This sources sought notice requests information from service-disabled veteran owned small businesses and veteran owned small businesses to provide preventative and repair maintenance for Hologic imaging equipment at the Phoenix VA Health Care System. Required services include maintenance for mammography tomography systems, fluoroscopy x-rays, PACS post image processors, and related equipment. Interested parties must submit company information and documentation by August 17, 2020. If a solicitation is issued, the award of a contract is contemplated for October 2020. The anticipated period of performance is a base year plus four option years from November 1, 2020 to October 30, 2025.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Contract Opportunity Sources Sought Notice

PRODUCT SERVICE CODE

SUBJECT

CONTRACTING OFFICE'S

ZIP-CODE

SOLICITATION NUMBER

RESPONSE

DATE/TIME/ZONE

ARCHIVE

DAYS AFTER THE RESPONSE DATE

RECOVERY ACT FUNDS

SET-ASIDE

NAICS CODE

CONTRACTING OFFICE

ADDRESS

POINT OF CONTACT

(POC Information Automatically Filled from User Profile Unless Entered)

DESCRIPTION

See Attachment

AGENCY'S URL

URL DESCRIPTION

AGENCY CONTACT'S EMAIL

ADDRESS

EMAIL DESCRIPTION

ADDRESS

POSTAL CODE

COUNTRY

ADDITIONAL INFORMATION

GENERAL INFORMATION

PLACE OF PERFORMANCE

Required Field Contract Opportunity Sources Sought Notice J065 Hologic Imaging Equipment preventative and repair maintenance.

85014 36C25821Q0003 08-17-2020

1:00PM

PACIFIC TIME, LOS ANGELES, USA

N 811219 Department of Veterans Affairs NCO22-Gilbert Network Contracting 777 E. Missouri, Suite 300 Phoenix AZ 85014 Eric Carroll, Contract Specialist Eric.Carroll@va.gov Department of Veterans Affairs Phoenix VA Healthcare System Carl T. Haydon Medical Center 650 E. Indian School Road Phoenix AZ 85012

USA

Eric.Carroll@va.gov Eric Carroll, Contract Specialist This request for information (RFI) is issued solely for information and planning purposes only and does not constitute a solicitation. All information received in response to this RFI that is marked as proprietary will be handled accordingly. In accordance with FAR 15.201(e), responses to this notice are not offers and cannot be accepted by the government to form a binding contract. Responders are solely responsible for all expenses associated with responding to this RFI.

This notice is published to conduct market research to determine if there is a sufficient number of certified Service-Disabled Veteran Owned Small Businesses (SDVOSB) and/or Veteran Owned Small Businesses (VOSB), capable of performing the requirements to warrant a socio-economic set-aside or full & open competition. It is strongly recommended that interested parties register in the US Department of Veterans Affairs Vet Biz database and the Small Business Administration (SBA) database. Registration may be accomplished at http://www.vip.vetbiz.gov/ and www.sba.gov/, respectively.

Interested parties responding to this Sources Sought request shall submit the following information at a minimum:

(1) company name and address, point of contact with phone number, and DUNS number;

(2) documentation relating to capability of performance

(3) intentions of subcontracting program to include set-asides if any;

(4) proof of registration in the VA Vet Biz or SBA websites or intention of registration prior to submission of offers to a solicitation;

(5) If you are a distributor a statement from the manufacturer that your company is an authorized distributor;

(6) a statement to inform the government if you are a manufacturer or distributor;

(7) any information regarding FSS Schedule availability;

(8) SAM Registration

(9) any other pertinent company documentation.

Responses are due by August 17, 2020 at 1:00 pm PST. It is anticipated that a solicitation may be issued in September 2020. If issued, award of a subsequent contract is contemplated in October 2020. Electronic submissions are acceptable via Eric.Carroll@va.gov, and are highly preferred.

NO ADDITIONAL INFORMATION IS AVAILABLE OTHER THAN WHAT IS NOTED HEREIN. ALL FIRMS ARE INSTRUCTED NOT TO CALL THE INDIVIDUAL VA MEDICAL CENTERS REQUESTING MORE INFORMATION.

This procurement is for a service agreement at the Phoenix VA Health Care System (PVAHCS) to provide annual preventive and repair maintenance on its Hologic imaging equipment that includes: Mammography tomography systems, fluoroscopy x-rays, and PACS post image processors. Reference enclosed equipment listing for quantities and details.

It is requested that responses include a rough estimate of pricing for completion of the requirements identified below.

PURPOSE:

The Phoenix VA Health Care System (PVAHCS) requires a service agreement to provide annual preventive and repair maintenance on its Hologic imaging equipment that includes: Mammography tomography systems, fluoroscopy x-rays, and PACS post image processors, reference enclosed equipment listing and PWS for quantities and details.

REQUIRED SERVICES:

a. Maintenance for Hologic Affirm Breast Biopsy Guidance System

i. Parts coverage

1. Biopsy gun

2. Affirm 3D license

3. Plastics coverage; includes paddle plastics and face shield

4. General replacement parts coverage

b. Maintenance for Hologic ATEC Pearl

i. Parts coverage

1. General replacement parts coverage

2. ATEC Advance Exchange Program

c. Maintenance for Hologic ATEC Sapphire

i. Parts coverage

1. General replacement parts coverage

2. ATEC Advance Exchange Program

d. Maintenance for Hologic Digital CAD software

i. General replacement parts coverage

e. Maintenance for Hologic Selenia Dimensions 3D 8000

i. Parts coverage:

1. Glassware coverage; includes x-ray tube, digital array detector

2. Plastics coverage; includes paddle plastics and face shield

3. C-View coverage

4. General replacement parts coverage

f. Maintenance for Hologic Fluoroscan Insight mini c-arm

i. Parts Coverage

1. Glassware coverage; includes x-ray tube, high voltage power supply assembly, and image intensifier or flat detector

2. Plastic coverage

3. General spare parts coverage

g. Maintenance for SecurView DX-400 workstations

i. General replacement parts coverage

h. Maintenance for Hologic SecurView RT workstations

i. General replacement parts coverage

i. Maintenance for Hologic Trident Specimen Radiography System

i. Parts coverage:

1. Glassware coverage; includes x-ray tube, digital array detector

2. General spare parts coverage

j. For SOW items A – I above the following Scope of Work and Technical Requirements shall be met.

i. Coverage period: M-F, 8:00am-5:00pm Arizona Local Time

ii. Phone response: 30 minutes

iii. On-site response: 8 hours

iv. Parts order requirement: 6:00pm Arizona Local Time

v. Parts delivery: next day

vi. Planned maintenance

vii. Quality assurance

viii. Updates

ix. Labor

x. Travel

xi. Phone and remote diagnostic and repair services

xii. Safety checks

k. The Contractor shall notify the COR to schedule a time and their intent to come on-site.

l. The Contractor shall verify that the equipment that was serviced is fully functional before checking out.

m. The Contractor shall deliver a hard copy or electronic copy of a Field Service Report that delineates the service performed and confirms quality assurance tests results to Agency.

n. The Contract shall inform the COR prior to adding or removal of equipment along with providing documentation for both equipment added or removed.

o. The Contractor shall not remove any hard drives off station. All hard drives shall be turned into the COR.

p. Service technicians shall be certified on/by Hologic equipment to perform any services/maintenance on the enclosed specified equipment. Contractor shall provide documentation/copy of certification with proposal submittals and upon award.

q. Noncontract Charges: The contractor shall not perform any services that shall result in additional charges without prior approval from the Contracting Officer.

r. Payment: Payment shall be made in arrears, and upon receipt of a properly prepared invoice which references the period for which payment is due, the contract number and/or assigned purchase order number. The invoice shall include the correct payment address.

s. Waste Removal: The contractor is responsible to remove and properly dispose of all waste related to the services provide on-site at PVAHCS.

Failure to carry out the above procedures shall be considered as a failure to document the service visit and shall constitute a failure to perform.

DELIVERABLES:

Filed Service Reports shall be provided for all service activities including preventive maintenance, emergency repair, and safety inspections, and Shall contain the following information:

A. Date(s) and time period of service.

B. Complete description of equipment serviced including model number and serial number.

C. Complete description of equipment services performed including upgrades, updates and software changes made.

D. Complete list of all materials, parts, or assemblies installed or replaced including serial numbers of parts where applicable.

E. A statement of the complaint that initiated the service call.

F. Total standard labor hours, total overtime/premium hours, and total travel hours expended during the repair.

EQUIPMENT INVENTORY

L/I
SYSTEM
S/N
EE#
LOCATION
FREQUENCY
A.
Affirm Breast Biopsy
820011611361RM
107865
1236-1
Annual
B.
ATEC Pearl
101372
62015
1236-1
Annual
C.
ATEC Sapphire
202442
79445
1237-1
Annual
D.
Cenova Digital CAD
C2215
87262
1609-1
Annual
E.
Cenova Digital CAD
C5166
87262
1609-1
Annual
F.
Dimensions 3D 8000
81006143576
87260
1236-1
Semi-Annual
G.
Dimensions 3D 8000
SDM130200116
101971
1237-1
Semi-Annual
H.
Insight-FD Fluroroscan
20-1014-06
89559
D314-1
Annual
I.
SecurView DX-400
2980309B2628
74097
A2242A-8
Annual
J.
SecurView DX-400
2980313B4688
74098
1201-1
Annual
K.
SecurView DX-400
2980816B6840
96496
1200-1
Annual
L.
SecurView Manager
2960313B0671
--
--
Annual
M.
SecurView RT
3980511B1055
65018
1216-1
Annual
N.
SecurView RT
3980514B1225

Annual

O.
SecurView RT
3980716B1283
97335
1233-1
Annual
P.
Trident Radiography System
85009140323
89855
1236-1
Annual

PERIOD OF PERFORMANCE:

The anticipated period of performance is a base plus four (4) contract award.

Base Year:11/1/2020-10/30/2021
Option Year 1:11/1/2021-10/30/2022
Option Year 2:11/1/2022-10/30/2023
Option Year 3:11/1/2023-10/30/2024
Option Year 4:11/1/2024-10/30/2025