This document contains follow-up questions and answers related to a Request for Quote (RFQ) for a dental equipment installation or upgrade. Three key technical questions were addressed: (1) input power configuration confirmation (3-phase power), (2) electrical breaker amperage adjustment (downgrade from 20amp to 15amp), and (3) equipment component replacement (removal of a clear solid collector and installation of an HG5 Amalgam Separator). The contractor demonstrated flexibility and capability to accommodate the clinic's technical requirements, specifically confirming they can modify the electrical setup and install the requested new amalgam separator.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C24926Q0053 0002.pdf | ||
| Existing Equipment Reference.pdf | ||
| 36C24926Q0053 0001 Questions and Answers .pdf | ||
| 36C24926Q0053 0001.pdf | ||
| WDOL 15-4647 R30 12.3.25.pdf | ||
| 36C24926Q0053 final.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Follow up Questions received.
Question 1: The RFQ is requesting an RV5-2 which is a 20amp breaker size but it appears from the photos the input power is set up as a 3phase, the RV5-2 is a 1 phase see chart below: Can you confirm if it is a 1 phase or 3 phase input power?
Answer: The current setup is 3 phase.
Question 2: If the clinic would like to stay with a 3 phase system, they will need to downgrade the amperage breaker from a 20amp to 15amp, see chart below:
Answer: We will stay with a 3 phase unit, and are capable of downgrading the breaker to a 15 amp in-house.
Question 3: In the photos, there is not an Amalgam Separator installed but a clear solid collector. Is the clinic requesting to replace the clear solid collector with a new HG5 Amalgam Separator?
Answer: Yes, the contractor to remove the small clear collector & have the contractor install an HG5 Amalgam Separator.