36C26025Q0178.docx
DOCX document 206 KB Posted
- Attached to
- 6640--FY25 – Spokane – CPRR Clinical Chemistry Laboratory Equipment Base Plus 4 Procurement
- Solicitation number
- 36C26025Q0178
About this file
This is a Request for Quotes (RFQ) issued by the VA Network Contracting Office 20 for clinical chemistry laboratory equipment and services at the Spokane VA Medical Center. The solicitation (36C26025Q0178) seeks a contractor to provide laboratory analyzers, reagents, controls, supplies, maintenance, and support services on a Cost Per Reportable Result (CPRR) basis for a base year (4/1/2025-3/31/2026) plus four one-year options through 3/31/2030.
Key requirements include providing random access chemistry/immunochemistry instruments with 98% uptime, continuous inventory of reagents and supplies, 24/7 technical support with 1-hour callback response, repair service within 24 hours, preventive maintenance, operator training, and middleware that interfaces with VA systems. The contractor must provide validation studies, correlation analysis, and quarterly sales reports. Quotes are due February 7, 2025 at 17:00 PST. This is an unrestricted Federal Supply Schedule procurement under NAICS 325413 with a $41.5M size standard. Technical factors are weighted more heavily than price in the evaluation criteria. The equipment must maintain barcode reader accuracy of less than 1% failure rate and meet Clinical Laboratory Improvement Act (CLIA) standards.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C26025Q0178 0003.docx | DOCX document | |
| Spokane Laboaratroy Layout.jpg | JPG image | |
| Edited 2-5-2025-36C26025Q0178.docx | DOCX document | |
| 36C26025Q0178 0002.docx | DOCX document | |
| Attachment A Spokane Estimated Test Volume.xlsx | XLSX spreadsheet | |
| Mann-Grandstaff Architectural Plans.pdf | ||
| Detailed counter dec 23 to dec 24.xlsx | XLSX spreadsheet | |
| Edited 1-30-2025-36C26025Q0178.docx | DOCX document | |
| Laboratory.png | PNG image | |
| 36C26025Q0178 0001.docx | DOCX document | |
| Attachment A Estimated Test Volume Spokane.xlsx | XLSX spreadsheet |
Show all 11
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
36C26025Q0178
PAGE 1 OF
1. REQUISITION NO.
2. CONTRACT NO.
3. AWARD/EFFECTIVE DATE
4. ORDER NO.
5. SOLICITATION NUMBER
6. SOLICITATION ISSUE DATE
a. NAME
b. TELEPHONE NO. (No Collect Calls)
8. OFFER DUE DATE/LOCAL
TIME
9. ISSUED BY
CODE
10. THIS ACQUISITION IS
UNRESTRICTED OR
SET ASIDE:
% FOR:
SMALL BUSINESS
HUBZONE SMALL
BUSINESS
SERVICE-DISABLED
VETERAN-OWNED
SMALL BUSINESS
WOMEN-OWNED SMALL BUSINESS
(WOSB) ELIGIBLE UNDER THE WOMEN-OWNED
SMALL BUSINESS PROGRAM
EDWOSB
8(A)
NAICS:
SIZE STANDARD:
11. DELIVERY FOR FOB DESTINA-
TION UNLESS BLOCK IS
MARKED
SEE SCHEDULE
12. DISCOUNT TERMS
13a. THIS CONTRACT IS A
RATED ORDER UNDER
DPAS (15 CFR 700)
13b. RATING
14. METHOD OF SOLICITATION
RFQ
IFB
RFP
15. DELIVER TO
CODE
16. ADMINISTERED BY
CODE
17a. CONTRACTOR/OFFEROR
CODE
FACILITY CODE
18a. PAYMENT WILL BE MADE BY
CODE
TELEPHONE NO.
UEI:
EFT:
PHONE:
FAX:
17b. CHECK IF REMITTANCE IS DIFFERENT AND PUT SUCH ADDRESS IN OFFER 18b. SUBMIT INVOICES TO ADDRESS SHOWN IN BLOCK 18a UNLESS BLOCK BELOW IS CHECKED
SEE ADDENDUM
19.
20.
21.
22.
23.
24.
ITEM NO.
SCHEDULE OF SUPPLIES/SERVICES
QUANTITY
UNIT
UNIT PRICE
AMOUNT
(Use Reverse and/or Attach Additional Sheets as Necessary)
25. ACCOUNTING AND APPROPRIATION DATA
26. TOTAL AWARD AMOUNT (For Govt. Use Only) 27a. SOLICITATION INCORPORATES BY REFERENCE FAR 52.212-1, 52.212-4. FAR 52.212-3 AND 52.212-5 ARE ATTACHED. ADDENDA
ARE
ARE NOT ATTACHED.
27b. CONTRACT/PURCHASE ORDER INCORPORATES BY REFERENCE FAR 52.212-4. FAR 52.212-5 IS ATTACHED. ADDENDA
ARE
ARE NOT ATTACHED
28. CONTRACTOR IS REQUIRED TO SIGN THIS DOCUMENT AND RETURN _______________
29. AWARD OF CONTRACT: REF. ___________________________________ OFFER
COPIES TO ISSUING OFFICE. CONTRACTOR AGREES TO FURNISH AND
DATED ________________________________. YOUR OFFER ON SOLICITATION
DELIVER ALL ITEMS SET FORTH OR OTHERWISE IDENTIFIED ABOVE AND ON ANY
(BLOCK 5), INCLUDING ANY ADDITIONS OR CHANGES WHICH ARE
ADDITIONAL SHEETS SUBJECT TO THE TERMS AND CONDITIONS SPECIFIED
SET FORTH HEREIN IS ACCEPTED AS TO ITEMS:
30a. SIGNATURE OF OFFEROR/CONTRACTOR 31a. UNITED STATES OF AMERICA (SIGNATURE OF CONTRACTING OFFICER) 30b. NAME AND TITLE OF SIGNER (TYPE OR PRINT) 30c. DATE SIGNED 31b. NAME OF CONTRACTING OFFICER (TYPE OR PRINT) 31c. DATE SIGNED
AUTHORIZED FOR LOCAL REPRODUCTION
(REV. NOV 2021)
PREVIOUS EDITION IS NOT USABLE
Prescribed by GSA - FAR (48 CFR) 53.212
7. FOR SOLICITATION
INFORMATION CALL:
STANDARD FORM 1449
SOLICITATION/CONTRACT/ORDER FOR COMMERCIAL PRODUCTS AND COMMERCIAL SERVICES
OFFEROR TO COMPLETE BLOCKS 12, 17, 23, 24, & 30
01-24-2025 Hector Gonzalez 360-553-7632 02-07-2025 17:00
PST
36C260 Network Contracting Office 20
ATTN: V4CONT
1601 E. Fourth Plain Blvd Bldg.17, Suite B428 Vancouver WA 98661
X
325413 $41.5 Million
N/A
36C668 Department of Veterans Affairs Spokane VAMC
4815 N Assembly Street Spokane WA 99205 36C260 Network Contracting Office 20
ATTN: V4CONT
1601 E. Fourth Plain Blvd Bldg.17, Suite B428 Vancouver WA 98661
Dept of Veterans Affairs
FMS-VA-2(101)
Financial Services Center
PO BOX 149971
Austin TX 78714-9971
See CONTINUATION Page
a. Location: Spokane VA Medical Center.
b. Supplies Being Solicited: Clinical Laboratory Equipment.
Cost per reportable result.
c. This is a Base Plus Four Federal Supply Schedule Order.
Base Year: 04/01/2025 - 03/31/2026 Option Year One (1): 04/01/2026 - 03/31/2027 Option Year Two (2): 04/01/2027 - 03/31/2028 Option Year Three (3): 04/01/2028 - 03/31/2029 Option Year Four (4): 04/01/2029 - 03/31/2030
d. Please see the Continuation Pages below for details regarding the submission of quotes.
e. For further information, please contact Hector Gonzalez at hector.gonzalez1@va.gov.
See CONTINUATION Page
One
Jose Ronstadt Contracting Officer
Table of Contents
| SECTION A | 1 |
| A.1 SF 1449 SOLICITATION/CONTRACT/ORDER FOR COMMERCIAL PRODUCTS AND COMMERCIAL SERVICES | 1 |
| SECTION B - CONTINUATION OF SF 1449 BLOCKS | 3 |
| B.1 CONTRACT ADMINISTRATION DATA | 3 |
| B.2 PRICE/COST SCHEDULE | 4 |
| ITEM INFORMATION | 4 |
| B.3 DELIVERY SCHEDULE | 5 |
| B.4 STATEMENT OF WORK | 6 |
| SECTION C - CONTRACT CLAUSES | 24 |
| C.1 FSS RFQ INTRODUCTORY LANGUAGE | 24 |
| C.2 52.252-2 CLAUSES INCORPORATED BY REFERENCE (FEB 1998) | 24 |
| C.3 52.217-8 OPTION TO EXTEND SERVICES (NOV 1999) | 24 |
| C.4 52.217-9 OPTION TO EXTEND THE TERM OF THE CONTRACT (MAR 2000) | 24 |
| C.5 VAAR 852.208-70 SERVICE-DISABLED VETERAN-OWNED AND VETERAN-OWNED SMALL BUSINESS EVALUATION FACTORS – ORDERS or BPAs (JAN 2023) (DEVIATION) | 25 |
| C.6 VAAR 852.208-71 SERVICE-DISABLED VETERAN-OWNED AND VETERAN-OWNED SMALL BUSINESS EVALUATION FACTOR COMMITMENTS-ORDERS AND BPAS (JAN 2023) (DEVIATION) | 26 |
| C.7 VAAR 852.215-70 SERVICE-DISABLED VETERAN-OWNED AND VETERAN-OWNED SMALL BUSINESS EVALUATION FACTORS (OCT 2019) | 26 |
| C.8 VAAR 852.219-70 VA SMALL BUSINESS SUBCONTRACTING PLAN MINIMUM REQUIREMENTS (NOV 2022) | 27 |
| C.9 VAAR 852.242-71 ADMINISTRATIVE CONTRACTING OFFICER (OCT 2020) | 27 |
| C.10 VAAR 852.232-72 ELECTRONIC SUBMISSION OF PAYMENT REQUESTS (NOV 2018) | 28 |
| SECTION D - CONTRACT DOCUMENTS, EXHIBITS, OR ATTACHMENTS | 30 |
| SECTION E - SOLICITATION PROVISIONS | 31 |
| E.1 52.204-24 REPRESENTATION REGARDING CERTAIN TELECOMMUNICATIONS AND VIDEO SURVEILLANCE SERVICES OR EQUIPMENT (NOV 2021) | 31 |
| E.2 52.252-1 SOLICITATION PROVISIONS INCORPORATED BY REFERENCE (FEB 1998) | 33 |
| E.3 52.212-2 EVALUATION—COMMERCIAL PRODUCTS AND COMMERCIAL SERVICES (NOV 2021) | 34 |
SECTION B - CONTINUATION OF SF 1449 BLOCKS
B.1 CONTRACT ADMINISTRATION DATA
1. Contract Administration: All contract administration matters will be handled by the following individuals:
a. CONTRACTOR:
b. GOVERNMENT: Contracting Officer 36C260 Network Contracting Office 20
ATTN: V4CONT
1601 E. Fourth Plain Blvd Bldg.17, Suite B428 Vancouver WA 98661
2. CONTRACTOR REMITTANCE ADDRESS: All payments by the Government to the contractor will be made in accordance with:
| [X] |
| 52.232-33, Payment by Electronic Funds Transfer—System For Award Management, or |
| [] |
| 52.232-36, Payment by Third Party |
3. INVOICES: Invoices shall be submitted in arrears:
| a. Quarterly | [] |
| b. Semi-Annually | [] |
| c. Other | [X] Monthly |
4. GOVERNMENT INVOICE ADDRESS: All Invoices from the contractor shall be submitted electronically in accordance with VAAR Clause 852.232-72 Electronic Submission of Payment Requests.
https://www.fsc.va.gov/einvoice.asp
ACKNOWLEDGMENT OF AMENDMENTS: The offeror acknowledges receipt of amendments to the Solicitation numbered and dated as follows:
| AMENDMENT NO |
| DATE |
B.2 PRICE/COST SCHEDULE
ITEM INFORMATION
| ITEM NUMBER |
| DESCRIPTION OF SUPPLIES/SERVICES |
| QUANTITY |
| UNIT |
| UNIT PRICE |
| AMOUNT |
| 1.00 |
| YR |
| __________________ |
| ______________ |
Cost Per Reportable Result FSS Order for Clinical Chemistry Laboratory Equipment.
Please see Attachment A for list of Tests and Reagents.
Contract Period: Base POP Begin: 04-01-2025 POP End: 03-31-2026 PRINCIPAL NAICS CODE: 325413 - In-Vitro Diagnostic Substance Manufacturing PRODUCT/SERVICE CODE: 6640 - Laboratory Equipment and Supplies
| 1.00 |
| YR |
| __________________ |
| ______________ |
Cost Per Reportable Result FSS Order for Clinical Chemistry Laboratory Equipment.
Contract Period: Option 1 POP Begin: 04-01-2026 POP End: 03-31-2027
| 1.00 |
| YR |
| __________________ |
| ______________ |
Cost Per Reportable Result FSS Order for Clinical Chemistry Laboratory Equipment.
Contract Period: Option 2 POP Begin: 04-01-2027 POP End: 03-31-2028
| 1.00 |
| YR |
| __________________ |
| ______________ |
Cost Per Reportable Result FSS Order for Clinical Chemistry Laboratory Equipment.
Contract Period: Option 3 POP Begin: 04-01-2028 POP End: 03-31-2029
| 1.00 |
| YR |
| __________________ |
| ______________ |
Cost Per Reportable Result FSS Order for Clinical Chemistry Laboratory Equipment.
Contract Period: Option 4 POP Begin: 04-01-2029 POP End: 03-31-2030
| GRAND TOTAL |
| ______________ |
B.3 DELIVERY SCHEDULE
| ITEM NUMBER |
| SHIPPING INFORMATION |
| QUANTITY |
| DELIVERY DATE |
| ALL |
| SHIP TO: |
| SPOKANE VAMC |
4815 N ASSEMBLY STREET
SPOKANE, WA 99205 6185
USA
| ALL |
| 4/1/25-3/31/30 |
Page 1 of
B.4 STATEMENT OF WORK
SCOPE OF PROCUREMENT:
1.1.1. The Clinical Laboratory Chemistry/Immunochemistry instrument shall have random access capability (if discrete testing is required) and be able to simultaneously perform the complete profile as described below meeting the performance characteristics for accuracy and precision as defined by the 1988 Clinical Laboratory Improvement Act (CLIA) and the Clinical and Laboratory Standards Institute (CLSI).
1.1.2. The Contractor is required to provide a continuously stocked inventory of reagents, standards, Vendor provided Controls, supplies, disposables, and any other materials, to include replacement parts, required to properly perform tests on the equipment such that equipment operations are not interrupted. These items shall be of the highest quality, sensitivity, specificity and tested to assure precision and accuracy. Expiration date must be clearly marked on reagent, standards, and control containers. Unexpected changes in methodology/technology shall be at the expense of the Contractor. Alert/Notification of any delays in shipment as well as any or all technical advisory/recalls/alerts, prior to or simultaneously with field alerts should be forwarded to the designated individuals determined at contract award.
1.1.3. Contractor is required to provide vendor controls for troubleshooting and/or routine use as needed.
1.1.4. The Contractor will provide a detailed list of all parts and supplies that are to be purchased by the facility and are not included in the CPRR contract.
1.1.5. Special handling for emergency orders of supplies: In the event that the supplies are found to be defective and unsuitable for use with the Contractor’s equipment, or the Contractor has failed to comply with the requirements for routine supply delivery, the Contractor is required to deliver the supplies within 24 hours of receipt of a verbal order for emergency delivery. If either circumstance has occurred, the Contractor shall deliver to the Government site in the most expeditious manner possible without additional cost to the Government, the necessary consumables in sufficient quantity as required to allow operation of the Contractor’s equipment for one week (under normal Government test load volume). If additional requests for emergency supply delivery are required by the Government, they shall be honored by the Contractor until the arrival at the laboratory of the monthly standing order/routine supplies delivery.
1.2. DEFINITIONS:
1.2.1. Cost per Patient Reportable Result (CPRR)- as defined in the Federal Supply Schedule FSC Group 66, Part III, Cost-Per-Test Clinical Laboratory Analyzers. The per patient reportable result price shall include costs covering five-year equipment use, all reagents, standards, vendor provided quality controls, supplies, consumable/disposable items, parts, accessories, and any other item required for the proper operation of the Contractor’s equipment and necessary for the generation of a patient reportable result. This per patient reportable result price shall also encompass all costs associated with dilution; repeat and confirmatory testing required to produce a single patient reportable result. It shall also include the material to perform as well as all other costs associated with quality control, calibration and correlation study testing that is prescribed by the Clinical and Laboratory Standards Institute (CLSI). All necessary maintenance to keep the equipment in good operating condition (This element includes both preventive maintenance and emergency repairs) and training for Government personnel. Contractors shall provide delivery, installation, and removal of equipment at no additional charge.
1.2.2. Cost per Test (CPT)- as defined in the Federal Supply Schedule FSC Group 66, Part III, Cost-Per-Test Clinical Laboratory Analyzers – Contractors are required to provide a price for each test that can be performed on its equipment. The per test price shall include costs covering five-year equipment use including the following: All reagents, standards, quality Vendor provided Controls, linearity material, calibration verification material, supplies, consumable/disposable items, parts, accessories. Including any other item required for the proper operation of the Contractor’s equipment and necessary for the generation and reporting of a test result. All necessary maintenance to keep the equipment in good operating condition (This element includes both preventive maintenance and emergency repairs) and training for Government personnel. Contractors are required to provide delivery, installation, and removal of equipment at no additional charge.
1.2.3. Business Associate Agreement (BAA)- A business associate is an entity, including an individual, company, or organization that, on behalf of VHA, performs or assists in the performance of functions or activities involving the use or disclosure of PHI, or that provides certain services involving the disclosure of protected health information (PHI). VHA is a covered entity under the HIPAA Privacy Rule (Privacy Rule). HIPAA regulations require VHA to execute HIPAA-compliant BAAs with certain entities that receives, uses, or discloses VHA PHI to perform some activity for VHA. These BAAs obligate VHA business associates to provide the same protections and safeguards to PHI that is required of VHA under the Privacy Rule.
1.2.4. Specimen Processing Automation Line – Pre-analytical processing equipment offered to that will automate these specimen processing functions, as indicated in the general requirements section.
1.2.5. Specimen Management System - A component of the Processing Automation Line that directs and manages the operation and components of the pre-analytical processing/automation system. A component of specimen archival post-analytical processing/automation system (applies only if selected).
1.2.6. Contractor Middleware Management System - A server installed with software that interfaces the testing instrumentation to the Laboratory Information system and is able to receive, process and send data following CLSI guidelines.
1.2.7 Throughput – The speed that the equipment processes and/or operates reported in units per hour.
1.2.7.1 Installation (applies only if selected): Contractor will provide an alternate site for validation of the instrumentation if space is not available for side-by-side validation with existing instrumentation. Contractor will perform, at their cost, new validation studies as defined by the customer when instrumentation is installed in the final workspace.
1.2.7.2 Provide the same type of coverage on the following (current) instrumentation until the new instrumentation has been installed and validated.
GENERAL REQUIREMENTS:
1.2.8. Primary analyzer(s) – Base equipment offered shall be new and shall fully support the scope of operations (minimal requirements). Depending upon the technical functionality and the capabilities of the individual manufacturer’s instrumentation, one analyzer or multiple analyzers may be required to meet the productivity specifications defined herein. Analyzers offered per facility shall provide sufficient capacity and throughput to meet the volume and service demands in those instances, the additional analyzer(s) shall, likewise, be considered primary and shall meet all of the technical specifications of this solicitation. Those additional analyzer(s) offered meeting the definition of a primary analyzer may serve as a back-up analyzer (see definition below) and shall replace the requirement for offering that category of equipment.
1.2.8.1. (Applies only if selected) Back-up Analyzer; equipment required in support of operations for the VA laboratories in the event the primary analyzer(s) becomes non-operational/non-functional. In the instance that the mirror imaged analyzers become unavailable the back-up analyzer will be used. As such, the requirements for consumable supplies, i.e. reagents, quality control material, calibrators, etc., shall be minimal and corollary to the successful operation of the primary instrumentation. Specific tests that require back-up performance. Additional primary analyzers required for the performance of daily workload are not considered back-ups for the purposes of consumables, reagents, etc.
1.2.8.2. Installation (applies only if selected): Contractor will provide an alternate site for validation of the instrumentation if space is not available for side-by-side validation with existing instrumentation. Contractor will perform, at their cost, new validation studies as defined by the customer when instrumentation is installed in the final workspace.
1.2.9. Operational and Technical Features- The instrumentation offered shall be approved by the Food and Drug Administration (FDA) and Contractor’s FSS Contract at the time of proposal submission and have the following:
1.2.9.1. Primary Processing Automation Line Instrumentation. Offering for an Integrated Clinical Laboratory Chemistry/Immunochemistry Instrumentation and Robotics System (Automated track or line) in accordance with the requirements of each respective laboratory shall have the following listed below. Please include a clear description regarding how samples are transported on the rack.
1.2.9.1.1. Sufficient capacity and throughput to meet the volume and service demands as indicated in Attachment A.
1.2.9.1.2. Includes a specimen management system to manage and track sample progress and position.
1.2.9.1.3. Specimen archival through mapping of specimen location; for easy retrieval once moved off-line (applies only if selected).
1.2.9.1.4. The ability, based on test requests, to sort and route specimens. (Line/track system/Specimen Management System)
1.2.9.1.5. The ability to send the primary tube or processed aliquots by means of a tracking system to the proper testing instrumentation to maximize efficiency and to maintain and standardize turnaround times of results. (Line/track system/Specimen Management System)
1.2.9.1.6. Minimal operator intervention to introduce STAT specimens or to change a routine specimen priority.
1.2.9.1.7. The ability to detect processing errors and provide error notification. (Specimen Management System)
1.2.9.1.8. The ability to separate the serum/plasma from the blood cells through the process of centrifugation. (Centrifuge)
1.2.9.1.9. Ability to detect short samples and clots (Aliquotter)
1.2.9.1.10. The ability to remove caps from blood collection tubes. (Decapper)
1.2.9.1.11. The ability to recap and/or reseal - tubes. (Recapper)
1.2.9.1.12. The ability to remove an aliquot of serum/plasma using disposable pipette tips to prevent contamination and move into another specimen tube or cup for analysis. (Aliquotter)
1.2.9.1.13. The ability to print bar code labels and label daughter (aliquot) tubes to maintain positive specimen identification.
1.2.9.1.14. The ability to connect all primary and back-up testing analyzers.
1.2.9.1.15. Ensure immediate availability of primary tube after initial testing is completed.
1.2.9.1.16. The ability to auto-archive, store and retrieve specimens.
1.2.9.1.17. Barcode reader stations must have the following capabilities:
1.2.9.1.17.1. The accuracy of the barcode reading must have less than a 1.0 % failure rate.
1.2.9.1.17.2. Equipment must be able to support multiple barcode formats (Code 39, Code 128) that may be enabled concurrently.
1.2.9.1.17.3. Equipment must accept, at a minimum, 10 characters in specimen identifier that is alpha, numeric, and/or alphanumeric concurrently.
1.2.9.2. Testing Instrumentation: The testing instrumentation must be an integrated platform with flexible, components that can be upgraded and or reconfigured on site and is approved by the Food and Drug Administration (FDA) and on Contractors’ FSS Contract at the time of proposal submission and shall have the following:
1.2.9.2.1. Sufficient capacity and throughput to meet the volume and service demands as indicated in Attachment A.
1.2.9.2.2. An instrument management system (internal to testing instrumentation) that provides/maintains the following:
1.2.9.2.3. Ability to monitor instrument performance.
1.2.9.2.3.1. Continuous monitoring of vital functions with immediate operator notification of failure(s) and on-board storage of these records.
1.2.9.2.3.2. Capability to detect and alert operator of out-of-range quality control results via flagged results on QC printout and visual alerts on display monitor.
1.2.9.2.3.3. Ability to store and retransmit patient records to the VA Laboratory Information system for a minimum 48 hours in case of interface outage.
1.2.9.2.3.4. Capability to record, store and print the following information:
1.2.9.2.3.4.1. Required quality control and instrument maintenance information.
1.2.9.2.3.4.2. Patient demographic information and specimen results.
1.2.9.2.3.5. On board reagent inventory system.
1.2.9.2.4. A bi-directional, bar-coded computer interface compatible with the current VA laboratory information system. The fully operational interface (both hardware and software) shall be immediately available for implementation to the VA computerized hospital information system.
1.2.9.2.4.1. The accuracy of the barcode reading must have less than a 1.0 % failure rate.
1.2.9.2.4.2. Equipment must be able to support multiple barcode formats (Code 39, Code 128) that may be enabled concurrently.
1.2.9.2.4.3. Equipment must accept, at a minimum, 10 characters in specimen identifier that is alpha and/or numeric depending on site that may be enabled concurrently.
1.2.9.3. Bar coding of reagents and the ability to track reagent containers throughout the testing process using bar code technology.
1.2.9.4. Ability to prioritize STAT testing without compromising existing programmed testing.
1.2.9.5. Minimal operator intervention to introduce STAT specimens or to change a routine specimen to a STAT specimen, as well as introduce STAT specimens during a test run without aborting a run. Vendor will outline STAT procedure.
1.2.9.6. On board reagent stability, sufficient to accommodate both high and low volume use.
1.2.9.7. The ability to detect and alert operator of low reagent levels and the potential of depletion.
1.2.9.8. The ability to load and unload all reagents from the equipment during operation without interrupting testing in progress or have sufficient capacity to enable a complete 24 hours of testing.
1.2.9.9. The ability to support multiple reagent lots of the same reagent on the equipment at the same time with active, valid calibrations.
1.2.9.10. The ability to maintain multiple calibrations per analyte.
1.2.9.11. The capability to calibrate assays during test run without aborting the run.
1.2.9.12. The capabilities to store, print, and retrieve calibration data.
1.2.9.13. For routine (general) chemistry tests, when more than one lot of a given reagent has a valid calibration and quality control material is programmed to run as a control (in the control mode):
1.2.9.13.1.1. Facility shall be notified if a new lot number is about to be run to ensure that quality control material is performed.
1.2.9.13.1.2. Quality control results will be easily distinguishable i.e., identified by reagent lot number or similar mechanism, on instrument printout or display monitor.
1.2.9.13.1.3. Operator may select to run a test on only a specified lot of reagent even though more than one lot has a valid calibration
1.2.9.14. The ability to continuously load patient specimens.
1.2.9.15. The ability to detect short samples and interfering indices i.e. hemolysis, lipemia, and icterus, on all systems including Chemistry, Immunochemistry, and pre-analytics.
1.2.9.16. Clot detection with alert notification.
1.2.9.17. Primary tube sampling from various manufacturers and sizes of evacuated tubes.
1.2.9.18. The instrumentation shall be capable of handling all routine sample collection tubes e.g.: 13x75, 13x100, 16x100. plus, other various sized sample containers i.e., sample cups (0.5, 1.0, and 2.0 ml), carrier tubes and tube inserts.
1.2.9.19. Automatically repeat testing when defined limits are exceeded via auto dilution or sample volume adjustment.
1.2.9.20. The capability to program a test as a repeat with interfacing of results to overlay initial result.
1.2.9.21. Safety features to avoid unnecessary exposure to biohazardous and chemical material. The exposure to and the volume of biohazardous and chemical material generated by the equipment must be minimal and require a minimum amount of handling.
1.2.9.22. For those sites requiring back up analyzers, it is desirable for the backup analyzer to be a mirror image or have the same reagent requirement as the primary analyzer.
1.2.9.22.1. Ability to store and retransmit records (48 hours of maximal instrument throughput) in case of interface outage.
1.2.9.23. Calibration stability adheres to manufacture’s specifications.
1.2.9.24. User defined/Open Channel test capability. In the event that the contractor cannot provide desired assays, the contractor will provide application development support for open channel test implementation.
1.2.10. Contractor Middleware Management System
1.2.10.1. The Contractor shall provide instrumentation and middleware that is compatible with VA’s Electronic Health Record or interfaces directly with Data Innovations Middleware.
1.2.10.2. The Middleware shall be able to communicate via HL7 messages version 1.5 or later.
1.2.10.3. Middleware for clinical chemistry systems should be open for bidirectional interfacing of front-end automation, reference laboratories, and laboratory /clinical information systems, third party middleware applications, and laboratory instrumentation, current and future generations, for all other vendors and/or laboratory sections at no additional cost.
1.2.10.4. The Contractor shall provide the server(s) and Data Innovations’ Instrument Manager Middleware software or other middleware that is compatible with VA’s Electronic Health Record and provide technical support for both the hardware and software for the duration of the contract.
1.2.10.5. All driver development shall be provided by the respective vender, free-of-charge.
1.2.10.6. The middleware should have the ability to interface with regulatory agency applications. i.e. CAP.
1.2.10.7. The Contractor shall assist in the provision of the appropriate software drivers and interfaces for effective use of automated regulatory agency reporting features.
1.2.10.8. The Contractor shall assist in the establishment, testing, deployment, and troubleshooting of the full capacity of the Middleware before implementation date.
1.2.10.9. The Middleware shall contain systems that facilitate the analysis of laboratory pre-analytic, analytic, and post-analytic processes and the generation of summary reports for quality improvement and monitoring efforts.
1.2.10.10. Summary reports shall be interactive, customizable, and accessible on-demand.
1.2.10.11. Reports shall have, at a minimum, the flexibility to display information organized by department, workflow unit, desired time frame, production phase and priority. Software shall have the capability to drill down within the summary reports, identify problematic data and generate corrective action plans.
1.2.10.12. Canned summary reports shall include, at a minimum: Turnaround time, specimen and test volumes, instrument quality control, specimen quality indicators, and % auto-verification, % (and number) repeat tests.
1.2.10.13. The Middleware shall include moving averages system. The Contractor shall establish, test, deploy and troubleshoot the functionality of the moving averages system.
1.2.10.14. The patient moving average system should have real-time, failure notification capability and an auto verification interrupt function at the analyte, specimen, and patient or instrument level.
1.2.10.15. The Middleware shall include Boolean logic rule writing applications and vendor developed drivers or functionality that enable the use auto-validation/auto-verification in accordance with CAP regulations.
1.2.10.16. The Middleware shall facilitate the development of compound, nested rules with multiple event actions.
1.2.10.17. The middleware rule-writing application should have a visual (point-&-click) graphical user interface consistent with Microsoft Windows application and not require complex programming or coding. It should provide sufficient data elements & granularity so criteria can be defined for patient, specimen, and test-level conditions.
1.2.10.18. The Middleware shall be able to query incoming orders and outgoing results and hold either for user review and action.
1.2.10.19. The Middleware shall contain a specimen management system that allows the user to quickly locate any specimen in the system.
1.2.10.20. The Middleware shall have the ability to transmit QC data directly to Bio-Rad Unity Real Time or any other comparable online Quality Control program.
1.2.10.21. The software shall have a system for inventory management that facilitates the tracking of consumables and ordering of supplies, and include pertinent information about the consumables, including, lot numbers, expiration dates, and item numbers. If required, management system is available on board the analyzer, and in the storage of room temperature and refrigerated reagents.
1.2.10.22. The server(s) must have sufficient memory to store all middleware records for a minimum of 14 days with downloading capability to an external medium for long term storage of patient records and other information.
1.2.10.23. The vendor shall provide an accessible online medium for long term storage for patient records, results, and other information with storage capacity to maintain 2 years’ worth of information. Information should be stored in a format that is readable by standard software.
1.2.10.24. Patient status display for technologist review and workflow management for all integrated (linked) testing instrumentation.
1.2.10.25. Ability to retransmit patient records to universal interface system in case of interface outage.
1.2.10.26. Technology to automatically repeat testing based on customer configurable testing criteria (repeat testing)
1.2.10.27. Technology to automatically direct additional specimen testing based on customer-configurable testing criteria (reflex technology).
1.2.10.28. Vendor shall provide middleware and server, a backup server capable of taking over system functionality in event of primary server failure is provided at no cost.
1.2.10.29. Test Environment or Test Instance of the Middleware in order to be utilized as an isolated environment for testing new systems, configuration changes, instrument interfaces, upgrades, etc. Test System should have the ability to interface with VA’s Electronic Health Record and VA’s Electronic Health Record test account.
1.2.10.30. It is preferable that the vendor utilize the VA national site-to-site VPN or work with the VA Officer of Cyber and Information Security and Information Security Officers to establish a client-based VPN.
1.2.10.31. Provide a completed copy of the Manufacturer Disclosure Statement for Medical Device Security (MDS2) and VA Form 6550.
1.2.10.32. Contractor shall collaborate with each lab to write/develop protocols to establish customer configurable rules to enhance workflow management and productivity. Vendors will define level and type of support offered.
1.2.10.33. Contractor shall assist customer with optimizing operation and utilization of the data management system to fully integrate desired testing instrumentation enhancing productivity and management of workflow.
1.2.10.34. The Contractor shall provide all hardware, software, lines, adapters, and devices required to connect all Laboratory instruments to the Middleware, including, but not limited to, Lantronix devices or equivalent that are capable of converting DB9 serial to RJ45 Ethernet following RS232 protocols.
1.2.10.35. The Contractor shall provide 24/7 technical support for all hardware and software of the Middleware and its connections.
1.2.10.36. The Contractor shall provide all updates, upgrades, revisions, patches, and fixes at no cost.
1.2.10.37. The Contractor shall provide all necessary information to complete an ISA/MOU, if necessary, to create a VPN for the contractor to directly connect to the Middleware for troubleshooting purposes.
1.2.10.38. The Contractor shall provide all necessary licenses and support for licenses in order for the Laboratory to optimize the utility of the Middleware; i.e. licenses for connections and Thin Client licenses.
1.2.10.39. If required, remote access to server for multiple VA users, either with dedicated thin client terminals or via Windows Remote Desktop Connection
1.2.10.40. The Contractor shall provide training for the Middleware prior to implementation, annually thereafter and anytime there is an upgrade where additional training would be beneficial.
1.2.11. Hardware Features- The instrumentation shall have the following:
1.2.11.1. A total equipment footprint that when installed in the laboratory shall not impact the functionality/operations of that laboratory.
1.2.11.2. All monitors/screens will clearly display information in all light conditions.
1.2.11.3. A printer(s) that has the capability of printing a patient report with patient demographic information that includes minimally the patient’s name and accession or unique identifier number (UID).
1.2.11.4. If required. an uninterruptible power supply (UPS) with line conditioner for each instrument provided as needed. (This includes UPS units for sites with automation lines, specimen management systems, data management systems, refrigerated archive storage, etc.,) Each UPS must provide electrical power for a minimum of 15 minutes after electrical power fails and the system must allow for an automatic controlled shutdown to prevent damage to the instrument and data records.
1.2.11.5. If required, the proposed instrument system requires an independent water system to operate, the vendor will include the water system and all maintenance of the system as part of the CPRR package.
1.2.12. Specific Equipment Requirements- -
1.2.12.1. Single lot of reagent for each test/analyte per shipment with a minimum dating of 90 days.
1.2.12.2. User defined option for patient testing to be disabled if QC failure occurs.
1.2.12.3. The printer should be user defined to print the results real-time or on demand as well as option to print exceptional reports held in the middleware for auto verification purposes.
1.2.12.4. Equipment installation and possible reinstallation should the equipment need to be moved due to construction or laboratory redesign at no additional cost for one (1) location within each site.
1.2.13. Method Performance/Validation Requirements
Method performance/comparison shall be at the expense of and performed by the Contractor, shall include linearity material and reagents, and be consistent with current CLSI guidelines and related documents, College of American Pathologists (CAP) standards and Federal regulations. All studies performed will be appropriate for the test menu of the respective laboratory to include serum, plasma, urine, and body fluids as applicable. All studies must be approved by the local Laboratory Medicine Medical Director and made electronically available. These requirements shall be in effect during installation and any future changes to the test menu and/or method updates.
1.2.13.1.1. Correlation studies for each analyte. A minimum of 20 samples spanning the reportable range shall be run by the present and the proposed method. In systems where multiple sampling modes exist, mode to mode correlation studies must also be performed. Contractor shall analyze results and provide statistical data to support acceptance of the new method for above studies. Statistics shall consist of at least mean, bias, slope, y-intercept, correlation coefficient, ROC analysis.
1.2.13.1.2. Analytical Measurement Range (AMR) Validation shall be performed on proposed instrument(s) for each analyte to validate the reportable range. The material must have values, which are near the low, mid, and high values of the AMR and be of appropriate matrix for the clinical specimens assayed by that method. A minimum 5-point linearity analysis that adheres to the Beer-Lambert Law and spans the entire range shall be performed as a minimum.
1.2.13.1.3. Precision study using normal and abnormal control material. This shall include, at a minimum, within run precision study of 10 normal and 10 abnormal Vendor provided Controls.
1.2.13.1.4. Sensitivity. Sensitivity may be validated concurrently with correlation studies. Mathematical calculations to determine efficiency, sensitivity, false positive rate, and false negative rate are applied.
1.2.13.1.5. Specificity Studies. A review of product literature and assay inserts to determine any adverse effects for increased bilirubin, hemolysis, lipemia, or other interfering substances. If required, validation of serum indices for lipemia, Bilirubin and Hemolysis.
1.2.13.1.6. Carryover Studies. Successful carryover studies shall be completed by the contractor on all analyzers during installation, if required. These studies shall be performed using either contractor developed program(s) or program(s) developed by a third party (CAP/CLSI). The programs shall be provided to each laboratory at no charge.
1.2.14. Reference Range- A reference range must be determined for each test following CLSI guidelines. Samples used for the reference range study must be representative of the patient population being tested. Reference range assessment must be performed for each lab. One of the following protocols shall be used:
1.2.14.1. A verification of the manufacturer’s suggested reference range may be performed as long as the suggested range is based on a comparable population of test subjects. The manufacturer shall provide specific information defining how the suggested range was determined. A minimum of 20 reference individuals shall be used to verify the manufacturer’s range. Any apparent outliers should be discarded, and new specimens obtained to provide a statistically valid verification.
1.2.14.2. If the suggested manufacturer’s range is not appropriate for the patient population, a reference range shall be established. Establishing a reference must follow CLSI guidelines. This requires a minimum of 120 reference individuals to be used to establish a reference range. The reference interval should be determined using the nonparametric method.
1.2.14.3. If a laboratory is currently using the proposed instrument/reagent system, the “in-use” reference range can be transferred to the “new” system if a method comparison study between the two systems proves to be acceptable. If comparison studies are not acceptable, one of the two above items must be performed.
1.2.15. Reports- The Contractor shall provide to the Contracting Officer and other individuals (designated post-award) a copy of a quarterly report of sales, by ordering facility, within 30 calendar days after the close of each quarter’s business. Reports are to reflect, at a minimum, total net sales amounts before discount, and discount amounts by ordering facility as well as the raw data used to develop these reports. These reports shall be used to monitor the commitment of each facility, reporting the savings realized and shall be shared with each participating facility, personnel associated with acquiring the products, and respective laboratory personnel. Additional invoice charges associated with reagent and/or supply wastage or repair parts included at no charge (per FSS awarded contract) shall not be accepted. There will be no additional charges for any reports required as part of the BPA.
1.2.16. Support Features-
1.2.16.1. Commercial marketing. The equipment models being offered shall be in current production as of the date this offer is submitted. For purposes of this solicitation, “current production” shall mean that the clinical laboratory analyzer model is being offered as new equipment. Discontinued models that are only being made available as remanufactured equipment are not acceptable.
1.2.16.2. Start-Up Reagents. The Contractor shall provide all reagents, calibrators, Vendor provided Controls, linearity materials, consumable/disposable items, parts, accessories, and any other item included on the list of supplies defined in the Federal Supply Schedule contract and required to validate instruments for operation for performance of acceptance testing. This applies to all equipment as well as additional or replacement equipment placed under the terms and conditions of this BPA. The Contractor shall perform/assist, to the satisfaction of the Government, all validation studies including precision, method comparison with current analyzer, accuracy (recovery), linearity (reportable range), calibration verification, verification of reference interval, and determination of sensitivity and specificity at no cost to the Government. The Contractor shall perform all of the statistical analysis as stated in the Method Performance/Validation section above and provide a hard-copy and electronic copy of data in an organized, clearly comprehensible format.
1.2.16.3. Equipment Preventative Maintenance/Repair Service. The Contractor shall be able to provide, at no cost, emergency equipment repair and preventative maintenance on all primary and back-up instrumentation, primary processing automation line instrumentation and any incremental support/ancillary equipment, e.g. water system, printers, UPS, etc. offered according to the following terms:
1.2.16.4.1. Service Requirements
1.2.16.4.1.1. A technical assistance center shall be available by telephone 24 hours per day, 7 days per week with a maximum call back response time of 1 hour
1.2.16.4.1.2. Equipment repair service shall be provided for each facility during operational hours. Most repairs will take place during normal business hours, certain circumstances may dictate the need for repair service to be conducted outside routine business hours. All such arrangements shall be coordinated between the Contractor and VA laboratory personnel. Equipment repair response time shall be no more than 24 hours.
1.2.16.4.1.3. Remote Access diagnostic trouble shooting capabilities available with a national ISO Internet Security agreement.
1.2.16.4.1.4. Preventative maintenance will be performed as frequently as published in manufacturer’s operator’s manual and within 2 weeks of the scheduled due date.
1.2.16.4.1.5. A malfunction incident report shall be furnished to the Laboratory upon completion of each repair call, within 24 hours of the repair. The report shall include, as a minimum, the following:
1.2.16.4.1.5.1.date and time notified 1.2.16.4.1.5.2.date and time of arrival 1.2.16.4.1.5.3.serial number, type, and model number of equipment 1.2.16.4.1.5.4.time spent for repair, and 1.2.16.4.1.5.5.proof of repair that includes documentation of a sample run of quality control verifying acceptable performance.
1.2.16.4.2. Each notification for an emergency repair service call shall be treated as a separate and new service call.
1.2.16.4.3. Uptime Guarantee: The vendor agrees that all equipment provided shall be operable no less than 98% of operational time. Each instrument at each site shall be computed separately.
1.2.16.4.3.1 Scheduled maintenance is excluded from downtime.
1.2.16.5. Upgrades – the Contractor shall provide upgrades to both the equipment hardware and software in order to maintain the integrity of the system and the state-of –the art technology, at no additional charge to the Government. These shall be provided as they become commercially available and at the same time as they are being provided to commercial customers. This requirement only applies to “system upgrades” that enhance the model of equipment being offered, i.e. new version of software, correction of hardware defect, upgrade offered to commercial customers at no additional charge, upgrade to replace model of equipment no longer Contractor supported, etc. This does not refer to replacing the original piece of equipment provided under the Order; however, it does refer to significant changes in the hardware operational capability.
1.2.16.6. Ancillary support equipment – the Contractor shall provide, install, and maintain through the life of the Order, as indicated, any and all ancillary support equipment to fully operate the analyzer as defined in these specifications, e.g. cabinetry to support/house the analyzer (if necessary), water systems (including consumable polishers, filters, preventative maintenance, and repair, etc.), printers and universal interface equipment, UPS, etc. In addition, the Contractor shall include all ancillary components that are customarily sold or provided with the model of equipment proposed, e.g. starter kits, tables/stands, etc.
1.2.16.7. Characterization of waste – The Contractor shall provide documentation that it has characterized the hazardous nature of all wastes produced by all equipment, devices, reagents, and discharges in accordance with the requirements of the Code of Federal Regulations Title 40 “Protection of the Environment” Part 261 et seq. and applicable state and local requirements. Documentation shall include a description of the characteristics of the hazardous waste produced as a byproduct of the instrument operations, Safety Data Sheets (SDS) meeting the requirements of the Occupational Safety and Health Administration (OSHA) and Environmental Protection Agency (EPA), the analytical process used to determine the hazardous nature and characteristics of the waste, and the analytical test results. Testing of hazardous waste is to be done in accordance with testing protocol specified for each individual waste as described in the Code of Federal Regulations Title 40 to make a determination if the waste is a hazardous waste or otherwise regulated.
1.2.16.7.1. The determination and description shall address the following:
1.2.16.7.1.1. Waste toxicity (Reference 40 CFR §261.11 and 40 CFR §261.24)
1.2.16.7.1.2. Waste ignitability (Reference 40 CFR §261.21)
1.2.16.7.1.3. Waste corrosivity (Reference 40 CFR §261.22)
1.2.16.7.1.4. Waste reactivity (Reference 40 CFR §261.23)
1.2.16.7.1.5. Hazardous waste from non-specific sources (F-listed) (Reference 40 CFR §261.31)
1.2.16.7.1.6. Discarded commercial products (acutely toxic or P-listed and toxic or U-listed) (Reference 40 CFR §261.33)
1.2.16.7.1.7. Solid Waste (Reference 40 CFR §261.2)
1.2.16.7.1.8. Exclusions (Reference 40 CFR §261.4)
1.2.16.7.2. The contractor shall provide written instructions and training material to ensure VHA laboratory staff are trained as needed to properly operate devices with special emphasis to managing and disposing of hazardous waste in accordance with EPA and state requirements. Additionally, the training provided by the contractor must fulfill Resource Conservation and Recovery Act (RCRA) requirements for training as applicable to devices.
1.2.16.7.3. Contractor shall provide a description of all wastes the process or equipment may discharge so that the facility can determine whether the discharge meets Local Publicly Owned Treatment Works (POTW), State and Federal discharge requirements. At a minimum, the characteristics of ignitability, corrosivity, reactivity and toxicity as defined in 40 CFR §261 must be determined and documented. Any mercury containing reagents must be identified in any concentrations. All test results shall be provided. All listed chemicals (F, U, K and P) found in 40 CFR §261 shall be provided in product information and their concentrations documented. For those materials with a positive hazardous waste determination, a mechanism for the laboratory to meet local discharge requirements (i.e. mercury, thimerosal, and formaldehyde) must be developed and SDS sheets must be provided in advance for review. At a minimum, documentation shall include, but not be limited to the concentration/measures of the elements and parameters listed below and must be included with vendor response:
1.2.16.7.3.1. Barium (Total)
1.2.16.7.3.2. Cadmium (Total)
1.2.16.7.3.3. Chromium (Total)
1.2.16.7.3.4. Copper (Total)
1.2.16.7.3.5. Cyanide (Total)
1.2.16.7.3.6. Lead (Total)
1.2.16.7.3.7. Mercury (Total)
1.2.16.7.3.8. Nickel (Total)
1.2.16.7.3.9. Silver (Total)
1.2.16.7.3.10. Zinc (Total)
1.2.16.7.3.11. Arsenic (Total)
1.2.16.7.3.12. Selenium (Total)
1.2.16.7.3.13. Tin (Total)
1.2.16.7.3.14. pH
1.2.16.7.3.15. Flash point (to higher than 200.F)
1.2.16.7.3.16. BOD; biochemical oxygen demand
1.2.16.7.4. The documentation the contractor provides will be used to work with the VAMC and the public and/or private organization (e.g., POTW) to determine whether or not the waste from each device can legally be disposed of via the sewerage system.
1.2.16.8. Implementation/transition timeframe - The implementation of the services/requirements described in this solicitation shall be completed as outlined by Network Medical Facility order.
1.2.17. Standard and Quality of Performance- This paragraph establishes a standard of quality performance that shall be met before any equipment listed on the delivery order is accepted by the Government. This also includes replacement, substitute instruments and instruments that are added or field modified after a system has demonstrated successful performance. The acceptance period shall begin on the installation date. It shall end when the equipment has met the standard of performance for a period of 30 consecutive calendar days by operating in conformance with the Contractor’s technical specification or as quoted in the Order at an effectiveness level of 95% or more.
1.2.17.1. If equipment does not meet the standard of performance during the initial 30 consecutive calendar days, the standard of performance tests shall continue a day-by-day basis until the standard of performance is met for a total of 30 consecutive days.
1.2.17.2. If the equipment fails to meet the standard of performance after 90 calendar days from the installation date, the user may, at his/her option, request a replacement or terminate the order in accordance with the provisions of FAR 52.212-4 entitled “Termination for cause.” (The Contractor shall receive revenue for tests reported during the…