Point of Care Ultrasound Salient Characteristics.pdf
PDF 67 KB Posted
- Solicitation number
- 36C24425Q0555_1
About this file
The document is a Salient Characteristics file for a Point of Care Ultrasound System, referencing the FujiFilm Sonosite PX as a brand name item. The solicitation requires an ultrasound-imaging equipment with specific technical specifications across three primary areas: 1) Imaging Capabilities, 2) Size and Design Requirements, and 3) Cybersecurity.
Key technical requirements include high-contrast imaging with advanced presets, support for multiple imaging modes (B Mode, M Mode, Color Doppler, etc.), compatibility with linear/phased/curved array transducers, and multiple measurement packages (Abdomen, Cardiac, OB/Gyn, Vascular). The system must be portable, with stand dimensions not exceeding 23"x26" and a total height of 65 inches, and include a cable management system. Cybersecurity requirements mandate built-in data encryption, secured patient data transmission/storage, and compliance with Federal Information Processing Standards (FIPS) for moderate security control selections. The opportunity is issued by the Philadelphia VA Medical Center, with quotes due by 4:30 PM EST on 4/25/2025, to be submitted via email to joanne.skaff@va.gov.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C24425Q0555 0001 Transducer Picture.docx | DOCX document | |
| 36C24425Q0555 0001.pdf | ||
| 36C24425Q0555_1.pdf | ||
| Point of Care Ultrasound Final SON.pdf | ||
| 36C24425Q0555 LineItems.xlsx | XLSX spreadsheet |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SALIENT CHARACTERISTICS
Brand Name Item for Reference: FujiFilm Sonosite PX
This acquisition is for ultrasound-imaging equipment for portable applications. Please ensure your offer meets or exceeds the salient characteristics listed below:
1. Imaging Capabilities
a. Must have high contrast resolution and image penetration
b. Must include advanced imaging presets enabling accurate visualizations for guided injections
c. Must be capable of operating in:
i. B Mode;
ii. M Mode;
iii. Color Doppler;
iv. Color Power Doppler;
v. Pulsed Wave; and
vi. Continuous Wave mode
d. Must support transducers with:
i. Linear;
ii. Phased; and
iii. Curved array formats
e. Must include a full array of measurement and calculation packages, including but not limited to:
i. Abdomen package
ii. Cardiac package
iii. OB/Gyn package
iv. Vascular package
2. Size and Design requirements
a. Must be portable
b. Must be adjustable with ability to transition between horizontal and vertical positions
c. Must include cable management system
d. Stand dimensions no greater than 23 inches x 26 inches (WxL)
e. Height maximum no greater than 65 inches (system and stand, monitor up)
3. Cybersecurity
a. Must include built-in data encryption
b. Must have secured patient data transmission and storage
c. Must meet Federal Information Processing Standards (FIPS) for moderate security control selections
File details come from the government source that posted it. Updated .