36C24822Q0952 -- Histology Staining -- Vendor Questions/Government Answers

QUESTION 7: Is Special Stain question "e - Must have the capacity to add/remove slides and reagents randomly without workflow interference" meant for IHC? If not, there is no instrument on the market that can accommodate "e"

And question "f - Must have independent slide pads for up to 40 slides for the Special Stain System with independent temperature control for each position."

Please clarify if both "e" and "f" are required or nice to have.

ANSWER: e. Not required, best value if available

f. Required

QUESTION 8: Is Special Stain question "g - Must have automated cover slipping" a requirement or a nice to have?

ANSWER: g. Not required, best value if available