This notice is a presolicitation for hazardous waste removal services at the Dallas VA Health Care System facilities in Texas. The contractor will be required to provide all labor, materials, equipment, and supervision to remove hazardous waste from the basement of Building 2 and the second floor of the Herzog Building. The work includes asbestos abatement. The NAICS code for this procurement is 236220 for contracting work with a small business size standard of $39.5 million. The project value is estimated between $500,000 and $1 million. This opportunity is 100% set aside for service-disabled veteran-owned small businesses. Interested contractors must be verified as SDVOSBs by the Department of Veterans Affairs Center for Verification and Evaluation prior to proposal submission. The point of contact for this project is Michael Walls, who can be reached by email.

View the file

Other files

Show all 22

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

VA MEDICAL CENTER

DALLAS, TEXAS

BUILDING 2

A.H.E.R.A. ASBESTOS INSPECTION

MANAGEMENT PLAN

3 - YEAR REASSESSMENT & UPDATE - 2014

PREPARED FOR:

VA NORTH TEXAS HEALTH CARE SYSTEM

ENGINEERING SERVICES

4500 S. LANCASTER ROAD

DALLAS, TX 75216

MICRO AIR PROJECT # B-1403

PREPARED BY:

MICRO AIR OF TEXAS, INC.

P.O. Box 1225 Dickinson, TX 77539

(281) 280-9965

DALLAS, TEXAS

AHERA

ASBESTOS INSPECTION

PREPARED FOR:

VA NORTH TEXAS HEALTH CARE SYSTEM

ENGINEERING SERVICES

4500 S. LANCASTER ROAD

DALLAS, TX 75216

PREPARED BY:

MICRO AIR OF TEXAS, INC.

P.O. Box 1225 Dickinson, TX 77539

(281) 280-9965

Original Inspection Dates: September through

December, 2010

Individual Asbestos Consultant: ________________________

DSHS License # 10-5436

TABLE OF CONTENTS

Page #

General Inventory Management Plan Format

Section I - AHERA Asbestos Inspection General Information

Introduction, Required Information, & Facility Description 1 - 3

Section II - Asbestos Methodology & Lab Analytical Report

Methodology 1 Classification 1 Limitations 2 Homogeneous Sampling Area Summary Tables 3 - 13 Laboratory Reports Sample Location Drawing(s) Homogeneous Area Location Drawing(s)

Section III - ACM Material Assessment & Quantification

General Information 1 - 2 Surfacing Materials 3 - 4 Thermal System Insulation Materials 4 - 11 Miscellaneous Materials 11 - 58 Photos of Asbestos-containing Homogeneous Areas

Section IV - Asbestos Material Response Action Recommendations

Introduction 1 Response Action Prioritization 2 Response Action Recommendations 3 - 7

Section V - Operations & Maintenance Program

Purpose 1 Scope 1 Facility ACM 2 - 3 Operations & Maintenance Recommendations 4 - 5 O&M Activities 6

Section VI - Facility Owner / Designated Person(s)

Licenses and Training Certificates

Section VII - Six Month Surveillance Documentation Section VIII - Three Year Reinspection Updates Section IX - Response Actions

BUILDING INSPECTION FOR FRIABLE AND NON-FRIABLE ACM

GENERAL INVENTORY

FACILITY ID: Building 2

ORIGINAL INSPECTION COMPLETION DATES: September through December, 2010 3 YR REASSESSMENT INSPECTION COMPLETION DATE: June through December, 2014

FACILITY ADDRESS: Building 2

VAMC

4500 S. Lancaster Road Dallas, TX 75216

TOTAL SQUARE FOOTAGE: APPROXIMATELY 441,649 SF

YEAR CONSTRUCTED: 1955

USE OF FACILITY: Main Hospital

DESIGNATED PERSON: Thomas Rattler or Ken Taylor

ENGINEERING PHONE: 214-857-1069 - Contact Charles Ballard

INSPECTION COMPANY: MICRO AIR OF TEXAS, INC.

P.O. BOX 1225

DICKINSON, TX 77539

PHONE: (281) 280-9965

NAME OF ACCREDITED PERSONS

PERFORMING INSPECTION: JENNIFER MCMILLIN (DSHS LICENSE # 60-1893)

ACCREDITATION AGENCY: TEXAS DEPT. OF STATE HEALTH SERVICES

ACCREDITATION #: N/A

THIS BUILDING CONTAINS:

1. FRIABLE ACBM X

2. NON-FRIABLE ACBM X

3. FRIABLE MATERIAL

(ASSUMED TO BE ACBM) X

4. NON-FRIABLE MATERIAL

(ASSUMED TO BE ACBM) X

5. NONE OF THE ABOVE

____________________________________ 1/21/11

ACCREDITED INSPECTOR DATE

DSHS LICENSE # 60-1893

BUILDING 2

DALLAS, TX

A.H.E.R.A. ASBESTOS INSPECTION &

MANAGEMENT PLAN

VOLUME I OF III

PREPARED FOR:

VA NORTH TEXAS HEALTH CARE SYSTEM

ENGINEERING SERVICES

4500 S. LANCASTER ROAD

DALLAS, TX 75216

MICRO AIR PROJECT # B-1403

PREPARED BY:

MICRO AIR OF TEXAS, INC.

P.O. Box 1225

DALLAS, TX

A.H.E.R.A. ASBESTOS INSPECTION &

MANAGEMENT PLAN

VOLUME II OF III

PREPARED FOR:

VA NORTH TEXAS HEALTH CARE SYSTEM

ENGINEERING SERVICES

4500 S. LANCASTER ROAD

DALLAS, TX 75216

MICRO AIR PROJECT # B-1403

PREPARED BY:

MICRO AIR OF TEXAS, INC.

P.O. Box 1225

DALLAS, TX

A.H.E.R.A. ASBESTOS INSPECTION &

MANAGEMENT PLAN

VOLUME III OF III

PREPARED FOR:

VA NORTH TEXAS HEALTH CARE SYSTEM

ENGINEERING SERVICES

4500 S. LANCASTER ROAD

DALLAS, TX 75216

MICRO AIR PROJECT # B-1403

PREPARED BY:

MICRO AIR OF TEXAS, INC.

P.O. Box 1225

MICRO AIR OF TEXAS, INC. BUILDING 2 - VA Medical Center; Dallas, TX

AHERA ASBESTOS INSPECTION & MANAGEMENT PLAN

P.O. Box 1225

MANAGEMENT PLAN FORMAT

SECTION I: This section contains general information regarding the AHERA survey and management plan, required information, and a facility description.

SECTION II: This section contains general information, sampling information, and Homogeneous Sample Area Summary. The laboratory analytical report for the bulk samples collected during the 2010 inspection has been included. Additionally, information on the survey report and classification of homogeneous areas is located in this section. Drawings depicting the locations of all bulk building material samples collected and asbestos-containing homogeneous area locations can be found in this sub-section.

SECTION III: This section contains material assessments and quantification which are used to develop the appropriate response actions for the management plan. Photographs of each asbestos-containing material identified has been included in this section.

SECTION IV: This section contains the recommended response actions for each type of ACM in the facility including recommendations for immediate abatement, if necessary. Immediate response action recommendations are prioritized.

SECTION V: This section contains an updated Operations and Maintenance Program for the facility.

This program includes recommendations for ongoing maintenance functions.

SECTION VI: This section has been reserved for informational purposes. This section includes the Facility Owner / Designated Person information, as well as the Re-inspection / Periodic Surveillance Schedule. All applicable licenses, certifications, and accreditations of the Inspector, Management Planner, and Laboratory performing services on this Management Plan are also included in this section.

SECTION VII: This section has been reserved for reports for the six month inspections, assessments, and any ambient air monitoring.

SECTION VIII: This section has been reserved for documentation obtained from three year re-inspections. Updates to the Management Plan are to be included in this section.

SECTION IX: This section has been reserved for the owner to reference abatement response action reports and building maintenance action (O&M) reports. All documentation from asbestos abatement activities should also be included in this section.

SECTION I

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 1

AHERA ASBESTOS INSPECTION

SECTION I - GENERAL INFORMATION

INTRODUCTION

Micro Air of Texas, Inc. (Micro Air) was contracted by the Veterans Administration North Texas Health Care System (VANTHCS) located in Dallas, Texas to perform an asbestos inspection and generate a management plan in accordance with guidelines established in 40 CFR 763, the Asbestos Hazard Emergency Response Act of 1987 (AHERA), for the Veterans Administration Medical Center (VAMC) located in Dallas, Texas. This facility consists of numerous buildings constructed between 1940 and 2009. This particular report pertains to Building 2, which houses the Main Hospital.

Micro Air performed a comprehensive asbestos survey of Building 2 on various dates from September through December, 2010. After the comprehensive asbestos survey results were compiled, a Management Planner accredited by an EPA-approved training provider and licensed as an Individual Asbestos Consultant in the state of Texas, reviewed the analytical results of the inspection and current physical condition assessments. The management planner then determined the potential hazard posed by the ACM and evaluated and selected control / response options based on all compiled information. This information was then composed and written into a final report.

During the 2014 3-Year Asbestos Assessment and Re-inspection, Micro Air noted that Building 2 had been remodeled since the original asbestos inspection on this facility was conducted in 2010. Between 2010 and 2014, the interior of multiple areas in Building 2 were demolished and new building materials were re-installed. Micro Air conducted an asbestos inspection on the interior of Building 2 in order to update the Management Plan. A total of 163 bulk building material samples were collected between June and December, 2014, and the results were included in this management plan.

The federal National Emission Standard for Asbestos (NESHAP) and the Texas Asbestos Health Protection Act (TAHPA) both require that prior to any demolition or renovation activity in a building, a thorough inspection of the affected facility or part of the facility where the demolition or renovation operation will occur, must be performed for the presence of asbestos, including Category I and Category II non-friable asbestos-containing material (ACM).

Additionally, TAHPA requires that a minimum of three samples be collected from each homogeneous sampling area (material type) identified in the building. The work area and immediately surrounding areas which could foreseeably be disturbed by the actions necessary to perform the project must be inspected and sampled as applicable prior to renovations or demolition. While the VANTHCS is not subject to the TAHPA regulation, due to its exclusion as a federal facility, Micro Air collected bulk samples in accordance with TAHPA and AHERA protocol at the request of the VANTHCS engineering personnel.

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 2

REQUIRED INFORMATION

In accordance with AHERA protocol, the VAMC located in Dallas, Texas is required to name a person who shall be responsible for implementation of the management plan (Designated Person - see Section V). Per the Occupational Safety and Health Administration (OSHA) 1926.1101, construction industry standard, maintenance personnel should be made aware of the types and locations of all facility ACM. These employees are required to have adequate asbestos awareness training (minimum 2 hours), and all maintenance personnel who may come in contact with or disturb ACM are required to receive a minimum of 16 hours of asbestos training in asbestos hazards and recognition of ACM. If the owner wishes to perform ACM activities with its own personnel, the owner is responsible for obtaining the required EPA training in accordance with the Model Accreditation Plan (MAP), which is located in Appendix C, Subpart E of 40 CFR, 763 (AHERA). Additionally, all OSHA requirements must be met, as well (i.e., asbestos physical, etc.). Otherwise, it is recommended that all ACM activities be performed by a Texas licensed Asbestos Contractor, and project oversight be conducted by a Texas licensed Asbestos Consultant Agency.

In accordance with the VA directive to comply with AHERA, the entire facility must be re-inspected and a reassessment of ACM conducted by an accredited inspector at least once every three years, and the management plan should be updated by a licensed Asbestos Inspector / Management Planner or Individual Asbestos Consultant. Additionally, visual inspections of the ACM must also be made by the Designated Person for the facility at least once every six months to determine whether any areas of the material have changed in condition, or if there are any other factors which may necessitate abatement actions.

FACILITY DESCRIPTION

Building 2 is approximately 441,649 square feet in size, and it serves as the Main Hospital for the Dallas VA. The construction date of this building is believed to be 1955. The building consists of a sub-basement, basement, and nine floors where Main Hospital activities are conducted, and the 10 and 11 floors, which are mechanical areas. The general exteriorth th construction of the building is a concrete slab foundation with red exterior brick and a flat tar and gravel roof.

The sub-basement consists of dirt floors and concrete walls and ceilings. Pipe insulation was abated from the eastern portion of the sub-basement in the early 1990s; removal of asbestos-contaminated soil was also completed at this time. Other areas of the abatement may have been abated prior to the 1990s; however, the VA has no documentation stating this; therefore, all other areas of the basement are assumed to have asbestos-containing pipe insulation and possibly, asbestos-contaminated soil.

The basement and 1 through 9 floors are typically finished with a variety of flooring materialsst th including vinyl floor tile, sheet vinyl flooring, ceramic tile, rubber flooring material, carpet and concrete. Floor tile and black floor tile mastic are present beneath the carpet in some locations. Walls are generally sheetrock or plaster. Ceilings include several types of lay-in

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 3 ceiling tile, sheetrock, plaster and concrete. Abatement of pipe insulation and floor tile and mastic was completed in the following areas: portions of the basement in 2001 through 2005 and 2007, portions of the 1 Floor in 2000 through 2004 and 2009, portions of the 2 Floorst nd in 2000 through 2003, portions of the 3 Floor in 1997, 1998, 2003 and 2005, portions of therd

4 Floor in 1999 through 2004, one area on the 5 floor in 2007, portions of the 6 Floor inth th th

1996 through 1998, portions of the 7 Floor in 2005, 2009 and 2010, portions of the 8 Floorth th in 2005 and 2006, and one portion of the 9 Floor in 2007. th

Asbestos-containing pipe and fitting insulation (steam, hot water, chilled water, etc.) has been abated from numerous areas inside Building 2. Additionally, HVAC duct insulation with black mastic and floor tile and black mastic have also been abated from various areas inside the building. See Section IX for more specific information regarding areas that have been abated.

HVAC air handlers were observed to have rubber vibration dampeners and friable asbestos-containing cloth vibration dampeners. HVAC ducts with insulation and black mastic were observed throughout the building. Remnant sections of original asbestos-containing steam pipe insulation were noted above the ceiling in numerous areas. Access to areas located above the plaster ceilings was limited; therefore, it is assumed that pipes exist above the solid ceilings and in wall chases. Recessed wall radiators are located primarily on the 7 , 8 , andth th

9 floors of the building, and they are assumed to have asbestos-containing insulation locatedth behind the radiator in the recessed wall cavity. It also appears that some of the recessed wall radiators may have been removed from some of the rooms on the 8 floor, as there is a metalth plate over the opening on the wall where the radiator(s) once existed.

2014 3-Year Asbestos Reassessment and Update

The interior of Building 2 was remodeled between 2010 and 2014. Based on information provided by the VAMC, the interior building materials of Building 2 were removed, and the interior of this building was remodeled at the following locations: Basement Suite BB02, 1st

Floor Suite 1C-421, 2 Floor Geropsych, 7 Floor Main Elevator Lobby, 7 Floor A-Wing andnd th th

C-Wing, and 9 Floor Elevator Lobby. th

SECTION II

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 1

SECTION II- ASBESTOS LABORATORY ANALYTICAL REPORT

METHODOLOGY

Micro Air of Texas, Inc. performed a comprehensive asbestos survey of Building 2 on several dates from September through December 2010 in accordance with AHERA and TAHPA sampling protocol and assessment factors. The inspection was performed by Melissa Kiser, an accredited asbestos inspector and management planner who is licensed as an Individual Asbestos Consultant by the state of Texas (license #10-5436). Ms. Kiser was assisted by Jennifer McMillin, David Easley, and Elizabeth Bargar, accredited asbestos inspectors who are licensed as Building Inspectors. Micro Air of Texas, Inc. is licensed as an Asbestos Consulting Agency (#10-0110). All samples collected during this inspection were analyzed by J Resources, Inc., whose laboratory is accredited by the National Institute of Standards3 and Technology (NIST) under the National Voluntary Laboratory Accreditation Program (NVLAP) program (accreditation #200525-0) and licensed by the state of Texas (license #30- 0273). Laboratory analysis of collected samples was performed in accordance with EPA and NVLAP prescribed methodologies for Polarized Light Microscopy of bulk material samples for the presence of asbestos. The results of laboratory analysis of collected bulk material samples are included in this section for reference.

A total of 807 bulk material samples were collected during this comprehensive survey, and 777 of the collected samples were analyzed by PLM. Federal regulations require that each identifiable layer of an individual bulk building material sample be analyzed separately for asbestos content. In addition, federal regulations state that if any one sample from a homogeneous area is found to be asbestos-containing, then the entire homogeneous area is considered asbestos-containing, regardless of the results from the remaining samples from that homogeneous area.

Between June and December , 2014, Micro Air of Texas conducted additional asbestos surveys in various locations of Building 2 that were renovated (see page 3 above for locations). The inspections were performed by Jennifer McMillin. A total of 163 bulk material samples were collected and analyzed by PLM during the asbestos surveys. All of the samples were analyzed by J Resources, Inc.3

CLASSIFICATION

All materials which were sampled as being suspect for containing asbestos or were assumed to be asbestos-containing were grouped by sample type into Homogeneous Sampling Areas for purposes of assessment and quantification. Homogeneous materials are those which have similar appearance, texture, and function. If the inspector believes these materials to be the same, even if they are located in separate areas of a building, they may be grouped into the same sampling set. For purposes of this inspection report, the Homogeneous Area is designated as : HA. Each Homogeneous Area is given a number to distinguish it from other Homogeneous Areas (Example: HA-01A).

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 2

A CAD drawing(s) demonstrating the location of each collected bulk building material sample has been included in this section, as well as a CAD drawing(s) depicting the location of asbestos-containing homogeneous areas. CAD drawings showing the approximate locations of areas that have been abated in the sub-basement, basement, and the 1 through 9 floorsst th can be found in Section IX of this Management Plan.

LIMITATIONS

Micro Air’s scope of services was limited primarily to accessible interior building materials.

While Micro Air made every effort to perform a thorough survey, the possibility exists that materials containing asbestos were not identified, inaccessible, or different from those identified at other locations. The findings and opinions presented are indicative of conditions on the date of our site inspection and should not be relied upon to represent conditions at substantially later dates, due to possible changes in the condition of building materials, renovations, and repairs. Additionally, most of the buildings located at this VA facility have steam pipes running underground to supply heat to each building. This material was inaccessible at the time of this inspection. All pipes located in wall chases, underground, in the crawlspace of a building, etc., is assumed to be ACM until tested and proven otherwise.

Micro Air was informed by the VAMC locksmith that there are several wooden doors located in Building 2 that have insulation inside the door. Micro Air did not sample any wooden doors during this survey; however, all wooden doors located inside this building should be considered suspect for containing asbestos-containing interior insulation. All doors should be sampled for ACM prior to being renovated or demolished.

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 3

VAMC - DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

THE FOLLOWING TABLE DENOTES THE HOMOGENEOUS AREAS

IDENTIFIED IN BUILDING 2. THE TABLE INDICATES THE NUMBER OF

BULK BUILDING MATERIAL SAMPLES COLLECTED FOR EACH

HOMOGENEOUS AREA AND WHETHER THE HOMOGENEOUS AREA WAS

ASBESTOS-CONTAINING. ADDITIONALLY, ESTIMATED QUANTITIES OF

THE VARIOUS BUILDING MATERIALS SAMPLED HAVE ALSO BEEN

INCLUDED IN THE TABLE.

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 4

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-01 Texture on Sheetrock No 6 880 SF

HA-02 Sprayed-On Fire Proofing No 7 6,018 SF

HA-03 Friable Steam Pipe, Hot Water Pipe, etc.

Insulation - Old

Appearance (Straight Runs and Fittings)

Yes 27 50,000 LF

HA-04 Yellow Fiberglass Pipe Insulation with Paper or Canvas Wrap and White

Mastic (Straight Runs and Hard Fittings)

Yes 14 20,000 LF

HA-05 Yellow Fiberglass Pipe Insulation with White Mastic Wrap (Straight

Runs and PVC Elbows)

No 105 55,744 LF

HA-06 Thin, Painted Pipe Insulation, Cream with

Rough Texture

No 6 350 LF

HA-07 HVAC Fiberglass Duct Insulation with Foil and

Canvas Wrap and White Mastic/Paint

No 26 14,836 SF

HA-08 HVAC Fiberglass Duct Insulation with Foil Wrap and Black Mastic

Yes 26 25,000 SF

HA-09 HVAC Fiberglass Duct Insulation with Black

Paper and Tar

No 3 240 SF

HA-10 HVAC Fiberglass Duct Insulation with Paper

Wrap and Black Mastic

Yes 3 8000 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 5

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-11 HVAC Vibration Dampener, Black

No 2 200 SF

HA-12 HVAC Vibration Dampener, Beige

No 2 200 SF

HA-95A HVAC Duct Seam Sealant, Gray

No 30 645 SF

HA-95B HVAC Duct Seam Sealant, Gray (Basement

Location)

Yes 1 5000 SF

HA-96 HVAC Duct Seam Sealant, White

No 3 2000 SF

HA-97 HVAC Duct Seam Sealant, Brown

Yes 6 5000 SF

HA-13 Radiators (Recessed) Assumed

ACM

Not Accessible

2,025 SF

HA-14 Thermal Pad Insulation Under Tanks

No 3 200 SF

HA-15 2' x 2' Ceiling Tile, White with Short Fissures and Pinholes Pattern, Lay-In

(Type I)

No 23 228,545 SF

HA-16 2' x 2' Ceiling Tile, White with Crow’s Feet Pattern, Lay-In (Type II)

No 6 18,011 SF

HA-17 2' x 2' Ceiling Tile, White with Tight Fissures and Pinholes Pattern, Lay-In

(Type III)

No 3 8,322 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 6

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-18 2' x 2' Ceiling Tile, White with Rough Texture, Lay-

In (Type IV)

No 5 2,704 SF

HA-19 2' x 2' Ceiling Tile, White Sheetrock with Vinyl

Covering, Lay-In (Type V)

No 7 11,382 SF

HA-20 2' x 2' Ceiling Tile, White with Shallow Fissures and Gray Brittle Backing, Lay-

In (Type VI)

No 3 1,230 SF

HA-21 4' x 2' Ceiling Tile, White with Tight Fissures and

Pinholes Pattern, Lay-In (Type VII)

No 5 22,140 SF

HA-22 2' x 2' Ceiling Tile, White, with Long Fissures and

Pinholes Pattern, Lay-In (Type VIII)

No 5 8,551 SF

HA-23 2' x 2' Ceiling Tile, White, with Tiny Fissures and

Pinholes Pattern, Lay-In (Type IX)

No 3 5,440 SF

HA-24 2' x 2' Ceiling Tile, White, with 1' x 1' Inset Pattern, Lay-In (Type X)

No 3 948 SF

HA-25 2' x 2' Ceiling Tile, White, with Tiny Pinholes

Pattern, Lay-In (Type XI)

No 3 532 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 7

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-26 2' x 2' Ceiling Tile, White, with Shallow Texture, Lay-

In (Type XII)

No 3 4,222 SF

HA-27 Plaster and Scratch Coat No 20 252,352 SF

HA-28A Sheetrock and Joint Compound

No 74 641,102 SF

HA-28B Sheetrock and Joint Compound (Portions of 3rd and 8 Floors)th

Yes 20 57,000 SF

HA-29 12" x 12" Vinyl Floor Tile, White with Pastel Specks and Yellow Mastic (Type I)

No 17 66,639 SF

HA-30 12" x 12" Vinyl Floor Tile, White with Beige Marble

Pattern and Yellow Mastic (Type II)

No 3 1,915 SF

HA-31 12" x 12" Vinyl Floor Tile, Dark Blue with Black

Mastic (Type III)

Yes (Mastic Only)

3 876 SF

HA-32 12" x 12" Vinyl Floor Tile, Orange with Black Mastic

(Type IV)

Yes (Mastic Only)

3 876 SF

HA-33 12" x 12" Vinyl Floor Tile, Light Blue with Black

Mastic (Type V)

Yes (Mastic Only)

3 288 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 8

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-34 12" x 12" Vinyl Floor Tile, Tan, Beige and Gray with

Black Mastic (Type VI)

Yes (Mastic Only)

9 19,881 SF

HA-35 12" x 12" Vinyl Floor Tile, Tan, Beige and Gray with

Yellow Mastic (Type VII)

No 11 14,049 SF

HA-36 12" x 12" Vinyl Floor Tile, Terrazzo Pattern (Black, Brown and White) with

Black Mastic (Type VIII)

Yes (Mastic Only)

1 115 SF

HA-37 Sheet Vinyl Flooring, White with Pastel Specks and Cream and Black Mastic

(Type IX)

No (<1%

Asbestos in the black mastic)

3 320 SF

HA-38 12" x 12" Vinyl Floor Tile, Cream, Tan and Yellow with Black Mastic (Type X)

Yes (Mastic Only)

3 2,539 SF

HA-39 12" x 12" Vinyl Floor Tile, Tan with Blue and Green Streaks and Black Mastic

(Type XI)

Yes (Mastic Only)

3 1,702 SF

HA-40 Sheet Vinyl Flooring, Gray with Multi-Colored Specks and Diamond Plate Design

(Type XII)

No 3 256 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 9

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-41 Sheet Vinyl Flooring, Green Foamy Pattern with

Canvas Backing (Type XIII)

No 3 160 SF

HA-42 Sheet Vinyl Flooring, White, Blue and Green with Brown Mastic (Type XIV)

No 3 128 SF

HA-43 12" x 12" Vinyl Floor Tile, White and Beige Marble

Pattern with Black Mastic (Type XV)

Yes (Mastic Only)

5 11,943 SF

HA-44 12" x 12" Vinyl Floor Tile, White with Pastel Specks and Black and Yellow Mastic

(Type XVI)

Yes (Mastic Only)

5 5,700 SF

HA-45 Sheet Vinyl Flooring, Beige with Gray Specks

(Type XVII)

No 4 854 SF

HA-46 12" x 12" Vinyl Floor Tile, White with Blue and Pink Short Streaks and Yellow

Mastic (Type XVIII)

No 3 2,008 SF

HA-47 12" x 12" Vinyl Floor Tile, Terrazzo Pattern (Black, Brown and White) with

Brown Mastic (Type XIX)

No 4 795 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 10

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-48 12" x 12" Vinyl Floor Tile, White with Black and

Gray Streaks and Black Mastic

(Type XX)

Yes (Both Floor

Tile and Mastic)

6 1,590 SF

HA-49 12" x 12" Vinyl Floor Tile, Grayish-Green with Cream Streaks and

Brown Mastic (Type XXI)

No 4 296 SF

HA-50 Stair Tread, Gray with Brown Mastic (Type XXII)

No 4 300 SF

HA-51 Wood Laminate, Medium Brown with Yellow Mastic

(Type XXIII)

No 5 2,859 SF

HA-52 12" x 12" Vinyl Floor Tile, Brown with Black Pinhole Pattern and Yellow Mastic

(Type XXIV)

No 3 7,184 SF

HA-53 Sheet Vinyl Flooring, Dark Blue with Black and Gray

(Type XXV)

No 3 7,004 SF

HA-54 12" x 12" Vinyl Floor Tile, White with Tan and Gray

Streaks and Yellow Mastic (Type XXVI)

No 3 2,612 SF

HA-55 9" x 9" Vinyl Floor Tile, Olive Green with White

Streaks and Black Mastic (Type XXVII)

Yes (Both Floor

Tile and Mastic)

3 64 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 11

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-56 9" x 9" Vinyl Floor Tile, Red with White Streaks and Black Mastic (Type XXVIII)

Yes (Both Floor

Tile and Mastic)

3 1,520 SF

HA-57 12" X 12" Vinyl Floor Tile, Cream with Small Tan

Streaks and Specks with Black Mastic (Type XXIX)

Yes (Both Floor

Tile and Mastic)

5 10,186 SF

HA-58 16" x 16" Vinyl Floor Tile, Green with Yellow Mastic

(Type XXX)

No 3 288 SF

HA-59 24" x 24" Rubber Squares, Cream with Blue and Green Specks and Cream Mastic (Type XXXI)

No 3 4,489 SF

HA-60 Sheet Vinyl Flooring, Tan with Tiny Green and

Brown Specks (Type XXXII)

No 3 1,780 SF

HA-61 12" x 12" Vinyl Floor Tile, White with Blue and Gray Streaks and Yellow Mastic

(Type XXXIII)

No 3 1,237 SF

HA-62 12" x 12" Vinyl Floor Tile, Tan with Blue and Gray

Streaks and Black Mastic (Type XXXIV)

Yes (Mastic Only)

8 21,652 SF

HA-63 12" x 12" Vinyl Floor Tile, Dark Blue with Brown

Mastic (Type XXXV)

No 3 1,340 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 12

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-64 Stair Tread, Cream With Brown Spots and Brown

Mastic (Type XXXVI)

No 3 276 SF

HA-65 12" x 12" Vinyl Floor Tile, Orange with Brown Mastic

(Type XXXVII)

No 3 1,152 SF

HA-66 12" x 12" Vinyl Floor Tile, Light Blue with Yellow

Mastic (Type XXXVIII)

No 3 2,976 SF

HA-67 12" x 12" Vinyl Floor Tile, White with Medium Blue Long Streaks and Black

Mastic (Type XXXIX)

Yes (Both Floor

Tile and Mastic)

3 1,875 SF

HA-68 12" x 12" Vinyl Floor Tile, White with Black Spots and Yellow Mastic (Type XL)

No 3 360 SF

HA-69 12" x 12" Vinyl Floor Tile, Tan with Tight Gold and

Brown Specks and Brown Mastic

(Type XLI)

No 3 80 SF

HA-70 12" x 12" Vinyl Floor Tile, Tan with Tiny Tight Black Specks and Yellow Mastic

(Type XLII)

No 3 750 SF

HA-71 12" x 12" Vinyl Floor Tile, White with Tan Specks and Yellow Mastic (Type XLIII)

No 3 2,292 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 13

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-72 12" x 12" Vinyl Floor Tile, White with Blue Long

Streaks, Green and Tan Streaks and Yellow Mastic

(Type XLIV)

No 3 42 SF

HA-73 12" x 12" Vinyl Floor Tile, White with Tight Green and Olive Specks with

Brown Mastic (Type XLV)

No 3 1,344 SF

HA-74 12" x 12" Vinyl Floor Tile, White with Tiny Black

Specks and Yellow Mastic (Type XLVI)

No 3 171 SF

HA-75 12" x 12" Vinyl Floor Tile, Gray with Tiny Black

Specks and Yellow Mastic (Type XLVII)

No 3 171 SF

HA-76 12" x 12" Vinyl Floor Tile, Green with Lighter Green Specks and Yellow Mastic

(Type XLVIII)

No 3 374 SF

HA-77 Sheet Vinyl Flooring, Blue and White Marbled

Pattern (Type XLIX)

No 3 576 SF

HA-78 12" x 12" Vinyl Floor Tile, White with Dark Blue and

Light Blue Specks and Yellow Mastic

(Type L)

No 3 3,858 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 14

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-79 12" x 12" Vinyl Floor Tile, Tan with Tight Gold and Brown Specks and Black

Mastic (Type LI)

Yes (Mastic Only)

3 1,312 SF

HA-80 12" x 12" Vinyl Floor Tile, Dark Rose with Yellow

Mastic (Type LII)

No 3 420 SF

HA-81 Mastic On Concrete, Black, Yellow and/or

Brown (Beneath Carpet in Some Locations)

Yes 27 31,366 SF

HA-82 Cove Base Mastic, Tan (Type I)

No 22 55,634 LF

HA-83 Cove Base Mastic, Yellow (Type II)

No 4 854 LF

HA-84 Cove Base Mastic, Tan with Dark Brown

(Type III)

No 7 6,620 LF

HA-85 Firestop, Red No 21 343 SF

HA-86 Sink Soundproofing, Gray No 7 12 SF

HA-87 Sink Soundproofing, Black

No 21 40 SF

HA-88A Sink Soundproofing, White

No 17 34 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 15

VAMC; DALLAS, TX

BUILDING 2

HOMOGENEOUS SAMPLE AREA SUMMARY

Homogeneous Area #

Sample Type Asbestos-containing?

# of Samples Collected

Est.

Quantity

HA-88B Sink Soundproofing, White

Yes 3 6 SF

HA-89 Sink Soundproofing, Pink Yes 1 2 SF

HA-90 Laboratory Counter Tops, Black

Yes 3 600 SF

HA-91 Transite Interior Vent Hood Wall Material, Black

Yes 1 64 SF

HA-92 Mirror Mastic, Black Yes 3 14 SF

HA-93 Mirror Mastic, Yellow No 3 15 SF

HA-94 Transite, Gray Assumed Asbestos

Not Accessible

312 SF

NOTE: SOME OF THE ABOVE BUILDING MATERIALS WERE REMOVED DURING

RENOVATION PROJECTS THAT OCCURRED DURING 2010-2014. INTERIOR COMPONENTS

OF BUILDING 2 WERE REMOVED AND NEW BUILDING MATERIALS WERE INSTALLED AT

THE FOLLOWING LOCATIONS IN BUILDING 2: BASEMENT SUITE BB02,1ST FLOOR SUITE

1C-421, 2 FLOOR GEROPSYCH, 7 FLOOR MAIN ELEVATOR LOBBY, 7 FLOOR A-WINGND TH TH

AND C-WING, AND 9 FLOOR ELEVATOR LOBBY. BELOW ARE THE SUMMARY TABLESTH

OF THE NEWLY INSTALLED BUILDING MATERIALS.

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 16

VAMC; DALLAS, TX

BUILDING 2 - BASEMENT - SUITE BB02

HOMOGENEOUS SAMPLE AREA SUMMARY

NOVEMBER, 2014 ASSESSMENT & RE-INSPECTION

Homogeneous Area #

Sample Type Asbestos-containing

# of Samples Collected

Est.

Quantity

HA-14-B-01 Red Fire Stop No 3 <10 SF

HA-14-B-02 Dark Brown Vinyl Flooring No 3 120 SF

HA-14-B-03 Tan Cove Base Adhesive No 3 350 LF

HA-14-B-04 2' x 2' White Ceiling Tile No 3 1,500 SF

HA-14-B-05 Sheetrock & Joint Compound

No 3 1,600 SF

VAMC; DALLAS, TX

BUILDING 2 - 1 FLOOR - SUITE 1C-421st

HOMOGENEOUS SAMPLE AREA SUMMARY

NOVEMBER, 2014 ASSESSMENT & RE-INSPECTION

Homogeneous Area #

Sample Type Asbestos-containing

# of Samples Collected

Est.

Quantity

HA-14-01-01 Tan Cove Base Adhesive No 3 1,000 LF

HA-14-01-02 Brown Vinyl Flooring No 3 2,500 SF

HA-14-01-03 2' x 2' White Ceiling Tile with Pinholes & Fissures

No 3 3,000 SF

HA-14-01-04 Red Fire Stop No 3 20 SF

HA-14-01-05 Sheetrock & Joint Compound

No 3 3,500 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 17

VAMC; DALLAS, TX

BUILDING 2 - 2 FLOOR - GEROPSYCHND

HOMOGENEOUS SAMPLE AREA SUMMARY

JUNE, 2014 ASSESSMENT & RE-INSPECTION

Homogeneous Area #

Sample Type Asbestos-containing

# of Samples Collected

Est.

Quantity

HA-14-02-01 Tan Adhesive on Blue Cove Base (Type I)

No 3 1,000 LF

HA-14-02-02 Yellow Adhesive on Gray Cove Base (Type II)

No 3 1,000 LF

HA-14-02-03 Tan Adhesive on Green Cove Base (Type III)

No 3 1,000 LF

HA-14-02-04 2' x 2' White Ceiling Tile with Pinholes & Fissures

No 3 10,000 SF

HA-14-02-05 Yellow Adhesive on Vinyl Wood Style Cove Base

(Type IV)

No 3 1,000 LF

HA-14-02-06 12" x 12" White Floor Tile With Pastel Specks &

Adhesive

No 3 5,000 SF

HA-14-02-07 Wood Style Vinyl Flooring with Adhesive

No 3 5,000 SF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 18

VAMC; DALLAS, TX

BUILDING 2 - 7 FLOOR TH

HOMOGENEOUS SAMPLE AREA SUMMARY

JUNE, 2014 ASSESSMENT & RE-INSPECTION

Homogeneous Area #

Sample Type Asbestos-containing

# of Samples Collected

Est.

Quantity

HA-14-07-01 Fiberglass Pipe Insulation with White Mastic & Paper

Wrap

No 9 1,000 LF

HA-14-07-02 Gray Seam Sealant on Metal HVAC Duct

No 3 300 SF

HA-14-07-03 2' x 2' White Ceiling Tile with Pinholes & Fissures at Main Elevator Lobby

No 3 2,000 SF

HA-14-07-04 Yellow Adhesive on Gray Cove Base at Main

Elevator Lobby

No 3 5,000 LF

HA-14-07-05 Sheetrock, Joint Compound, & Tape at Main Elevator Lobby

No 3 5,000 SF

HA-14-07-06 Wood Style Vinyl Flooring with Adhesive at Main

Elevator Lobby

No 3 1,000 SF

HA-14-07-07 Red Fire Stop at Main Elevator Lobby

No 3 50 SF

HA-14-07-08 Fiberglass Pipe Insulation with Paper Wrap & Mastic on Straight Runs

No 6 2,000 LF

HA-14-07-09 Gray Seam Sealant on Metal HVAC Return Ducts

No 6 300 SF

HA-14-07-10 Seamless Wood Style Vinyl Flooring with

Adhesive In C-Wing

No 3 5,000 SF

HA-14-07-11 Wood Style Vinyl Flooring with Adhesive In C-Wing

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 19

VAMC; DALLAS, TX

BUILDING 2 - 7 FLOOR TH

HOMOGENEOUS SAMPLE AREA SUMMARY

JUNE, 2014 ASSESSMENT & RE-INSPECTION

Homogeneous Area #

Sample Type Asbestos-containing

# of Samples Collected

Est.

Quantity

HA-14-07-12 12" x 12" White Floor Tile With Pastel Specks & Adhesive in C-Wing

No 3 5,000 SF

HA-14-07-13 Sheetrock, Joint Compound, & Tape

No 15 30,000 SF

HA-14-07-14 Red Fire Stop in A & C Wings

No 6 100 SF

HA-14-07-15 2' x 2' White Ceiling Tile with Pinholes & Fissures in C-Wing

No 3 10,000 SF

HA-14-07-16 No Drip Tape No 1 10 LF

HA-14-07-17 Fiberglass HVAC Duct Insulation with White

Wrap and Mastic

No 3 2,000 LF

HA-14-07-18 12" x 12" White Floor Tile With Pastel Specks & Adhesive in A-Wing

No 3 5,000 SF

HA-14-07-19 Tan Adhesive on Gray Cove Base

No 3 5,000 SF

HA-14-07-20 2' x 2' White Ceiling Tile with Pinholes & Fissures in A-Wing

No 3 10,000 SF

HA-14-07-21 Gray Sink Soundproofing No 1 50 SF

HA-14-07-22 Wood Style Vinyl Flooring With Adhesive in A-Wing

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 20

VAMC; DALLAS, TX

BUILDING 2 - 7 FLOOR TH

HOMOGENEOUS SAMPLE AREA SUMMARY

JUNE, 2014 ASSESSMENT & RE-INSPECTION

Homogeneous Area #

Sample Type Asbestos-containing

# of Samples Collected

Est.

Quantity

HA-14-07-23 Seamless Wood Style Vinyl Flooring with

Adhesive in A-Wing

No 3 5,000 SF

HA-14-07-24 Yellow Adhesive on Gray Cove Base in C-Wing

No 3 5,000 LF

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 21

VAMC; DALLAS, TX

BUILDING 2 - 9 FLOOR - ELEVATOR LOBBYTH

HOMOGENEOUS SAMPLE AREA SUMMARY

JUNE, 2014 ASSESSMENT & RE-INSPECTION

Homogeneous Area #

Sample Type Asbestos-containing

# of Samples Collected

Est.

Quantity

HA-14-09-01 Fiberglass Pipe Insulation with White Mastic on

Straight Runs

No 2 1,000 LF

HA-14-09-02 Sheetrock, Joint Compound, & Tape

No 3 3,000 SF

HA-14-09-03 Tan Vinyl Flooring with White Adhesive

No 3 2,000 SF

HA-14-09-04 2' x 2' White Ceiling Tile with Fissures & Pinholes

No 3 3,000 SF

HA-14-09-05 Tan Cove Base Adhesive No 3 1,000 LF

HA-14-09-06 Red Fire Stop No 3 10 SF

SECTION III

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 1

SECTION III - ACM Material Assessment and Quantification

Each area where ACM was identified in the asbestos management plan was visually inspected and assessed as to physical condition and potential for airborne asbestos fiber release.

Included in the assessment factors are such items as friability, accessibility, amount of current damage, amount of airflow, vibration, or abrasion to which the material is subject, potential for water damage, type and number of personnel normally located within each homogeneous sampling area, proximity to normal maintenance activities and potential for disturbance of such materials under normal operating conditions, potential for deterioration, and overall health risk of the material.

Final hazard assessment ratings are based upon the system designed in the EPA AHERA protocol for asbestos building inspectors. The material assessment categories in order of decreasing critical importance are as follows:

1. Damaged or Significantly Damaged TSI ACM

2. Damaged Friable Surfacing ACM

3. Significantly Damaged Friable Surfacing ACM

4. Damaged or Significantly Damaged Friable Miscellaneous ACM

5. ACBM with the Potential for Damage

6. ACBM with the Potential for Significant Damage

7. Any Remaining Friable ACBM or Friable Suspected ACBM

These assessment categories were used to determine appropriate response actions for all ACM, which are included in Section II of this report.

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 2

The following accredited Inspectors / Licensed Individual Asbestos Consultant has assessed the ACM located within this facility in accordance with the above AHERA requirements.

Name: Jennifer McMillin Signature

License #: 60-1893 January 21, 2011 Date

Accreditation #: N/A

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 3

HOMOGENEOUS AREA SUMMARY AND MATERIAL

ASSESSMENT - Building 2

SURFACING MATERIALS

DESCRIPTION OF MATERIAL(S) INCLUDED IN THIS GROUP: TROWELED-ON OR

SPRAYED ON SURFACING MATERIALS

HA-01: DESCRIPTION OF MATERIAL: TEXTURE ON SHEETROCK

This sampling area includes a thin, painted texture with a cream color and rough surface located on the 8 Floor inth

Room A845, 2 Floor in Rooms 2C-403E and 2C-403N,nd and in the Basement in Room BB15 (storage).

NUMBER OF SAMPLES TAKEN: 6

SAMPLE ID NUMBERS: 02-92, 02-93, 02-94, 02-672, 02-673 & 02-

QUANTITY OF MATERIAL: APPROX. 880 SF

MATERIAL ASSESSMENT: This material was found not to be asbestos-containing, therefore no assessment or response action is necessary.

HA-02: DESCRIPTION OF MATERIAL: SPRAYED-ON FIRE PROOFING

This sampling area includes gray, sprayed-on fire proofing material located in a vertical chase located behind the ABC / Telecare Rooms on floors 2 through 9. It can also be found on the 2 floor in the Geropsych Addition. nd

NUMBER OF SAMPLES TAKEN: 7

SAMPLE ID NUMBERS: 02-24, 02-25, 02-26, 02-407, 02-408, 02-409,

and 02-804 through 02-807

QUANTITY OF MATERIAL: APPROX. 6,018 SF

MATERIAL ASSESSMENT: This material was found not to be asbestos-containing, therefore no assessment or response action is necessary.

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 4

THERMAL SYSTEM INSULATION

DESCRIPTION OF MATERIAL(S) INCLUDED IN THIS GROUP: PIPE AND FITTING

INSULATION & HVAC DUCT INSULATION MATERIALS

HA-03: DESCRIPTION OF MATERIAL: FRIABLE STEAM PIPE, HOT WATER PIPE,

ETC. INSULATION, OLD APPEARANCE

(STRAIGHT RUNS AND FITTINGS)

This sampling area includes all friable steam pipe, hot water pipe, etc. insulation (straight runs and fittings) located throughout the building, including the Sub-Basement and all inaccessible locations, such as chases, above or behind plaster ceilings / walls, and in crawl space areas under the building.

This ACM has been abated from various locations of Building 2. Some abatement has been small scale (i.e., glovebag removal of a fitting, etc.), and some of the abatement has been on a large scale basis (i.e., several rooms or entire wings). Please see Section IX of the Management Plan to view the large scale locations where this ACM was abated.

NUMBER OF SAMPLES TAKEN: 27

SAMPLE ID NUMBER: 02-01, 02-02, 02-03, 02-10 through 02-15, 02-127, 02-160, 02-161, 02-232, 02-233, 02- 276, 02-277, 02-278, 02-331, 02-333, 02- 334, 02-335, 02-614, 02-615, 02-616, 02- 617, 02-770 & 02-780

QUANTITY OF MATERIAL: 50,000 LF

MATERIAL ASSESSMENT: This material was found to be asbestos-containing. This material is classified as friable ACM, and the physical condition of the material was noted to be significantly damaged in localized areas. This material has been assessed and determined to present a significant fiber release potential if disturbed or further damaged in any way.

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 5

HA-04: DESCRIPTION OF MATERIAL: YELLOW FIBERGLASS PIPE INSULATION

WITH PAPER OR CANVAS WRAP AND

WHITE MASTIC (STRAIGHT RUNS AND

HARD FITTINGS)

This sampling area includes fiberglass pipe insulation with paper or canvas wrap, white mastic and hard mud fittings located throughout the building. Some of the samples tested negative for asbestos content, while some tested positive for asbestos (mastic wrap and/or fittings). It is impossible to determine which white or beige mastic with hard fittings is negative and which is positive for asbestos;

therefore, all of the fiberglass pipe insulation with white or beige mastic wrap and hard fittings located throughout the building is considered to be asbestos-containing and included in this homogeneous area.

This ACM has been abated from various locations of Building 2. Some abatement has been small scale (i.e., glovebag removal of a fitting, etc.), and some of the abatement has been on a large scale basis (i.e., several rooms or entire wings). Please see Section IX of the Management Plan to view the large scale locations where this ACM was abated.

NUMBER OF SAMPLES TAKEN: 14

SAMPLE ID NUMBER: 02-27, 02-28, 02-29, 02-82, 02-83, 02-207,

02-208, 02-332, 02-410, 02-411, 02-412, 02- 413, 02-414, 02-517, 02-518, 02-621, 02- 634, 02-635 & 02-636

QUANTITY OF MATERIAL: 20,000 LF

MATERIAL ASSESSMENT: This material was found to be ACM. This material is classified as Category II, non-friable ACM, and it is considered to be ACM with the potential for damage. This material does not present a significant potential for asbestos fiber release in its present condition nor under normal operating circumstances.

Therefore, no abatement response action is recommended for this non-friable material at this time. All ACM has the potential, however, for significant damage and release of airborne asbestos fibers if improperly

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 6 disturbed in any way. This material should be continuously assessed and managed under the Management Plan until such time as their condition changes or plans for renovation or demolition require their removal.

HA-05: DESCRIPTION OF MATERIAL: YELLOW FIBERGLASS PIPE INSULATION

WITH WHITE MASTIC WRAP (STRAIGHT

RUNS AND PVC ELBOWS)

This sampling area includes yellow fiberglass pipe insulation with white mastic wrap (straight runs and PVC elbows) located on Floors 1 through 9, and in the Basement and Sub-Basement (abated portions). This insulation is a newer variety, and it appears to have been used to re-insulate pipes that were previously abated, and / or it has been used in new additions / renovated areas in Building 2.

NUMBER OF SAMPLES TAKEN: 105

SAMPLE ID NUMBERS: 02-30 through 02-38, 02-84, 02-85, 02-86, 02-124, 02-125, 02-126, 02-164 through 179, 02-212 through 219, 02-279 through 286, 02-336 through 340, 02-417 through 02-423, 02-510 through 516, 02-519, 02-618, 02-619, 02-620, 02-622 through 02- 633, 02-637 through 02-644, 02-767, 02-768, 02- 769, 02-771 through 02-779, 02-781, 02-782 & 02-

QUANTITY OF MATERIAL: APPROX. 55,744 LF

MATERIAL ASSESSMENT: This material was found not to be asbestos-containing, therefore no assessment or response action is necessary.

HA-06: DESCRIPTION OF MATERIAL: THIN, PAINTED PIPE INSULATION,

CREAM WITH ROUGH TEXTURE

This sampling area includes thin, painted insulation with a cream color and rough texture located on the 7 Floor inth

Room B731, 5 Floor in the Equipment Room and B535,th and on the 3 Floor in Room B307C. rd

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 7

SAMPLE ID NUMBERS: 02-128, 02-129, 02-130, 02-209, 02-210 &

02-211

QUANTITY OF MATERIAL: APPROX. 350 LF

MATERIAL ASSESSMENT: This material was found not to be asbestos-containing, therefore no assessment or response action is necessary.

HA-07: DESCRIPTION OF MATERIAL: HVAC FIBERGLASS DUCT INSULATION

WITH FOIL AND CANVAS WRAP AND

WHITE MASTIC / PAINT

This sampling area includes HVAC fiberglass duct insulation with foil and canvas wrap and white mastic / paint located in the Basement and Floors 2, 3, 4, 5, 6, and Floors 8, 9 and 10.

NUMBER OF SAMPLES TAKEN: 23

SAMPLE ID NUMBER: 02-16, 02-17, 02-18, 02-39, 02-40, 02-41, 02-

72, 02-73, 02-154, 02-155, 02-220, 02-221, 02-222, 02-223, 02-224, 02-288, 02-289, 02- 345, 02-346, 02-424, 02-425, 02-426, 02- 427, 02-428, 02-612 & 02-613

QUANTITY OF MATERIAL: 14,836 SF

MATERIAL ASSESSMENT: This material was found not to be asbestos-containing, therefore no assessment or response action is necessary.

HA-08: DESCRIPTION OF MATERIAL: HVAC FIBERGLASS DUCT INSULATION

WITH FOIL WRAP AND BLACK MASTIC

This sampling area includes HVAC fiberglass duct insulation with foil wrap and black mastic located in the Basement and Floors 1 through 9.

NUMBER OF SAMPLES TAKEN: 26

SAMPLE ID NUMBER: 02-42, 02-43, 02-44, 02-74, 02-75, 02-131,

02-156, 02-157, 02-158, 02-159, 02-225, 02- 226, 02-227, 02-228, 02-229, 02-287, 02- 341, 02-342, 02-343, 02-344, 02-429, 02-

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 8

500, 02-608, 02-609, 02-610 & 02-611

QUANTITY OF MATERIAL: 25,000 SF

MATERIAL ASSESSMENT: This material is classified as Category II, non-friable ACM, and it is considered to be ACM with the potential for damage. This material does not present a significant potential for asbestos fiber release in its present condition nor under normal operating circumstances.

Therefore, no abatement response action is recommended for this non-friable material at this time. All ACM has the potential, however, for significant damage and release of airborne asbestos fibers if improperly disturbed in any way. This material should be continuously assessed and managed under the Management Plan (O&M) until such time as their condition changes or plans for renovation or demolition require their removal.

HA-09: DESCRIPTION OF MATERIAL: HVAC FIBERGLASS DUCT INSULATION

WITH BLACK PAPER AND TAR

This sampling area includes HVAC fiberglass duct insulation with black paper and tar located on the 8 Floorth in Room A851 and on the 1 Floor in Room 1C-303st

(storage).

NUMBER OF SAMPLES TAKEN: 3

SAMPLE ID NUMBERS: 02-76, 02-77 & 02-78

QUANTITY OF MATERIAL: APPROX. 240 SF

MATERIAL ASSESSMENT: This material was found not to be asbestos-containing, therefore no assessment or response action is necessary.

HA-10: DESCRIPTION OF MATERIAL: HVAC FIBERGLASS DUCT INSULATION

WITH PAPER WRAP AND BLACK MASTIC

This sampling area includes HVAC fiberglass duct insulation with paper wrap and black mastic located on the

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 9

1 Floor in Corridors 1DA, 1DC and 1DD, Waiting Roomsst

1D-308, 1D-400C, 1D-400D and 1D-603A, Rooms 1C-307 and 1C-310 and throughout this clinic section located on the 1 floor. st

SAMPLE ID NUMBERS: 02-501, 02-502 & 02-503

QUANTITY OF MATERIAL: APPROX. 8,000 SF

MATERIAL ASSESSMENT: This material is classified as Category II, non-friable ACM, and it is considered to be ACM with the potential for damage. This material does not present a significant potential for asbestos fiber release in its present condition nor under normal operating circumstances.

Therefore, no abatement response action is recommended for this non-friable material at this time. All ACM has the potential, however, for significant damage and release of airborne asbestos fibers if improperly disturbed in any way. This material should be continuously assessed and managed under the Management Plan (O&M) until such time as their condition changes or plans for renovation or demolition require their removal.

HA-11: DESCRIPTION OF MATERIAL: HVAC VIBRATION DAMPENER, BLACK

This sampling area includes black vibration dampener material associated with the HVAC system located in mechanical rooms on the 10 and 5 Floors.th th

NUMBER OF SAMPLES TAKEN: 2

SAMPLE ID NUMBERS: 02-19 & 02-234

QUANTITY OF MATERIAL: APPROX. 200 SF

MATERIAL ASSESSMENT: This material was found not to be asbestos-containing, therefore no assessment or response action is necessary.

M ICRO AIR OF TEXAS, INC. VAMC; Dallas, Texas - Building 2 10

HA-12: DESCRIPTION OF MATERIAL: HVAC VIBRATION DAMPENER, BEIGE

This sampling area includes beige vibration dampener material associated with the HVAC system located in mechanical rooms on the 5 and 10 Floors of Building 2.th th

SAMPLE ID NUMBERS: 02-20 & 02-235

QUANTITY OF MATERIAL: APPROX. 200 SF

MATERIAL ASSESSMENT: This material was found not to be asbestos-containing, therefore no assessment or response action is necessary.

HA-95A: DESCRIPTION OF MATERIAL: HVAC DUCT SEAM SEALANT, GRAY

This sampling area includes gray seam sealant associated with HVAC ducts on the 9 Floor in Room A907A, 8 Floorth th in Corridor 8A at Room 8C-124, Corridor 8B and Room…