ATTACHMENT III - Doppler_SOW.docx

DOCX document 1 MB Posted

Attached to
6515--ABI Doppler and Vascular Doppler Indy East Activation
Solicitation number
36C25027Q0089
Issued by

About this file

This is a Statement of Work (SOW) for the procurement of handheld vascular Doppler systems supporting the East Indianapolis VA Clinic's Initial Outfitting, Transition, and Activation (IOT&A) efforts. The Department of Veterans Affairs, VISN 10, Network Contracting Office 10 requires five M4106 ABI Doppler systems, each featuring an 8 MHz bidirectional probe, and one X2127 Vascular Doppler with a 4 MHz probe for the clinic's Multi-Specialty and Patient Aligned Care Team (PACT) departments. The M4106 basis of design is the Cooper Surgical L250AB LifeDop 250 ABI System, which includes a handheld Doppler unit, blood-pressure cuffs, aneroid pressure gauge, waveform printer, and carrying case; the unit operates on AA alkaline batteries and features an LCD display for battery status and blood-flow indicators. The X2127 basis of design is the Cooper Surgical L150R-SD4 LifeDop 150 Doppler, a rechargeable handheld unit with a 4 MHz nondirectional probe and automatic shutdown after three minutes of inactivity.

Delivery is scheduled between April 5 and April 16, 2027, to the East Indianapolis VA Clinic at 9402 E 21st Street, Indianapolis, Indiana 46229, with coordination required through the IOT&A Logistics Manager or VA Representative. Products must be unused, not refurbished, manufactured no more than 24 months before the purchase order date, and include batteries manufactured within 12 months of purchase order issuance. All equipment must be delivered during normal loading dock business hours (8:00 AM to 3:30 PM, Monday through Friday, excluding Federal Holidays) with proper shipping labels referencing the VA PO number. The contractor must provide service manuals in electronic format and warranty information for parts and labor to both the IOT&A contractor and VA HTM Department within three days of delivery. No assembly, installation, or calibration is required, and no training is mandated. This is an equipment purchase with delivery only; no connection to the VA network is required, and contractor access to VA patient information is not authorized.

View the file

Other files

File Type Posted
36C25027Q0089.pdf PDF
ATTACHMENT I - ABI Doppler_Spec Sheets .pdf PDF
ATTACHMENT II - Vascular Doppler_Spec Sheets.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

PROJECT TITLE: East Indianapolis VA Clinic 21 OCTOBER 2026

VETERAN HEALTH INDIANA

EAST INDIANAPOLIS VA CLINIC (INDY EAST)

INITIAL OUTFITTING, TRANSITION & ACTIVATION

STATEMENT OF WORK (SOW)

FOR

M4106 ABI DOPPLER & X2127 VASCULAR DOPPLER

Richard L. Roudebush Veterans Administration Medical Center 1481 West Tenth Street Indianapolis, IN 46202-2884

Introduction The Department of Veterans Affairs (VA), VISN 10, Network Contracting Office 10 (NCO 10), requires five handheld vascular Doppler ankle-brachial index (ABI) systems, each with an 8 MHz bidirectional probe, and one handheld vascular Doppler with a 4 MHz probe for the new East Indianapolis VA Clinic’s Multi-Specialty and Patient Aligned Care Team (PACT) departments. This acquisition supports the clinic’s Initial Outfitting, Transition, and Activation (IOT&A) efforts. The basis of design is outlined in Section 3, Products and Services, and Section 4, Specifications/Salient Characteristics.

General Conditions Site Address Indy East Clinic 9402 E 21st St Indianapolis, IN 46229 Delivery Schedule Products must be delivered no earlier than the estimated delivery start date and no later than the delivery end date specified below.

Building Name
Est. Delivery Start Date
Est. Delivery End Date
East Indianapolis VA Clinic
5 April 2027
16 April 2027

The delivery schedule detailed above is based on the current construction timeline, considering the nature of activation efforts. It may change as dates fluctuate, so delivery dates must remain flexible.

The delivery of all products depends on the completion of relevant areas within the building. If any delay occurs with the construction or building acceptance, the vendor may need to adjust the delivery date at no cost to the Government. The Contractor will coordinate with the on-site points of contact prior to delivery.

The VA activations team will confirm the awardee’s actual delivery dates upon award.

Post-award, the vendor will coordinate delivery identified in this document with the VA before beginning any work (See Section 2.3). Once the dates have been confirmed, modifications to the schedule are subject to written approval by the CO, COR, and the vendor.

Delivery and Receiving The delivery and receipt of the proposed items are expected to be directed to the location specified below.

☒ East Indianapolis VA Clinic (detailed in section 2.1) ☐ IOT&A contractor provided warehouse: Address TBD Confirmation of the delivery location will be provided after the award. To coordinate delivery, please contact:

· Primary: IOT&A Logistics Manager, Soren Nicholson at (719) 761-4387, Soren.Nicholson@sme-solutions.co

· Secondary: IOT&A Project Manager, Joe Santinoceto at (509) 227-9134, Joe.Santinoceto@sme-solutions.co

· Alternate: VA Representative, Julie Webb at (317) 988-2747, Julie.Webb@va.gov The delivery of items identified in this document shall occur during normal loading dock business hours, which are defined as 8:00 AM to 3:30 PM Local Time, Monday through Friday, excluding Federal Holidays.

Delivery trucks are not permitted to remain at the receiving area. Trucks shall be unloaded, moved from the receiving area, and then brought back to the receiving area if necessary to haul out any waste, tools, or excess materials.

Shipping Label The shipping label on the outside of the box and the packing slip for all products must include at a minimum the basic shipping address information and the following additional details:

· East Indianapolis VA Activations

· VA PO#*******

Below is an example of an acceptable shipping label:

Attn: East Indianapolis VA Activations (VA PO# ********) Indy East Clinic 9402 E 21st St Indianapolis, IN 46229 Assembly and Installation Required (Yes ☐ No ☒) Calibration Required (YES ☐ NO ☒) Training Required (Yes ☐ No ☒) Additional Requirements The product manufacturer’s date cannot exceed 24 months (2 years) before the purchase order date. The equipment must be unused and not refurbished, and it should have the latest software and hardware updates installed by the manufacturer.

Batteries included with the unit must have been manufactured within 12 months (1 year) of the purchase order issuance date.

Service Manuals, including the theory of operation, maintenance requirements, repair information, schematics, part lists, and troubleshooting details, will be provided to the VA HTM Department in electronic format.

Warranty information regarding parts and labor must be provided for all procured products in electronic format. Documentation must be sent to both the IOT&A contractor and the VA HTM Department within three (3) days of delivery.

Confidentiality and Non-Disclosure The purchased equipment will not connect to the VA network. This acquisition does not require connection of contractor-owned IT devices or remote contractor systems to the VA internal network. Accordingly, under Question 6 of VA Handbook 6500.6, Appendix A, the C&A requirements do not apply, and a Security Accreditation Package is not required.

This is an equipment purchase with delivery only. The contractor’s activities will not involve storing, generating, transmitting, or exchanging VA-sensitive information. Accordingly, the response to Question 7 of VA Handbook 6500.6, Appendix A, is No. The ABI system temporarily buffers a waveform for printing; VA personnel shall control the equipment during clinical use and safeguard printed results under applicable VA privacy procedures. Contractor access to VA patient information is not authorized.

Specifications / Salient Characteristics

M4106 - ABI DOPPLER SYSTEM

General Requirements

· Total Quantity: 5

· Basis of design: Cooper Surgical - L250AB LifeDop 250 ABI System w/8 MHz Bi-directional Probe

Specifications and Salient Characteristics Portable vascular Doppler system for noninvasive assessment of peripheral circulation and measurement of ankle and brachial systolic pressures for manual ankle-brachial index calculation. Includes a handheld Doppler, 8 MHz bidirectional vascular probe, two blood-pressure cuffs, aneroid pressure gauge, compact waveform printer, adhesive-backed paper, carrying case, and ABI reference materials.

· Supports manual measurement of ankle and brachial systolic pressures using the included cuffs and aneroid for manual ABI calculation. LCD provides battery status, blood-flow direction, and relative signal-strength indicators.

· 8 MHz bidirectional probe for detecting peripheral blood flow toward and away from the probe.

· Audible Doppler signals and visual indication of flow direction and relative signal strength.

· Two-cuff and aneroid configuration for obtaining arm and ankle systolic pressures.

· Automatic waveform scaling and printing on adhesive-backed paper for attachment to assessment forms.

· Handheld Doppler powered by three 1.5 V AA alkaline batteries; rated for approximately 100 four-minute examinations.

· Printer powered by a rechargeable 3.7 V lithium-ion battery; rated for up to 500 prints. Supplied charging adapter requires 100–120 VAC, 50/60 Hz.

· Handheld unit approximately 140 × 70 × 35 mm.

· Weight approx. 320g.

X2127 - VASCULAR DOPPLER

General Requirements

· Total Quantity: 1

· Basis of Design: Cooper Surgical - L150R-SD4 LifeDop 150 Doppler w/4 MHz probe

1.1.1 Specifications and Salient Characteristics

Rechargeable handheld vascular Doppler for noninvasive detection of blood flow in peripheral veins and arteries. Includes a non-display Doppler unit, interchangeable 4 MHz nondirectional vascular probe, rechargeable batteries, and wall charging adapter. The probe’s broad beam facilitates vessel location and maintenance of probe position during vascular assessment.

· 4 MHz nondirectional vascular probe with a broad, flat beam for locating peripheral vessels.

· Audible blood-flow signals through an integrated speaker with adjustable volume.

· LED indicators for power, low battery, and charging status.

· Three rechargeable AA nickel-metal hydride batteries; manufacturer-rated battery life of approximately 1,000 one-minute examinations per charge.

· In-unit charging through the supplied wall adapter; charging cycle ends upon reaching the maximum battery voltage or after 14 hours.

· Automatic shutdown after approximately three minutes of inactivity.

· Handheld unit approximately 140 × 70 × 35 mm.

· Weight approx. 320g.

Statement of Work Page 1 of 8 image1.png