PRODUCT DESCRIPTION

For

NON-EXPENDABLE EQUIPMENT:

Aspirators: Airway

Prepared By:

Strategic Acquisition Center Department of Veterans Affairs 10300 Spotsylvania Avenue, Suite 400 Fredericksburg, VA 22408

1. SCOPE OF WORK:

The Veterans Health Administration (VHA) Non-Expendable Equipment National Program has identified the Aspirators: Airway product line as a candidate for a VA-wide (otherwise referred to as “national”) single Requirements contract award Hospitals use Aspirators: Airway on crash carts and in areas that do not have access to the central vacuum system. Portable Aspirators: Airway are used to remove obstructing secretions, blood, or vomitus from a patient's airway to keep air passages to the lungs open and to allow spontaneous or mechanical ventilation. Suctioning can be either oropharyngeal (to prepare for emergency intubation or to remove secretions from the upper airway above the glottis) or tracheal (to remove secretions and to maintain the airway of an intubated patient).

The Contractor must demonstrate the ability to meet all requirements for the solicitation. The objective is to provide Aspirators: Airway to be used by clinicians throughout the VA medical centers and facilities. The period of performance is for one (1) 12-month base period with four (4) twelve (12) month option periods.

2. REQUIREMENT

This requirement will be IAW FAR clause 52.211-6, “Brand Name or Equal” which requires the quoter indicate that each product being offered as an “equal” product to the Aspirators: Airway. For each equal product, the offeror must include a description reflecting the salient characteristics (SC) and level of quality that will satisfy the salient physical, functional, or performance characteristics of the equal product(s) specified in the solicitation. The quoter must also clearly identify the item by brand name (if any) and make/model number. Finally, the quoter must include descriptive literature, such as illustrations, drawings, or a clear reference to previously furnished descriptive data or information available to the contracting officer, and clearly describe any modifications it plans to make to a product to make it conform to the solicitation requirements.

Contract Line Items
Manufacture
Part Number
Description

SSCOR

2314
SSCOR Duet Flat Back (Model 2314) - portable countertop model (includes Canister, 6’ patient tube, HI-D® suction tip, AC power cord and QuickStart Guide)

SSCOR

2314B
SSCOR Duet with Retention Bracket (Model 2314B) – portable mountable model (includes Canister, 6’ patient tube, HI-D® suction tip, AC power cord and QuickStart Guide)

The Department of Veterans Affairs (VA) is seeking vendors who can provide Aspirators: Airway as listed above or equal commodities which meet all the following salient characteristics:

CLIN 0001 and CLIN 0002 apply to the following SC’s:

SC #
SALIENT CHARACTERISTICS
METHOD OF EVALUATION
SC – Literature Map

SSCOR

Duet Allied OptiVac

SC 1
The unit should be portable for intrahospital use.
Literature Review
Pg. 1
Pg. 1
SC 2
The unit should be able to provide suction of at least ≥ 400 mm Hg at maximum vacuum settings.
Literature Review
Pg.2
Pg. 2
SC 3
The unit should have an adjustable external vacuum gauge display and LED battery charging status indicator lights.
Literature Review
Pg. 1 & 2
Pgs. 1 & 2
SC4
The unit should have a performance/vacuum (air) flow rate of at least ≥ 30 lpm.
Literature Review
Pg. 4
Pg. 2
SC 5
The unit should support a disposable collection canister of at least ≥1,000 mL capacity.
Literature Review
Pg. 3
Pg. 4
SC 6
The unit should be equipped with an internal automatic battery charger and minimum 100V – 240V AC Electrical Systems.
Literature Review
Pg. 2
Pg. 2
SC 7
The unit should be equipped with a battery depletion (bad battery) indicator warning.
Literature Review
Pg. 5
Pg. 2
SC 8
Fully charged batteries should power the unit at maximum/full capacity vacuum for ≥30 minutes.
Literature Review
Pg. 2
Pg. 2