FedBizOpps Special Notice
CLASSIFICATION CODE
SUBJECT
CONTRACTING OFFICE'S
ZIP-CODE
SOLICITATION NUMBER
ARCHIVE
DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
NAICS CODE
CONTRACTING OFFICE
ADDRESS
POINT OF CONTACT
(POC Information Automatically Filled from User Profile Unless Entered)
DESCRIPTION
See Attachment
AGENCY'S URL
URL DESCRIPTION
AGENCY CONTACT'S EMAIL
ADDRESS
EMAIL DESCRIPTION
GENERAL INFORMATION
ADDITIONAL INFORMATION
* = Required Field FedBizOpps Special Notice Rev. March 2010 J PerkinElmer Service Maintenance Agreement 78229 36C25719Q0252 N 811219 Department of Veterans Affairs VISN17 Network Contracting Activity 7400 Merton Minter Blvd. (10N17/90C) San Antonio TX 78229 Juan Martinez Juan.Martinez4@va.gov Juan.Martinez4@va.gov Special Notice – Notice of Intent to Award a Single-Source Procurement
The Department of Veterans Affairs, Veterans Health Administration, Veterans Integrated Service Network (VISN) 17, Network Contracting Office (NCO) 17, located at 5441 Babcock Road, Ste. 302, San Antonio, TX 78240 intends to solicit and award a single-source firm-fixed priced service maintenance agreement for one (1) base year and four (4) option years to PerkinElmer Health Sciences, Inc. for the following PerkinElmer equipment:
| DESCRIPTION |
| MAKE/MODEL |
| LOCATION |
| Liquid Scintillation Analyzer |
| Tri-Carb 2910 TR |
| Radiation Safety, Lab room K203.2 |
| Epson Printer |
| Model B251A |
| Radiation Safety, Lab room K203.2 |
| Wizard2 Automatic Gamma Counter |
| 2480 |
| Radiation Safety, Lab room K203.2 |
| Lexmark Printer |
| 4514-220 |
| Radiation Safety, Lab room K203.2 |
| Liquid Scintillation Analyzer |
| Tri-Carb 2910 TR |
| Research, room S202 |
| Brother Printer |
| HL-L2320D |
| Research, room S202 |
| Wizard2 Automatic Gamma Counter |
| 2480 |
| Research, room S202 |
| Lexmark Printer |
| 4514-220 |
| Research, room S202 |
| Wizard2 Automatic Gamma Counter |
| 2470 |
| Radiation Safety Hot Lab |
Brief Service Description: The Contractor must provide all materials, tools, labor, travel, replacement parts, and test equipment necessary to provide preventive maintenance and repairs for the equipment listed above. Contractor shall maintain or restore the listed equipment to operating specifications according the manufacturer's recommendations. Preventive maintenance must include, but is not limited to, the following:
· Note all possible problems the customer has had with the instruments.
· Check the I-129 peak position (2470 WIZARD2). Adjust if necessary.
· Check the Cs-137 main peak and X-ray peak position (2480 WIZARD2). Adjust if necessary.
· Run I-129 normalization and record the peak channel for each detector.
· Calculate and record the I-129 counting efficiency for each detector.
· Record the I-129 resolution.
· Run I-125 normalization and record the peak channel for each detector.
· Calculate and record the I-125 counting efficiency for each detector.
· Record the I-125 resolution
· Run background normalization and print the result. Record the Open window background values. Use 600s counting time for the background normalization.
· Clean the detectors and sample lifting forks with alcohol.
· Clean and lubricate the arm and lift mechanism.
· Clean the conveyor and check the condition and tightness of the belts and chains.
· Clean the rack sensor reflective mirror.
· Apply silicon grease onto the conveyor motor drive belts if the conveyor is noisy.
· Run a test with short counting time and check the rack and lifting fork stop positions. Adjust if necessary.
· While running a test check the operation of the lid sensor.
Preventive maintenance on the Tri-Carb shall include, but not limited to the following:
· Note all possible problems the customer has had with the instrument.
· Number of cycles logged
· Inspect the lead base area
· Clean PMT faces
· Clean PMT reflectors
· Inspect elevator spring assembly
· Clean/Inspected elevator pedestal
· Clean/Inspected light seal rings on elevator body assembly
· Cleaned the detector enclosure area
· Clean the fan filter on instrument back
· Clean the sample changer deck and both sides of lid
· Blow dust out of static controller
· Clean optical readers with MeOH or distilled water
· Check the following adjustments
· Load end drive adjustments
· Transfer drive adjustments
· Sample changer belt tension
· Shutter sensors
· Tab puller sensors
· Hold down rail adjustments
· All hardware for tightness
· Elevator up adjustment
· Elevator down adjustment
· Protocol plug and flag actuators
· Inspect the Barium source for wear
· Check the operation of the following mechanical assemblies
· All sensors
· Transfer drives
· Tab pullers
· Shutters
· Verified sample changer operation with large and small vials
· Inspect damaged cassettes
· Test the temperature if a chill pack if installed
· Perform a System Normalization and Calibration (SNC) on the system
Preventive maintenance on the various printers shall include, but not be limited to:
· Inspect paper pickup rollers for dust, clean if necessary
· If rollers are shiny/glazed replace them
· Inspect separation pads for dust, separation pads should be replaced at the same time when rollers are replaced
· Clean transfer rollers and printer heads
· Inspect mirrors for dust and grime buildup, clean if necessary
· Inspect fuser assembly rollers for marks and replace if there any marks of blemishes present
Services will be performed at the South Texas Veterans Health Care System, Audie L. Murphy VA Hospital, 7400 Merton Minter St., San Antonio, Texas 78229.
The NAICS Code for this procurement is 811219 – “Other Electronic and Precision Equipment Repair and Maintenance” with has a Small Business Size Standard of $20.5M. This action is accomplished under FAR 13.106-1(b)(1)(i), “Soliciting from a Single-Source”. This notice is not to be considered a request for quotations or proposals. No contract will be awarded based on offers received in response to this notice; however, any offers received may be used to determine if a comparable source is available and more advantageous to the government. All responsible sources may submit a response; which, if timely received, will be considered by the agency. To avoid Gray Market purchases, any response received from other than the OEM for any medical equipment must contain documentation showing OEM training certification on the equipment listed and a letter from the OEM stating the company is an authorized distributor and/or service provider for the medical equipment listed. If parties choose to respond, any cost associated with preparation and submission of data or cost incurred in response to this announcement shall be the sole responsibility of the contractor and will not be reimbursed by the Government. A determination by the Government not to compete this proposed requirement based upon responses to this notice is solely within the discretion of the Government. If no affirmative written response is received by the response date and time, a purchase order will be issued to PerkinElmer Health Sciences, Inc. without any further notice. Interested firms must respond by email directly to Juan.Martinez4@va.gov no later than 12:00 P.M. CST on 18 January 2019.