSIN, Tympanometer (Middle Ear Analyzer), BKB-SIN

• General Specifications o Size: 20.1"(W) x 14.6"(D) x 13.2"(H), Weight: 17 lb.

o Test Types: Tympanometry, Acoustic Reflex Threshold, Reflex Decay, Eustachian Tube Function (Intact and Perforated) o Protocols: Diagnostic, Screening, Two-Component Tympanometry, User-defined o Display: Internal Color Touchscreen and optional eternal HDMI monitor o Interface: USB (keyboard, mouse, Flash Drive, PC communications) o Printout: Internal Printer o Probe Tone:

• 226 Hz (85 dB SPL ± 1.5 dB) 678 Hz (72 dB SPL ± 1.5 dB) 1000 Hz (69 dB SPL ± 1.5 dB)

• Admittance Measurements Range: 226 Hz (-7.0 to +7.0 mmho) 678 Hz (-21.0 to +21 mmho)1000 Hz (-32.0 to +32 mmho) Sensitivity Scale: Auto Scales to Appropriate Range, Manual selection also possible in Reflex Modes only

• Pressure Measurements (load volume of 0.2 to 7.0 ml) Range: Normal = +200 to -400 daPa Wide = +400 to -600 daPa Sweep Rate: 12.5, 50.0, 200, 600 and 600/200 daPa/sec.

Sweep Accuracy: 10% of nominal rate Maximum limits (in 0.5cc cavity): -800 daPa and +600 daPa

• Reflex Measurements Stimuli: 250, 500, 1k, 2k, 4k, BBN, LBN, HBN, Click, External Input, Non-acoustic

Audiometer Control Panel – Brand Name or Equal Manufacturer: Madsen Model: ASTERA HIGH FREQUENCY 2

•Must be compatible with Otometrics
•PC based two channel audiometer-Two separate and identical channels
•Integrates with all Otosuite products, NOAH and ability to import data to CPRS
•Frequency range: TDH39 Earphones: 125-12500 Hz
•HDA200earphones: Standardfrequencies:125-20000 Hz
•Insert Earphones: 125-8000 Hz
•SF: Standard frequencies:125-20000 Hz
•Bone Conduction: 250-8000 Hz
•FRESH noise stimulus: Available in entire frequency range within the transducer specified range. (for SF125-12500 Hz). Accuracy0.3%
•Narrow Band Noise masking: Available in entire frequency range, Speech Weighted Noise, White Noise.
•Frequency resolution: 1/48,1/24,1/12, and 1/6oct,1 Hz step
•Stimulus types – Tone, Warble, Pulsed and Pulsed Warble
•Built in recorded calibrated speech material
•USB powered audiometer control panel for easy and fast operation
•1.3 kg (2.85 lbs.)
•5 Free Field outputs
•Mini XLR connectors
•105 dB HL output
•40 Watt built-in amplifier
•VRA controls
•Touch screen mode
•Cycle buttons
•Configurable Masking Assistant
•Unique tests
•CUNY Test
•MLST-A, MLST-C
•Dedicated tinnitus test modality
•Pediatric test modality
•SAL Test
•Automatic SNR calculation
•Premium QuickSIN
•Phonemes counting feature
•Tinnitus Assessment

Middle Ear Analyzer (Zodiac Diagnostic Stand Alone Tympanometer - Brand Name or Equal)

•Test Types: Tympanometry, Acoustic Reflex Threshold, Reflex Decay, Eustachian Tube Function (Intact and Perforated).
•Probe Tone:

226 Hz (85 dB SPL ± 1.5 dB) 678 Hz (72 dB SPL ± 1.5 dB) 1000 Hz (69 dB SPL ± 1.5 dB) Accuracy: ± 1% Harmonic Distortion: < 1%

• Admittance Measurements Range: 226 Hz (-10 to +10 mmho) 678 Hz (-21.0 to +21 mmho) 1000 Hz (-32.0 to +32 mmho) Sensitivity Scale: Auto Scales to appropriate range, manual selection in reflex modes.

Accuracy: (226 Hz):

Tymp Mode: ± 5% of reading or ± 0.1 mmho, whichever is greater Reflex Mode: ± 5% of reading or ± 0.02 mmho, whichever is greater

• Pressure Measurements (load volume of 0.2 to 7.0 ml) Range: Normal + +200 to -400 daPa / Wide = +400 to -600 daPa Accuracy: ± 10% of reading or ± 10 daPa, whichever is greater Sweep Rate: 12.5, 50.0, 200, 600, and 600/200 daPa and manual Sweep Accuracy: 10% of nominal rate Maximum limits (in 0.5cc cavity): -800 daPa and +600 daPa

• Reflex Measurements Stimuli: 250,500, 1k, 2k, 4k, BBN, LBN, HBN, Click, External Input, Non-acoustic Frequency Accuracy: ± 3% Harmonic Distortion (THD): < 5% Noise Signals: 3 dB bandwidths Low Band: 400 – 1,600 Hz High Band: 1,600 – 4,000 Hz Broad Band: 400 – 4,000 Hz Intensity Range: 35 to 120 dB HL Step Size: 5 dB, 1 dB, and 2 dB Calibration Accuracy: ± 3 dB Step Accuracy: ± 0.5 dB On/Off Ratio: 70 dB minimum Must be in compliance with ANSI S3.39, IEC 60601-1, ISO 389

•Aurical Probe Microphone Measurement w HI-PRO
oTechnical Data

Interface: Wireless Bluetooth data transfer to PC, version 2.0, + EDR, class 2 (10 meters/33ft).

Frequency resolution: 1/24 octave or 1/3 octave, in 4 channels Intensity accuracy: ± 3 dB Measurement intensity range: Probe microphone 40 - 130 dB SPL Probe microphone: Intensity: 40 - 130 dB Reference microphone: Intensity: 40 - 110 dB

Dynamic range: > 80 dB Calibration data: Stored electronically inside the probe Battery types: Rechargeable, Ni-MH, AA (R6) 1.2V, 1 pc. Can also be used with Alkaline AA (R6) 1.2V, 1 pc.

Estimated battery life: 5 hours of continuous use.

Probe Microphone Measurements module o Probe Microphone Measurements module Stimuli: Integrated custom stimuli, Pink Noise, White Noise, Speech ANSI, ISTS, ICRA, Rainbow Passage, Tone Sweep, FreeStyle sound library with favorites list, live, recorded signals, recordings, CD material (wav, mp3).

Tests available: REUR, REUG, REOR, REOG, REAR, REAG, REIG, Noise Reduction, RECD, Dynamic REM, FreeStyle.

Counseling and Simulations module o Counseling and Simulations module Functions: Hearing Loss Simulator, Hearing Instrument Simulator, Speech Mapping, Predicted Aided Audiogram.

Stimuli: Sound library, live, recorded signals, CD material (wav, mp3).

o Standards Real Ear Measurement: EN 61669, ISO 12124 Safety: IEC 60601-1, UL 60601-1, CAN/CSA -C22.2 NO 601.1-90 AURICAL FreeFit: IEC 60601-1, Internally Powered, Type BF, IPX0

EMC: IEC 60601-1-2, EN 300 328-2, EN 301 489-17

Required Individual Line Items & Quantity of Items

· (2 each) Trave, Zone 1

· (2 each) 1066 CD Quicksin Material, US

· (2 each) 1096 Zodiac SA Clinical w/o Printer with Classic Probe (US only)

· (2 each) 1096 Zodiac Insert Phone

· (2 each) NOAH System 4 License

· (6 each) Service, Hour

· (2 each) 1081, PMM CS, W HI-PRO US ONLY

· (2 each) 1066 MADSEN Astera 2, US Only

· (2 each) 1066, ACP ASTERA 2

· (2 each) KIT, TOUCHSCREEN, WIN7, HDMI

· (2 each) 1081, MOUNTING PLATE Optional Line Items

· Trade-In Items (include make, model and serial # of all items to be used as a trade-in) - *Consult Logistics

· Product Specifications (CD, PDF or Word)

· Discount for Volume Purchase

· Expendable Supplies – *Consult Logistics

· Maintenance and Full-Service Repair Contracts - *Consult BioMed

· Warranty Information - *Consult BioMed

· In Service Training

· BioMed Training – *Consult BioMed

· Installation and/or De-Installation Services – *Consult BioMed/Engineering/IRM as needed

· Delivery Requirements:

· All products should be ordered to arrive on or about January 1, 2024. Please include any shipping costs that are not included in the cost of the line items.

*= Required Field
Sources Sought Notice

Sources Sought Notice

File details come from the government source that posted it. Updated .