Jan 2022 Consolidation - Schedule of Line Items with QA - FINAL 4-3-22.xlsx

XLSX spreadsheet 54 KB Posted

Attached to
6525--January 2022 Consolidation
Solicitation number
36A79722Q0010
Issued by

View the file

Other files

File Type Posted
573B25017 - 635B20005 TS.zip ZIP file
640B24001 - 757B29001 TS.zip ZIP file
36A79722Q0010 0001.docx DOCX document
541B20003 - 573B25015 TS.zip ZIP file
504B20001 - 539B20002 TS.zip ZIP file
Amendment 0001 - Continuation Page.pdf PDF
Vendor Folders.zip ZIP file
RFQ Instructions to Offerors - January 2022 Consolidation.pdf PDF
January 2022 Consolidation Schedule.pdf PDF
Memorandum for Turnkey.pdf PDF
605B20005 - 688B20003 BM.zip ZIP file
Vendor Workup Spreadsheet - 36A79722Q0010.xlsx XLSX spreadsheet
436B20003 - 605B20004 BM.zip ZIP file
573B25008 - 573B25019 TS.zip ZIP file
VA Directive 6550 - New Appendix A - published Jun. 2019.pdf PDF
516B20004 - 541B20003 TS.zip ZIP file
Jan 2022 Consolidation - Schedule of Line Items - Updated 3-2-2022.xlsx XLSX spreadsheet
Memorandum for Training.pdf PDF
504B20001 - 516B20003 TS.zip ZIP file
642B20005 - 757B29001 TS.zip ZIP file
541B20014 - 561B20005 TS.zip ZIP file
580B20019 - 640B24001 TS.zip ZIP file
36A79722Q0010_1.docx DOCX document
Show all 23

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Schedule of Line Items

HTME VISNStationRequisition Number in ActionEquipment TypeQtyBenchmark (BM) or Technical Specification (TS)
436VISN 19: VA Rocky Mountain Network436 BILLINGS, MT436B20003XR US1BM
499VISN 23: VA Midwest Health Care Network499 FARGO, ND499B29001XR US1BM
499VISN 23: VA Midwest Health Care Network499 DES MOINES, IA499B29002XR US6BM
504VISN 17: VA Heart of Texas Health Care Network504 AMARILLO, TX504B20001XR PORT RAD1TS
504VISN 17: VA Heart of Texas Health Care Network504 AMARILLO, TX504B27040XR PORT C-ARM1TS
509VISN 7: VA Southeast Health Care Network509 AUGUSTA, GA509B22006XR CT1TS
509VISN 7: VA Southeast Health Care Network509 AUGUSTA, GA509B22007XR CT1TS
512VISN 5: VA Capitol Health Care Network512 BALTIMORE, MD512B21007XR RF1TS
512VISN 5: VA Capitol Health Care Network512 BALTIMORE, MD512B21008RT LINAC1TS
512VISN 5: VA Capitol Health Care Network512 BALTIMORE, MD512B21009XR CT1TS
516VISN 8: VA Sunshine Healthcare Network516 ST PETERSBURG, FL516B20001XR MAMMO1TS
516VISN 8: VA Sunshine Healthcare Network516 ST PETERSBURG, FL516B20002XR MOBILE CT1TS
516VISN 8: VA Sunshine Healthcare Network516 ST PETERSBURG, FL516B20003XR US3TS
516VISN 8: VA Sunshine Healthcare Network516 ST PETERSBURG, FL516B20004XR MRI1TS
520VISN 16: South Central VA Health Care Network520 BILOXI, MS520B20011XR US1TS
520VISN 16: South Central VA Health Care Network520 BILOXI, MS520B20012XR US1TS
520VISN 16: South Central VA Health Care Network520 BILOXI, MS520B20013XR US1TS
520VISN 16: South Central VA Health Care Network520 PENSACOLA, FL520B20014XR US1TS
520VISN 16: South Central VA Health Care Network520 BILOXI, MS520B20015XR US1TS
521VISN 7: VA Southeast Health Care Network521 BIRMINGHAM, AL521B12034XR CATH LAB1BM
534VISN 7: VA Southeast Health Care Network534 CHARLESTON, SC534B22007XR CT1TS
534VISN 7: VA Southeast Health Care Network534 CHARLESTON, SC534B22008XR CT1TS
539VISN 10: VA Healthcare System of Ohio539 CINCINNATI, OH539B20002XR CT1TS
541VISN 10: VA Healthcare System of Ohio541 CLEVELAND, OH541B20003XR MRI1TS
541VISN 10: VA Healthcare System of Ohio541 CLEVELAND, OH541B20014XR PORT C-ARM1TS
541VISN 10: VA Healthcare System of Ohio541 CLEVELAND, OH541B20017XR PORT C-ARM1TS
541VISN 10: VA Healthcare System of Ohio541 CLEVELAND, OH541B20018XR US6BM
541VISN 10: VA Healthcare System of Ohio541 CLEVELAND, OH541B20026XR PORT C-ARM1TS
541VISN 10: VA Healthcare System of Ohio541 CLEVELAND, OH541B20027XR US2BM
542VISN 4: VA Stars & Stripes Healthcare Network542 COATESVILLE, PA542B20006XR CT1TS
544VISN 7: VA Southeast Health Care Network544 COLUMBIA, SC544B12070XR US2BM
546VISN 8: VA Sunshine Healthcare Network546 MIAMI, FL546B20008XR MRI1TS
549VISN 17: VA Heart of Texas Health Care Network549 DALLAS, TX549B20214XR MRI1TS
549VISN 17: VA Heart of Texas Health Care Network549 DALLAS, TX549B20215XR MRI1TS
549VISN 17: VA Heart of Texas Health Care Network549 DALLAS, TX549B20220NM BONE DEN1TS
550VISN 12: VA Great Lakes Health Care System550 DANVILLE, IL550B20013XR CT1TS
561VISN 2: VA Healthcare Network Upstate New York561 EAST ORANGE, NJ561B20005XR US1TS
573VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE, FL573B25008NM SPECT CT1TS
573VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE, FL573B25009XR CT1TS
573VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE, FL573B25010XR CT1TS
573VISN 8: VA Sunshine Healthcare Network573 TALLAHASSEE, FL573B25012XR RAD1TS
573VISN 8: VA Sunshine Healthcare Network573 LAKE CITY, FL573B25013XR RF1TS
573VISN 8: VA Sunshine Healthcare Network573 THE VILLAGES, FL573B25015XR RAD1TS
573VISN 8: VA Sunshine Healthcare Network573 JACKSONVILLE, FL573B25016XR PORT C-ARM1TS
573VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE, FL573B25017XR US2TS
573VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE, FL573B25018XR RAD1TS
573VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE, FL573B25019XR US1TS
578VISN 12: VA Great Lakes Health Care System578 HINES, IL578B28002XR US3BM
580VISN 16: South Central VA Health Care Network580 HOUSTON, TX580B20019XR MRI1TS
580VISN 16: South Central VA Health Care Network580 HOUSTON, TX580B20020XR MRI-LINAC1TS
586VISN 16: South Central VA Health Care Network586 JACKSON, MS586B20002XR US1BM
589VISN 15: VA Heartland Network589 KANSAS CITY, MO589B20044XR US1BM
589VISN 15: VA Heartland Network589 KANSAS CITY, MO589B20045RT LINAC1BM
589VISN 15: VA Heartland Network589 KANSAS CITY, MO589B20046XR CT1TS
593VISN 21: VA Sierra Pacific Network593 LAS VEGAS, NV593B20005XR CATH LAB1TS
596VISN 9: VA Mid South Healthcare Network596 LEXINGTON, KY596B20014XR US1TS
600VISN 22: VA Desert Pacific Healthcare Network600 LONG BEACH, CA600B20007WORKSTATION1BM
603VISN 9: VA Mid South Healthcare Network603 LOUISVILLE, KY603B23007WORKSTATION1BM
603VISN 9: VA Mid South Healthcare Network603 LOUISVILLE, KY603B23008XR PORT C-ARM1BM
605VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B20004XR PORT RAD1BM
605VISN 22: VA Desert Pacific Healthcare Network605 LOMA LINDA, CA605B20005XR PORT RAD1BM
612VISN 21: VA Sierra Pacific Network612 MCCLELLAN, CA612B29000XR RAD1TS
620VISN 2: VA Healthcare Network Upstate New York620 MONTROSE, NY620B21005XR US1BM
620VISN 2: VA Healthcare Network Upstate New York620 MONTROSE, NY620B21006XR US2BM
631VISN 1: VA New England Healthcare System631 LEEDS, MA631B21001XR US1TS
635VISN 19: VA Rocky Mountain Network635 OKLAHOMA CTY, OK635B20004XR MAMMO1TS
635VISN 19: VA Rocky Mountain Network635 OKLAHOMA CTY, OK635B20005XR US1TS
640VISN 21: VA Sierra Pacific Network640 PALO ALTO, CA640B24001XR MAMMO1TS
642VISN 4: VA Stars & Stripes Healthcare Network642 PHILADELPHIA, PA642B20005XR PORT C-ARM1TS
646VISN 4: VA Stars & Stripes Healthcare Network646 PITTSBURGH, PA646B22006XR US1TS
654VISN 21: VA Sierra Pacific Network654 RENO, NV654B20005XR US3BM
655VISN 10: VA Healthcare System of Ohio655 SAGINAW, MI655B20005NM BONE DEN2BM
657VISN 15: VA Heartland Network657 ST LOUIS, MO657B20006XR CT1TS
657VISN 15: VA Heartland Network657 MARION, IL657B22003XR CT1TS
662VISN 21: VA Sierra Pacific Network662 SAN FRANCISCO, C662B20005XR MRI1TS
662VISN 21: VA Sierra Pacific Network662 SAN FRANCISCO, C662B20006XR US1TS
663VISN 20: VA Northwest Network663 SEATTLE, WA663B23003XR PORT C-ARM3TS
663VISN 20: VA Northwest Network663 SEATTLE, WA663B23004XR US1TS
663VISN 20: VA Northwest Network663 SEATTLE, WA663B23005XR US1TS
667VISN 16: South Central VA Health Care Network667 SHREVEPORT, LA667B29002XR RAD1TS
671VISN 17: VA Heart of Texas Health Care Network671 SAN ANTONIO, TX671B20019XR CT1TS
672VISN 8: VA Sunshine Healthcare Network672 SAN JUAN, PR672FC2001RT LINAC1BM
673VISN 8: VA Sunshine Healthcare Network673 TAMPA, FL673B20022XR PORT RAD1BM
675VISN 8: VA Sunshine Healthcare Network675 ORLANDO, FL675B27001XR US7BM
675VISN 8: VA Sunshine Healthcare Network675 ORLANDO, FL675B27002XR US7BM
675VISN 8: VA Sunshine Healthcare Network675 ORLANDO, FL675B27003XR US1BM
688VISN 5: VA Capitol Health Care Network688 WASHINGTON, DC688B20003RT LINAC1BM
757VISN 10: VA Healthcare System of Ohio757 COLUMBUS, OH757B29001XR MRI1TS

Q&A

VISNStationRequisition NumberQUESTIONSGOVT Response
VISN 23499 FARGO, ND499499B29001Please confirm station name for 499B29001Ultrasound is for FARGO VA, Station 437.
VISN 23499 CENTRAL IOWA499499B29002Please confirm station name for 499B29002Ultrasounds are for CENTRAL IOWA VA, Station 636A6.
VISN 17504 AMARILLO, TX504504B20001Technical Specifications matrix:

1d. Minimum tube rotation from axis in both directions (degrees): 300 Is this axis meant to describe column rotation?

To allow for open and fair competition, please change to ± 180°.

1e. Minimum fixed detector size (in x in): 14x17 Please remove or revise - there are no fixed detectors for portable x-ray systems

2a. Quantity of wireless detectors: 2 Detector quantity labeled as 2. Specifications only specify 14x17. Does the site require 2 - 14x17 or are there other sizes required?

1d. Tech spec updated. Changed to 180.

1e. Tech spec updated. Revised to wireless detector

2a. Tech spec updated. Revised to 1 wireless detector VISN 17 504 AMARILLO, TX 504 504B20001 1a - Please reduce to 32 kW to ensure maximum competition 1c - To ensure maximum competition, please change to 0.1 - 320. There are no portable x-ray exams that should ever require 500 mAs given the efficiency (DQE and TQF) of today's detectors.

1e - Please change to "wireless" instead of "fixed" 1f - Please change to "wireless" instead of "fixed" 2 - Please better define what the site is looking for with respect to requesting 2 detectors with a minimum size of 14"x17". Does the site want qty 1 14"x17" and qty 1 greater than 14"x17"? Qty 2 14"x17" detectors? Qty 2 17"x17" detectors?

3y - Please remove the spec as only one vendor offers this software. 1a Tech spec updated. Changed to 30kW 1c - Tech spec updated. Changed to .1-320 1e - Tech spec updated. Changed to wireless 1f - Tech spec updated. Changed to wireless 2 - Tech spec updated. Changed to 1 wireless 3y - Tech spec updated - removed.

VISN 17 504 AMARILLO, TX 504 504B20001 1.b - 65kV max HVG power is too low for clinical use. Recommend 150kVp requirement since 150kVp is getting required in clinical use for obese patient.

2.d - Detector distribution weight of 300lb is too low for obese patient in portable use. Recommend max 881lb(400kg) requirement for obese patient.

3.x - Pls add comments as dual laser alignment for SID positioning. Single laser cannot guide SID, but dual laser can provide it. 1b Tech spec updated. Changed to 150kVP 2d - Tech spec updated. Changed to equal to greater than 600 3x - Tech spec updated. Changed to single laser VISN 17 504 AMARILLO, TX 504 504B20001 1A. High Frequency Generator Power - Q: Please change this specification to 30 kW to allow for more fair competition.

3G. Algorithms to Reprocess For Tubes and Lines - Q: Please elaborate on this specification regarding reprocessing algorithms, would a list of ALL image processing algorithms on the system be sufficient?

3I. Ability To Move Portable From Tube Head - Q: Please make optional, as this feature limits open competition

3K. Trickle Charge Method For Detector Battery - Q: Is this specification looking to confirm whether or not the detector is charged while it is still in the bucky?

3M. Adjustable Motor Drive - Q: Can you please elaborate on the clinicial / operatoinal goal for this specification?

1A. Tech spec updated. Changed to 30 kW

3G. This is the portable unit’s ability to have additional technology to reprocess certain exams when looking for central lines, feeding tubes, etc. This ability is to be able to enhance such lines for better viewing. We will not need a list of ALL image processing algorithms

3I. This is a clinically relevant, desired feature

3K. This is a descriptive equipment specification simply asking whether it is desired or not. Trickle charging is assumed that the portable will continue to charge the detector while plugged in to a power source but not actively shut down for a full charge.

3M. This is a descriptive equipment specification simply stating whether this feature is desired or not. Low vs. High speed typically comes in to factor when the tube head is not parked sufficiently, and a clinician needs to move at a heightened speed to quickly address patient needs.

VISN 17504 AMARILLO, TX504504B27040Can you get the actual ‘serial number’ for the 9900 listed below.Assuming reference to the trade-in GE unit: Serial # of the base of the unit: E9-0881
VISN 07509 AUGUSTA, GA509509B22006Requirement Overview: Dual-source CT is a vendor specific CT design. Multiple vendors accomplish the same acquisition speeds with a single, wide-area detector. Please make dual-source CT a desired characteristic and not a specific requirement in order to allow for full and fair competition.

4. Advanced Applications - There are multiple references to "FORCE" software applications. These are vendor specific trade name which refer to Siemens specific product catalogs. This requirement for a specific vendors catalog names to be included in an open bid is unduly restrictive as only one vendor can meet this specification. Please remove these trademark Siemens catalog names or make them optional to allow full and fair competition

1. a. This is a vendor specific specification. Please change to 256 slices to allow full and fair competition

1. c. Please make dual detectors optional to allow for full and fair competition

1. e. Please change to .625mm to allow full and fair competition

1. j. This is a vendor specific requirement which would only allow a single vendor to bid. Please change the requirement for 2 tubes to optional to allow full and fair competition

1. l. This is a vendor specific requirement which would only allow a single vendor to bid. Please make this requirement optional to allow for full and fair competition

1. n. "Turbo Flash" is a vendor trade name and is unduly restrictive. Plesae make "Turbo Flash" optional to allow full and fair competition Dual-Source was moved to the added value section E.

All references to FORCE were removed from Tech Specs

1a.Changed to 256

1c. Dual Detectors moved to Added Value Section E.

1e. Changed to .625 1j. 2 Tubes moved to Added Value Section E.

1l was removed

1n “with Turbo Flash” was removed VISN 05 512 BALTIMORE, MD 512 512B21007 Question #60- consider 500lbs to promote fair competition?

Question #61- consider 350lbs to promote fair competition?

Question #64- please clarify full articulation definition?

Questions #71 and #72- consider 19” monitors to promote fair competition? Minimum patient table load capacity (static) has been updated to 500lbs to promote competition

Minimum patient table load capacity (dynamic) has been updated to 350lbs to promote competition

Vertical, transverse, and longitudinal movement of the patient table

Minimum control room monitor size has been updated to 19” to promote competition

Minimum in-room monitor size has been updated to 19” to promote competition VISN 05 512 BALTIMORE, MD 512 512B21007 Technical Specifications matrix:

Added Value, Question d - Wider patient table Please specify the required measurements for a wider table.

The minimum table width was defined in the table requirements section as 31”. A table width greater than 31” would be wider than the minimum requirement.
VISN 05512 BALTIMORE, MD512512B210081.Will you consider a comparable item the meets or exceeds the proposed benchmarked system?
2.Under B. Technical Requirements, b. Minimum microwave power [MW], could you please elaborate more on this request?
3.Under B. Technical Requirements, j. Minimum patient table load capacity [lbs], should the table capacity be 500lb maximum?1.Yes, all offers will be considered
2.The microwave power source can be either a magnetron or a klystron, but must provide a minimum 2.5 MW.
3.The patient table must be able to withstand loads of at least 500 lbs
VISN 08516 ST PETERSBURG, FL516516B200021.a 256 slice is not a configuration typically installed in mobile trailers. Would the site consider modifying the specification to either 64 or 128 slices, which are our only available mobile CT slice options.

1.b 80 cm bore is not available on a mobile configuration. Would the site consider modifying to 70 cm to allow for a greater range of options to consider.

1.c 8 cm detector is not available on a mobile. Would the site consider revising to 38.4 mm to allow for a greater range of options to consider

2.a/b Fluoroscopy / intervention is not performed on mobile units. There would be challenges with maintaining a sterile environment for these types of procedures. Would the site consider removing this spec. 1.a Changed to 128 slices

1.b Changed to 70 cm

1.c Changed to 40mm

2.a/b Removed requirements for fluoroscopy VISN 08 516 ST PETERSBURG, FL 516 516B20002 Requirement Overview: The project requests a "mobile" CT. Is the intention of the facility to routinely move the system or will the system remain in place. If the system will remain in place, would a modular building similar to the existing MRI be an acceptable or desired solution?

1. Unit Physical Specifications e. .5 mm is unduly restrictive. Please change this specification to .625mm to allow full and fair competition

1. h. CT scanners routinely utilize rotations above .3 seconds. Please change maximum rotation time to 1 second to accommodate routine CT protocols

1. j. 1900 lbs is unduly restrictive. Please change this specification to "maximum system weight must accommodate placement in a mobile environment" to allow full and fair competition

2. k. Mobile CT trailers are not designed to accommodate the required space for a full UPS back up. The normal mobile trailer will accommodate a partial UPS to power the CPU's during a power outage. Please change the requirement to a partial UPS to safely save patient data during a power outage. Must be a trailer; will be moved eventually

1.e Changed to 0.625 mm

1.h Changed to 1 second

1.j Removed Maximum System Weight requirement

2.k is now 2.i Changed to Partial UPS VISN 08 516 ST PETERSBURG, FL 516 516B20004 1. c. Current MRI technology is designed to minimize bore depth. Please change to 105cm to allow full and fair competition.

1. f. The industry standard for RF Channels is the number of Analog- to- Digital Converters (ADC) for a system. No commercial MR system has 200 ADC. Please change to 64 to allow full competition.

1. g. maximum system dimensions 60 inches x 30 inches - Please clarify as these dimensions are too small for a 1.5T MR system

2. c. Please change to 500 lbs to allow full competition

3. f. UPS - maintains full functionality of the system for a minimum of 30 minutes during a power outage - Please change to 6 minutes to align with industry standards and clarify if UPS is for full MR system or system console 1.c Changed to 105 cm

1.f Changed to 64

1.g Changed to 235 cm x 200 cm x 235 cm

2.c Changed to 500 lbs

3.f Changed UPS requirement to 7 minutes

VISN 16520 BILOXI, MS520520B20014Please confirm station name for 520B20014Gulf Coast Veterans Health Care System Station 520 (400 Veterans Ave, BLDG 26 Biloxi, MS) as listed in original tech specs
VISN 16520 BILOXI, MS520520B20015XR US Biloxi
Request 1.k. Recommend k. Minimum battery life of 24 hours be reduced to 40 minutes.1.k Updated tech spec
VISN 07534 CHARLESTON, SC534534B220071.a Would the site consider alternative technologies / slice counts that would achieve or exceed the following stated objectives: "high slice count ... better image quality, better spatial resolution, larger imaging area, lower radiation dose, significant reduction of artifacts, and an increase in both dependability and reliability. Ultimately, reducing procedure time, image processing/viewing time, and increasing clinic workflows." Recommendation to allow for 256 and/or 384 slice variations.

1.b Would the site consider allowing for 78 cm gantry opening to allow for a greater range of options that can achieve or exceed the stated clinical objectives

1.c Would the site consider an alternative detector size that would allow for a greater range of options that can achieve the stated clinical objectives?

2.l Is a full system UPS needed or can a partial UPS for the acquisition workstation / reconstruction computer suffice?

2.p Would the site consider removing the requirement for gantry tilt to allow for a greater range of options that can achieve the stated clinical objectives?

2.r Would the site consider removing the requirement for gantry tilt to allow for a greater range of options that can achieve the stated clinical objectives?

1.a: No, High slice CT is a clinical need. More than one vendor can meet the 512 requirement

1.b: No, the largest gantry aperture is required by our facility. We have a large bariatric patient population. More than one vendor can meet the requirement

1.c:.No, We have developed a cardiac imaging program and hope to further it by providing quality cardiac scans to a wider range of veterans to include those with higher heart rates and arrythmias. The 16cm detector size allows for the acquisition of the entire heart in one rotation, mitigating motion artifact and misregistration artifact caused when a smaller detector size is present

2.l Yes, Partial UPS is acceptable, however please also include a full UPS as an option

2.p:: No, Gantry tilt can be performed mechanically or digitally. More than one vendor can meet the requirement.

2.r: No, Gantry tilt can be performed mechanically or digitally. More than one vendor can meet the requirement

VISN 07534 CHARLESTON, SC534534B220071. o. maximum non-segmented reconstruction time 150 msec - Q: Please clarify the clinical need being asked with this specification.
1. g. minimum high-contrast spatial resolution .20 mm - Please change to .23mm to allow full competition1.o Low reconstruction time allows the technologist to assess image quality as close to the acquisition time as possible, allowing for split-second decision making in instances where a second acquisition must be obtained in order to prevent a patient from receiving a second dose of contrast. This applies primarily with CT angiography

1.g Yes, Changed to .23 mm to allow for more competition

VISN 07534 CHARLESTON, SC534534B220071.Will the VA allow the vendors to quote a 78cm gantry , as the 80cm gantry aperture minimum requirement is not met by most of the vendors.1.b: No, the largest gantry aperture is required by our facility. We have a large bariatric patient population. More than one vendor can meet the requirement
VISN 07534 CHARLESTON, SC534534B220081.Minimum Acquisition rate 70 f/s? This must be a typo no system can do 70 frames per second as a maximum much less a minimum.
2.Minimum rotation speed (degrees/second) 55. This is either a typo or do they mean minimum rotation speed for 3D?
3.Maximum number of slices acquired simultaneously 128? They must mean minimum
4.Maximum 360 degree rotation time, .30. Will .35 sec be acceptable?
5.Minimum generator size kW output, 100. Will 72kW be acceptable?
6.Remove PET image fusion to allow for more competition
7.Remove live 2K imaging for more competition.1.Updated to reference 3D acquisition

2. Updated to reference 3D rotation speed

3. Yes, updated to say minimum

4. No, the fastest rotation time allows for less noise when performing cardiac imaging procedures. The VA hopes to hope to further it by providing quality cardiac scans to a wider range of veterans to include those with higher heart rates and arrythmias.

5. No, more than one vendor can meet the need.

6. CT and MR image fusion is a requirement. PET image fusion was moved to added value section

7. No, Live 2K provides improved visibility and resolution of procedures in real time. The VA is looking for a High resolution suite with real time image processing

VISN 10539 CINCINNATI, OH539539B2000239.m. Gantry Tilt - Please make gantry tilt optional or consider either gantry tilt or digital tilt for similar clinical outcomes.
55.f. Other advanced applications Dose Tracking? Integration to Advanced Visualization System - Q: Please clarify what Advanced Visualization System requires integration.39m. Gantry tilt is a clinical requirement. No change has been made.

55f. Advanced Visualization System Integration has been removed from the technical specification.

VISN 10 541 CLEVELAND, OH 541 541B20003 1) On the solicitation there is no reference to the trade in of any MRI coils. Please see the attached trade in request form and please list the coils that will be traded in. The following coils are available with the MRI trade-in:

Canon/Toshiba Coils: NV Head, Cervical, Spine, Hand/Wrist, Shoulder, 2 Flex, Abdomen, Knee, 3 round surface coils (small, medium, large), rectangle surface coil (for extremities), Breast, and a T/R head coil.

VISN 10541 CLEVELAND, OH541541B20017Regarding the Matrix Line k: The requirement is for a 30KW generator. Would the requirement be lowered to a 25KW generator for more competition?
A 30KW generator is clinically required and will not be updated.
VISN 10541 CLEVELAND, OH541541B20018a.What specific software packages are needed
b.What specialty will be using these units
c.What procedures do they intend to perform with this ultrasound.Application: POCUS program CLE has an active POCUS program in place using Sychronicity application: required SW packages listed in the benchmark and correspond to the procedures the US units required to perform.
VISN 10541 CLEVELAND, OH541541B20027a.What specific software packages are needed
b.What specialty will be using these units
c.What procedures do they intend to perform with this ultrasound.Application: Anesthesia/Cardiology OR; There are no specific software packages, the description of everything is included under line item one of benchmark including needle visualization of software applications
VISN 04542 COATESVILLE, PA542542B20006Will neuro perfusion imaging be performed and if so will perfusion maps be created by the scanner or by a third-party workstation?Neuro perfusion imaging will not be performed with this equipment.
VISN 04542 COATESVILLE, PA542542B20006i. UPS – maintains full functionality of the system for a minimum of 30 minutes during a power outage - Q: Please change to 6 minutes to align with industry standards and clarify if UPS is for full MR system or system consoleUPS – maintains full functionality of the full CT system for a minimum of 6 minutes during a power outage. This changed has been made to the technical specifications.
VISN 04542 COATESVILLE, PA542542B20006Seeking clarification on the following specs: the stated numbers appear to be reversed on line C & D. Please clarify if Min detector width should be 30mm and Min FOV should be 50cm

In section 1: Technical Requirements Line C: Minimum detector width z-axis [mm] 50

Line D: Minimum scan field of view [cm] 30Confirmed. Detector width 30mm minimum and FOV 50 cm minimum. This changed has been made to the technical specifications.
VISN 04542 COATESVILLE, PA542542B20006What specific types of dual-energy studies do you expect to perform?Dual-energy studies to be performed with this equipment include detection of pulmonary embolism, differentiation of contrast from calcified plaques for vessel stenosis evaluation on CT angiograms, differentiating the composition of kidney stones (uric acid vs calcium), gouty urate from other calcific deposition.
VISN 04542 COATESVILLE, PA542542B20006What type of ‘CT Liver Analysis’ will be performed and what clinical decisions are to be supported?CT liver analysis will not be performed with this equipment. This has been removed from the technical specifications.
VISN 08546 MIAMI, FL546546B200081.The benchmark asks for a mobile trailer, with no specifications. Can the VA provide trailer size/specs ?The 546B20008 is not for a trailer MRI request; it replaces an existing MRI installed on the building’s first floor.
VISN 17549 DALLAS, TX549549B202141.Specs call for an optional ceiling mount injector. This done in CT, but not in MRI as the ceiling mount means that the contractor has to penetrate the RF shield. To my knowledge injector companies do not offer MRI ceiling mounts for this reason.

2. Specs ask for a fixed or wand type metal detector. Would they prefer a fixed? We offer a fixed type that will record the event with real time video, and generate a report.

3. Specs call for a MRI Compatible Anesthesia Ventilator. The Anesthesia Vent is part of the anesthesia machine. We would suggest that they purchase a MRI Compatible anesthesia machine with vent. We do not furnish either the ventilator, or the machine with ventilator.

4.We would like more information on the trade in of the GE 450 W system.
•Is it a GEMS system?
•What coils are included?
•What software?
•Will the hard drive be removed?
•Is the system currently operational?
•Is the system currently filled with helium and is the cold head running?1. Multiple vendors can meet this requirement and is needed by the clinical staff to meet their workflows.

2. Either option would be acceptable

3. No mention of “Anesthesia Ventilator” in tech spec. A MRI-Compatible Ventilator is a value-add option which does not affect vendor eligibility.

4.

• Is it a GEMS system? Yes

• What coils are included?

- Full Coil Package

- Abdominal

- Large Flex

- Small Flex

- Long leg run-off

- Breast

- Neurovascular Head

- Shoulder

- Extremities

• What software?

- 3D applications

- Syngo.via

- Neuro Package

- Cardiac Package

• Will the hard drive be removed? - YES

• Is the system currently operational? - YES

• Is the system currently filled with helium and is the cold head running? - YES VISN 17 549 DALLAS, TX 549 549B20214 1. For row 54, will a mobile MRI compatible injector be acceptable?

2. For row 114, will security/password software be sufficient?

3. For row 124, Per NAC rules of engagement off-site training can be offered only in the “option” section, do you still want this included or more on site hours?

4. Please clarify id these items listed in the “Added Value” section should be listed as options and not included in the body of the quotation?

-MRI compatible stretcher, MRI compatible ventilator, MRI compatible fusion pump, User presents to standardize workflows, Advanced software/applications to improve or create efficient clinical workflows to decrease patient wait times and increase case throughput, Continuous clinical training opportunities to combat against staff overturn and new hires, Human factors system and area design to improve patient and staff experience 1. Yes

2. Yes

3. Yes, add more hours for on-site hours.

4. Yes, added values are shall be listed as options.

VISN 17 549 DALLAS, TX 549 549B20214 1f. System weight 6000lbs - Please increase maximum weight to 9000lbs to allow fair competition 3v. MRI Compatible injector ceiling mounted -Please change to MRI compatible injector pedestal or ceiling mounted to allow fair competition.

4c Workstations 2 - Please clarify if you would like a total of 4 workstations between to the 2 MRI systems or would you like 1 workstation per MRI? 1f. Tech spec will not change. Multiple vendors meet this requirement.

3v. Tech spec will not change. Multiple vendors meet this requirement.

4c. 2 workstations are required per MRI VISN 17 549 DALLAS, TX 549 549B20215 1.Specs call for an optional ceiling mount injector. This done in CT, but not in MRI as the ceiling mount means that the contractor has to penetrate the RF shield. To my knowledge injector companies do not offer MRI ceiling mounts for this reason.

5. Specs call for a MRI Compatible Anesthesia Ventilator. The Anesthesia Vent is part of the anesthesia machine. We would suggest that they purchase a MRI Compatible anesthesia machine with vent. We do not furnish either the ventilator, or the machine with ventilator.

6. Specs ask for a fixed or wand type metal detector. Would they prefer a fixed? We offer a fixed type that will record the event with real time video, and generate a report.

7.The specs call for a workstation.
•What software is required on this workstation? Cardiac, 3D or any other software packages?

8. Off site applications are requested for three technologists. This is not a problem for us to furnish, but normally the VA is not able to allow off site applications.

9.We would like more information on the trade in of the Siemens Avanto system
•Is it a TIM system?
•What coils are included?
•What software?
•Will the hard drive be removed?
•Is the system currently operational?
•Is the system currently filled with helium and is the cold head running?1. Two or more vendors can meet this requirement and is needed by the clinical staff to meet their workflows.

5. No mention of “Anesthesia Ventilator” in tech spec. A MRI-Compatible Ventilator is a value-add option which does not affect vendor eligibility.

6. Please provide whatever option is available. There is currently no preference to a fixed or wand type.

7. The software needed for the workstation is outlined in the Advanced Applications section.

8. It is something VA North Texas would like to have. It has been converted to on-site training if things change.

9. • Is it a TIM system? - YES

• What coils are included?

-Full Coil Package
-Abdominal
-Large Flex
-Small Flex
-Long leg run-off
-Breast
-Neurovascular Head
-Shoulder
-Extremities

• What software?

-3D applications
-Syng.via
-Neuro Package
-Cardiac Pacakge

• Will the hard drive be removed? - YES

• Is the system currently operational? - YES

• Is the system currently filled with helium and is the cold head running? - YES VISN 17 549 DALLAS, TX 549 549B20215 1f. System weight 6000lbs - Please increase maximum weight to 9000lbs to allow fair competition 3v. MRI Compatible injector ceiling mounted -Please change to MRI compatible injector pedestal or ceiling mounted to allow fair competition.

4c Workstations 2 - Please clarify if you would like a total of 4 workstations between to the 2 MRI systems or would you like 1 workstation per MRI? 1f. Multiple vendors meet this requirement.

3v. Multiple vendors meet this requirement.

4c. 2 workstations are required per MRI.

VISN 17 549 DALLAS, TX 549 549B20220 a. Minimum scan time at 1% CV for whole body [s] 40 - Please change minimum standard scan time to 339 seconds for whole body to allow fair competition

g. Exposure time [s]/entrance dose [mGy] for lumbar spine 10/0.040 - Please change minimum spec to 10x/0.047mGy to allow fair competition

h. Exposure time [s]/entrance dose [mGy] for proximal femur 10/0.040 - Please change minimum spec to 10x/0.047mGyto allow fair competition

j. Exposure time [s]/entrance dose [mGy] for IVA option HD 15/0.025 - Please change minimum spec to 15/0.0269mGy (183s/.329mGy)to allow fair competition

l. Minimum patient table load capacity [lbs]- 450 - Please change Minimun patient weight to 350 to allow fair competition

p. Maximum system dimensions [in]: 40x100x50 - Please chnage system dimensions to 113x52x49 to allow fair competition a. Tech spec will not change, clinical requirement

g. Tech spec will not change, clinical requirement

h. Tech spec will not change, clinical requirement

j. Tech spec will not change, clinical requirement

i. Tech spec will not change, clinical requirement

p. Tech spec will not change, clinical requirement VISN 12 550 DANVILLE, IL 550 550B20013 1.c Would the site consider modifying to a detector width of 38.4 mm to allow for a greater range of options that can achieve or exceed the stated spec of 256 slices (1a) and allow for a greater range of options that can all achieve or exceed the stated clinical objectives?

1. Lines 18 and 20 you are requesting minimum 256 acquired slices per rotation and an 8 cm wide CT detector. Will you accept a 128 acquired slice system with a 4 cm detector?

2. Line 24 – benchmark for table height range from floor is 55-103. Our table height has a range of 67-104. This is a difference of 12 mm or ~ 0.5 inch. Will you accept this difference?

1c. Danville will consider a detector size of 38.4 so long as it achieves or exceeds stated clinical objectives.

1. Danville will consider a 128 acquired slice system with a 4cm detector.

2. Table height range is a categorical difference and it will be considered comprehensively. Danville will consider a table with this range.

VISN 12550 DANVILLE, IL550550B200132.h. UPS – maintains full functionality of the system for a minimum of X minutes during a power outage. - Q: Do you want the proposal to include a partial UPS or would you like to use your existing full UPS?Current UPS allows for a 10-20 minutes of functionality during a power outage. It is the vendors responsibility to ensure the submitted quote takes UPS compatibility into consideration. Danville will use the exising UPS if compatible. If it is not compatible the cost of a replacement should be included in the quote.
VISN 02561 EAST ORANGE, NJ561561B20005Questions for NAC regarding Line #37:
1)What is the application for “additional specifications” line f. Shear wave and strain-based elastography? If specific transducer is required, should that transducer be included?
2)Total number of transducers should be 3each (2each C10-4ec and 1each L15-7io)1.) This was moved to "Added Value"

2.) Adjusted the total number of transducers. The need is for the Curved Array transducers to be able to be used for Transrectal Ultrasound and Prostate Biopsies. A total of 3 Curved Array and 1 Linear Array transducers are needed.

VISN 08 573 GAINESVILLE, FL 573 573B25008 3. h. automatic collimator exchanger - Please change to "semi-automated collimator exchange" as this requirement is exclusive to one vendor and is unduly restrictive based on available technology

4. a. minimum workstation monitor size - Please change to 17 inches to allow full and fair competition

6. l. Auto QC - Please make this requirement optional to allow full and fair competition 3h. Semi-Auto collimator is acceptable 4a. 17in is acceptable 6I. Auto QC is required

VISN 08573 GAINESVILLE, FL573573B25008a.Can we get more information on the Siemens Symbia T6 trade-in? Information requested is:
i.Collimators – Will they be included? What type are they?
ii.Acquisition & Processing Stations – Will they be included? What is the Software Version?
iii.What is the Operating system & software?
iv.Has the camera been under a service contract?
v.Are you able to provide photos of your trade in? If yes, please send with the responses.
vi.Will the original moving equipment be with the system?i. LEHR, LEUHR, MEGP, HEGP

iii. Software version Syngo VB10E/ Operating system Windows 10

iv. Yes

v. NO

vi. Not available on site VISN 08 573 GAINESVILLE, FL 573 573B25009 1.c Site requested 64 slices (1.a) and 160 mm detector. This configuration is not possible. Would the site consider revising the specification to be 38.4 mm to allow for a greater range of technology that delivers 64 slice scanning and meets or exceeds the stated clinical objectives?

2.j Site requested a full system UPS to allow for 20 minutes. Is a full-system UPS actually needed or only a partial UPS for the acquisition workplace and reconstruction computers? If the answer is "Yes, full system" would the site consider revising the number of minutes of uptime to allow for ~7 minutes of scanning. 20 minutes of scanning time is not possible even on a full system UPS.

1.c Yes 2j. UPS was removed from the specification - only a partial UPS for the acquisition workplace and reconstruction computers is needed.

VISN 08 573 GAINESVILLE, FL 573 573B25010 1.c Site requested 64 slices (1.a) and 160 mm detector. This configuration is not possible. Would the site consider revising the specification to be 38.4 mm to allow for a greater range of technology that delivers 64 slice scanning and meets or exceeds the stated clinical objectives?

2.j Site requested a full system UPS to allow for 20 minutes. Is a full-system UPS actually needed or only a partial UPS for the acquisition workplace and reconstruction computers? If the answer is "Yes, full system" would the site consider revising the number of minutes of uptime to allow for ~7 minutes of scanning. 20 minutes of scanning time is not possible even on a full system UPS. 1.c Yes 2j. UPS was removed from the specification - only a partial UPS for the acquisition workplace and reconstruction computers is needed.

VISN 08 573 GAINESVILLE, FL 573 573B25012 Technical Specifications matrix:

2c. Minimum generator voltage range (kV): 10 - 120kV To allow for open and fair competition, please change to 40 - 150kV

2j. Minimum table height from floor (in): 15in To allow for open and fair competition, please change the requirement to (22 in 2c. 40-150kV is acceptable 2j. Specification changed to 22in

VISN 08573 GAINESVILLE, FL573573B250122j - Please change to 21" to maximize competition.2j. Specification changed to 22in
VISN 08573 GAINESVILLE, FL573573B250122.j - Minimum table height from floor [in] 15 --> Please change from 15" to 21.5" for fair competition. It appears to be a lockout spec eliminating other vendors from competing.
3.l - Pls remove Motor-driven tabletop movent requirement. It is function of fluoroscopy table and no DR product provide motorized X-Y table movement.2j. Specification changed to 22in

3l. Specification removed - please note specification 3k - Tabletop should be free floating

VISN 08573 GAINESVILLE, FL573573B250122. j. minimum table height from floor - please change to 20 inches to allow full and fair competition
4. c. touchscreen acquisition workstation monitor - please make this requirement optional to allow full and fair competition2j. Specification changed to 22in

4c. Touch screen is optional VISN 08 573 GAINESVILLE, FL 573 573B25013 Technical Specifications matrix:

2f. Minimum SID range (in): 34" - 48" Please revise the minimum SID range of 34" as this is not recommended for best practices of radiation protection. Please revise to 45".

4a. Maximum lowest table height from floor (in): 18in To allow for open and fair competition, please change the requirement to 20 in 2f. Specification has been updated 4a. Specification changed to 20in

VISN 08573 GAINESVILLE, FL573573B250131.Removable lead curtains with counter weights for when curtains are not in use- Please remove multipurpose type R/F systems do not have the ability to attach lead curtain to the flat panel detector
2.UPS – maintains full functionality of the system for a minimum of 20 minutes during a power outage. Please consider a single phase computer only UPS. A full system UPS will add significant cost to the system and to the site preparation.
3.Tilt range- please consider changing to +89/-89 degrees.1. Specification has been updated to require Lead shielding in the form of rolling/ceiling mounted.

2. Full UPS specification was removed 3.+/-89 is acceptable VISN 08 573 GAINESVILLE, FL 573 573B25015 Technical Specifications matrix:

2c. Minimum generator voltage range (kV): 10 - 120kV To allow for open and fair competition, please change to 40 - 150kV

2j. Minimum table height from floor (in): 15in To allow for open and fair competition, please change the requirement to (22 in) 2c. Specification changed to 40-150kV 2j. Specification changed to 22in

VISN 08573 GAINESVILLE, FL573573B250152.j - Minimum table height from floor [in] 15 --> Please change from 15 to 21.5 for fair competition. It appears to be a lockout spec eliminating other vendors from competing.
3.l - Pls remove Motor-driven tabletop movent requirement. It is function of fluoroscopy table and no DR product provide motorized X-Y table movement.2j. Specification changed to 22in

3l. Specification removed - please note specification 3k - Tabletop should be free floating

VISN 08573 GAINESVILLE, FL573573B250152. j. minimum table height from floor - please change to 20 inches to allow full and fair competition
4. c. touchscreen acquisition workstation monitor - please make this requirement optional to allow full and fair competition2j. Specification changed to 22in

4c. Touch screen moved to the value added section

VISN 08573 GAINESVILLE, FL573573B25017XR US Gainesville
Can you verify 2.b. Navigation Software is required? This system would not typically be used to Fuse CT & MR imaging together and Navigate the with the Ultrasound Image.Specificatoin for naigation software was removed.
VISN 08573 GAINESVILLE, FL573573B25018Technical Specifications matrix:

2c. Minimum generator voltage range (kV): 10 - 120kV To allow for open and fair competition, please change to 40 - 150kV

2j. Minimum table height from floor (in): 15.7in To allow for open and fair competition, please change the requirement to (22 in) 2c. Specification changed to 40-150kV 2j. Specification changed to 22in

VISN 08573 GAINESVILLE, FL573573B250182j - Pease change to 21" to maximize competition.2j. Specification changed to 22in
VISN 08573 GAINESVILLE, FL573573B250182.j - Minimum table height from floor [in] 15.7 --> Please change from 15.7 to 21.5 for fair competition. It appears to be a lockout spec eliminating other vendors from competing.
3.l - Pls remove Motor-driven tabletop movent requirement. It is function of fluoroscopy table and no DR product provide motorized X-Y table movement.2j. Specification changed to 22in

3l. Specification removed - please note specification 3k - Tabletop should be free floating VISN 08 573 GAINESVILLE, FL 573 573B25018 1. b. If the detectors in both the wall bucky and the table are wireless do you need an additional 3rd wireless detector?

4. c. Please make touchscreen optional to allow full and fair competition

5. a. Are stitching capabilities required at the table and wall bucky?

5. b. Ability to apply multiple image processing algorithms both pre and post-acquisition to allow for soft tissue and/or bone enhancement - Is the requirement for Dual Energy to be part of the offer system/offer? 1b. Yes, an additional wireless detector is required 4c. touchscreen was moved to the "Added Value" section 5a. yes 5b. Dual Energy is not needed.

VISN 16 580 HOUSTON, TX 580 580B20019 1. Can the Vendor quote a 3rd Party mobile rental trailer for the site , or does this need to be done by the site outside of this award ?

2. Specs call for an optional ceiling mount injector. This done in CT, but not in MRI as the ceiling mount means that the contractor has to penetrate the RF shield. To my knowledge injector companies do not offer MRI ceiling mounts for this reason.

3. In the specs, a MRI Compatible Stretcher is desired. Due to the inpatient load, and heavy use of the system, would the VA Houston prefer a second table instead? This would allow the patient to be moved to the table, prepped, and then brought into the room in one easy step, instead of transferring from a stretcher in two steps. The second table would increase throughput and make it much easier on the staff in terms of moving patients.

4.We would like more information on the trade in of the GE extremity system.
•What coils are included?
•What software?
•Will the hard drive be removed?
•Is the system currently operational?
•Is the system currently filled with helium and is the cold head running?1. 3rd party rental can be included in offer

2. Updated tech spec

3. please quote a stretcher and include a table as an and added value option for consideration

4.

Coils: unknown SW version: unknown HDD will be removed Yes It is currently filled and running VISN 16 580 HOUSTON, TX 580 580B20019 Technical requirements:

c. Minimum bore depth [cm] 105cm. How is this measurement calculated? True Bore size or cover to cover?

d. Minimum field of view[cm]. Please adjust minimum field of view requirements to 50x50x45cm?

3. e. UPS – maintains full functionality of the system for a minimum of 30 minutes during a power outage. Please adjust to 20 minutes?

3. k. RF shielding – Please explain RF Shielding and who would be responsible for installing the shielding? Typically included with the room construction and not by the magnet vendor.

3. l. Magnetic shielding. – Please explain what Magnet Shielding statement? In the Room design or internal within the Magnet?

3.s Ceiling-mounted – Will a mobile MRI compatible injector be acceptable?

3. w. Patient Comfort/Experience package. Could this be included in with the Design and Construction Contractor? If not, please consider moving to added value section.

6. Safety Requirement

a.Safety system for… Wouldn’t the Oxygen monitor in gantry room be supplied and included in the construction costs if needed?
b.Fire suppression system… Wouldn’t the Fire Suppression system be supplied and provided as part of the construction costs if needed?

c. cover to cover

d. Updated tech spec

3e. Updated tech spec

3k. RF shielding will be designed, procured and installed as part of site prep scope. MRI vendor must provide recommended specs with bid

3l. After survey if it is determined that Magnetic shielding is required, this will be designed, procured and installed as part of site prep scope. If shielding internal to the magnet is available, it should be quoted. If external to the magnet, MRI vendor must recommend specs

3s. Updated tech spec - pedestal mounted/mobile injector acceptable 3w. No, this must be through by MRI vendor

6. Fire suppression and life safety devices will be designed, procured and installed as part of site prep scope. Technical specifications changed to add “compatibility with” to statement. Please provide recommendations/specifications if applicable.

VISN 16 580 HOUSTON, TX 580 580B20019 5n. Endorectal Coil - Do you specifically need an endorectal coil or do you need the clinical capability to perform a prostate exam?

3s. MRI injector- ceiling mounted - Please change MRI injector to ceiling OR pedestal mounted to allow fair competition 3w. Patient comfort/experience package - Please give specific information regarding custom cabinet/storage requirements and exact lighting needs in order to give a comprehensive quote meeting facility needs and expectations Please clarify if you would like post processing included to support advanced clinical applications? 5n.clinical capability and coils(if applicable))

3s. Updated tech spec - pedestal injector acceptable

3w. cabinet storage should be sufficient to store provided coils. Vendor is recommended to make the best judgement based on facility conditions.