S02 36C24125Q0745 0002 Vaccine Refrigerators_9-3-2025.pdf
PDF 217 KB Posted
- Attached to
- Vaccine Refrigerators
- Solicitation number
- 36C24125Q0745
About this file
This document is an Amendment (0002) to a Request for Quote (RFQ) for Vaccine Refrigerators issued by the Department of Veterans Affairs. The solicitation seeks two refrigerator models with specific requirements: Item 001 must store up to 1,300+ vaccine doses and maintain CDC recommended temperature range (2°C to 8°C) for 6 days during power outages, while Item 002 must store up to 3,600+ vaccine doses with similar temperature and power outage specifications. The VA emphasized critical needs such as self-sustaining power backup for remote locations, where past power outages have caused vaccine degradation, and external backup sources like battery packs or ice are not acceptable. The RFQ closing date remains unchanged at 09-04-2025 14:30 EDT, and the amendment includes a detailed Q&A section addressing technical specifications, storage capacity, certification considerations, and delivery expectations for 11 appliances across medication storage rooms in VA Community-Based Outpatient Clinics (COBCs).
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
5. PROJECT NUMBER (if applicable)
CODE 7. ADMINISTERED BY
2. AMENDMENT/MODIFICATION NUMBER
CODE
6. ISSUED BY
8. NAME AND ADDRESS OF CONTRACTOR
4. REQUISITION/PURCHASE REQ. NUMBER 3. EFFECTIVE DATE
9A. AMENDMENT OF SOLICITATION NUMBER
9B. DATED
PAGE OF PAGES
10A. MODIFICATION OF CONTRACT/ORDER NUMBER
10B. DATED
BPA NO. 1. CONTRACT ID CODE
FACILITY CODE CODE
Offers must acknowledge receipt of this amendment prior to the hour and date specified in the solicitation or as amended, by one of the following methods:
The above numbered solicitation is amended as set forth in Item 14. The hour and date specified for receipt of Offers
E. IMPORTANT:
is extended,
(a) By completing Items 8 and 15, and returning __________ copies of the amendment; (b) By acknowledging receipt of this amendment on each copy of the offer submitted; or (c) By separate letter or electronic communication which includes a reference to the solicitation and amendment numbers. FAILURE OF YOUR ACKNOWLEDGMENT TO BE RECEIVED AT THE PLACE DESIGNATED FOR THE RECEIPT OF OFFERS PRIOR TO THE HOUR AND DATE SPECIFIED MAY is not extended.
12. ACCOUNTING AND APPROPRIATION DATA
(REV. 11/2016)
is required to sign this document and return ___________ copies to the issuing office. is not, A. THIS CHANGE ORDER IS ISSUED PURSUANT TO: (Specify authority) THE CHANGES SET FORTH IN ITEM 14 ARE MADE IN THE CONTRACT ORDER NO. IN ITEM 10A.
15C. DATE SIGNED
B. THE ABOVE NUMBERED CONTRACT/ORDER IS MODIFIED TO REFLECT THE ADMINISTRATIVE CHANGES SET FORTH IN ITEM 14, PURSUANT TO THE AUTHORITY OF FAR 43.103(b).
RESULT IN REJECTION OF YOUR OFFER. If by virtue of this amendment you desire to change an offer already submitted, such change may be made by letter or electronic communication, provided each letter or electronic communication makes reference to the solicitation and this amendment, and is received prior to the opening hour and date specified.
C. THIS SUPPLEMENTAL AGREEMENT IS ENTERED INTO PURSUANT TO AUTHORITY OF:
D. OTHER
Contractor
16C. DATE SIGNED
14. DESCRIPTION OF AMENDMENT/MODIFICATION
16B. UNITED STATES OF AMERICA
Except as provided herein, all terms and conditions of the document referenced in Item 9A or 10A, as heretofore changed, remains unchanged and in full force and effect.
15A. NAME AND TITLE OF SIGNER 16A. NAME AND TITLE OF CONTRACTING OFFICER
15B. CONTRACTOR/OFFEROR
STANDARD FORM 30 PREVIOUS EDITION NOT USABLE
Prescribed by GSA - FAR (48 CFR) 53.243
(Type or print) (Type or print)
(Organized by UCF section headings, including solicitation/contract subject matter where feasible.)
(Number, street, county, State and ZIP Code)
(If other than Item 6)
(Specify type of modification and authority)
(such as changes in paying office, appropriation date, etc.)
(If required)
(SEE ITEM 11)
(SEE ITEM 13)
(X)
CHECK
ONE
13. THIS ITEM APPLIES ONLY TO MODIFICATIONS OF CONTRACTS/ORDERS,
IT MODIFIES THE CONTRACT/ORDER NO. AS DESCRIBED IN ITEM 14.
11. THIS ITEM ONLY APPLIES TO AMENDMENTS OF SOLICITATIONS
AMENDMENT OF SOLICITATION/MODIFICATION OF CONTRACT
(Signature of person authorized to sign) (Signature of Contracting Officer)
1 3
0002 09-03-2025
36C241
Department of Veterans Affairs (NCO 1) VAMC Leeds MA 421 North Main Street Leeds, MA 01053
36C241
Department of Veterans Affairs VAMC Manchester
718 Smyth Road Manchester NH 03104
To all Offerors/Bidders
36C24125Q0745
X
X X
X 1
1. The purpose of this Amendment is to answer questions submitted by prospective offerors
2. The RFQ closing date and time remains unchanged as: 09-04-2025 14:30 EDT.
3. In addition, all other terms and provisions described in the RFQ number of block 9A remain unchanged.
CONTINUATION: 36C24125Q0745 AMENDMENT 0002
VACCINE REFRIGERATORS
Q1: Why is it necessary for your current solution to maintain stable temperatures for 6 days without power. Don’t most patient care facilities have emergency generator backup power?
A1: This back-up power outage requirement must be self-sustaining for up to 6 days in case of power outage, specifically at remote VAMC locations. In the past, there have been instances of power outages in remote areas and power outages were not discovered in time to maintain temperature range consistency, which caused samples and vaccines to become degradated and items had to be destroyed. In addition, during power outages, this temp range cannot be maintained with external backup power sources such as battery packs, ice packs, or dry ice.
Q2: Is the uniformity data of +/- 1°C without Phase Change Material (PCM) sufficient to meet your specifications?
A2: YES.
Q3: Have you reviewed the uniformity data with PCM for a more accurate comparison?
A3: No – your company can provide a more accurate comparison.
Q4: Have you verified the storage capacity of the unit, some manufacturers use external dimensions for volume as opposed to actual storage capacity? For example a typical 20 cu ft upright pharmacy refrigerator stores up to 24,000 doses.
A4: Yes storage capacity has been verified
Q5: What specific items will be stored in the refrigeration units you are considering?
A5: Pharmaceuticals that require refrigeration. (Temp range – 36F to 46F)
Q6: When storing critical medications or vaccines, have you factored in the importance of NSF/ANSI 456 certification standards in your purchasing decision?
A6: It is not required for the refrigerator to have NSF/ANSI 456 certification, but it is considered.
Q7: Do the refrigeration units you are considering need to be ADA compliant to ensure accessibility for all users?
A7: For the location the refrigerator is located ADA is not relevant.
Q8: Do you prefer products that are manufactured in the US?
A8: No preference.
Q9: What considerations have been made for cleaning and managing potential spills within the refrigeration unit?
A9: The Clinical staff follow VA procedures on cleaning and managing spills.
CONTINUATION: 36C24125Q0745 AMENDMENT 0002
VACCINE REFRIGERATORS
Q10: It is 1300 and 3600 doses not vials for the capacity correct? (Reason I ask because to use the Covid vaccine as example 400 vials was equal to 6000 doses. So you need options that can old 400 vials more or less?)
A10: Yes
Q11: Do you require an inside delivery of all 11 appliances? Placing inside the rooms and removing all packing material? If so, please let me know and I can provide a quote for this service – note that refrigerators with standard delivery on a pallet:
RFQ: Item 001
a. Storage Capacity: Stores up to 1,300+ vaccine doses. b. Extended Hold Time Without Power: Maintains CDC recommended temperature range (2°C to 8°C) for up to 6 days during power outages (external backup power source such as ice packs, or dry ice not acceptable). The BFRV36 refrigerator meets most of the qualifications however the holdover time for this particular appliance is 4 day.
A11a: The VA will evaluate each quote based on the longest available hold-over/back-up power outage time available in the current market for vaccine refrigerators.
Item 002
a. Storage Capacity: Stores up to 3,600+ vaccine doses. b. Extended Hold Time Without Power: Maintains CDC recommended temperature range (2°C to 8°C) for up to 6 days during power outages (external backup power source such as ice packs, or dry ice not acceptable). The BFRV84 refrigerator meets most of the qualifications however the capacity is 200 of the 10x0.5ml vials = approximately 2000 doses.
A11b: This is acceptable for medication storage rooms in VA COBCs
End of Amendment 0002