36C24825Q0546.docx

DOCX document 19 KB Posted

Attached to
W065--6 Month Lease C-Arm XRAY System BRAND NAME OR EQUAL
Solicitation number
36C24825Q0546
Issued by

About this file

This Sources Sought Notice is for the Department of Veterans Affairs (VA) Network Contracting Office 8, seeking a 6-month lease of a portable C-Arm X-Ray System for the CW Bill Young VA Medical Center. The solicitation (36C24825Q0546) is set aside for Service-Disabled Veteran-Owned Small Businesses (SDVOSBC) and requires a Siemens Cios Alpha VA30 System or equivalent to support Electrophysiology (EP) and Pacemaker procedures during a Cath Lab renovation.

The technical requirements specify detailed specifications for the C-Arm system, including physical dimensions (30x30 cm detector, 78x33x65 inch maximum dimensions), advanced imaging capabilities such as Digital Subtraction Angiography, roadmap technology, and vascular imaging features. Additional requirements include DICOM compatibility, PACS integration with Philips Intellispace and VA VistA Imaging, encrypted hard drive, and on-site clinical training for two technologists and two physicians. The response deadline is March 20, 2025, at 14:00 Eastern Time, with interested vendors asked to submit a capability statement via email to the contract specialist, Michael A. Shook.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sources Sought Notice Sources Sought Notice

SUBJECT*
6 Month Lease C-Arm XRAY System BRAND NAME OR EQUAL

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
33744
SOLICITATION NUMBER*
36C24825Q0546
RESPONSE DATE/TIME/ZONE
03-20-2025 14:00 EASTERN TIME, NEW YORK, USA
ARCHIVE
15 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N
SET-ASIDE
SDVOSBC
PRODUCT SERVICE CODE*
W065
NAICS CODE*
334510
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs

Network Contracting Office 8 (NCO 8)

10,000 Bay Pines Blvd Bay Pines FL 33744

POINT OF CONTACT*

Contract Specialist Michael A Shook Michael.Shook@va.gov 813-972-2000 Ext 2819

PLACE OF PERFORMANCE

ADDRESS
10,000 Bay Pines Blvd

Bay Pines FL 33744

POSTAL CODE

COUNTRY

ADDITIONAL INFORMATION

AGENCY’S URL
http://www.va.gov
URL DESCRIPTION
Department of Veterans Affairs
AGENCY CONTACT’S EMAIL ADDRESS
michael.shook@va.gov
EMAIL DESCRIPTION
Contracting Officer

DESCRIPTION

Background: The CW Bill Young VA Medical Center has procured a Cath Lab Imaging System to replace the system currently used for Electrophysiology (EP) and Pacemaker procedures. During the construction and replacement of this system, Cardiology Section will be performing these procedures in an Operating Room in the Main OR. The purpose of this contract is to provide the CW Bill Young VA Medical Center (VAMC) with a short-term rental of a Siemens Cios Alpha VA30 System -Portable C-Arm X RAY System NAME BRAND OR EQUAL that will allow the continuation of EP and Pacemaker procedures in the Operating Room during renovation of the EP Cath Lab.

Please direct reply in the affirmative via email if you are interested and can provide the required product for this requirement. Please provide capability statement, I will accept responses for this Sources Sought until the 20th of March 2025 14:00, thanks.

3. Technical Requirements:

I. Unit physical specifications

a. R/F unit, mobile type
Full
b. Image type
Flat panel detector
c. Minimum detector size [cm]
30 x 30
d. Minimum central resolution [lp/mm]
2.5
e. Minimum range of motion – orbital rotation [deg]
148
f. Minimum range of motion – pivot rotation [deg]
225
g. Minimum range of motion – panning motion [deg]
12
h. Minimum range of motion – horizontal travel [cm]
20
i. Minimum range of motion – vertical travel [cm]
45
j. Minimum c-arm depth [in]
28.7
k. Minimum generator output power [kW]
25
l. Minimum capacity for images [# of images]
300,000
m. Minimum image matrix size [pixels]
1952 x 1952
n. Maximum system dimensions [in]
78 x 33 x 65
o. Maximum system weight [lbs.]
375

p. Other physical specifications

II. Additional specifications

☒
a. Flat surface beneath detector to allow for efficient table-top positioning
☒
b. Image magnification with a minimum of 3 settings
☒
c. Consistent field of view during rotation
☒
d. Tube head controls for use by clinician
☒
e. Foot pedal
Wireless
☒
f. Centering/alignment light for tube head
☒
g. Reference monitor
☒
h. Motion and metal detection
☒
i. Measurement software
☒
j. Automatic video playback
☒
k. User-configurable threshold dose warning
☒
l. Digital image rotation, reversal, and image invert
☒
m. Selectable mode settings for high-level pulse and low dose
☒
n. Touchscreen monitor
☒
o. UPS on board
☒
p. Automatic brightness stabilization (ABS) control
☒
q. Remote control for entire unit, mountable on the patient table
☒
r. Motorization of orbital, angular, and vertical movement.

III. Advanced applications

☒
a. Digital Subtraction Angiography (DSA)
☒
b. Pulsed cine Bolus chase imaging with single contrast injection
☒
c. Roadmap technology for vascular imaging, to include.

Increased frame rate from base model 30fps or greater Real-time digital subtraction Road mapping Increased image storage (1000 images or greater) Bolus tracking Vessel measurement tools and calculations Catheter/guidewire visualization enhancement Preset vascular profiles

IV. Security/Connectivity requirements

☒
a. OEM-supported operating system
☒
b. Latest DICOM print, store, commit, structured report (SR) and modality worklist
☒
c. Encrypted hard drive
☒
d. PACS compatibility – Philips Intellispace
☒
e. PACS compatibility – VA VistA Imaging
☒
f. OR table compatibility – Philips NCVA791 Xper Table Tilt

V. Clinical Training

☒
a. On-site clinical applications training for 2 technologists during go-live
☒
b. On-site clinical applications training for 2 physicians during go-live
*= Required Field