36C25018Q9342-0001007.docx
DOCX document 18 KB Posted
- Solicitation number
- 36C25018Q9342
About this file
36C25018Q9342 0001 Attachment 6.docx
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C25018C0229-000.docx | DOCX document | |
| 36C25018Q9342-0002000.docx | DOCX document | |
| 36C25018Q9342-0001000.docx | DOCX document | |
| 36C25018Q9342-0001003.pdf | ||
| 36C25018Q9342-0001002.pdf | ||
| 36C25018Q9342-0001004.pdf | ||
| 36C25018Q9342-0001005.pdf | ||
| 36C25018Q9342-0001001.docx | DOCX document | |
| 36C25018Q9342-0001006.pdf | ||
| 36C25018Q9342-008.docx | DOCX document | |
| 36C25018Q9342-010.docx | DOCX document | |
| 36C25018Q9342-003.pdf | ||
| 36C25018Q9342-009.pdf | ||
| 36C25018Q9342-002.docx | DOCX document | |
| 36C25018Q9342-007.doc | DOC document | |
| 36C25018Q9342-004.pdf | ||
| 36C25018Q9342-001.docx | DOCX document | |
| 36C25018Q9342-006.docx | DOCX document | |
| 36C25018Q9342-011.pdf | ||
| 36C25018Q9342-005.pdf | ||
| 36C25018Q9342-012.pdf | ||
| 36C25018Q9342-000.docx | DOCX document |
Show all 22
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DEPARTMENT OF VETERANS AFFAIRS, DAYTON MEDICAL CENTER 552-18-508, 18-508 Radiology Special Procedure Room and Urology Exam Rooms - BUILDING 330
SECTION 13 49 00 RADIATION PROTECTION
PART 1 –
GENERAL 1.1 DESCRIPTION A. This section specifies lead radiation shielding. B. Construction of products and assemblies used for radiation shielding complying with applicable requirements of NCRP Reports 147 and 102. C. This section includes the following items: 1. Lead Lined Wood Doors 2. Lead Glass 3. Lead Lined Frames 4. Lead Lined Gypsum Wallboard
1.2 RELATED WORK A. Wood Veneer finish for doors: Section 08 14 00, WOOD DOORS, and Section 09 06 00, SCHEDULE FOR FINISHES. B. Steel door frames: Section 08 11 13, HOLLOW METAL DOORS AND FRAMES. C. Hardware for doors: Section 08 71 00, DOOR HARDWARE. D. Installation of Doors and Hardware: Section 08 11 13, HOLLOW METAL DOORS AND FRAMES / Section 08 14 00, INTERIOR WOOD DOORS / Section 08 71 00, DOOR HARDWARE
1.3 MANUFACTURERS QUALIFICATIONS A. Approval by COR is required of product or service of proposed manufacturer and suppliers, and will be based upon submission by Contractor of certification that: 1. Manufacturer regularly and presently manufactures lead radiation shielding as specified as one of its principal products. 2. Manufacturer's product submitted has been in satisfactory and efficient operation or three installations similar and equivalent to this project for three years. 3. Manufacturer submits list of installations.
1.4 TESTS A. Lead radiation shielding will be tested at the expense of the Government after X-ray equipment is installed. B. Any additional testing required due to correction and replacement of defective work will be done by the Government at Contractor's expense. NOTE: Lead glass, lead lined gypsum wallboard will not be tested prior to installation.
1.5 SUBMITTALS A. Submit in accordance with Section 01 33 23, SHOP DRAWINGS, PRODUCT DATA, AND SAMPLES. B. Shop Drawings: Each lead radiation shielding item specified showing thickness of lead, details of construction and installation. C. Samples: 1. Lead lined gypsum wallboard. 2. Bottom corner section of lead lined door, 300 mm (12 inches) square showing bottom and side edge strips. D. Manufacturers' Literature and Data: Each lead radiation shielding item specified.
1.6 WARRANTY A. Warranty lead lined doors against defects in workmanship and materials subject to terms of "Warranty of Construction" Article in GENERAL CONDITIONS, except that warranty period shall be two years. B. Warp or twist of lead lined flush veneered doors may not exceed 6 mm (1/4 inch) in any face dimension of door (including full diagonal), measured not less than six months after doors have been hung and finished.
1.7 APPLICABLE PUBLICATIONS A. The publications listed below form a part of this specification to the extent referenced. The publications are referenced in the text by the basic designation only. B. Federal Specifications (Fed. Spec.): QQ-L-201F(2)..........Lead Sheet C. American Society for Testing and Materials (ASTM): A167-99(2009)...........Stainless and Heat Resisting Chromium-Nickel Steel Plate, Sheet and Strip C1396/C1396M-04.........Gypsum Wallboard/Gypsum Lath C90-11..................Load-Bearing Concrete Masonry Units C1002-07................Steel Drill Screws for the Application of Gypsum Board or Metal Plaster Bases D1187-97(R2002).........Asphalt-Base Emulsions for Use as Protective Coatings for Metal D. United States Department of Commerce Product Standard (PS): FED PSI 83-84...........Construction and Industrial Plywood E. Military Specifications (Mil. Spec.): MIL-C-36373.............Cabinet, Cassette Transfer, Wall Mounted F. National Council on Radiation Protection and Measurements (NCRP):
Report 147..............Structural Shielding Design for Medical X-Ray Imaging Facilities (2004) Report 102..............Medical X-Ray, Electron Beam and Gamma-Ray Protection for Energies up to 50 MeV (Equipment Design, Performance and Use), (1989)
PART 2
PRODUCTS 2.1 MATERIALS A. Lead Sheet: Fed. Spec. QQ-L-201, Grade C, of thickness shown on drawings. B. Lead Lined Gypsum Wallboard: 1. ASTM C1396, Type X, 13 mm (1/2 inch) thick. 2. Factory bond sheet lead to one side of wallboard. 3. Apply sheet lead in thicknesses shown, unpierced and in one piece. C. Stainless Steel: ASTM A167. D. Lead Glass: Laminated, Safety, Clear, X-ray proof, of sufficient thickness to provide X-ray protection equivalent to that provided by partition or door in which glass occurs. E. Lead Control Windows: Cast lead, rigid, single unit type without joints, with or without voice passage as shown and with lead stop beads and lead glass. 1. Stainless steel thresholds over lead lining F. Fasteners: 1. Cadmium or chromium plated steel screws for securing lead louvers. 2. Standard steel drill screws, ASTM C1002, with lead washers for application of lead lined sheet materials to metal studs and attach washers in accordance with shielding manufacturer's instructions. 3. Nails: a. Use barbed lead head nails for application of lead lined materials to wood furring strips. b. Long enough to penetrate furring strips not less than 25 mm (one inch). c. Cast-lead head sufficiently thick to equal lead shielding of room provided. G. Lead Discs: Same thickness as lead lining, diameter 25 mm (1 inch) larger than fastener.
2.2 FABRICATION A. General: Lead lining of frames, doors and other items occurring in partitions shall provide an X-ray absorption equivalent to that of partitions in which they occur. B. Clearance
1. Jambs and Heads: A maximum 3 mm (1/8 inch) clearance. 2. Bottom of door to finish floor: Maximum 19 mm (3/4 inch) clearance. C. Lead Lined Wood Doors: 1. Flush veneered construction. 2. Construct doors of two separate solid wood cores with a single sheet of lead lining through center. 3. Doors shall have filler strips, crossbanding, face veneers and hardwood edge strips, all glued together with unextended urea resin glue applied under heavy pressure. 4. Extend sheet lead lining to all door edges, providing X-ray absorption equal to partition in which door occurs. 5. Fasten wood cores together with either countersunk steel bolts through lead with bolt heads and nuts covered with poured lead, or with poured lead dowels. 6. Bolts or dowels shall be located 38 mm (1-1/2 inches) from door edges, and at not more than 200 mm (eight inches) on center in each direction over door area. 7. Finish face of dowels and lead covering of bolt heads and nuts flush with wood cores. 8. Edge strips: a. Same species of wood as face veneer. b. Minimum thickness of edge strips shall be 38 mm (1-1/2 inches) at top edge and 63 mm (2-1/2 inches) at bottom edge. c. Glue strips to cores before face veneer is applied. d. Extend vertical edge strips full height of door and bevel 3 mm (1/8 inch) for each 50 mm (two inches) of door thickness. e. Give top and bottom edges of doors to receive transparent finish two coats of water resistant sealer before shipment to site. 9. Crossbanding of hardwood: a. Not less than 2 mm (1/12 inch) thick and face veneer not less than 1 mm (1/28 inch) thick, after sanding. b. When straight grain stock such as Basswood, Aspen or Poplar is used for crossbanding, its thickness may be 1.6 mm (1/16 inch) in lieu of 2 mm (1/12 inch). 10. Face veneer for doors specified in Section 09 06 00, SCHEDULE FOR FINISHES to have transparent finish, shall be rotary cut premium grade, uniform light, Birch. D. Hardware: 1. Hardware for doors is specified in Section 08 71 00, DOOR HARDWARE.
2. Stagger bolts to door pulls on plates which penetrate lead lining relative to opposite plate and recess on side of door opposite pull. 3. Provide lead plugs or discs over recessed nut ends of such bolts, unless otherwise shown. 4. Nut ends of bolts for surface applied hinges, door closures and automatic door operators shall be countersunk and covered with lead lined 16 gage stainless steel pans.
5. Provide round head screws with dull chromium plated finish to secure stainless steel pans.
6. Provide mortises for flushbolts, floor hinge arms, and top pivots with sheet lead on each side. Enclose floor boxes of floor hinges with sheet lead at sides and bottom.
7. Make recesses for lock and latch cases at mill and line with lead butted tightly to lead in door.
8. Make total thickness of sheet lead used for lining hardware, equivalent to thickness of sheet lead core of door.
9. Lead Lining of Frames: 1. Line or cover steel frames, stops for doors, and corner type control windows with sheet lead. 2. Install sheet lead free of waves, lumps and wrinkles with as few joints as possible. 3. Make joints in sheet lead to obtain X-ray absorption equivalent to adjacent sheet lead. Finish joints smooth and neat.
PART 3
EXECUTION 3.1 LEAD LINED GYPSUM WALLBOARD PANELS A. Apply lead lined gypsum wallboard to wood furring strips and/or to metal studs as shown. B. Predrill or drill pilot holes for nails or screws as necessary to prevent deformation of the fastener and lead shielding and to prevent distortion of the wallboard. C. Apply wallboard vertically with lead linings placed next to supports. D. Install sheet lead strips behind joints not less than the thickness used for the wallboard.
1. The lead strips: 45 mm (1-3/4 inches) wide, except at corner joints, 45 mm by 45 mm (1-3/4 by 1-3/4 inch) lead angles shall be used. 2. Secure the lead strips to supports at outer edges of strips. E. Wallboard: 1. Nail to supports with nails fastened to supports with screws and lead washers or discs at approximately 250 mm (ten inches) on centers. 2. Make provisions for connection with lead lined door frames and for cutouts for vision panels. 3. Joint treatment of lead lined gypsum board panels and fastening depressions shall be as specified for wallboard in Section 09 29 00, GYPSUM BOARD.
3.2 SUPPLEMENTAL LEAD SHIELDING A. Line or cover penetrations of wall lead, pipe chases, columns fasteners and elsewhere where shown with sheet lead. Install sheet lead free of waves, lumps and wrinkles and with as few joints as possible. Joints in sheet lead shall provide X-ray absorption equivalent to adjacent sheet lead finished smooth and neat.
B. Where plaster finish is required over columns or other vertical surfaces covered with sheet lead, drive bolts or other fasteners securing the sheet lead to backing surface half way, and wrap an 18 gage stainless steel tie wire around fasteners. Both ends of each tie wire shall be of sufficient length so that when fastener is fully driven, fastening of metal lath may be accomplished. Locate fasteners not over 400 mm (16 inches) on centers both ways and cover heads with lead strips or discs if washers are not used.
C. Provide sufficient lead shielding for spaces around outlet boxes, junction boxes, film illuminators, and pipes, to obtain a net radiation protection at these spaces equaling net radiation protection specified for wall or partition in which they occur.
3.3 SIGNS: FURNISH SIGNS AS FOLLOWS: A. One for each lead insulated partition in room in which not all partitions are insulated (or in which partitions have been insulated differently from other partitions of room), located on partition and lettered as follows: THIS PARTITION HAS BEEN INSULATED TO HEIGHT OF __8____ FEET, WITH SHEET LEAD OF ____1/32__ inch THICKNESS, PROVIDING A TOTAL LEAD EQUIVALENT PROTECTION OF 1/32 inch. B. One for each door to which only the door is insulated, lettered as follows:
THIS DOOR HAS BEEN INSULATED WITH SHEET LEAD OF __1/32_inch THICKNESS, PROVIDING A TOTAL LEAD EQUIVALENT PROTECTION OF 1/32 inch. C. Signs: 1. Heavy white paper or cardboard. 2. Height of lettering and number not less than 3 mm (1/8 inch).
3. Mount in stainless steel or extruded aluminum frames (with acrylic plastic, 3 mm (1/8 inch) thick over sign) and fasten with suitable screws, one to each corner of each frame. 5. Provide manufacturer's standard stainless steel frame, to hold card size 100 mm by 150 mm (four by six inches).
- - - E N D - - -