Copy of CFM JUNE 2026 SPECIAL EQ CONSOLIDATION - Schedule of Line Items with QA.xlsx

XLSX spreadsheet 208 KB Posted

Attached to
6525--CFM SPECIAL EQ Consolidation
Solicitation number
36A79726Q0018
Issued by

About this file

This is a Schedule of Line Items with Quality Assurance documentation for VA medical equipment procurement. The document outlines equipment requirements across multiple VA VISNs (Veterans Integrated Service Networks) and consolidates vendor questions with government responses regarding technical specifications.

The procurement involves medical imaging equipment for five line items: XR CT equipment for VISN 9 Nashville (PO 626B64043), XR RAD equipment for VISN 9 Nashville (PO 626B64044), XR RAD equipment for VISN 15 Topeka Kansas (PO 589B62001), and XR RAD equipment for VISN 20 Walla Walla Washington (POs 687B63002 and 687B63007). The document documents vendor inquiries seeking to increase competition through specification modifications and government responses. Key modifications approved include: wireless detector weight limits increased to 9.25 lbs, minimum gantry aperture reduced from 75 cm to 70 cm, minimum patient table load capacity dynamic reduced to 660-680 lbs, pixel size changed to 3.4 lp/mm, and minimum tube height adjusted to 12.5 inches. The government rejected certain requests, including removal of UPS requirements (maintaining 10-minute full functionality during power outages for both CT and RAD systems), retention of fluoroscopy capability with remote-controlled solutions, and retention of dual energy imaging specifications. Technical specifications were clarified regarding stitching location (wall only), grid ratio preferences (12:1), and tabletop movement requirements. Amendment 0001 reduced equipment quantity for 687B63002 from 2 to 1 unit and removed equipment 687B63007 from the consolidation. Updated technical specifications and compliance matrices with dates after 7/27/2026 are required for proposal submissions.

View the file

Other files

Show all 12

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SCHEDULE OF LINE ITEMS

HTME VISNSTATIONEQUIPMENT POEQUIPMENT TYPEQTY
General
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64043XR CT1NOTE: Row 2 is intentionally skipped.
VISN 9: VA Mid South Healthcare Network626 NASHVILLE TN626B64044XR RAD1The List of Requirements start with Row 3.
VISN 15: VA Heartland Network589 TOPEKA KS589B62001XR RAD1
VISN 20: VA Northwest Network687 WALLA WALLA687B63002XR RAD1
VISN 20: VA Northwest Network687 WALLA WALLA WA687B63007XR RAD1REMOVED

QA TEMPLATE

USE DROP DOWN MENU for Column A, then Columns B, C, and D will auto populate
EQUIPMENT POVISNSTATIONEQUIPMENT TYPEVENDOR EQUIPMENT QUESTIONSGOVERNMENT EQUIPMENT RESPONSE
626B64044VISN 9: VA Mid South Healthcare Network626 NASHVILLE TNXR RADSystem Digital Detector Configuration:

Can the maximum wireless wall detector weight and maximum wireless table detector weight, with battery installed [lbs] be changed to 9.25lb to increase competition?

Yes; tech spec updated to 9.25 lbs
626B64044VISN 9: VA Mid South Healthcare Network626 NASHVILLE TNXR RADUnit Physical Specifications:

Can the maximum pixel size [µm] be changed to 3.4 lp/mm to increase competition?

Can the Minimum tube height from floor when used with wall stand [in] be changed to 12.5 inches to increase competition?

Can the minimum patient table load capacity dynamic be changed to 660 to increase competition?

Yes; tech spec updated to 295 µm

Yes; tech spec updated to 12.5 in

Yes; tech spec updated to 660 lbs 626B64044 VISN 9: VA Mid South Healthcare Network 626 NASHVILLE TN XR RAD Additional Specifications:

Can the UPS – maintains full functionality of the entire modality / system for a minimum of 10 minutes during a power outage – be removed as this is not necessary for the entire system in radiography?

No, UPS needed to allow for proper shut down of the system in the event of a power outage
626B64043VISN 9: VA Mid South Healthcare Network626 NASHVILLE TNXR CTUnit Physical Specifications:

Can the minimum gantry aperture be revised from 75 cm to 70 cm to increase competition?

Yes; tech spec updated to 70 cm
626B64043VISN 9: VA Mid South Healthcare Network626 NASHVILLE TNXR CTAdditional Specifications:

Is emergency power available? If so is a full system UPS required or is it just needed for the work station?

Emergency power will not be available for the CT; tech spec for full functionality updated to 10 minutes to allow for proper shut down of device during outage
626B64043VISN 9: VA Mid South Healthcare Network626 NASHVILLE TNXR CTAdditional Specifications:

Is emergency power available? If so is the surge protector with acquisition workstation on sufficient?

Emergency power will not be available for the CT; tech spec for full functionality updated to 10 minutes to allow for proper shut down of device during outage
589B62001VISN 15: VA Heartland Network589 TOPEKA KSXR RAD5 g: Auto Rotate – ability to automatically rotate acquired images: As only two vendor can meet this requirement, please consider changing this to added value to increase competition.This is a clinical requirement and more than one vendor can meet it.
589B62001VISN 15: VA Heartland Network589 TOPEKA KSXR RAD3.j. Motor-driven tabletop movement in the X and Y axis at stated maximum weight capacity - Please remove this specification as this is more appropriate/applicable for RF system tables (x/y axis motorized movement)Tech spec updated
687B63002VISN 20: VA Northwest Network687 WALLA WALLAXR RAD#3 Additional Specifications, Letter O – "Have fluoroscopy capability" - Does the VA prefer a patient-side/tableside fluoroscopy system or a remote-controlled fluoroscopy systemWe don’t want a tableside/tower fluoro unit. We want a remote solution with capability to fluoro in control room.
626B64043VISN 9: VA Mid South Healthcare Network626 NASHVILLE TNXR CTLine 23 Minimum gantry aperture 75 cm: Can the minimum be
lowered to 72 cm for free and fair competition?Yes, tech spec updated to 70 cm
626B64043VISN 9: VA Mid South Healthcare Network626 NASHVILLE TNXR CT2k. Would a partial UPS that allows for safe shutdown and

getting the patient off the table suffice?

If a full UPS is required is there adequate citing available?Emergency power will not be available for the CT; tech spec for full functionality updated to 10 minutes to allow for proper shut down of device during outage
589B62001VISN 15: VA Heartland Network589 TOPEKA KSXR RAD5a. Do you need to do stitching at the wall and table or wall only?Tech spec updated. Wall only
589B62001VISN 15: VA Heartland Network589 TOPEKA KSXR RAD3n-q) Please edit to a 12:1 ratio grid to maximize vendor response, as several different grid ratios are available

5b) Does the station prefer Dual Energy Imaging instead of Bone Suppression Software?

"3n-q) Tech spec updated.

5b) The station prefers Dual Energy. Tech spec updated."

687B63002VISN 20: VA Northwest Network687 WALLA WALLAXR RAD3o) Please remove 'fluroscopy' capability as this is a vendor specific feature
Ed) Please remove 'integrated patient lift' as this is a vendor specific feature3o) No, there are at least two vendors with solutions that meet this requirement.

Ed) This is listed as an added value not a requirement so this will not be removed.

589B62001VISN 15: VA Heartland Network589 TOPEKA KSXR RAD2.e Can the minimum patient table load capacity (dynamic)
be reduced to 680lbs to allow for more competition?Tech spec updated
589B62001VISN 15: VA Heartland Network589 TOPEKA KSXR RAD3.j Can this request be removed, as there is also a
request for a free-floating tabletop?Tech spec updated