Attachment 2 - AMSCO 600 Electric CAB.pdf

PDF 237 KB Posted

Attached to
Dental Equipment Purchase and Install (Amended, 1)
Solicitation number
36C25921Q0393
Issued by

About this file

This document package includes technical specifications and drawings for an AMSCO 600 electric sterilizer as well as details of a related federal contract opportunity posted by the Department of Veterans Affairs.

The technical specifications provide requirements for the AMSCO 600 electric sterilizer including dimensions, capacity, utilities needs, noise levels, weights, and installation requirements. Utilities include cold water, reverse osmosis water, compressed air, steam, and electricity. The opportunity posting from the VA seeks proposals for the purchase and installation of dental equipment but provides no further details on specific products or services required. The documents include drawings and diagrams of the sterilizer but no pricing or response deadline information for the federal contract.

View the file

Other files

File Type Posted
Attachment 3 - AMSCO 7053 HP DRAWING.pdf PDF
Attachment 4 - SPS Equipment CA.pdf PDF
36C25921Q0393 Amendment 1.docx DOCX document
36C25921Q0393.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

'A'

42-1/4 [1073]

UNIT WIDTH

89-1/16 [2263]

UNIT

HEIGHT

31-1/2 [800]

LOAD HT.

4 [102] TO 8 [203]

WALL THK.

CHAMBER LENGTH UNIT LENGTH-DIM. 'A'

39" [991] 53-3/8 [1356]

51" [1295] 65-3/16 [1656]

63" [1600] 77 [1956]

T I T L E

EQUIPMENT DRAWING, AMSCO 600 , 26.5" X 26.5" X 39", 51" & 63" LG.

SLIDING, SD, CABINET, ELECTRIC STEAM GENERATOR

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:

A - Gielink, Bob 10778442 11008531 B

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

SHEET

TOP VIEW

SIDE VIEWFRONT VIEW

NOTES:

1. THE FLOOR AREA (IN SERVICE CLEARANCE AREAS) MUST BE SMOOTH AND HAVE A LEVEL TOLERANCE WITHIN 1/2 deg OR 1/16" PER FOOT, WITHOUT ANY OBSTRUCTIONS SUCH AS PIPING, WIRING, FLOOR SINKS, ETC.

2. SERVICE AREA IS THE WIDTH OF THE UNIT. LOADING AND SERVICE AREA REQUIRED IN FRONT OF THE LOAD SIDE IS EQUAL TO 2x THE LENGTH OF CHAMBER.

3. ALL DIMENSIONS ARE INCHES AND MILLIMETERS [MM].

STEAM

GENERATOR

SERVICE ACCESS

DOOR

FINISHED

FLOOR

FINISHED

WALL

STEAM

GENERATOR

16-9/16 [420]

28-1/16 [713]

13-1/2 [343]

3-3/4 [95]

17-5/16 [440]

8-3/4 [223]

89-1/16 [2263]

24-7/16 [621]

76-3/4 [1949]

3-9/16 [91]

21-1/2 [546] 32-1/8

[816]

CW COLD WATER

DiW DIONIZED WATER

D DRAIN

A COMPRESSED AIR

GD STEAM GENERATOR DRAIN

T I T L E

EQUIPMENT DRAWING, AMSCO 600 , 26.5" X 26.5" X 39", 51" & 63" LG.

SLIDING, SD, CABINET, ELECTRIC STEAM GENERATOR

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:

A - Gielink, Bob 10778442 11008531 B

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

SHEET

TOP VIEW

FRONT VIEW SIDE VIEW

NOTE: THIS UNIT HAS THE CAPABILITY TO BE CONNECTED TO

FACILITY STEAM BY REMOVING THE PLUG LOCATED IN THE STEAM

INLET PORT AND INSTALL A STERIS SUPPLIED SHUT-OFF VALVE.

SHUT-OFF VALVE TO BE PLACED BETWEEN THE STERILIZER AND THE

BUILDING STEAM CONNECTION.

CW

DiW

D

GD

A

DiW-A-CW

D GD D

CW

ADiW

GD

[838]

3-1/2 [88]

6-3/16 [157]

36-13/16 [935]

87-5/8 [2226]

20-3/8 [518]

3-7/16 [87]

41-3/4 [1060]

T I T L E

EQUIPMENT DRAWING, AMSCO 600 , 26.5" X 26.5" X 39", 51" & 63" LG.

SLIDING, SD, CABINET, ELECTRIC STEAM GENERATOR

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:

A - Gielink, Bob 10778442 11008531 B

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

SHEET

E STERILIZER ELECTRIC

GE STEAM GENERATOR ELECTRIC

GE1 STEAM GEN. CONTROL ELECTRIC

N NETWORK

TOP VIEW

FRONT VIEW SIDE VIEW

E N

GE-GE1

E N N E

GE

GE1

MECHANICAL

UTILITY REQUIREMENTS CHART

UTILITY PARAMETER REQUIREMENT NOTES

STERILIZER

COLD WATER

(CW)

CONNECTION SIZE/TYPE 3/4" NPT

INSTALLATION SPECIFICATIONS:

1. PIPE SIZES LISTED UNDER PLUMBING REQUIREMENTS

INDICATE THE EQUIPMENT TERMINATION SIZES ONLY.

SIZE PIPING TO EQUIPMENT DEPENDING ON LENGTH OF

PIPE RUN FROM PRESSURE REGULATING STATION FOR

THE STEAM LINE AND MAIN WATER HEADERS. SUPPLY

THE SPECIFIED SERVICE PRESSURE AND FLOW RATE AT

EQUIPMENT TERMINALS, INCLUDE EFFECT OF COINCIDENT

DRAW OF MULTIPLE UNIT INSTALLATIONS.

2. PROVIDE PIPING, SHUTOFF VALVE, PIPE PLUGGED TEE,

AND UNION IN STEAM AND WATER SUPPLY CONNECTIONS

BETWEEN EQUIPMENT AND STUB OUTS. PLUGGED TEE

CAN LATER BE USED FOR TEST PRESSURE GAUGE

CONNECTION. ARRANGE CONNECTION PIPING TO ALLOW

ACCESS TO MACHINE COMPONENTS AND ELECTRICAL

CONTROL PANEL.

3. RECOMMENDED PROVISION OF BLOW DOWN VALVE AT

EACH STEAM AND WATER STRAINER TO ENABLE STRAINER

CLEAN OUT.

4. BLOW DOWN BUILDING (FACILITY) STEAM AND WATER

SUPPLY LINES BEFORE FINAL CONNECTION TO THE

EQUIPMENT.

5. THE STERILIZER IS NOT SUPPLIED WITH A VACUUM

BREAKER OR BACKFLOW PREVENTER. WHERE REQUIRED

BY LOCAL CODES, INSTALLATION OF A VACUUM BREAKER

IN THE WATER LINE IS SUPPLIED BY OTHER PARTIES.

STERIS OFFERS A BACKFLOW PREVENTER AS AN OPTION.

6. PLACEMENT OF PIPING SHUTOFFS WHEN INSTALLING.

SHUTOFF VALVES MUST BE LOCATED IN A SUITABLE

POSITION WITHIN THE LINE OF SIGHT, AND CLEAR OF ANY

OBSTRUCTIONS THAT PLACE THE SERVICE PERSONNEL IN

HARMS WAY, IN ORDER TO TURN THE VALVE OFF.

7. REQUIRED PIPING PRACTICES FOR RELIEF VALVE PIPING

CAN BE FOUND IN ASME BOILER CODE AND PRESSURE

VESSEL CODE, SECTION VIII (8), PER PARAGRAPH UG-135.

8. SEE FEED WATER QUALITY TABLE. REFERENCE:

STERIS DRAWING NUMBER 62941-091.

MINIMUM RESISTIVITY: 1M -cm (1 MEGOHM-cm), PER ASME PEB-5.3

9. IN CYCLE CONSUMPTION DATA BASED ON PREVAC OF

270 F, 4 MINUTE STERILIZE, 30 MINUTE DRY WITH METAL

LOAD.

PRESSURE RANGE 20-50 DYNAMIC PSIG (1.4-3.4 BAR G)

TEMPERATURE RANGE 40 -70 F (4 -21 C)

MIN. FLOW RATE 12 GPM (45 LPM)

AVERAGE

CONSUMPTION

39" LG - PEAK: 12 GPM (45 LPM)

PER CYCLE: 97 GAL (367 L)

OUT OF CYCLE: 20 GPH (75 LPH)

51" LG - PEAK: 12 GPM (45 LPM)

PER CYCLE: 112 GAL (424 L)

OUT CYCLE: 25 GPH (95 LPH)

63" LG - PEAK: 12 GPM (45 LPM)

PER CYCLE 128 GAL (484 L)

OUT OF CYCLE: 30 GPH (114 LPH)

GENERATOR

DIONIZED and/or

REVERSE

OSMOSIS

WATER

(DiW)

SEE NOTE 8

CONNECTION SIZE/TYPE 1/2" NPT

SUPPLY TEMP. 70-140 DEG. F (21-60 DEG. C)

PRESSURE RANGE 20-50 DYNAMIC PSIG (1.4 - 3.4 BARG)

FLOW RATE PEAK: 0.6 GPM 2.3 LPM)

AVERAGE

CONSUMPTION

39" LG - PER CYCLE: 4.2 GAL (16 L)

OUT OF CYCLE: 3 GPH (11 LPH)

51" LG - PER CYCLE: 5.4 GAL (20 L)

OUT OF CYCLE: 4 GPH (15 LPH)

63" LG - PER CYCLE: 6.6 GAL (25 L)

OUT OF CYCLE: 5 GPH (19 LPH)

DRAIN (TO

FLOOR SINK)

(D)

CONNECTION SIZE/TYPE 1-1/2" ODT OPEN

AVERAGE

OPERATING

FLOW RATE

GRAVITY DRAIN MUST BE ABLE TO

HANDLE PEAK WATER

CAPACITY OF 15 GPM (57 LPM)

MAX. DISCHARGE TEMP. 140 F (60 C)

FLOOR SINK MIN.

SIZE & ELEVATION

12" X 12" X 6" DEEP MIN. SIZE

INSTALL AT FLOOR GRADE

COMPRESSED

AIR

(A)

CONNECTION SIZE/TYPE 1/2" NPT

PRESSURE RANGE 80-100 PSIG DYNAMIC

5.5 - 6.9 BAR G

FLOW RATE 3 SCFM (5 SCMH)

AIR QUALITY OIL FREE, DRY (DEHUMIDIFIED)

GENERATOR

DRAIN

(GD)

CONNECTION

SIZE/TYPE 1/2 NPT

T I T L E

EQUIPMENT DRAWING, AMSCO 600 , 26.5" X 26.5" X 39", 51" & 63" LG.

SLIDING, SD, CABINET, ELECTRIC STEAM GENERATOR

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:

A - Gielink, Bob 10778442 11008531 B

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

SHEET

ELECTRICAL

UTILITY REQUIREMENTS CHART

UTILITY PARAMETER REQUIREMENT NOTES

STERILIZER

ELECTRIC

(E)

208-240 VAC

THREE PHASE

60 HZ

CURRENT 10 AMPS,

MAX. PROTECTION 20 AMPS,

#12 [3.31] AWG

ELECTRICAL UTILITY REQUIREMENTS

1. ALL CONNECTIONS SHOULD BE IN ACCORDANCE WITH

LOCAL CODES.

2. ALWAYS FOLLOW LOCAL ELECTICAL CODES AND SAFETY-

RELATED WORK PRACTICES TO DETERMINE THE VALUE

OF PRODUCTION.

3. DISCONNECT SWITCHES WITH OFF POSITION LOCKOUT

(SUPPLIED BY OTHERS) MUST BE INSTALLED IN ELECTRIC

SUPPLY LINES THAT ARE NOT SUPPLIED BY STERIS.

4. STERIS RECOMMENDS A LAMP IN SERVICE AREA TO

ILLUMINATE THE SURROUNDING SPACE.

INSTALLATION SPECIFICATIONS:

1. PROVIDE GROUPED OR GANGED CIRCUIT PROTECTION AND

DISCONNECT FOR STERILIZER POWER AS REQUIRED BY

CODES AND STANDARDS. INDIVIDUAL POWER SHUTOFFS

REQUIRED PER STANDARDS. INDIVIDUAL POWER SHUTOFFS

REQUIRED NEAR MACHINE FOR SERVICING.

2. PROVIDE GROUNDED METAL CONDUIT AND WIRING

BETWEEN EQUIPMENT TERMINALS AND STUB OUT OR

DISCONNECTS. CHECK LOCAL CODES FOR MINIMUM AWG.

WIRE SIZE.

3. PLACEMENT OF ELECTRICAL DISCONNECTS: WHEN

INSTALLING; DISCONNECTS MUST BE LOCATED IN SUITABLE

LOCATIONS WITHIN LINE OF SIGHT AND CLEAR OF ANY

OBSTRUCTIONS THAT WOULD PUT THE SERVICE PERSON

OR OPERATOR IN HARMS WAY IN ORDER TO TURN IT OFF.

ALSO, THE LOCATION OF THE DISCONNECTS SHOULD

ALLOW THE SERVICE PERSON TO SHUTOFF POWER FROM

THE SIDE TO PREVENT POSSIBLE ARC FLASH.

4. CAUTION: DO NOT USE GROUND FAULT CURRENT

INTERRUPTER (GFCI).

5. ATTENTION: THE ELECTRICAL CLEARANCES REQUIRED BY

THE NEC ARE THE RESPONSIBILITY OF THE INSTALLER.

ALSO, ADHERENCE TO LOCAL CODES AND PROCUREMENT

OF PERMITS ARE THE RESPONSIBILITY OF THE CUSTOMER

UNLESS AGREED TO IN WRITING WITH STERIS.

6. MAIN INCOMING POWER WIRING TO HAVE 220 F

INSULATION RATING MINIMUM, COPPER ONLY.

440-480 VAC

THREE PHASE

60 HZ

CURRENT 7 AMPS,

MAX. PROTECTION 20 AMPS,

#12 [3.31] AWG

STEAM

GENERATOR

ELECTRIC

(GE)

208-220 VAC

THREE PHASE

50/60 HZ

CURRENT 125 AMPS,

MAX. PROTECTION 170 AMPS,

#1 [42.4] AWG

230-240 VAC

THREE PHASE

50/60 HZ

CURRENT 113 AMPS,

MAX. PROTECTION 145 AMPS,

#2 [33.6] AWG

380-415 VAC

THREE PHASE

50/60 HZ

CURRENT 59 AMPS,

MAX. PROTECTION 95 AMPS,

#6 [13.3] AWG

440-480 VAC

THREE PHASE

50/60 HZ

CURRENT 57 AMPS,

MAX. PROTECTION 75 AMPS,

#6 [13.3] AWG

STEAM

GENERATOR

CONTROL

(GE1)

120 VAC

50/60 HZ

CURRENT 10 AMPS,

MAX. PROTECTION 20 AMPS,

#12 [3.31] AWG

NETWORK

(N)

AREA NETWORK

10/100 BASE T SUPPORTING

TCP/IP PROTOCOL

RJ45F CONNECTION

T I T L E

EQUIPMENT DRAWING, AMSCO 600 , 26.5" X 26.5" X 39", 51" & 63" LG.

SLIDING, SD, CABINET, ELECTRIC STEAM GENERATOR

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:

A - Gielink, Bob 10778442 11008531 B

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

SHEET

REQUIREMENT NOTES:

CIRCUIT BREAKER IS REQUIRED. INSTALL NEAR EQUIPMENT WITHIN EASY REACH OF THE OPERATOR AND MARKED AS

'DISCONNECTING DEVICE' FOR THE EQUIPMENT.

NETWORK CONNECTION (IF APPLICABLE):

- ACTIVE NETWORK DROP WITH CAT 5/6 CABLE WITHIN 10' OF UNIT

- STATIC OR DHCP RESERVED IP ADDRESS (FOR EACH UNIT)

- NETWORK CABLE SUPPLIED BY FACILITY

MAX CONSUMPTION PER CYCLE: 10.6 KW-HR

PHYSICAL

SPECIFICATIONS CHART

CURB WEIGHT

39" LG - 2090 LBS (948 KG)

1. LEVELING FEET ARE PROVIDED FOR PROPER

INSTALLATION. STERILIZER MUST BE LEVELED DURING

INSTALLATION. CHECK LEVELNESS FRONT TO BACK,

SIDE TO SIDE, AND DIAGONAL.

2. STERIS DOES NOT ASSUME ANY RESPONSIBILITY FOR

CHANGES MADE NECESSARY THROUGH FAILURE TO

OBSERVE THE SPECIFICATIONS ON THE EQUIPMENT

DRAWING AND NOTE SHEETS. SPECIFICATIONS AND

DESCRIPTIONS ARE SUBJECT TO CHANGE WITHOUT

NOTICE.

51" LG - 2350 LBS (1066 KG)

63" LG - 2605 LBS (1182 KG)

MAXIMUM OPERATING

WEIGHT BASED ON

CHAMBER FULLY LOADED

WITH INSTRUMENTS

39" LG - 2315 LBS (1050 KG)

51" LG - 2650 LBS (1202 KG)

63" LG - 2980 LBS (1352 KG)

SHIPPING WEIGHT

39" LG - 2640 LBS (1197 KG)

51" LG - 2900 LBS (1315 KG)

63" LG - 3195 LBS (1449 KG)

SHIPPING DIMENSIONS

W x H x L

39" LG - 62 1/2" W x 89 1/4" H x 79 1/2" L [1588mm W x 2267mm H x 2019mm L]

51" LG - 62 1/2" W x 89 1/4" H x 79 1/2" L [1588mm W x 2267mm H x 2019mm L]

63" LG - 62 1/2"W x 89 1/4" H x 91 1/2" L [1588mm W x 2267mm H x 2324mm L]

HEAT LOSS AT 70 F (21 C)

BTU & (kJ) /HR PREVACUUM HEAT LOSS

26.5" X 26.5" X 39" LG

[673 X 673 X 991 LG]

26.5" X 26.5" X 51" LG

[673 X 673 X 1295 LG]

26.5" X 26.5" X 63" LG

[673 X 673 X 1600 LG]

13380 BTU (14117 kJ)

13580 BTU (14328 kJ)

13780 BTU (14539 kJ)

NOISE LEVEL 64 dB A

ADDITIONAL INSTALLATION SPECIFICATIONS NOTES

A. INSTALLATION SPECIFICATION IS LISTED AS ENGINEERING AND INSTALLATION GUIDES. REFERENCED COMPONENTS AND SERVICE CONNECTIONS ARE NOT FURNISHED AS PART OF EQUPMENT UNLESS INSTALLATION IS QUOTED SEPERATELY

AND TO BE PERFORMED BY STERIS.

B. PROVIDE GROUPED OR GANGED CIRCUIT PROTECTION AND DISCONNECT FOR STERILIZER POWER AS REQUIRED BY CODES AND STANDARDS. INDIVIDUAL POWER SHUTOFFS REQUIRED NEAR EACH MACHINE FOR SERVICING.

C. PROVIDE GROUNDED METAL CONDUIT AND WIRING BETWEEN EQUIPMENT TERMINALS AND STUB OUTS OR DISCONNECTS.

CHECK LOCAL CODES FOR MINIMUM AWG. WIRE SIZE, #16 AWG. MINIMUM RECOMMENDED.

D. FOR GENERAL INSTALLATION INFORMATION SEE STERIS ADDENDUM (DRAWING 62941-091) ATTACHED TO THE END OF THIS DOCUMENT. (THIS DRAWING SHOULD ALWAYS ACCOMPANY THE EQUIPMENT DRAWINGS) IF DRAWING IS NOT INCLUDED, CONTACT

STERIS SERVICE ENGINEERING AT 1-800-333-8848 FOR A COPY.

T I T L E

EQUIPMENT DRAWING, AMSCO 600 , 26.5" X 26.5" X 39", 51" & 63" LG.

SLIDING, SD, CABINET, ELECTRIC STEAM GENERATOR

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:

A - Gielink, Bob 10778442 11008531 B

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

SHEET

9-11/16 [246]

42-1/4 [1073]

UNIT WIDTH

'A'

SEE SHEET 1

12 [305] SQ

3-1/2 [89]

TYP

4-5/16 [110]

39-3/8 [1000]

FRAME

22 [559]

11 [279]

8-11/16 [221]

12-5/8 [321]

TYP

1-7/16 [36,5]

12-15/16 [329]

T I T L E

EQUIPMENT DRAWING, AMSCO 600 , 26.5" X 26.5" X 39", 51" & 63" LG.

SLIDING, SD, CABINET, ELECTRIC STEAM GENERATOR

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:

A - Gielink, Bob 10778442 11008531 B

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

SHEET

FLOOR SINK AND SEISMIC ANCHORS

NOTES:

1. ENSURE INSTALLED FLOOR SINK IS ABLE TO HANDLE PEAK

WATER CAPACITY PER SHEET 4. DRAIN FIXTURE AND SINK

TO BE PROVIDED BY OTHERS.

2. FLOOR DRAIN OR SINK IS RECOMMENDED FOR EACH

CUBICLE STERILIZER UNIT, AND SHOULD BE PROVIDED

WITHIN THE CONFINES OF THE STERILIZER FRAMEWORK.

SEE DIMENSIONS SHOWN IN TOP VIEW.

3. USE SEISMIC KIT PART NUMBER 11012047.

4. VERIFY THAT THE CONCRETE SLAB, TO WHICH THE EQUIPMENT

IS SEISMIC ANCHORED, MEETS THE REQUIREMENTS OF THE

APPLICABLE ICC ESR AND OPM CERTIFICATION.

5. ASSURE ACCESS IS AVAILABLE FOR SEISMIC INSTALLATION.

6. ASSURE AT LEAST 18" OF CLEARANCE IS AVAILABLE FROM

BACK OF UNIT FOR SEISMIC ANCHORS INSTALLATION.

TOP VIEW

SEISMIC ANCHOR

6 PLCS

SEE NOTE 3 & 4

STERILIZER

FRAME CORNER

4 PLCS

FLOOR SINK

SEE NOTES 1 & 2

Sheet 1
Views
new_view_1
new_view_2
bottom_3
right_4
Zones
A1
Sheet 2
Views
new_view_15
bottom_16
right_17
Zones
A1
Sheet 3
Views
new_view_21
bottom_22
right_23
Zones
A1
Sheet 4
Zones
A1
Sheet 5
Zones
A1
Sheet 6
Zones
A1
Sheet 7
Views
new_view_24
Zones
A1

File details come from the government source that posted it. Updated .