36C24418Q0081-00001001.pdf
PDF 2 MB Posted
- Attached to
- Pharmacy Fume Hood Certification
- Solicitation number
- 36C24418Q0081
About this file
36C24418Q0081 00001 Fan Powered HEPA 5 reviewed.pdf
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C24418D0189-000.docx | DOCX document | |
| 36C24418Q0081-00002001.docx | DOCX document | |
| 36C24418Q0081-00001002.pdf | ||
| 36C24418Q0081-00001000.docx | DOCX document | |
| 36C24418Q0081-00001003.pdf | ||
| 36C24418Q0081-00001006.pdf | ||
| 36C24418Q0081-00001004.pdf | ||
| 36C24418Q0081-00001005.pdf | ||
| 36C24418Q0081-001.docx | DOCX document | |
| 36C24418Q0081-002.pdf | ||
| 36C24418Q0081-000.docx | DOCX document |
Show all 11
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PHONE: 570/825-7760 FAX: 570/825-9001
SUITE 722 67 PUBLIC SQUARE WILKES-BARRE, PA 18701AENGINEERING
GROUP LTD.
RRIS
SHOP DRAWING REVIEW
CONTRACT: HVAC Sub Contractor: Marx Mechanical
PROJECT: VAMC Wilkes-Barre HVAC Upgrade Building 1 Phase 1, 693-07-136
ITEM: Pharmacy HEPA Fan Filter Unit
SUBMITTAL NO. 5
DATE: 2-18-11
Construction Submittal Review Arris Engineering
NO EXCEPTION TAKEN
MAKE CORRECTIONS NOTED X
AMEND & RESUBMIT
REJECTED – SEE REMARKS
Review is only for General Confirmation with the current design of the project non-conformities and any errors found are noted.
The CONTRACTOR is solely responsible for the correctness of dimensions and quantities and fitting to other work:
verification of physical interrelation of elements of the work from plans and specifications and in the field fabricating, procedures, construction methods, technologies and sequences. This review does not relieve the contractor from these responsibilities.
This review does not constitute approval or acceptance of deviations from Contract Documents, such deviations must be requested in writing or clearly identified as deviations in accordance
PROJECT: 854.01 SUB: # 5
DATE: 2/18/11 BY: L. Marchetti
PE
Comments:
1. HEPA-3 Unit to be 570 cfm per drawing.
2. Verify HEPA unit fan adjustment can be accomplished either from the room side of the unit as the area will be provided with a hard ceiling.
3. Are the 30% pre-filters accessible from the room side? If not, delete these filters, since the air from the air handling unit is being filtered with 30% and 95% filters prior to entering the HEPA fan unit.
Detailed, Grouped by Each Number
Submittal Transmittal
Marx Sheet Metal & Mechanical, Inc.VAMC Bldg 1 HVAC Replacement Phase 1 Project # 10/374 Marx Sheet Metal & Mechanical, Inc.
373 High Street Wilkes Barre, PA 18702
Tel: 570-824-4737 Fax: 570-824-4947
Date: 1/18/2011 Reference Number: 0005
Transmitted To: Transmitted By: Ryan MarxPaul Hamilton Hamilton Pacific Revis Marx Sheet Metal & Mechanical, Inc.
9475 Lottsford Rd.
Largo, MD 20774
PO Box 890 373 High Street Wilkes Barre, PA 18703-0890 Tel: 570-824-4737 Fax: 570-824-4947
Submittal Package No Description Due Date Package ActionQty 2/1/2011Fan Powered HEPA Diffusers0001 - 233700 - 01
Transmitted For Delivered Via Tracking Number E-MailApproval
Notes Item ActionItems DescriptionQty 1 FAn Powered HEPA Diffusers1
Contact Name Copies NotesCompany NameCc:
Remarks Submitted for Approval.
CC: 374-015
Signature Signed Date
Prolog Manager Printed on: 1/18/2011 Marx DB Page 1 ryanmarx Ryan ryanmarx Text Box 01/18/11
Marx Sheet Metal & Mechanical, Inc.
373 High Street, PO Box 890 Wilkes Barre, PA 18703 Ph: 570-824-4737 Fax: 570-824-4947 E-Mail: Info@MarxHVAC.com
SHOP DRAWING COVER SHEET
VAMC Bldg 1 HVAC Replacement Phase 1
1/18/2011
Submittal No.: 0001-233700-0
Contractor/Vendor:
General Aire Systems, Inc.
PO Box 712 Doylestown, PA 18901 Ph: 215-348-1497 Fax: 215-348-1582 Contact: Cory Lock
Specification Section Description 233700 Fan Powered HEPA Diffusers
Marx Sheet Metal & Mechanical, Inc. hereby states we have reviewed the information herewith submitted for details, dimensions and schedules and unless corrected to the contrary, have been found to be in compliance with the contractual drawings and specifications as we interpret them.
For Marx Sheet Metal & Mechanical, Inc. ________________________________________
374-015 ryanmarx Ryan
General Aire Systems, Inc.
Air Filtration Products Engineered Products Group HVAC Products Group
Representative: Cory Lock Office # :215-348-1497 Office Fax # :215-348-1582 Cell Phone: 610-496-8918
E-Mail: clock@genairesys.com
Submittal Summary
PO BOX 712, DOYLESTOWN, PA 18901
215-348-1497 Fax: 215-348-1582
Date of Submission: Monday, January 17, 2011
Project Name: V.A. Medical Center, Pharmacy
Location: Wilkes-Barre, PA
Mechanical Engineer: Arris engineering
Location: Wilkes-Barre, PA
Contractor: Attention: Rvan Marx Marx Sheet Metal, Inc.
373 High Street P.O. Box 890 Wilkes-Barre, PA 18702 Phone: 570-824-4737 Ext:
Fax #: 570-824-4947
Please visit our Web Site www.genairesys.com
Representatives For:
CamfillFarr
Camfil/Farr High Performance Products Clean Rooms International
ESI
Fiberbond Inc.
Purafil Steril-Aire Inc.
General Aire System, Inc. Service Division EnviraHealth
Products & Services ASHRAE, HEPA, ULPA & Carbon Filters/Filter Housings and Systems, Clean Room Products Bag in/Bag-out Containment Systems Clean room walls, ceilings, passthru's, HEPAIULPA Filtered Ceiling Modules, Laminar Flow Cleanroom Ceiling Systems ASHRAE Replacement Filters Gas Phase Filtration
Germicidal Ultraviolet LiQhts. UVC Duct Cleaning & Sanitizing. Indoor Air Quality Testing IAQ & Industrial Hygiene testing services, analysis & remedies for Corporate, Educational, & Industrial Environments
Prepared on 1/17/2011
AIR FILTRATION / SYSTEM SUBMITTAL SCHEDULE
General Aire Systems, Inc.P. O. Box 712, Doylestown, PA 18901Phone #: 215-348-1497 Fax # 215-348-1582Job Name: VA Med Ctr, Pharmacy
Engineer:Arris PREPARED BY: Cory lock Contractor Marx Sheet Metal
Address:Wilkes-Barre, PA
Address:
Wilkes-Barre, PA Phone #: Spare Filters: No Contact:
Joe Marx Fax #:
Phone #:
570-824-4737 Magnehelic Gages: No Fax # :
Contractor P.O. #:374-015
I SIZE OF Fil TERFil TER HOUSINGPERFORMANCERESISTANCE
ITEM
SIZE OF HOUSINGFil TER TYPE RATEDOPERATING
NUMBER
UNIT#CFMQTYHXWXD DESCRIPTIONEFFMERVWEIGHTINITIAL FINALINITIAL FINAL
HEPA-1450124X48X SAM MicroSound, 1299.9984
1XX 12" duct collar adaptor
HEPA-2450124X48X SAM MicroSound99.9984
1XX 12" duct collar adaptor
HEPA-3450124X48x SAM MicroSound99.9984
1xx 12" duct collar adaptor
SAM MicroSound OS units feature a HEPA or ULPA filter replaceable from rooms ide while the filter hous ing remains in place.
Housings are fabricated from steel with a white powdercoat finish and include seismic clip suspension points.
The energy efficient N310 motor blower is a low watt backward curved impeller, single phase, permanent split capacitor type motor blower assembly with ther mal overload protection. Units are available in 1I5V (60 Hz), 208-230V (50/60 Hz) and 277V (60 Hz).
SAM MicroSound OS fan filter units are ideal for pharmaceutical companies, manufacturers of medical devices and companies in highly regulated indus tries. The SAM OS units feature a gel channel on the upstream side of the final filter which provides a leak free seal between the filter and the housing.
Clean Rooms IrneroatiOrJai 'De4i9".i#9 7tc.dte Soi«~
Ideal for Pharmaceutical Companies Gel Seal Style HEPA or ULPA Filter
Replaceable From Roomside
SAM MicroSound Fan Filter Units include a 99.99% efficient HEPA filter, 30% ASHRAE pleated type prefilter, seismic clip suspension points, safety switch which disables the unit while servicing, and solid state variable speed control with RFI suppression, allowing the airflow to be fine tuned.
• Energy Efficient Motor: Operating Amps Low As .6 For 277 Volt And 1.6 For 115 Volt
• Safe!)' Switch: Disables Unit While Servicing
• Variable Speed Control: Allows Air Flow To Be Fine Tuned For Room Requirements
Final filters are protected with a white epoxy diamond pattern grille and are tested according to accepted pro cedures as described by the Institute of Environmental Sciences (lEST) and/or ISO standards. The stan-dard HEPA filter is 99.99% eff. and an ULPA grade 99.9995% eff. filter is available as an option.
A polystyrene eggcrate grille in an anodized alumi num frame is hinged to open, permitting removal of the filter without removing the unit from the ceiling.
Optional perforated stainless steel or aluminum grilles are available.
Other popular options include an 8' power cord (110 Volt only), 99.9995% eff. ULPA grade final filter, UL 900 Class I filters and duct collar adapters. Refer to the back of this section for more options and acces sones.
For horizontal flow applications consult with one of our cleanroom experts in your area, or call the factory for more information.
SAM is a registered trademark owned by Clean Rooms International
March/2009 Phone: 616452-8700 • Fax: 616- 452-2372 Email: sales@cleanroomsint.com www.cleanroomsint.com
Clean Rooms InterrJatiorJaJ 'De4¥«9 ?teWlte SoMioIPJ
SAM 22 MicroSound GS
GS Style Filter Data: The HEPA Filter is designed for 90 FPM average face velocity @ initial 0.47 w.g. and has a rated efficiency of99.99% @ 0.3 micron or larger.
An anodized aluminum fiame holds the 53mm media which is protected by a diamond pattern white epoxy coated steel grille.
Final filters are replaceable ITom roomside.
Access Port: This port is located in the center of the filter element face and is used for introducing an upstream challenge or measur ing static pressure.
Standard Ceiling Size: All 2' x 4' models are designed to fit a nominal 2' x 4' ceiling grid with a 22.5" x 46.5" standard opening.
Air Flow: Airflow ranges from 340 to 640 CFM for a 2 x 4 unit and 220 to 340 CFM for a 2 x 2 unit. At 90 FPM, CFM is nominal 510 on a 2 x 4 and 220 on a 2 x 2 unit respectively. Airflow for units running at 50 Hz will be about 20% less.
Sound Level: At average face velocity of 90 FPM measured 30" from face offilter, sound level is approximately 54 dba with less than 45 dba ambient sound level. Field conditions, voltage and method of testing could produce different results.
SAM 23 MicroSound GS Popular Options: Refer to next page for a list of popular options.
Filter Clamp Detail Exploded View
Swing DO\"11 ~/
Grille
Filter
Filter
Clamp
Filter
Filter Clamp Bolt
Gel Seal
Channel and
Knife Edge Seal
SAM 24 MicroSound GS (Swing Down Grille Allows Easy
Access To Filter)
Motor Data
Motor I VoltageIHzI Watts I RunningStart-Up
Amps Amps
N310
115601851.64.1
N310
208-23050/601650.8-0.7(1)1.5(2)
N310
27760165.61.5
(1) Running amps will be Jess on 50Hz
(2) Line voltage can affect actual start-up amperage
Mfg.
Code
302201
302204
302214
302215
302200
302203
302190
Replacement Filters Description
2 x 2 GS Style HEPA Filter, 99.99% elf. @ 0.3 micron
2 x 2 GS Style ULPA Filter, 99.9995% elf. @ 0.12 micron
2 x 3 GS Style HEPA Filter, 99.99% elf. @ 0.3 micron
2 x 3 GS Style ULPA Filter, 99.9995% eff. @ 0.12 micron
2 x 4 GS Style HEPA Filter, 99.99% eff. @ 0.3 micron
2 x 4 GS Style ULPA Filter, 99.9995% elf. @ 0.12 micron
Prefilter 20" x 20" xl", 30% ASHRAE efficient, pleated type
410 Phone: 616-452-8700 • Fax: 616- 452-2372 Email: sales@cleanroomsint.com
July/2007
Clean Rooms loteroatiofJaI'DC4~ ?ee.dte s.~
. Base Model (Ordering Information)Mfg. Code
Model or
Style SizeVoltageMotor
Variable
HEPA
WeightDescription Speed ControlFilter
12800'
SAM 22 MS GSGS2x2 115N31070308430220168 Ibs. (30.8 kg.)
12801
SAM 22 MS GSGS2x2208-230N31070391230220168 Ibs. (30.8 kg.)
12802-
SAM 22 MS GSGS2x2 277N31070391530220168 Ibs. (30.8 kg.)
12803-
SAM 23 MS GSGS2x3 115N31070308430221476 Ibs. (34.5 kg.)
12804-
SAM 23 MS GSGS2x3 277N31070391530221476 Ibs. (34.5 kg.)
12805"
SAM 24 MS GSGS2x4 115N31070308430220084 Ibs. (38.1 kg.)
12806
SAM24MSGSGS2x4208-230N31070391230220084 1bs. (38.1 kg.)
12807"
SAM 24 MS GSGS2x4 277N31070391530220084 Ibs. (38.1 kg.)
• UL Listed Prefilter and Safety Switch are Standard on all Models
Popular Options: (Please choose from the following part numbers when ordering options)
Final Filters:
0302204 2 x 2 GS ULPA Filter 99.9995% eff.
0302214 2 x 3 GS ULPA Filter 99.9995% eff.
0302203 2 x 4 GS ULPA Filter 99.9995% eff.
o 000000 2 x 2 GS HEPA UL 900 Class I o 000000 2 x 4 GS HEPA UL 900 Class I o 000000 2 x 4 GS ULPA UL 900 Class 1 o 000000 2 x 2 GS ULPA UL 900 Class I Prefilters:
o 309802 Super Clean Prefilter Suitable for Class 1-100
Duct Collar Adapters:
o 19950 10" Duct Collar Adapter
)( 19951 12" Duct Collar Adapter o 19952 Side Load with 10" Duct Collar o 19953 Side Load with 12" Duct Collar o 19955 10" Top Load Duct Collar Adapter o 19956 12" Top Load Duct Collar Adapter
Grilles 1/8" Round Perforated: (40% open area) o 19506 2 x 2 CRF Anodized Alum. Grille and Frame o 19507 2 x 4 CRF Anodized Alum. Grille and Frame o 19508 2 x 2 CRF Round Perfw/l/8" Holes, White Powdercoated Grille and Frame o 19509 2 x 4 CRF Round Perfw/1/8" Holes, White Powdercoated Grille and Frame o 19511 2 x 4 CRF S.S. Rnd. Perf Grille w/Anodized Alum. Frame o 19512 2 x 2 CRF S.S. Rnd. Perf Grille w/Anodized Alum. Frame o 19514 2 x 4 CRF S.S. Perf Grille w/S.S Frame
Air Conditioning Mixing Box:
o 19958 Slanted Top with 10" Duct Collar o 19959 Slanted Top with 12" Duct Collar o 19947 Square with (2) 10" Duct Collars o 19948 Square with (2) 12" Duct Collars
Roomside Variable Speed Control:
o 19933 CRF Style 110-120 Volt o 19934 CRF Style 208-230 Volt o 19935 CRF Style 277 Volt
Pilot Light:
o 19702 GS/CRF Style Red Pilot Light 110-120 Volt Only (Light Indicates Power to Motor Only) o 19703 GS/CRF Style Red Pilot Light w/Pressure Switch 110-120 Volt Only (Light Indicates Power to Motor Only)
Radial Diffuser Grille:
o 19300 2 x 2 Radial DIffuser Grille o 1930 I 2 x 4 Radial Diffuser Grille
Miscellaneous:
o 703902 8' Power Cord 11O-120V Only o 603322 Gasket on Downstream Edge of Unit o Magnehelic Gage: Refer to the Back of this Section o Static Control: Refer to the Back of this Section
Comments:
NOTE: Duct col/ar adapters, air conditioning boxes, pilot lights and adding other electrical components negate...•· UL listing.
March/2009
For more information and i/Instrations please refer to the back of this section