S06 AMENDMENT 36C26026Q0991 0002.pdf

PDF 139 KB Posted

Attached to
H266 Portland VAHCS Hood Certifications and Testing Services
Solicitation number
36C26026Q0991
Issued by

About this file

This is Amendment 0002 to Solicitation 36C26026Q0991, issued by NCO 20 Network Contracting Office in Vancouver, Washington on September 1, 2026. The amendment provides answers to ten questions submitted by offerors regarding the scope of work for biosafety cabinet and cleanroom testing and certification services.

Key technical clarifications include: the contract follows a previous 6-month bridge contract (36C26026P0357); two of 100 biological safety cabinets (BSCs) are located in Portland Pharmacy Chemo Mixing Suite, and two of 18 laminar flow hoods are in Portland IV Mixing with a third planned for Vancouver; pharmacy-related equipment is subject to USP 797/800 semi-annual testing requirements; the Vancouver IV Mixing suite contains one RABS (Restricted Access Barrier System) isolator requiring semi-annual testing; viable samples total 80 per semi-annual period; and monthly samples are to be divided across months rather than performed in full each month. Pharmacy cleanrooms have room-side HEPA filters with local injection capability for leak testing. The government confirmed it will not accept alternative industry accreditations (such as CETA/CNBT or NEBB) in lieu of the required NSF-49 accreditation, and contractors must demonstrate a minimum of 18 months of BSC testing experience. The response date remains unchanged, and all other terms and conditions of the solicitation remain in full force and effect.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

5. PROJECT NUMBER (if applicable)

CODE 7. ADMINISTERED BY

2. AMENDMENT/MODIFICATION NUMBER

CODE

6. ISSUED BY

8. NAME AND ADDRESS OF CONTRACTOR

4. REQUISITION/PURCHASE REQ. NUMBER 3. EFFECTIVE DATE

9A. AMENDMENT OF SOLICITATION NUMBER

9B. DATED

PAGE OF PAGES

10A. MODIFICATION OF CONTRACT/ORDER NUMBER

10B. DATED

BPA NO. 1. CONTRACT ID CODE

FACILITY CODE CODE

Offers must acknowledge receipt of this amendment prior to the hour and date specified in the solicitation or as amended, by one of the following methods:

The above numbered solicitation is amended as set forth in Item 14. The hour and date specified for receipt of Offers

E. IMPORTANT:

is extended,

(a) By completing Items 8 and 15, and returning __________ copies of the amendment; (b) By acknowledging receipt of this amendment on each copy of the offer submitted; or (c) By separate letter or electronic communication which includes a reference to the solicitation and amendment numbers. FAILURE OF YOUR

ACKNOWLEDGMENT TO BE RECEIVED AT THE PLACE DESIGNATED FOR THE RECEIPT OF OFFERS PRIOR TO THE HOUR AND DATE SPECIFIED MAY is not extended.

12. ACCOUNTING AND APPROPRIATION DATA

(REV. 11/2016)

is required to sign this document and return ___________ copies to the issuing office. is not, A. THIS CHANGE ORDER IS ISSUED PURSUANT TO: (Specify authority) THE CHANGES SET FORTH IN ITEM 14 ARE MADE IN THE CONTRACT ORDER NO. IN ITEM 10A.

15C. DATE SIGNED

B. THE ABOVE NUMBERED CONTRACT/ORDER IS MODIFIED TO REFLECT THE ADMINISTRATIVE CHANGES

SET FORTH IN ITEM 14, PURSUANT TO THE AUTHORITY OF FAR 43.103(b).

RESULT IN REJECTION OF YOUR OFFER. If by virtue of this amendment you desire to change an offer already submitted, such change may be made by letter or electronic communication, provided each letter or electronic communication makes reference to the solicitation and this amendment, and is received prior to the opening hour and date specified.

C. THIS SUPPLEMENTAL AGREEMENT IS ENTERED INTO PURSUANT TO AUTHORITY OF:

D. OTHER

Contractor

16C. DATE SIGNED

14. DESCRIPTION OF AMENDMENT/MODIFICATION

16B. UNITED STATES OF AMERICA

Except as provided herein, all terms and conditions of the document referenced in Item 9A or 10A, as heretofore changed, remains unchanged and in full force and effect.

15A. NAME AND TITLE OF SIGNER 16A. NAME AND TITLE OF CONTRACTING OFFICER

15B. CONTRACTOR/OFFEROR

STANDARD FORM 30 PREVIOUS EDITION NOT USABLE

Prescribed by GSA - FAR (48 CFR) 53.243

(Type or print) (Type or print)

(Organized by UCF section headings, including solicitation/contract subject matter where feasible.)

(Number, street, county, State and ZIP Code)

(If other than Item 6)

(Specify type of modification and authority)

(such as changes in paying office, appropriation date, etc.)

(If required)

(SEE ITEM 11)

(SEE ITEM 13)

(X)

CHECK

ONE

13. THIS ITEM APPLIES ONLY TO MODIFICATIONS OF CONTRACTS/ORDERS,

IT MODIFIES THE CONTRACT/ORDER NO. AS DESCRIBED IN ITEM 14.

11. THIS ITEM ONLY APPLIES TO AMENDMENTS OF SOLICITATIONS

AMENDMENT OF SOLICITATION/MODIFICATION OF CONTRACT

(Signature of person authorized to sign) (Signature of Contracting Officer)

1 3

0002 09-01-2026

36C260

NCO 20 Network Contracting Office

ATTN: V4CONT

1601 E. Fourth Plain Blvd

Bldg.17, Suite B428

Vancouver WA 98661

36C260

NCO 20 Network Contracting Office

ATTN: V4CONT

1601 E. Fourth Plain Blvd

Bldg.17, Suite B428

Vancouver WA 98661

To all Offerors/Bidders

36C26026Q0991

X

X X

X 1

1. The purpose of this amendment is to provide answers to questions received.

2. See continuation page below for Q&A 1-10 below.

3. No other answers will be provided at this time. Response date remains unchanged.

4. No other changes. All other terms and conditions remain unchanged.

CONTINUATION PAGE

Q&A 1-10

Q.1 We would like to know if this is a brand new contract OR if there is (was) an incumbent performing these services. If not brand new, could you please provide the current / previous contract number?

A.1 Previous contract 36C26026P0357 (6-month bridge).

Q.2 Are any of the listed 100 BSCs, Pharmacy usage?

A.2 Yes – 2 are BSCs located in Portland Pharmacy Chemo Mixing Suite.

Q.3 Are any of the listed 18 Laminar Flow Hoods, Pharmacy usage?

A.3 Yes – 2 are Laminar flow in Portland IV Mixing. There will be a 3rd added in

Vancouver in the next year.

Q.4 All Pharmacy related Hoods and Cleanrooms are required by USP 797/800, to be tested semi-annually. Please clarify, is there one set of Pharmacy Cleanrooms – x1 Mix (Non-HD

Sterile Compounding)Room, x1 Ante Mix ( Sterile Anteroom), x1 Chemo ( HD Sterile

Compounding) Room and x1 Ante Chemo ( Sterile Anteroom) for a total of four rooms tested semi-annually or eight rooms – two sets of rooms tested semi-annually? Also, which hoods, primary engineering controls, are in the rooms?

A.4 Portland has a Chemo Compounding Suite, IV Mixing Suite, and both rooms have ante rooms. Vancouver has an IV Mixing Suite with a glove box hood and will be remodeled to have a Laminar Flow Hood in the future.

Portland IV:

• LAFW (laminar airflow workbench): BAKER EG-6320 S/N: 109320 (6 ft)

• LAFW (laminar airflow workbench): BAKER EG-6320 S/N: 109328 (6 ft)

Portland Chemo:

• BSC (biological safety cabinet): 109316 (BAKER SG403A-HE) (4ft)

• BSC (biological safety cabinet): 109315 (BAKER SG403A-HE) (4ft)

Vancouver:

• RABS (Restricted Access Barrier System) isolator -MIC Single: 0910-EMPAB-

• Future: LAFW (laminar airflow workbench)

Q.5 There are two Isolator units in the list, are these Pharmacy hoods? Are these to be tested semi-annually? Are these standard CAI/CACI Isolators? Can you provide documentation for these?

A.5 The isolator is in the Vancouver IV Mixing suite and needing 1 test and certification

2x (Semi-annual).

Q.6 Line 0007: Is 160 the number of viable samples to be performed total for the year ( 80 semi-annual) or 160 every six months.

A.6 80 Semi Annual.

Q.7 Line 0008: Is this the monthly trip or number of samples?

A.7 Number of sampling instances.

Q.8 Lines 0012, 0013 0014: Monthly samples are listed. Is this total to be performed every month or total to be divided by months?

A.8 Divided by months.

Q.9 Do Pharmacy Cleanrooms have HEPAs room side in the ceiling? Is there a system to locally inject the HEPAs for leak testing?

A.9 Yes, cleanrooms have room side HEPAs and can be locally injected for leak test.

Q.10 In review of the Scope of Work, Section 3 – Qualifications, it states that the contractor shall be NSF-49 accredited and have a minimum of 18 months of BSC testing experience.

Would the Government accept other recognized industry accreditations or certifications, such as

CETA/CNBT or NEBB, as an equivalent alternative to the NSF-49 accreditation requirement, provided the contractor can demonstrate the required BSC testing experience and capability to perform all specified testing and certification procedures?

A.10 Not at this time.