36C24226Q0078.docx

DOCX document 21 KB Posted

Attached to
H165--FY26 - Annual Calibration of the Hospital's Survey Meters (B+4)
Solicitation number
36C24226Q0078
Issued by

About this file

This document is a Request for Quote (RFQ) for annual calibration of radiation survey meters at the VA New York Harbor Healthcare System in Brooklyn, NY. The solicitation (36C24226Q0078) requires calibration of 3 area monitors and 8 handheld survey meters, including instruments from manufacturers like Ludlum Measurements, Fluke Victoreen, and SE International. The calibration must comply with Nuclear Regulatory Commission requirements (NUREG 1556, Appendix K), with some instruments requiring on-site calibration and others to be calibrated in the contractor's laboratory.

Key details include a response deadline of 10-30-2025 at 14:00 Eastern Time, performance at the VA Brooklyn Campus, and a 30-day contract duration. The project involves specialized calibration services, including efficiency determination for 14 survey meters, extra scale calibration, ion chamber fee assessment, and battery replacement. The contractor must provide calibration certificates for all radiation monitoring equipment. The procurement is classified under NAICS code 541690 and Product Service Code H165, with no specified set-aside status. The contracting office is located in Canandaigua, NY, with Lisa Harris as the primary point of contact.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SUBJECT*

FY26 - Annual Calibration of the Hospital's Survey Meters (B+4)

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
14424
SOLICITATION NUMBER*
36C24226Q0078
RESPONSE DATE/TIME/ZONE
10-30-2025 14:00 EASTERN TIME, NEW YORK, USA
ARCHIVE
5 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N

SET-ASIDE

PRODUCT SERVICE CODE*
H165
NAICS CODE*
541690

CONTRACTING OFFICE ADDRESS

Department of Veterans Affairs 400 Fort Hill Ave.

Canandaigua NY 14424

POINT OF CONTACT*

Contract Specialist Lisa Harris Lisa.Harris4@va.gov Lisa.Harris4@va.gov

PLACE OF PERFORMANCE

ADDRESS

VA New York Harbor Healthcare System

Brooklyn Campus

800 Poly Place NY

POSTAL CODE
11209
COUNTRY
USA

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS

EMAIL DESCRIPTION

Sources Sought Notice Sources Sought Notice

DESCRIPTION

*= Required Field
Sources Sought Notice

Sources Sought Notice

STATEMENT OF WORK

Calibration of the Hospital Survey Meters

Introduction/Background

The VA New York Harbor Healthcare System provides quality health care to veterans using the abilities of all employees supported by our commitment to education and research.

We are recognized locally, regionally, and nationally as a leader in quality patient care, positive customer service, medical/allied health education, health-related research and employment opportunities. We seek to be the provider of choice of veterans and the community by offering an efficient, integrated quality health care system capable of providing a full range of primary, specialty and chronic health care services in a system that is readily accessible and responsive to change.

Our key business drivers are quality, financial integrity, patient satisfaction and employee satisfaction.

Because of potential hazards involved in the use of radioactive materials, the VA recognizes the need for a Radiation Safety Program. The program is intended to protect personnel from unnecessary radiation exposure, to prevent contamination of our natural resources and to meet the state and federal regulations governing the possession and use of radioisotopes. Radiation can be an invaluable tool in healthcare and research and properly used can provide great benefits to mankind with little or no attendant risk. However, improper use can bring risk of high radiation exposure resulting in chronic illness, injury or even death.

The key objectives of the Radiation Safety Programs are:

· To make every reasonable effort to maintain radiation exposures, and releases of radioactive material in effluents to unrestricted areas As Low As is Reasonably Achievable (ALARA). The concept of ALARA is applied taking into account the state of technology and the economics of improvements in relation to benefits to public health and safety, and other societal and socioeconomic considerations

· To ensure control of the possession and use of "Sources of Radiation" in VA operations, research and operating programs to minimize, insofar as practicable, hazards to personnel and loss of property arising from the use of such materials

· To ensure compliance with all Federal, State and local laws covering the use of such materials, machines and devices.

Objectives/Requirements In accordance with Nuclear Regulatory Commission the hospital must calibrate hands held radiation survey meters and area monitors annually. The requirements are listed below:

The calibration must be performed in accordance with NRC requirement, Consolidated Guidance about Material License NUREG 1556, Appendix K.

List of survey meters and required actions:

CALIBRATION OF RADIATION MONITOR ON SITE: 3 AREA MONITORS:

# of units
Manufacturer
Model
Detector Type
Location and other comments
1
SE International
GM 1749

Area /monitor GM Waste room G. floor Needs to be calibrated in the contractor‘s laboratory. The unit screwed into the wall.

1
Ludlum Measurement Inc.
(Nal) 375
Portal Monitor

(Nal)

G. Floor, near Trash compactor.

Must be calibrated on site.

1
Ludlum Measurement Inc.
375P
Portal Monitor

Nal G-Floor near trash compactor.

Must be calibrated on site.

1
Berthhold
LB1210B
Xenon
Nuclear Med Room 205

LIST OF HANDHELD SURVEY

# of units
Manufacturer
Model
Detector Type
Locations and comments
3
Ludlum Measurements
3
GM
RSO office 13-E floor room 200.
1
Victreen fluke
451B-RYR
Ionization Chamber
RSO office 13-E floor room 200.
2
Ludlum Measurements
3
Gamma Scintillation
RSO office 13-E floor room 200
1
Fluke Victoreen
ASM-990
Gamma Scintillation
RSO office 13-E floor room 200
1
Eberline
ASP-1
Gamma Scintillation
RSO office 13-E floor room 200
3
Ludlum Measurements
14C
GM
Nuclear Med 2floor room -
1
Victoreen/Nuclear

Associate

470A
Ion Chamber
Nuclear Med 2 floor Room -2-524
1
Ludlum Measurements
3
GM
Radiation therapy ground floor Room-

G-116

Note-1: Time Duration: Thirty days.

Note-2 The contractor must travel one time for calibration.

Instrument Calibration with special I-125 efficiency determination: 14 survey meters Calibration of Extra Scale 14C: 3 survey meters Extra Ion Chamber Fee 470A: One survey meter Battery replacement for above instruments: For 14 survey meters

Other requirement: The contractor must provide certificate of calibration for the above radiation monitors.

Place of Performance VA New York Harbor Healthcare System Brooklyn Campus Radiation Safety Service 800 Poly Place Brooklyn, NY 11209

Calibration of the Hospital Survey Meters