This statement of work outlines requirements for a portable operating microscope to support activation of the VA Nebraska-Western Iowa Healthcare System Ambulatory Care Center. The Department of Veterans Affairs seeks one Carl Zeiss OPMI Lumera 700 System microscope and associated accessories from authorized vendors. Delivery is required between May 15th and June 20th 2020 to the Ambulatory Care Center address listed. Installation, training, calibration and documentation are also specified.
The related federal contract opportunity notice announces a forthcoming request for quotes under solicitation number 36A77620Q0259 for the Lumera microscope. Responses are due within 99 days of issuance with performance at the Omaha VA facility. The opportunity is unrestricted and conducted according to FAR 13.106-1(c)(1).
36A77620Q0259 P09-FINAL LUMERA SOW 4.7.20 M8535 - SOW (Rev01).docx
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36A77620Q0259-001.pdf | ||
| 36A77620Q0259-000.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
VA Nebraska-Western Iowa Healthcare System Initial Outfitting, Transition & Activation Statement of Work M8535 – Microscope, Operating, Portable
VA Nebraska-Western Iowa Healthcare System 4101 Woolworth Avenue Omaha, NE 68105 Ambulatory Care Center Project
Introduction Veteran Affairs Nebraska-Western Iowa Healthcare System (VANWIHS) has a requirement to purchase M8535 – Microscope, Operating, Portable with an authorized vendor to supply and deliver the item(s) identified in Table 1 below. See salient characteristics below. This requirement is in support of the activation of VA Nebraska-Western Iowa Healthcare System Ambulatory Care Center. The Ambulatory Surgery department has the need for a portable operating microscope for use within the Operating Rooms. The current manufacturer is the basis for design.
| CLIN |
| Item Number |
| Item Description |
| Quantity |
| 0001 |
| M8535 |
| CARL ZEISS MEDITEC, INC. | FLOOR STAND |
- OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | 6M CABLE FOR
14FCT. FCP - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | WIRELESS
FOOT CONTROL PANEL - OPMI LUMERA 700
SYSTEM
CARL ZEISS MEDITEC, INC. | XENON
SUPERLUX EYE - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | LED LIGHT
SOURCE - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | OBJECTIVE
LENS F=200MM OCT - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | OVERHEAD
DISPLAY - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. |
TRANSP.BAGGAGE(FLOOR,S/U) - OPMI
LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | ASSISTANT
SCOPE W/EL.ZOOM - OPMI LUMERA 700
SYSTEM
CARL ZEISS MEDITEC, INC. |
000000-2111-61430AL128D - OPMI LUMERA
700 SYSTEM
CARL ZEISS MEDITEC, INC. | 60D ASPHERIC
LENS - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | INT.DATA
INJECT.SYS W/SHUTTER - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | RESIGHT
METAL TRAY - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | RESIGHT 700 -
OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | RESCAN 700 -
INTEGR.OCT CAMERA - OPMI LUMERA 700
SYSTEM
CARL ZEISS MEDITEC, INC. | SET DOVETAIL
MOUNT RESIGHT - OPMI LUMERA 700
SYSTEM
CARL ZEISS MEDITEC, INC. | 1 LENS
HOLDER F200 - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | TRIO610
CAMERASYSTEM - OPMI LUMERA 700
SYSTEM
CARL ZEISS MEDITEC, INC. | CALLISTO
EYE,V3.2 - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | INT.CALLISTO
EYE FLOOR STAND - OPMI LUMERA 700
SYSTEM
CARL ZEISS MEDITEC, INC. | BASIC SET -
OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | EYEPIECE
10X,2PC - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. | EYEPIECE
10X,2PC - OPMI LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. |
INVERTERTUBE E INCL. ASEPSIS - OPMI
LUMERA 700 SYSTEM
CARL ZEISS MEDITEC, INC. |
INVERTERTUBE E MOT. W/ASEPSIS - OPMI
LUMERA 700 SYSTEM
| Installation and Calibration |
| 1 |
| On-Site Training |
| 1 |
Table 1 – Package Items Summary
Additional Item Requirements For all Operating Microscope systems, the vendor shall provide user manuals (quantity 2) that describe proper use, care and cleaning for users of the equipment.
In addition, the vendor will provide service manuals (quantity 2) that includes theory of operation, maintenance requirements, repair information, schematics, part lists and troubleshooting information in electronic format for use by the VA Biomedical Engineering Department.
The equipment must not be used or refurbished and must have the latest manufacturer’s software/hardware updates installed. Date of manufacture must not be greater than 2 years before the purchase order date. Any batteries provided with the unit will not be manufactured more than 12 months before the purchase order issue date.
General Conditions Site Address The address of the VA Nebraska-Western Iowa Healthcare System Ambulatory Care Clinic is provided below.
4101 Woolworth Avenue Omaha, NE 68105
Delivery Schedule Items shall be delivered no earlier than the estimated delivery start date and no later than the delivery end date listed in Table 2 below.
| Building Name |
| Building # & Letter |
| Acronym |
| Est. Delivery Start Date |
| Est. Delivery End Date |
| Ambulatory Care Center |
| 4101 |
| 15 May 2020 |
| 20 June 2020 |
Table 2 – Anticipated Building Outfitting Schedule Schedule and Order Management
| 2.3.1 | The above delivery schedule is based upon the current construction schedule. The vendor shall not increase prices for a change in delivery date for up to 30+ days due to construction delays. |
| 2.3.2 | The awardees actual delivery dates will be confirmed by VA upon award. |
| 2.4.3 | Post-award, the vendor will coordinate delivery identified in this document with the VANWIHS prior to beginning any work (See Section 2.4). Once the dates have been confirmed, modifications to the schedule are subject to written approval by the CO, COR and the vendor. Schedule changes that are beyond the dates specified in the contract will require approval by the Contracting Officer and a contract modification. |
Delivery and Receiving
2.4.1 Delivery and receipt of the proposed items is anticipated to be directed to the location identified below. Confirmation of delivery location will be provided post-award. To coordinate delivery, contact Government POC Mark Stockstell at (402) 995-5847, mark.stockstell@va.gov, Colin Beal at (402) 930-7775, colin.beal@va.gov, Dan Thompson at (785) 477-2820, dan.thompson@sme-solutions.co, and Derick Lacky at (301) 602-9109, dlackey@hcrmoves.com.
☒ VANWIHS Omaha VA Ambulatory Care Clinic (Direct Delivery to address detailed in paragraph 2.1.1) ☐ IOT&A Contractor-Provided Warehouse Local to the VANWIHS SME Solutions Warehouse: Attn: SME Solutions-VA Project, Gratton Warehouse Co. 11005 E Circle Omaha, NE 68137
| 2.4.2 | The vendor is required to define the need for a staging area to accommodate item delivery, if needed. Post-award, this information will be confirmed with the vendor. |
| 2.4.3 | The delivery of items identified in this document shall take place during normal business hours which are defined as: 8:15 AM to 3:00 PM Central Standard Time, Monday through Friday, and excluding Federal Holidays. After hour deliveries may need to be scheduled. Those hours are from 6:00 PM to 5:30 AM. There is no loading dock available during this time. All delivery vehicles must be equipped with a “Tommy Gate” type lift to facilitate unloading to ground level. |
| 2.4.4 | Delivery trucks will not be permitted to remain at the loading dock. Trucks shall be unloaded, moved from the dock, and then brought back to the dock if required to haul out any waste, tools, or excess materials. |
| 2.4.5 | Labeling of delivered items shall include the VA facilities contract number and VA purchase order number for identification and reference upon receipt of product. |
| Use of Warehouse |
If the vendor requires the use of the Government’s activation contractor’s warehouse to meet the requirements of this contract the vendor must provide a certificate of insurance prior to the delivery and offloading of the item(s). This insurance certificate must be completed and presented to the activation contractor prior to delivery.
The vendor shall communicate through the VA and IOT&A POC any required coordination requirements.
Clean Up and Disposal
2.6.1 There are no dumpsters available for vendor use.
2.6.2 The removal of waste and/or excess material shall be conducted through the loading dock area.
Assembly, Installation, and Calibration Required (Yes ☒ No ☐)
2.7.1 The vendor is required to manage and coordinate installation at the VANWIHS Omaha VA Ambulatory Care Clinic with the COR or his representative IOT&A POC.
2.7.2 On-site assembly and installation of items, and performance of services identified in this document shall take place during normal business hours which are defined as: 8:00 AM to 4:30 PM Central Standard Time, Monday through Friday, and excluding Federal Holidays. Proposed installation date will occur after delivery of equipment. Current timeframe of delivery is 1 April 2020 – 30 May 2020.
2.7.3 The vendor is required to define the need for a staging area to accommodate item assembly and installation, if needed. Post-award, this information will be confirmed with the vendor.
2.7.4 The vendor is required to provide tools, labor and materials to complete assembly and installation of the items detailed in this document.
2.7.5 The vendor shall protect all finished spaces and surfaces as required from delivery and installation damage. The vendor shall use covering and protection to the extent necessary to prevent damage to finished spaces. Any damage occurred during delivery and installation is the responsibility of the vendor. The vendor will be responsible for paying for and repairing any damage or noted deficiencies to finished spaces and surfaces that occur as a result of the vendor’s (or associated sub-contractors) installation.
2.7.6 During the entire duration of assembly and installation, the vendor will have a competent representative on-site as the vendor’s contact, and to serve as the interface between the vendor and VANWIHS. On-site representative can be the installer. All instructions provided from COR to the representative will be binding as if given to the vendor’s main contact. The COR may provide specific instructions, however, only the Contracting Officer may change the terms or conditions of the contract.
2.8 Training Required (Yes ☒ No ☐)
2.8.1 The vendor is required to provide live clinical training on set up, operation, shut down, and troubleshooting procedures to all Operating Room nursing staff.
2.8.2 The vendor shall coordinate training dates with the on-site Project Manager. Post-award, these dates will be confirmed with the vendor to meet the clinical schedule. Proposed training date will occur after installation of equipment. Current timeframe of training is July 2020.
2.8.3 The vendor shall provide documentation and/or certification to each staff member that has successfully completed the training.
Scope of Work
VA Nebraska-Western Iowa Healthcare System M8535
4101 Woolworth Avenue Omaha, NE 68105
Total Quantity: 1 ☐ Finish Sample Required if Checked ☒ Manufacturer Certified Installation Required if Checked ☐ BMET Training Required if Checked ☒ End User Training Required if Checked
JSN Nomenclature: M8535 - Microscope, Operating, Portable (Basis of Design is Carl Zeiss Meditec, Inc OPMI Lumera 700 System) Salient Characteristics:
· Motorized zoom system with apochromatic lens
· Zoom ratio 1:6
· Magnification factor = 0.4–2.4
· Focusing: electric/motorized, Focus range: 70 mm
· Objective lens: f = 200 mm (optionally also f = 175 mm or f = 225 mm with support ring)
· Binocular tube: Invertertube E (options: Invertertube, 180° tiltable tube, f = 170 mm, inclined tube, f = 170 mm)
· Widefield eyepieces 10x (option: 12.5x)
· SCI: red reflex illumination and field illumination, both are adjustable
· Halogen fiber optic illumination
· Wavelength 610 nm
· Slit width: 0.2 mm, 2 mm, 3 mm, 4 mm
· Slit height: 12 mm
· Travel range: max. 61 mm x 61 mm
· Automatic centering at push of button
· 17” LCD display, Resolution:1280 x 1024 or equivalent
· Maximum permissible load on the articulated arm: with mounted surgical microscope (without tube, eyepieces, objective lens) and mounted XY coupling, the additional load on the arm resulting from accessories must not exceed 9 kg
· Complies with the safety requirements of EN 60601-1, 3rd issue image2.png