36A77620Q0214-002.docx
DOCX document Posted
- Solicitation number
- 36A77620Q0214
About this file
36A77620Q0214 NIO attachment - Salient Characteristics Kent Sci SomnoSuite Anesthsia Sys.docx
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36A77620Q0214-001.docx | DOCX document | |
| 36A77620Q0214-004.docx | DOCX document | |
| 36A77620Q0214-000.docx | DOCX document | |
| 36A77620Q0214-005.pdf | ||
| 36A77620Q0214-003.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
Kent Scientific SomnoSuite Low Flow Anesthesia System (Brand Name)
General The Kent Scientific SomnoSuite is a low-flow anesthesia system designed specifically for mice and rats. It is engineered with a precision syringe pump and integrated digital vaporizer which uses either room air or compressed gas to deliver anesthesia drugs at low flow rates that is proportionate to the animal’s size. This equipment’s emphasis is on administering just the right amount of anesthesia because 1) it reduces the risk of exposure to lab personnel from wasted (escaped) anesthesia gases and 2) it benefits the animals from over exposure to the anesthesia during the procedures.
Similar to human anesthesia systems, animal anesthesia also requires all types of vital signs monitoring once the drugs are administered. This system offers a variety of software modules for the different types of vital signs monitoring as described below.
Warranty Requirements Manufacturer must offer a (1) one-year minimum warranty that product is free from defects in materials and workmanship. All defects will be replaced or repaired by the manufacturer at no cost to the buyer.
Features
a. Integrated Digital Vaporizer automatically calculates flow rate based on the animal’s weight.
b. Uses less than 1 mL/hour of anesthetic with flow rates from 25mL to 1L
c. Built-in air compressor
d. Precise digital dial for fine control of gas delivery
e. Oxygen flush function minimizes waste anesthetic gas exposure
f. Compatible with isoflurane and sevoflurane – no need for retrofitting
g. Built-in display for real-time observation of anesthetic parameters, such as flow rate and percent anesthetic concentration.
h. Internal memory for data storage
i. Mini-USB for data output
j. Designed with anti-spill, anti-vapor bottle top adapter
k. Includes (2) accessory adapters to connect (2) mice to anesthesia masks simultaneously.
l. Includes scavenging filter
m. Includes 1 charcoal canister
n. Modular system with optional temperature monitoring and homeothermic warming, vital signs monitoring, automatic ventilation, and end-tidal CO2 monitoring
Specifications Low-Flow Anesthesia System
| Flow Rate: | 0 to 1L/min. | |||
| 4 Modes of Operation: | Vaporizer: Internal Pump or Compressed Air | |||
| Ventilator: Internal Pump or Compressed Air | ||||
| Max. Pusher Force: | Greater than 36kg | |||
| Carriage Position Detection: | +/- 0.2mm | |||
| Anesthetics: | Isoflurane and Sevoflurane | |||
| Multiple-Animal Capability: | Yes |
Temperature Monitoring & Homeothermic Control Module
| Resistance: | 2252 ohm | |||
| Tolerance: | +/- 0.2°C | |||
| Accuracy: | 0 - 70°C | |||
| Operating Temp. | -60°C to 150°C | |||
| Sensor Diameter (mouse) | tip/ball diameter: 0.077” | |||
| Shaft diameter: 0.062” | ||||
| Time Constant: | 15.0 seconds | |||
| Dissipation Constant: | 0.75m/W (° C) |
Automatic Ventilator Module
· Fully automatic with preprogrammed settings based on animal weight
· Includes both Volume Control & Pressure Control, need for separate ventilators
| Control Modes: | Volume Controlled, Pressure Controlled | |
| I:E Ratio: | 1:1 to 1:5 | |
| Sigh Pressure: | 2 to 30 cmH20 | |
| Respirator Rate Range: | 20 to 350 bpm | |
| Max. Inspiratory Flow: | 1.5 SLPM | |
| Weight Range: | 3g to 1,250g | |
| Multi-Animal Capability: | Yes |
Pulse Oximeter & Heart Rate Monitor Module
| Heart Rate: | Up to 900 bpm | |
| Resolution: | 1% | |
| SpO2 Range: | 70 to 100% | |
| Alarms: | 5, Programmable | |
| Data Export: | Yes | |
| SpO2 Sensors: | Soft Touch Paw Sensors, Large Animal Clip Sensors, Pup Sensors |
End-Tidal CO2 Monitor Module
| Transducer Type | Sidestream Sensor | ||
| Sample Rate | 50 mL/min. +/- 10 mL/min. | ||
| Initialization Time: | < 15 seconds | ||
| Measurement Range: | 0 to 165, 0 to 120 mmHg | ||
| Rise Time: | < 60 ms | ||
| Resolution: | 0.1 mmHg 0 to 69 mmHg | ||
| 0.25 mmHg 70 to 150 mmHg | |||
| Accuracy: | 0 to 150 mmHg, +/-2 mmHg |
File details come from the government source that posted it. Updated .