This solicitation is for the purchase and installation of dental equipment for a Veterans Affairs medical center. The VA seeks to procure a washer/disinfector, along with associated drawings and specifications. Vendors must RSVP by June 14th to attend an optional pre-bid site visit on June 16th. Proposals are due by responding directly to this notice, as no separate solicitation will be issued. The equipment must meet the detailed technical requirements provided in the attachment regarding dimensions, utilities, performance standards, and warranty terms.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment 2 - AMSCO 600 Electric CAB.pdf | ||
| Attachment 4 - SPS Equipment CA.pdf | ||
| 36C25921Q0393 Amendment 1.docx | DOCX document | |
| 36C25921Q0393.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
80-3/4 [2051]
UNIT
HEIGHT
31 [787]
LOADING
HEIGHT
26-1/4 [667]
LOADING
CLEARANCE
HEIGHT
14-1/2 [368]
[381]
91 [2311]
RAISED DOOR
HEIGHT
40 [1016]
MAX LENGTH
38 [965]
UNIT LENGTH
90 [2286]
SERVICE AREA
HEIGHT
45 [1143]
WALL OPENING
*(NOTE F)
1-1/2 [38]
MINIMUM
94 [2388] *(NOTE G)
RECOMMENDED
98 [2489] *(NOTE H)
MAX 1 [25]
ADJUSTMENT WITH
WASHER LEGS
(UNIT HEIGHT 81-3/4 [2077] MAX)
MINIMUM
94 [2388] *(NOTE G)
RECOMMENDED
96 [2438] *(NOTE H)
42 [1067]
UNIT WIDTH
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7053HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10288795
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SERVICE AREA
REQUIRED FOR
DRYING MOTOR
REMOVAL
TYPICAL INSTALLATION
NOTES:
(A) UNLESS OTHERWISE SPECIFIED, ALL DIMENSIONS ARE IN INCHES [mm].
(B) SERVICE ACCESS SIDE OF UNIT IS RECOGNIZED BY THE TOUCH SCREEN AT THE RIGHT OF
WASH CHAMBER. ELECTRICAL AND CONTROL COMPONENTS, AS WELL AS MOST OF
SERVICABLE PARTS ARE ACCESSIBLE FROM THAT SIDE OF UNIT. NON-SERVICE SIDE IS
EQUIPPED WITH THE TOUCH SCREEN ON THE LEFT OF THE WASH CHAMBER.
(C) CUSTOMER MUST BE SURE THAT MACHINE STANDS ON A NON-COMBUSTIBLE FLOOR.
(D) WEIGHTED EQUIVALENT SURFACE SOUND PRESSURE LEVEL: 64.0 dB(A) CALCULATED AS DESCRIBED IN ISO 3746.
(E) VERTICAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN
UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 3-1/2 in [89 mm] WIDTH X 80 in [2032 mm] LONG.
(F) HORIZONTAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN
UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 49 in [1245 mm] WIDTH X 16 in [406 mm] HEIGHT.
(G) MINIMUM REQUIRED DIMENSION FOR DOOR REMOVAL WITH LIMITED ADJUSTMENT AVAILABLE: +0 in [0 mm] THE STANDARD BARRIER WALL FLANGES CAN'T BE USED AT THAT DIMENSION. -3/4 in [19.1 mm]
(H) RECOMMENDED DIMENSION FOR DOOR REMOVAL WITH FULL LEVELING ADJUSTMENT AVAILABLE: +1 in [25.4 mm] THE STANDARD BARRIER WALL FLANGES CAN BE USED AT THAT DIMENSION. *(NOTE E AND F) -3/4 in [19.1 mm]
FRONT VIEW
SERVICE ACCESS
SIDE OF UNIT
*(NOTE B)
TOP VIEW
PASS THROUGH
DOOR
SIDE VIEWLEVEL FLOOR
*(NOTE C)
W A
LL
O
P E
N
IN
G H
E
IG
H T
C E
IL
IN
G H
E
IG
H T
FO
R
D O
O R
R E
M O
V A
L
Page 1 of 10 10288795.C
Released
2X 33-3/4 [856]
[517] 20-3/8
7-1/8 [181]
4X 15 [381]
2-1/8 [54]
5-11/16 [144]
33-7/16 [849]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7053HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10288795
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
*NOTES:
(A) CONCRETE FLOOR MUST REMAIN FREE OF DEFECTS IN THESE AREA, AS PER ESR-1917 & ESR-3187.
FOR INSTALLATION OF SEISMIC ANCHORS (NOT PROVIDED BY STERIS).
(B) THE FLOOR SINK AND DRAIN SHALL NOT OCCUPY THESE AREAS. 5 in x 5 in [127 mm x 127 mm] FLOOR SINK WITH Ø4 in [102 mm] DRAIN IS RECOMMENDED WHEN USING SEISMIC MOUNTS.
SEISMIC MOUNTS
SEISMIC ANCHORAGE
FOR DETAIL SEE FD354
QTY 4 ANCHORS
FINISHED WALL
PREFERENCE
LOCATION OF
FLOOR SINK
*(NOTE A AND B)
Page 2 of 10 10288795.C
80-3/4 [2051]
7-1/8 [181]
[381]
31-1/8 [791]
[686]
26-1/32 [661]
23-17/32 [597]
22-1/2 [572]
3-7/8 [99]
[533]
18-1/2 [470]
32-11/32 [822]
HW HOT WATER
CW COLD WATER
PW PURE WATER
A COMPRESSED AIR
D FLOOR DRAIN LOCATION
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7053HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10288795
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (PLUMBING)
CONNECTION POSITION
FRONT VIEW
A
D
PW
HW
CW
SIDE VIEW
HW
PW
CW
A
D
CR
S
TOP VIEW
HW
CW
PW
A
CR
S
D
S STEAM
CR CONDENSATE RETURN
S
CR
Page 3 of 10 10288795.C
79-1/4 [2013]
35-13/16 [910]
28-3/8 [721]
[788]
38-1/8 [967]
39-5/8 [1007]
E ELECTRICAL
N NETWORK (SEE NOTE)
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7053HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10288795
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (ELECTRICITY)
CONNECTION POSITION
NOTE:
NETWORK CONNECTION (IF APPLICABLE)
- ACTIVE NETWORK DROP WITH CAT 5/6 CABLE WITHIN 10 ft [3 m] OF UNIT
- STATIC OR DHCP RESERVED IP ADDRESS (FOR EACH UNIT)
- NETWORK CABLE (SUPPLIED BY OTHER)
FRONT VIEW SIDE VIEW
E
N
E
N
TOP VIEW
E
N
Page 4 of 10 10288795.C
6-5/32 [156]
[254]
81-29/32 [2080]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7053HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10288795
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
EX VENTILATION EXHAUST
UTILITY (VENTILATION)
CONNECTION POSITION (OPTIONAL)
FRONT VIEW SIDE VIEW
EX
SUGGESTED DUCTWORK
(BY OTHERS)
EX
DRAIN 3/4 in O.D. [20 mm]
(ROUTED TO WASHER DRAIN)
DRY CONTACT (SEE PAGE 10 *(NOTE 3))
VAPOR CONDENSATE
COLLECTING POINT
TOP VIEW
EX
VENTILATION CONNECTION (OPTIONAL)
(SEE PAGE 9 *(NOTE 1))
EX
BY OTHERS
Page 5 of 10 10288795.C
60 [1524]
REF
8-1/2 [216]
132 [3350]
MAX.
17 [432]
INSTRUMENT
(With PUC HP)
UTENSIL
(With PUC HP)
CHEMICAL
(CH)
ENZYME CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]
DETERGENT CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]
LUBRICANT CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.2 oz [6 ml] N/A
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7053HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10288795
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
*NOTES:
(A) MAXIMUM CONTAINER HEIGHT FROM FLOOR IS 24 in [610 mm]. AMBIENT TEMPERATURE SHALL NOT EXCEED 40 °C [104 °F]
DOSING STATION
CEILING
ROUTE CHEMICAL LINES
AND SIGNAL CABLE
THROUGH PIPE TO REACH
BOTTOM OF UNIT
CHEMICAL CONTAINERS
*(NOTE A)
12 ft [3.6 m] OF INJECTION LINE
(FROM CONTAINERS TO
DOSING STATION)
DOSING STATION
WEIGHT : 15 lb [6.8 kg]
2X FD349 SHOWN
UP TO 5 TUBES
100 ft [30 m] OF INJECTION LINE
AND SIGNAL CABLE
DETERGENT LINES SHALL NOT BE IN CONTACT WITH HOT SURFACES.
STERIS RECOMMENDS CUSTOMER TO ROUTE CHEMICAL LINES AND
SIGNAL CABLE THROUGH A 3 in [76 mm] ID CONDUIT.
EXCESS LENGTH MUST BE CUT BEFORE CONNECTING TO
WASHER/STATION TO AVOID FORMING BUNDLES THAT COULD
AFFECT THE CHEMICAL INJECTION SYSTEM'S PERFORMANCE
(DO NOT CUT SIGNAL CABLE).
Page 6 of 10 10288795.C
UTILITY REQUIREMENTS CHART (PLUMBING)
UTILITY PARAMETER REQUIREMENT NOTES
HOT WATER
(HW)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
1. STERIS REQUIRES TO INSTALL WATER HAMMER
ARRESTOR, UNION & PRESSURE GAUGE.
2. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
*STERIS OFFERS FLEXIBLE HOSES
ACCESSORY (FD088 OR FD037)
WITH INSTRUCTIONS.
3. SAMPLING VALVES INSTALLATION AS CLOSE AS
POSSIBLE TO WASHER WATER CONNECTION IS
RECOMMENDED.
PRESSURE RANGE
15 - 50 psig DYNAMIC
[103-345 kPa] MAX. 90 psig STATIC [620 kPA]
OPERATING FLOW RATE 6.8 US GPM [25.7 L/min] TO
14.1 US GPM [53.4 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4)
9.9 US GAL [37.5 L]
WITH ACU-WASH RESERVOIR:
6.6 US GAL [25.0 L]
TEMPERATURE RANGE 110 °F [43 °C] TO 150 °F [65.5 °C]
WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
COLD WATER
(CW)
*(NOTES 6)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
4. CONSUMPTION BASED ON INSTRUMENT CYCLE WITH
PROLYSTICA ULTRA CONCENTRATE HIGH
PERFORMANCE CHEMICALS (PUC HP) WITH
THE FOLLOWING UTILITY CONDITIONS:
THREE-LEVEL MANIFOLD RACK
(CW) COLD WATER PRESSURE: 30 psig DYNAMIC [207 kPa]
(CW) COLD WATER TEMPERATURE: 70 °F [21 °C]
(HW) HOT WATER TEMPERATURE: 150 °F [65 °C]
(PW) PURE WATER TEMPERATURE 70 °F [21 °C]
(S) STEAM SUPPLY: 80 psig DYNAMIC [552 kPa]
UTENSIL CYCLE CONSUMPTION:
TWO-LEVEL MANIFOLD RACK
CW: 10.4 US GAL [39.4 L]
HW: 9.6 US GAL [36.3 L]
PW: 6.4 US GAL [24.2 L]
S: 19.3 lb [8.8 kg]
5. STERIS RECOMMENDS THE FOLLOWING WATER
QUALITY PARAMETERS:
TOTAL DISSOLVED SOLIDS: 100-200 ppm (MAX. 500 ppm) TOTAL ALKALINITY (CaCO3) : 70-120 ppm (MAX. 180 ppm) pH : 6.8-7 (MAX. 7.5) TOTAL SILICA : 0.1-0.5 ppm (MAX. 1 ppm)
PRESSURE RANGE
30 - 50 psig DYNAMIC
[206-345 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE 10.7 US GPM [40.5 L/min] TO
14.1 US GPM [53.4 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4)
FOR SUMP FILLING: 18.9 US GAL [71.5 L]
(ADD THE FOLLOWING CONSUMPTION
FOR OPTIONS/CONFIGURATIONS)
CONDENSATE RETURN COOL DOWN:
8.1 US GAL [30.7 L]
NON-VENTED SYSTEM:
13.3 US GAL [50.3 L]
DRAIN COOLDOWN:
15.8 US GAL [59.8 L]
DRAIN ACU-COOLDOWN:
12.1 US GAL [45.8 L]
TEMPERATURE RANGE 70 °F [21 °C] MAXIMUM
WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
PURE WATER
(PW)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
6. BACK FLOW PREVENTER AGAINST BACK
SIPHONAGE MAY BE REQUIRED (LOCAL
PLUMBING CODE)
7. PW IS OPTIONAL. IF UNAVAILABLE, UNIT WILL
PROCESS THERMAL RINSE USING HOT WATER
SUPPLY.
PRESSURE RANGE 5 - 30 psig DYNAMIC [30 - 206 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE 5.5 US GPM [21 L/min] TO
12.6 US GPM [48 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4 AND NOTE 7)
6.6 US GAL [25.0 L]
TEMPERATURE RANGE 70 °F [21.1 °C] TO 150 °F [65.5 °C]
RESISTIVITY
(CONDUCTIVITY)
0.1-0.5 M -cm (2-10 µS/cm) pH 6.8 - 7.5
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7053HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10288795
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 7 of 10 10288795.C
UTILITY REQUIREMENTS CHART (PLUMBING)
COMPRESSED
AIR
(A)
CONNECTION SIZE / TYPE 1/8 in / FEMALE NPT [1/8 in / MALE BSPT]
1. AIR PURITY PER ISO-8573-1: 2010 [7:4:4]
MAX. PARTICLE DENSITY: 10 mg/m³
MAX. DEW POINT FOR WATER CONTENT:
37 °F [3 °C]
MAX. OIL CONCENTRATION FOR THE OIL
CONTENT: 5 mg/m³
2. STERIS RECOMMENDS A REFRIGERATED AIR
DRYER WHERE ENVIRONMENTAL DEW POINT
CONDITIONS ARE HIGHER THAN
RECOMMENDED.
3. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
PRESSURE RANGE MIN 80 psig DYNAMIC [552 kPa] MAX. 125 psig STATIC [860 kPa]
OPERATING FLOW RATE 1.75 SCFM [50 L/min]
DRAIN
(D)
FLOOR DRAIN SIZE MIN 4 in [102 mm] O.D. WITH 8 in x 8 in [204 mm x 204 mm] FLOOR SINK
4. WASHER DRAIN OUTLET SIZE 1-1/2 in / MALE NPT
5. MAX OFFSET OF 3/4 in [19 mm] FROM
OUTLET (PAGE 3, D).
6. FOR DRAIN REQUIREMENTS WHEN USING
SEISMIC MOUNTS, SEE PAGE 2.
OPERATING FLOW RATE GRAVITY DRAIN : 50 US GPM
[189.3 L/min]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7053HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10288795
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
CHEMICAL
DOSING
SYSTEM
CEILING ROUTING
CONDUIT 3 in CONDUIT
11. STERIS RECOMMENDS TO ROUTE CHEMICAL
LINES AND SIGNAL CABLE THROUGH A
3 in [76 mm] ID CONDUIT (BY OTHERS).
100 ft [30 m] OF INJECTION LINE AND
SIGNAL CABLE ARE PROVIDED WITH THE UNIT.
STEAM INLET
(S)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
7. STERIS REQUIRES AT LEAST 97% SATURATED
DRY PLANT STEAM, FREE OF DIRT, RESIDUES
AND CONTAMINANTS.
8. STERIS RECOMMENDS TO INSTALL A STEAM
TRAP AS CLOSE AS POSSIBLE TO UNIT
CONNECTION.
9. PROPER OPERATION OF THE WASHER
REQUIRES A MINIMUM PRESSURE DIFFERENTIAL
OF 21.5 psig [148 kPa] BETWEEN STEAM
INLET AND CONDENSATE RETURN. IN CASES OF
LOW STEAM INLET PRESSURE, CONDENSATE
CAN BE ROUTED TO DRAIN TO INCREASE
DIFFERENTIAL.
*STERIS OFFERS CONDENSATE RETURN
TO DRAIN ACCESSORY (FD353)
10. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
PRESSURE RANGE
30 - 80 psig DYNAMIC
[206-552 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING
MASS FLOW RATE
169 lb/h [76.5 kg/h] AT 30 psig [207 kPa]
271 lb/h [123 kg/h AT 80 psig [550 kPa]
TOTAL CONSUMPTION
PER INSTRUMENT CYCLE
(SEE PAGE 7, *(NOTE 4))
19.3 lb [8.8 kg]
CONDENSATE
RETURN
(CR)
PRESSURE RANGE MAX. 8.5 psig [68.9 kPa] BACK PRESSURE (17 ft [5.2 m] ELEVATION)
OPERATING FLOW RATE PEAK : 0.54 US GPM [2.05 L/min]
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
Page 8 of 10 10288795.C
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7053HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10288795
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY REQUIREMENTS CHART (ELECTRIC)
UTILITY
PARAMETER AND REQUIREMENT NOTES
CONNECTION
SIZE/TYPE
NOMINAL
AMPERAGE
RECOMMENDED
CIRCUIT
ELECTRICITY
(EL)
200-208V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
36 A 50 A
1. ALL CONNECTIONS SHOULD BE IN
ACCORDANCE WITH LOCAL CODES.
2. NEUTRAL WIRE NOT REQUIRED.
3. ALWAYS FOLLOW LOCAL ELECTRICAL CODES
AND SAFETY-RELATED WORK PRATICES TO
DETERMINE THE VALUE OF PROTECTION.
4. WIRING ON THE EQUIPMENT TERMINATES AT
A JUNCTION BOX. WIRING BETWEEN JUNCTION
BOX AND BUILDING SERVICE LINES ARE NOT
SUPPLIED BY STERIS.
5. DISCONNECT SWITCHES WITH OFF POSITION
LOCKOUT (SUPPLIED BY OTHERS) MUST BE
INSTALLED IN ELECTRIC SUPPLY LINES NEAR THE
EQUIPMENT.
6. STERIS RECOMMENDS A LIGHT IN SERVICE
AREA.
200-208V 60 Hz, 3-Ph
WITH 40A
MIN. CIRCUIT
PROTECTION
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
28.5 A 40 A
200-208V 50 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
35 A 50 A
460-480V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
16.5 A 30 A
380-400V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
19 A 30 A
380-400-415V 50 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
18.5 A 30 A
Page 9 of 10 10288795.C
UTILITY REQUIREMENTS CHART
VENTILATION
EXHAUST
(EX)
OPTIONAL
CONNECTION SIZE 3 in [76 mm]
OUTSIDE DIAMETER
1. NOT REQUIRED WITH NON VENTED UNIT.
2. STERIS REQUIRES CORROSION-RESISTANT AND
WATER-TIGHT DUCT:
- MIN OPERATING TEMPERATURE 210 °F [99 °C]
- SLOPED TOWARDS THE WASHER AND NO DEAD LEGS
- INSTALL DRAIN PART AT LOWEST POINT NEAR UNIT
3. EXHAUST FAN NORMALLY OPEN-AUXILIARY CONTACTS
IS AVAILABLE TO SUPPLY A SIGNAL TO THE FACILITY HVAC
CONTROL SYSTEM WHENEVER THE WASHER VENTILATION
EXHAUST FAN IS IN OPERATION.
PRESSURE RANGE 1/2 in [50 mm]
VACUUM AT CONNECTION
OPERATING FLOW RATE 60 SCFM [1.7 m³/min]
TEMPERATURE RANGE MAX. 240 °F [116 °C]
PHYSICAL SPECIFICATIONS
SHIPPING DIMENSION W x H x L 50 in x 91 in x 48 in [1270 mm x 2311 mm x 1219 mm]
1. LEVELING FEET ARE PROVIDED FOR PROPER INSTALLATION.
2. PERMISSIBLE ENVIRONMENTAL CONDITIONS.
THIS EQUIPMENT IS DESIGNED TO GIVE OPTIMAL RESULTS
UNDER THE FOLLOWING CONDITIONS:
- INDOOR USE ONLY;
- ALTITUDE OF OPERATION UP TO 6562 ft [2000 m];
- TEMPERATURE OF 41 TO 104 °F [5 to 40 °C];
- MAXIMUM RELATIVE HUMIDITY IS 80% FOR TEMPERATURE UP
TO 88 °F [31 °C], DECREASING LINEARLY TO 50% RELATIVE
HUMIDITY AT 104 °F [40 °C];
- POLLUTION DEGREE 2: EQUIPMENT MUST BE INSTALLED IN AN
ENVIRONMENT WHERE NORMALLY ONLY NON-CONDUCTIVE
POLLUTION OCCURS BUT WHERE OCCASIONAL, TEMPORARY
CONDUCTIVITY CAUSED BY CONDENSATION CAN BE EXPECTED
(AS DETERMINED ACCORDING TO INTERNATIONAL STANDARD
EN/IEC 61010-1).
3. STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE TO,
OR FAILURE TO OBSERVE THE SPECIFICATIONS ON EQUIPMENT
DRAWING AND NOTE PAGES. SPECIFICATIONS AND
DESCRIPTIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE.
SHIPPING WEIGHT 1350 lb [612 kg]
OPERATING DIMENSION W x H x L 42 in x 80.75 in x 38 in [1067 mm x 2051 mm x 965 mm]
OPERATING WEIGHT
(WITH WATER AND FULLY LOADED) 1336 lb [606 kg]
AVERAGE FLOOR LOAD
OVER WASHER FOOTPRINT 0.84 psi [5.77 kPa]
MAX. FLOOR LOAD
AT WASHER FOOT PADS
(3 in [76.2 mm] DIAMETER PADS) 71 psi [491 kPa]
NOISE LEVEL:
A-WEIGHTED EQUIVALENT
SURFACE SOUND PRESSURE LEVEL
60.6 dB(A)
CALCULATED AS DESCRIBED
IN ISO 3746.
OPENING AREA FROM
CLEAN SIDE TO SOILED SIDE
46.5 in² [300 cm²]
*(NOTE 4)
4. OPENING AREA CAN BE REDUCED TO 13 in² [84 cm²]
WITH INSTALLATION OF AIR MANAGEMENT PANELS
ACCESSORY (FD352)
CHAMBER DIMENSION W x H x L
VOLUME
26.5 in x 26.25 in x 31.5 in [673 mm x 667 mm x 800 mm]
12.7 ft³ [0.36 m³]
TYPICAL HEAT LOSS AT 75 °F [24 °C]
CONTINUOUS OPERATION
NON-VENTED UNIT
WITHOUT ACU-RINSE RES.
VENTED UNIT (OPTIONAL)
WITHOUT ACU-RINSE RES.
CLEAN SIDE 2186 Btu/h [640 W]
2186 Btu/h [640 W]
SOILED SIDE 7069 Btu/h [2070 W]
5259 Btu/h [1540 W]
TOTAL 9255 ± 854 Btu/h [2710 ± 250 W]
7445 ± 854 Btu/h [2180 ± 250 W]
VALUES INCLUDE HEAT LOSS DURING DOOR OPENING FOR LOADING AND UNLOADING OPERATIONS. "TOTAL" WILL VARY DEPENDING ON DISINFECTION TIME AND TEMPERATURE, DRYING TIME AND TEMPERATURE AND USE RATE OF UNIT.
AN ADDITIONAL HEAT LOSS FOR LOAD COOL DOWN IN CLEAN SIDE CAN BE ADDED TO THE TOTAL HEAT LOSS OF THE WASHER:
MAXIMUM 1713 Btu [1807 kJ]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7053HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10288795
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 10 of 10 10288795.C
| Sheet 1 |
| Views |
| new_view_35 |
| new_view_37 |
| new_view_57 |
| new_view_66 |
| Zones |
| A1 |
| Sheet 2 |
| Views |
| new_view_55 |
| Zones |
| A1 |
| Sheet 3 |
| Views |
| new_view_38 |
| new_view_39 |
| new_view_40 |
| Zones |
| A1 |
| Sheet 4 |
| Views |
| new_view_44 |
| new_view_45 |
| new_view_46 |
| Zones |
| A1 |
| Sheet 5 |
| Views |
| new_view_41 |
| new_view_42 |
| new_view_43 |
| new_view_65 |
| Zones |
| A1 |
| Sheet 6 |
| Views |
| new_view_47 |
| Zones |
| A1 |
| Sheet 7 |
| Zones |
| A1 |
| Sheet 8 |
| Zones |
| A1 |
| Sheet 9 |
| Views |
| new_view_64 |
| Zones |
| A1 |
| Sheet 10 |
| Zones |
| A1 |
File details come from the government source that posted it. Updated .