This is an Amendment of Solicitation/Modification of Contract for replacement dental equipment issued by the Department of Veterans Affairs Network Contracting Office (NCO) 15 in Leavenworth, Kansas. Amendment 0001, effective January 23, 2026, updates standard warranty information in CLIN 0001 of the Statement of Work and addresses vendor questions. The solicitation number is 36C25526Q0155, and all offerors must acknowledge receipt of this amendment prior to the specified offer deadline or risk rejection.
The Wichita VA Medical Center (Robert J. Dole Medical Facility) requires eight dental chairs with minimum 500-pound patient capacity, eight delivery systems with six handpiece holders compatible with Kavo motors and Dentsply scalers, eight assistant packages, eight LED dental lights with cure-safe mode, six doctor stools, and six assistant stools. The contractor must also provide blanket wrap service and complete delivery, installation, on-site cleanup, and on-site training within 30 days after receipt of order (ARO) at the facility located at 5500 East Kellogg Avenue, Wichita, Kansas, 67218. Work hours are Monday through Friday, 8:00 AM to 4:30 PM CST, excluding federal holidays. The standard warranty requires a minimum of five years on components and two years on upholstery, with vendor response time for service calls within eight hours and shipment of replacement parts within 48 hours. Extended warranties beyond the minimum will be evaluated as exceeding the standard, though additional costs will be weighed against overall price for best value determination. No incumbent contractor currently services the facility, and past performance references should preferably include three examples weighted equally regardless of whether they involve VA medical centers or other federal healthcare facilities.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| S02 36C25526Q0155_JB.pdf | ||
| S02 36C25526Q0155_JB.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
5. PROJECT NUMBER (if applicable)
CODE 7. ADMINISTERED BY
2. AMENDMENT/MODIFICATION NUMBER
CODE
6. ISSUED BY
8. NAME AND ADDRESS OF CONTRACTOR
4. REQUISITION/PURCHASE REQ. NUMBER 3. EFFECTIVE DATE
9A. AMENDMENT OF SOLICITATION NUMBER
9B. DATED
PAGE OF PAGES
10A. MODIFICATION OF CONTRACT/ORDER NUMBER
10B. DATED
BPA NO. 1. CONTRACT ID CODE
FACILITY CODE CODE
Offers must acknowledge receipt of this amendment prior to the hour and date specified in the solicitation or as amended, by one of the following methods:
The above numbered solicitation is amended as set forth in Item 14. The hour and date specified for receipt of Offers
E. IMPORTANT:
is extended,
(a) By completing Items 8 and 15, and returning __________ copies of the amendment; (b) By acknowledging receipt of this amendment on each copy of the offer submitted; or (c) By separate letter or electronic communication which includes a reference to the solicitation and amendment numbers. FAILURE OF YOUR ACKNOWLEDGMENT TO BE RECEIVED AT THE PLACE DESIGNATED FOR THE RECEIPT OF OFFERS PRIOR TO THE HOUR AND DATE SPECIFIED MAY is not extended.
12. ACCOUNTING AND APPROPRIATION DATA
(REV. 11/2016)
is required to sign this document and return ___________ copies to the issuing office. is not, A. THIS CHANGE ORDER IS ISSUED PURSUANT TO: (Specify authority) THE CHANGES SET FORTH IN ITEM 14 ARE MADE IN THE CONTRACT ORDER NO. IN ITEM 10A.
15C. DATE SIGNED
B. THE ABOVE NUMBERED CONTRACT/ORDER IS MODIFIED TO REFLECT THE ADMINISTRATIVE CHANGES SET FORTH IN ITEM 14, PURSUANT TO THE AUTHORITY OF FAR 43.103(b).
RESULT IN REJECTION OF YOUR OFFER. If by virtue of this amendment you desire to change an offer already submitted, such change may be made by letter or electronic communication, provided each letter or electronic communication makes reference to the solicitation and this amendment, and is received prior to the opening hour and date specified.
C. THIS SUPPLEMENTAL AGREEMENT IS ENTERED INTO PURSUANT TO AUTHORITY OF:
D. OTHER
Contractor
16C. DATE SIGNED
14. DESCRIPTION OF AMENDMENT/MODIFICATION
16B. UNITED STATES OF AMERICA
Except as provided herein, all terms and conditions of the document referenced in Item 9A or 10A, as heretofore changed, remains unchanged and in full force and effect.
15A. NAME AND TITLE OF SIGNER 16A. NAME AND TITLE OF CONTRACTING OFFICER
15B. CONTRACTOR/OFFEROR
STANDARD FORM 30 PREVIOUS EDITION NOT USABLE
Prescribed by GSA - FAR (48 CFR) 53.243
(Type or print) (Type or print)
(Organized by UCF section headings, including solicitation/contract subject matter where feasible.)
(Number, street, county, State and ZIP Code)
(If other than Item 6)
(Specify type of modification and authority)
(such as changes in paying office, appropriation date, etc.)
(If required)
(SEE ITEM 11)
(SEE ITEM 13)
(X)
CHECK
ONE
13. THIS ITEM APPLIES ONLY TO MODIFICATIONS OF CONTRACTS/ORDERS,
IT MODIFIES THE CONTRACT/ORDER NO. AS DESCRIBED IN ITEM 14.
11. THIS ITEM ONLY APPLIES TO AMENDMENTS OF SOLICITATIONS
AMENDMENT OF SOLICITATION/MODIFICATION OF CONTRACT
(Signature of person authorized to sign) (Signature of Contracting Officer)
1 5
0001 01-30-2026
36C255
Department of Veterans Affairs Network Contracting Office (NCO) 15 3450 S 4th Street Trafficway
Leavenworth KS 66048
36C255
Department of Veterans Affairs Network Contracting Office (NCO) 15 3450 S 4th Street Trafficway
Leavenworth KS 66048
To all Offerors/Bidders
36C25526Q0155
01-23-2026
X
X X
X 1(ONE)
The purpose of this amendment is to update standard warranty information in CLIN 0001 of the statement of work and respond to questions received from a vendor.
Jessica D. Barton Contracting Officer
STATEMENT OF WORK (SOW)
REPLACEMENT DENTAL CHAIRS
1. INTRODUCTION
The Wichita VA Medical Center requires new and unused dental chairs, delivery units, assistant’s package, dental lights, clinician stools and accompanying equipment to perform general dentistry and oral surgery procedures within the dental department.
2. TECHNICAL REQUIREMENT -
• Dental chair capacity must be at least 500 pounds.
• Dental chairs must be adjustable to accommodate high percentage of disabled and mobility-impaired patients.
• Delivery units will be chair mounted with six handpiece holders that include one syringe, two Kavo electric motor tubing, two air driven tubing and one Cavitron ultrasonic tubing.
• Delivery unit needs to be compatible with current Kavo motors and hand pieces and
Dentslply scaler tips without any modifications or additional equipment needed.
• The dental light needs to be LED, stadium style, has a cure safe mode, and is touchless.
• Assistant packages need to be chair mounted.
3. SCOPE OF WORK
Contractor will provide eight dental chairs, eight delivery units, eight assistant packages, eight dental lights, six doctor stools and six assistant stools. All equipment will be delivered to the Wichita VAMC and will include installation and cleanup. On-site training will be provided by the contractor after installation is completed and will be scheduled through the dental department. The contractor will provide the standard warranty for equipment.
4. DELIVERABLES
CLIN 0001: Dental Chairs
QTY: 8 EA
Required:
o Dental chairs support a patient weight capacity of no less than 500 pounds (227 kg).
o Minimum 50-degree swivel with brake.
o Adjustable height range:
o Minimum: At least 12 inches from the ground (exceeding this requirement would be <12in from the ground) o Maximum: 30 inches from the ground (exceeding this requirement would be >30in from the ground) o Dual articulating headrest with lever or electric release.
o Must have armrests and armrests must support patients during entry and exit to accommodate impaired mobility.
o Power supply to support peripheral equipment o Controls— o Doctors’ touchscreen All controls (to include air and water adjustments, forward and reverse drive, presetting, and endo/surgical settings) integrated into one interface Touchscreen compatible with cover to ensure VA infection control guidelines are upheld and to prevent clinicians from re-gloving o Assistants control pad
• Standard Warranty: Minimum 5 years on components, 2 years on upholstery o Response time for service calls within 8 hours, shipment of replacement parts within 48 hours
CLIN 0002: Dental Delivery Systems
QTY: 8 EA
Required:
• A system with 6 positions to include 2 air driven with 5 holes, 2 Kavo electric, 1 Dentsply scaler, 1 air/water syringe
• Centrally located ambidextrous delivery system handle and touchscreen allows switching between left- and right-handed operators.
• Balanced flex arm with air brake and positive positioning.
• Minimum 6-inch glass touchscreen control.
• Touchscreen must support a minimum of 15 user profiles.
• Water indicator lights correspond with the handpiece in use.
• No less than 2-liter water bottle.
• Radius mounted for left/right use.
CLIN 0003: Assistant packages
Required:
• 4 position holders: 1 air/water syringe, 1 saliva ejector, 2 HVE
• Chair control pad, Chair mounted, rear.
CLIN 0004: Dental Lights
Required:
• LED, low heat led
• Cure-safe mode
• Touchless sensor and auto on-off and touchpad control
• Light head must rotate around 3 axes with a minimum of 180º rotation for each axis
• 3 light intensity settings: low setting minimum of 15,000 lux
• Lifespan expectancy of no less than 30,000 hours
• Chair mounted.
CLIN 0005: Doctor's Stools
QTY: 6 EA
Required:
• 5-wheel design
• Adjustable height
• Tilt seat
• Backrest height control
• Upholstery matches dental chairs
• Smooth, quiet caster wheels.
CLIN 0006: Assistant Stools
QTY: 6 EA
Required:
• (5)-wheel design
• Adjustable height
• Height-adjustable foot ring
• Upholstery matches dental chairs
• Smooth, quiet caster wheels
CLIN 0007: Blanket Wrap Service
QTY: 1 EA
CLIN 0008: Delivery, Installation, On-site Clean up, and On-site training for dental department personnel
QTY: 1 JB
5. DELIVERY– all equipment shall be delivered and installed NLT 30 Days ARO at the following location:
Robert J. Dole Medical Facility 5500 East Kellogg Ave Wichita Kansas, 67218
6. SPECIAL WORK REQUIREMENTS - Hours: The regular work hours for this facility are Monday through Friday, from 8:00 am to 4:30 pm (CST), excluding Federal Holidays.
Vendor Questions
1. For the dental chair weight capacity requirement of 'at least 500 pounds,' does exceeding this specification (e.g., 600 lbs.) provide any evaluation advantage, or is meeting the minimum sufficient for full technical credit? Yes, significantly higher weight capacities than the minimum requirement will be evaluated as exceeding the standard.
2. The minimum warranty is 5 years on components and 2 years on upholstery. Will proposals offering extended warranties (e.g., 7-year components, 3-year upholstery) receive preferential evaluation consideration under the technical criteria? Yes, an extended warranty will be evaluated as exceeding the standard. However, the additional cost of the extended warranty will be weighed against the overall price to determine if it offers the government the best value.
3. What response time expectations does the VA have for warranty service calls? Should we include guaranteed response times in our technical proposal?
• Vendor shall provide respond to service calls within 8 hours.
• Vendor shall ship replacement parts within 48 hours of identifying necessity by Dole VA
Healthcare Technology Management staff.
• Yes, please provide guaranteed response time
• Statement of work has been revised to include above information
4. Is there a current or recent incumbent contractor for dental equipment at Wichita VAMC? If so, does the government have any performance concerns we should be aware of? No
5. Will the government provide detailed operator floor plans, electrical specifications, and plumbing connection locations prior to award, or should we plan to conduct a site visit before finalizing the installation plan? This is not applicable to this requirement.
6. Are there any specific VA facility access requirements, background check processes, or security clearances our installation technicians must obtain prior to beginning work? No
7. How much advance notice will the government provide for scheduling installation after contract award and delivery? Please see page 6 – delivery information.
8. Does the 30-day timeline include both delivery and complete installation with training, or are these separate timelines? Please see page 6 – delivery information.
9. What is the typical payment timeline from invoice submission to payment for VA contracts of this type? Payment terms are outlined on page 3 of the solicitation.
10. For past performance evaluation, is there a preference for VA medical center experience over other federal healthcare facilities, or are they weighted equally? No, they are weighted equally.
11. How many past performance references should we provide for optimal evaluation? Is there a maximum number that will be considered? Preferably 3, but please provide as applicable.
12. Can you share how many responses or offers the government typically receives for dental equipment solicitations of this type and value? No
13. Is there an incumbent contractor currently servicing Wichita VAMC dental equipment? No