Salient Characteristics

Multiphoton Confocal System, A1R HD-Si-MP

1. General Multiphoton confocal laser microscope imaging system. A high-resolution galvanometer scanner and an ultra-high speed resonant scanner.

2. Warranty Requirements

a. One year warranty for parts and labor, including MP Lasers, 5 year mechanical “brass and glass”, and one-year electrical warranty. Additional years can be purchased separately after initial warranty expires. Extended warranty, beyond the first 12 months, for lasers can be purchased separately from the laser manufacturer.

3. Technical Specification and Code Requirements

a. Power Requirements: Minimum three (3) 115V 20A Outlets [NEMA5-20R type] on separate circuits. To include four (4) receptacles.

b. Input Voltage: +/- 5%; frequency 60Hz +/- 5%; harmonic distortion: <3%. Power free of surges and harmonics. NEMA-20R standard.

c. Include in-line conditioners.

d. Include battery backup to protect the system in the event of a power failure, to prevent data loss with long-term experiments.

4. Essential Features / Must Have…

a. 1300 nm dual beam multiphoton confocal microscope configured on the unique large format microscope platform for flexibility.

b. Stage that is appropriately sized to accommodate live animals, from mice to rabbits.

c. Minimum 25mm field of vision, to accommodate larger images per frame.

d. Scanhead and Controller

e. Dual Incident Optical Units for Dual Beam IR laser with automatic laser alignment capability

f. Coherent Dual Beam 1300nm IR Laser

g. High Definition Resonant Scanner capable of 1024x1024 resolution or up to 720FPS

h. Non-Descanned Detector Unit with 3 GaAsP and 1 High Sensitivity PMTs

i. Channel Detector with 2 GaAsP and 2 High Sensitivity PMTs

j. HD Spectral Detector with 2 GaAsP PMTs

k. Alignment free laser units with 405/445/488/514/561/640 laser lines

l. Large Format Open Frame Microscope Platform with Closed Loop Encoded XYZ Stage (see item 5.b on this sheet)

m. Nosepiece Piezo Z for Ultra fast triggered Z stack imaging

n. Plan Apo LWD 10x/ 0.5 NA Glycerin immersion objective with correction collar and 5.5 mm working distance

o. Apo 25x/ 1.10 NA Water immersion objective with correction collar, 1.43-2.04 mm working distance, and improved IR transmission to 1300 nm

p. Apo 20x/ 1.00 NA Glycerin immersion objective with correction collar and 8.2mm working distance

q. Lumencor Sola Epi Light Source with DAPI, FITC and TRITC cubes

r. Workstation and monitor with Confocal software package, providing 2D/3D Deconvolution and High Content Capabilities

s. Vibration Isolation Table (5’ x 6’) with shelving.