Questions and Answers.pdf
PDF 128 KB Posted
- Attached to
- Q702--Medical Gas Inspection and Certification
- Solicitation number
- 36C24126Q0061
About this file
This document is a Questions & Answers (Q&A) file for solicitation 36C24126Q0061 for Medical Gas Inspection and Certification at WRJ. The Q&A provides details about medical gas system inspection requirements across three buildings: Building 1, Building 31, and Building 39. Key details include a requirement to test 100% of medical gas outlets annually, conduct cross-connection verification following NFPA 99 standards, and provide a comprehensive inventory of medical gas infrastructure. The inventory breakdown shows Building 31 has the most extensive medical gas system with 9 area alarms, 43 zone valves, and 441 outlets/inlets, while Building 1 has 6 area alarms, 24 zone valves, and 68 outlets/inlets. For the newly renovated area in Building 31's first floor, the medical gas system manufacturer is Beacon Madaes, with a Total Alert Infinity model/type.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment 1.pdf | ||
| 36C24126Q0061 0001.docx | DOCX document | |
| Attachment 2.pdf | ||
| 36C24126Q0061.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
36C24126Q0061 | Medical Gas Inspection and Certification | WRJ Questions & Answers
1. Can you please provide an inventory to include quantities as this will be helpful for us to determine time on site? An inventory of compressors, pumps, manifolds and oxygen supplies and which buildings they serve (if more than one building). Number of zone valves and area alarm panels, a count of master alarms, quantity of outlets and inlets. This information can also be found in a previous inspection report. Attachment 1 is last years report, and an inspection report from the recently renovated space in Granite. Below are totals of my count. Refer to attachments for further info.
Building Area Alarm Zone Valves Outlets/Inlets
1 6 24 68
31 9 43 441
39 2 27 70
2. What percentage of outlets are to be tested per year? 100%
3. Will there be a requirement for the outlets to require cross connection verification and/or tested for purity? Requirements should follow NFPA 99 standards. In my research, yes, cross connection verification is required for medical gas systems annually.
4. Can you provide the manufacturer, model and serial number for the med gas system? For the newly renovated area, building 31 1st floor, manufacturer is Beacon Madaes.
Model/type – Total Alert Infinity.
5. Can you provide the previous report so that we can verify the equipment list? Attachment
File details come from the government source that posted it. Updated .