36C25026Q0162.docx
DOCX document 17 KB Posted
- Solicitation number
- 36C25026Q0162
About this file
This is a Sources Sought Notice for the Indianapolis VA Medical Center's Optical Lab Air Compressor Replacement Project. The Department of Veterans Affairs is seeking information from potential contractors for the procurement of two Ingersoll Rand RS55I, 75HP, 138 PSI Contact Cooled Rotary Screw air compressors or equivalent. Key requirements include two independent units capable of providing 335 CFM of air demand, integrated air-drying systems, and a fully turn-key project management and installation process that involves removing existing air compressors, transportation, rigging, utility hookups, and system integration.
Contractors must submit responses by December 17, 2025, at 4:30 PM EST to Brooke Hansen, referencing solicitation number 36C25026Q0162. Submissions should include business details such as name, GSA contract number, point of contact information, DUNS number, business size (small or large), small business type (if applicable), manufacturer information, and country of origin. The project is classified under NAICS code 333912 and Product Service Code 4310, with the place of performance at 5610 Dividend Rd, STE B, Indianapolis, IN 46241. An optional preventative maintenance program may be solicited in a future procurement.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sources Sought Notice Sources Sought Notice
| SUBJECT* |
| Optical Lab Air Compressor Replacement Project, IND |
GENERAL INFORMATION
| CONTRACTING OFFICE’S ZIP CODE* |
| 49037 |
| SOLICITATION NUMBER* |
| 36C25026Q0162 |
| RESPONSE DATE/TIME/ZONE |
| 12-17-2025 4:30PM EASTERN TIME, NEW YORK, USA |
| ARCHIVE |
| 15 DAYS AFTER THE RESPONSE DATE |
| RECOVERY ACT FUNDS |
| N |
SET-ASIDE
| PRODUCT SERVICE CODE* |
| 4310 |
| NAICS CODE* |
| 333912 |
| CONTRACTING OFFICE ADDRESS |
| Department of Veteran Affairs |
Network Contracting Office 10
5500 Armstrong Rd.
Battle Creek MI 49037
POINT OF CONTACT*
Contract Specialist Brooke Hansen brooke.hansen@va.gov Phoned inquiries will not be accepted.
PLACE OF PERFORMANCE
| ADDRESS |
| Indianapolis VA Medical Center - Optical Lab |
5610 Dividend Rd, STE B
Indianapolis IN
| POSTAL CODE |
| 46241 |
| COUNTRY |
| US |
ADDITIONAL INFORMATION
| AGENCY’S URL |
| https://www.va.gov |
| URL DESCRIPTION |
| Department of VA |
| AGENCY CONTACT’S EMAIL ADDRESS |
| brooke.hansen@va.gov |
| EMAIL DESCRIPTION |
| Contract Specialist |
DESCRIPTION
This is a REQUEST FOR INFORMATION and SOURCES SOUGHT for TWO Industrial Air Compressors Contractors that deem themselves capable of meeting the requirement shall provide the below information to Brooke Hansen, Contracting Officer, at Brooke.Hansen@va.gov no-later-than Wednesday, December 17, 2025 at 4:30PM, EST, referencing (36C25026Q0162). Responses shall include: (1) Business Name and Address, (2) GSA Contract Number and expiration date, if applicable (3) Point of Contact Name, Phone Number and E-mail Address (4) DUNS and NAICS code (5) Business Size SMALL or LARGE (6) Type of Small Business (if applicable): service disabled veteran owned, veteran owned small business, 8a, HUBZone, woman-owned, etc. (7) A concern that is not a manufacturer of the supply shall provide the manufacturers name and size standard (8) Country of Origin. If the concern is the manufacture and the sole distributor, please provide documentation to that effect.
Please see requirements below:
Indianapolis VA Medical Center, located at 5610 Dividend Rd, STE B, Indianapolis, IN 46241, is requesting the purchase of TWO Ingersoll Rand RS55I, 75HP, 138 PSI Contact Cooled Rotary Screw with Total Air System or equivalent.
Project minimal specification requirements:
| a. | ON-SITE pre-proposal visitation by potential vendor is available, to ensure accuracy in the vendor’s proposal to the government. |
| b. | Two (2) units, each unit independently able to continuously provide the minimally acceptable air demand requirements (335 CFM) of the VA Optical Lab, installed in tandem, to allow continuous, uninterrupted supply of air, regardless of which air compressor is operational. The VA currently owns a 500-gallon (approximate) air storage tank. |
| c. | Each air compressor unit shall include an integrated air-drying system. |
| d. | Variable speed/rotary design is preferred, but not mandatory. |
| e. | Fully turn-key project management/installation |
| - | Removal and scrapping of two (2) existing units IR model IRN40H-CC-HV air compressors |
| - | Transportation and all rigging of outgoing & incoming equipment |
| - | Integration and exhaust duct work required to place both new air compressor units in the most acceptable location in the VA’s mechanical room- core ductwork is present. |
| - | All utilities hook-up and/or modification required to optimize new air compressor requirements. Presently, the VA has one (1) 100A/60V safety switch, and one (1) 200A/240V safety switch. |
| - | All drain requirements of the new air compressors. |
| - | All plumbing & material requirements to tie into or modify present air supply lines (4” copper) to optimize new equipment replacement/location. |
| - | Relocation of existing 500-gallon (approximate) air storage tank to optimize new equipment replacement/location. |
| f. | Awarded vendor must be able to offer an optional, Preventative Maintenance program (to be solicitated and awarded later) to support this equipment for the OEM estimated life of these air compressors. |