HVAC Services for CT Scanner Room Modification
Solicitation number
246-2012-Q-0034
Issued by

About this file

CT-Mechanical

Text of this file

FW: Pawnee IHC - X-Ray Modification for CT on Mech & HVAC SOW

From
Gonzalez, Hugo (IHS/OKC)
To
Wallace, Joseph W. (IHS/OKC)
Cc
Perkins, Judy A (IHS/OKC)
Recipients
Joseph.Wallace@ihs.gov; Judy.Perkins@ihs.gov

Joe

Re-sending to your IHS e-mail because I accidently sent to your personal e-mail.

Hugo Gonzalez, M.S., P.E.

CDR, USPHS - Engineer

Division of Facilities Management

Office of Environmental Health & Engineering

Oklahoma City Area Indian Health Service

701 Market Drive Oklahoma City, OK 73114

Work: (405) 951-3857 Fax: (405) 951-3932 hugo.gonzalez@ihs.gov

From: Gonzalez, Hugo (IHS/OKC) Sent: Friday, August 17, 2012 12:52 PM To: Perkins, Judy A (IHS/OKC); Joe Wallace Wallace (joew@sbcglobal.net) Cc: Standingwater, Robert (IHS/OKC); Santiago, Michael (IHS/OKC); Wilson, Steven J (IHS/OKC); Gonzalez, Bobbie (IHS/OKC) Subject: Pawnee IHC - X-Ray Modification for CT on Mech & HVAC SOW

Judy and Joe, Attached is the SOW for the Pawnee IHC - X-Ray Modification (CT) for the Mechanical & HVAC SOW.

The approved UFMS REQ is # IHS1178528 for $50K.

Thanks for your support, Hugo Gonzalez, M.S., P.E.

CDR, USPHS - Engineer

Division of Facilities Management

Office of Environmental Health & Engineering

Oklahoma City Area Indian Health Service

701 Market Drive Oklahoma City, OK 73114

Work: (405) 951-3857 Fax: (405) 951-3932 hugo.gonzalez@ihs.gov

Pawnee X-Ray Modification (CT) Mech & HVAC SOW.doc Pawnee X-Ray Modification (CT) Mech & HVAC SOW.doc

Oklahoma City Area

Indian Health Service

OEHE - Division of Facilities Management

701 Market Drive

Oklahoma City, Oklahoma 73114

Phone: (405) 951-3857 Fax: (405) 951-3932

SCOPE OF WORK

Indian Health Service (IHS)

Pawnee Indian Health Center

Project # OK1PAS01C6

X-Ray Area Modification – Mechanical & HVAC

Pawnee, Oklahoma

I.

GENERAL:

A. The contractor shall provide mechanical HVAC services for modification of the existing mechanical HVAC systems inside the X-Ray and Lab department at the Pawnee Indian Health Center located in Pawnee, Oklahoma. An existing room G407 will be modified and converted in to two enclosed procedure rooms. Due to this, the existing HVAC system will have to be modified to account for the increased Cooling Load requirements with the installation of a CT Scanner system. The existing HVAC duct work layout and geothermal heat pump tonnage unit size will have to be modification due to these room layout design changes to account for the cooling load increases and end usage changes.

B. The contractor shall provide mechanical plumbing services for the installation of a fully operational sink unit (2ft x 2ft) inside a new G407A CT Scanner room.

C. Highly recommended that the contractor perform an initial on-site visit. The contractor is solely responsible for reviewing verifying the scope of work.

D. Work shall be done while the health center is occupied and operational. Work may be required to be performed during and after normal working hours. Normal working hours are Monday-Friday from 8:00 a.m. to 5:00 p.m. Coordination and scheduling of work is critical in this health center environment and should be coordinated with the IHS COTR and Facility Manager.

E. The contractor is fully responsible for verifying and validating all information contained within this scope of work. Contractor must fully evaluate the condition of the existing conditions prior to the start of any work.

F. Contractor shall abide by all Safety, Security and Infection Control Policies and Procedures of the Pawnee Indian Health Center. Before any work is to be performed a Start of Construction letter notice shall be provided to the assigned IHS COTR at least 7 days in advance for review and approval.

G. When egress corridors, fire alarm detection and suppression systems are impaired during construction, the Contractor shall inform the Facility Manager to ensure that an Interim Life Safety Measure is conducted in coordination with the IHS COTR, Pawnee Indian Health Center Safety Office and Facility Manage & Engineer.

H. The Pawnee Indian Health Center is a tobacco free campus. No smoking is permitted in the health center and grounds. The contractor is responsible for their subcontractors to obey and be in compliance of the IHS tobacco free campus.

I. Signing in/out and scheduling of contractors and sub-contractors must be coordinated with the Pawnee Indian Health Center Facilities Management Department in order to maintain traffic-flow, parking, access to building and locked areas.

J. All applicable documents shall be distributed as follows:

Project Officer / IHS COTR

Hugo Gonzalez, P.E.

Senior Staff Engineer, Mechanical Engineer

IHS - Oklahoma City Area Office

Phone: (405) 951-3857 Fax: (405) 951-3932

Hugo.Gonzalez@ihs.gov

OCAO Contracting

Joe Wallace, Contract Specialist

IHS - Oklahoma City Area Office

Phone: 405-951-3929

Joseph.Wallace@ihs.gov

Facility Manager:

Joe Jones, Facility Manager

Pawnee Indian Health Center

1201 Heritage Circle

Pawnee, OK 74058

Phone:

(918) 762-6711

Joseph.Jones@ihs.gov

K. All work must conform to, and comply with, the latest edition/version of ALL applicable OSHA, NFPA, BIC, BOCA, AMSE, SMACNA, NEC and any other required federal, state, and local laws, rules, regulations, codes, and standards.

L. Contractor shall use Energy efficient equipment and material when replacing defective government equipment and materials.

M. Upon completion of work, Contractor shall provide training on the cleaning and maintenance of new equipment that is installed.

N. All new materials and equipment shall be furnished with an owner’s manual, parts list, local authorized dealer contact information, and warranty ending date.

O. All new materials and equipment shall meet specifications listed below and shall have approved certificates of installation if required by ALL applicable federal, state, and local codes, standards, or regulations.

P. All construction produced waste and shipping containers shall be disposed of at an offsite suitable landfill or be recycled per federal regulations..

Q. The photos and drawings provided are for reference only. An investigation and verification of all the existing construction areas are to be verified and validated by the contractor.

R. Contractor shall conduct periodic project meeting with the Contract Specialist, Facility Engineer, and IHS COTR. Contractor shall provide approved engineering designs and construction documents and specifications of all material used during construction for approval or disapproval..

S. Contractor shall provide manufacture warranties on the manufacturer’s standard form as required.

T. Contractor shall provide written quality assurance reports to the IHS COTR for review and approval by the COTR. The reports shall address all work in progress during construction.

U. Contractor shall provide any applicable technical care training to the Pawnee Indian Health Center facility technician staff and engineering for maintenance and preventive maintenance.

WORK PLAN

SERVICES TO BE PERFORMED BY THE CONTRACTOR

Before any services are to start within a healthcare facility, all Safety and Infection Control Measures are to be put in place and approved by IHS COTR. Implementing all Safety and Infection Control Measures is the sole responsibility of the Contractor to ensure these measures meet OSHA & NFPA Codes and ASHRE & Joint Commission Standards.

General Mechanical HVAC Services Requirements:

· This work plan is for Mechanical HVAC services consisting of making modification to the existing HVAC system servicing both the X-Ray and Lab Department which are side by side connected departments. For additional detail reference Appendix A.

· The mechanical contractor in order for IHS to ensure the contractor is following good mechanical trade practices and to ensure quality workmanship is to follow all current ASHRE, SMACNA, ANSI, ASME Standards and Guidelines.

· Original 2004 Mechanical drawings are in metric units. These are being provided as references to the Mechanical contractors with easy to find color highlighted key areas for their full review and reference. Sole responsibility of the Mechanical contractor to verify and validate existing conditions against the 2004 mechanical drawings.

· The original 2004 HVAC zoning was design by using the following Carrier Geothermal Heat Pump unit for cooling and heating located in a large open attic area on a mezzanine

Carrier AQUAZONE™ 5-Ton Air to Water Source Heat Pump (WSHP)

Model 50RVR060LPT60120

· The X-Ray department is being served by a 5-Ton WSHP #M-11 located on the southwest attic mezzanine

· The Lab department is being served by a 5-Ton WSHP #M-22 located on the southeast attic mezzanine

· There is easy access and open area to work on the 5-Ton Water Source Heat Pump and easy access for making modifications to the duct work from above the ceiling tile and in the attic.

· Due to modification being made to the HVAC system, NEBB accredited Test & Balance will be required based on IHS provided CFM and Cooling Load requirements.

HVAC Modification Requirements:

Existing HVAC modifications are to be made as followed:

5-Ton WSHP #M-11 Modify to be use Only for the New G407A CT Scanner Room

· The 5-Ton WSHP #M-11 unit is to stay as-is and is to serve only the new G407A CT Scanner Room (20ft x 15.5ft) with the following Toshiba Aquilion 64 Slice CT system specification requirements.

SHAPE \* MERGEFORMAT

· The existing #M11 HVAC main supply and return duct truck is to be separated and modified to serve only the new G407A CT Scanner Room.

· For the following below existing room on #M-11HVAC system is to be separated and modified as indicated. An on-site field review and inspection above the ceiling tiles and attic will be required of the mechanical contractor to determine the most sensible and economical areas for detachment points of the duct branch in keeping as many branch duct lines together. Before any detachment IHS will require the mechanical contractor to provide a submittal denoting the detachment area for review and approval by the IHS COTR. The existing supply and return vents are to be kept in the same location.

G407A New CT Scanner Room (20ft x 15.5ft))

* Add four (4) new duct branch lines with four (4) 2ft x 2ft supply diffusers matching existing inside room for proper ventilation as provided by IHS COTR.

* Relocate one (1) existing supply diffusers inside room for proper ventilation as provided by IHS COTR.

* Relocate one (1) return vent existing by the corridor 4ft door and add one (2) new return vent. The total will be two (2) return vents being required in the new G407A CT Scanner Room.

· For the remaining #M-11HVAC system duct branch lines and supply diffuser & return vents that is outside the new G407A CT Scanner Room are to be modified and tied in to the main duct trunk of Lab department #M-22 HVAC system the new G407A CT Scanner Room.

5-Ton WSHP #M-22 to be Increase to 10-Ton WSHP

· The existing 5-Ton WSHP #M-22 HVAC system will be modified to server both the Lab and remaining X-Ray department that does not include the new G407A CT Scanner Room.

· Replace the existing 5-Ton WSHP with a Carrier AQUAZONE™ 10-Ton Air to Water Source Heat Pump (WSHP).

· The existing #M-22 HVAC system main supply and return duct truck is to be modified to tie into the separated #M-11HVAC system duct branch lines and supply diffuser & return vents that are outside the new G407A CT Scanner Room.

· For the following below existing rooms on #M-11HVAC system are to be separated and tied into #M-22 HVAC system main supply and return duct trunk lines. An on-site field review and inspection above the ceiling tiles and attic will be required of the mechanical contractor to determine the most sensible and economical areas for detachment points of the duct branch in keeping as many branch duct lines together. Before any detachment will require, the mechanical contractor is to provide as submittal denoting the detachment area for review and approval by the IHS COTR. The existing supply and return vents are to be kept in the same location.

G401 Waiting Area for Lab & X-Ray

G403 Hallway for X-Ray

G406 Procedure Room for Digital X-Ray

G408 Dressing Area changed to CT Control Room

G409 Procedure Room for Mammogram

G411 Storage Room changed to Tech Office Room

G412 Reception Open Area

G407B New Bone Densitometer & Ultrasound (BD&U) Room (14ft x 15ft)

* Relocate the two (2) existing supply diffusers inside room for proper ventilation as provided by IHS COTR.

* Add one (1) return vent by entrance door and tie into #M-22 main return trunk matching existing.

General Mechanical Plumbing Services Requirements:

For additional detail reference Appendix A.

· Mechanical plumbing services will consist of providing a fully operational pre-built combo sink & above storage cabinet unit (2ft x 2ft) to be installed inside the new G407A CT Scanner room.

· The style, color, sink type, and faucet type for the pre-built combo sink & above storage cabinet is to be selected by IHS. The mechanical contractor is to provide a submittal based on a pre-built combo sink & above storage cabinet for review and approval by the IHS Pawnee and IHS COTR.

· Original 2004 Mechanical drawings are in metric units. Being provided to the Mechanical contractors are mechanical plumbing drawings with easy to find key yellow highlighted areas for full review and reference. Sole responsibility of the Mechanical contractor to verify and validate existing conditions against the mechanical plumbing drawings.

· The mechanical contractor is to provide full plumbing connection and hook-up of the new sink hot and cold and hot plumbing and tie-in the existing near-by main water lines to match existing plumbing installation.

· Under the sink, install localized shut-off valves for the domestic cold and hot water lines. For safety, install hot water temperature control mixing valve to eliminate scolding of user.

· The mechanical contractor is to provide full plumbing connection and hook-up of the new sink drain and tie-into the near-by sewer system to match existing plumbing installation.

Example of pre-built combo sink & above storage cabinet is to be selected by IHS.

Appendix A

Pawnee Indian Health Center X-Ray & Lab Department HVAC Layout

Pic A1: Referencing 2004 M2.02 Mechanical Drawing of Existing HVAC Duct Layout for X-Ray & Lab Dept. using Heat Pump # M-11 & M-22

Pic A2: Referencing 2004 M2.03 Mechanical Drawing of Existing HVAC Heat Pump # M-11 & M-22 Attic Location

Pic A3: Reference Drawing for Modification of Existing HVAC Heat Pump # M-11 & M-22 for Change of Room G407A CT to only use 5-Ton WSHP #M-11 Only and move Room G407B and Remaining X-Ray Duct Work to #M-22 and increase to 10-Ton WSHP Unit

Pic A4: Reference 3D Drawing for Modification of Existing HVAC Heat Pump #M-11 & #M-22 for Change of Room G407A CT to only use 5-Ton WSHP #M-11 Only and move Room G407B and Remaining X-Ray Duct Work to #M-22 and increase to 10-Ton WSHP Unit

Pic A5: Room G407 East Side Supply Diffusers - Reference 2004 M2.02 HVAC Drawing

Pic A6: Room G407 East Side Supply Diffusers – Reference 2004 M2.02 HVAC Drawing

Pic A7: Room G407 South Side Supply Diffusers & Exhuast - Reference 2004 M2.02 HVAC Drawings

Pic A8: Room G407 Nort Side Supply Diffusers & Exhuast - Reference 2004 M2.02 HVAC Drawings

Pic A9: HVAC Duct above Room G407 - - Reference 2004 M2.02 HVAC Drawings

Appendix B

Pawnee Indian Health Center X-Ray & Lab Department Plumbing Layout

Pic B1: Referencing 2004 M1.01 Mechanical Drawing of Existing Plumbing Location on the Lower Floor under the X-Ray Department

Pic B2: Referencing 2004 M1.02 Mechanical Drawing of Existing Plumbing on the X-Ray Department

Pic B:3 Lower Floor under the X-Ray Department Existing Plimbing

Reference 2004 M1.01 Mechanical Drawing

Pic B:3 Plumbing in Lower Floor under the X-Ray Department Existing Plimbing

Reference 2004 M1.01 Mechanical Drawing

SOW CT - 2004 New Pawnee IHC Mechanical Combinedi.pdf SOW CT - 2004 New Pawnee IHC Mechanical Combinedi.pdf

PATIENT/VISITOR PARKING

STAFF PARKING (108)

FUTURE EXPANSION

GOVERNMENT VEHICLES

PARKING

BUS

C

LI

KS

F

IE

LD

R

O

AD

PROPOSED ELDERLY HOUSING DEVELOPMENT

EXISTING

WATER TANK

LOOP ROAD

PATI0

SERVICE DRIVE

EMPLOYEE &

NEW ENTRY DRIVE

NEW BUILDING

SHEET:

DRAWN:

CHECKED:

DATE:

FILE:

PROJECT:

PHASE:

REVISION: DATE:

M0.01

SITE PLAN

MECHANICAL PLAN

medi-01j1 wgm bak

MechCD-Site

N

ORIGINAL SCALE: 1 = 500

PAWNEE HEALTH CENTER, SITE PLAN

5 m 15 m

CW DOMESTIC COLD WATER

HW DOMESTIC HOT WATER

HWR DOMESTIC HOT WATER RETURN

V WASTE VENT

W WASTE

VTR WASTE VENT THROUGH ROOF

G NATURAL GAS

D CONDENSATE DRAIN

RD ROOF DRAIN

SD STORM DRAIN

FS FIRE SPRINKLER

PLUMBING ABBREVIATIONS

SW SANITARY WASTE

FIRE SPRINKLER SYSTEM

TO BE FULLY SPRINKLED. CONTRACTOR TO VERIFY

CONTRACTOR TO VERIFY SERVICE CAPACITY AND UPGRADE

AS REQUIRED. SPECIFICATION IS A PERFORMANCE SPEC.

CO WASTE LINE CLEAN OUT

SIZE OF SERVICE SPRINKLER LINE TO SERVE AREA.

ALL AREAS WITHIN LIMITS OF BUILDING CONSTRUCTION

GEOTHERMAL

REVERSE RETURN HEADER SYSTEMSWEEPING

ELBOWS

91,440 MINIMUMTYPICAL

WELL

FR

M P

U

"G

2"

3"

ONLY LOOP #1 SHOWN

GRADE OR SLAB

PVC SLEEVE (TYP)

1"

13 10 7 151434 1 1 2

DITTO - LOOPS

2,5,8,11,14

DITTO - LOOPS

3,6,9,12,15

1-1WELL NO. 1-2 1-3

(TYPICAL)

ISOLATION VALVE

TRANSITION

PE PIPE TO

STD HVAC PIPE

AT MECH SPACE

INSULATE ALL PIPE FROM

AND INCLUDING MANIFOLD.

INSIDE BLDG. SEE SPECS

1524 MINIMUM

THROUGH CRUSHED ROCK FOUNDATION STABILIZER

GEOTHERMAL FIELD FLOW DIAGRAM

PAWNEE HEALTH CENTER NOM 200 TON GEOTHERMAL SYSTEM

SUPPLY

RETURN

SCALE: NONE

GEOTHERMAL WELL

2134 RADIUS

MINIMUM (TYP)

1500 MIN.

ORANGE WARNING RIBBON

YELLOW WARNING RIBBON

SUPPLY HEADER LINE

RETURN HEADER LINE

450 MIN

GEOTHERMAL WELL,

FINAL GRADE

HEADER &

LOOP LINES

WITH IDENTIFICATION

OF WELL STAMPED IN TOP.

NO SPLICE

RADIUS BEND ONLY

OR ELBOWS.

SCALE: NONE

GEOTHERMAL PIPE DETAIL

#10 BARE COPPER TRACER WIRE

ALL WELL PIPING 25mm

GEOTHERMAL 10 WELL PIPE HEADER DETAIL

SEE SITE PLAN

FOR SPACING

91,440mm DEEP

LOCATE EACH SIDE OF WELL

(SHOWN FOR CLARITY)

32 40

BASED ON 4.5 GPM FLOW PER WELL @ 300 FT DEPTHD

B

WELLS 6850mm ON

CENTER, TYPICAL ALL

75 75 75 75 75 75 75 75 75 75

FLUSH CONN

FLUSH CONN

100 200

TO

S

YS

TE

75 LINE

75 LINE

GEOTHERMAL PIPES

15 - 75mm SUPPLY

TEMP SENSOR

THERMOMETER

10 WELL HEADER

RE D/M0.01 (TYP)

GEOTHERMAL PIPES

15 - 75mm RETURN

Record Drawing

April 15, 2004

FIREMAN'S CONNECTION

POST (COORDINATE WITH CIVIL PLANS)

PA

W

EE

, O

KL

AH

A

H

EA

LT

EN

TI

E

G

IN

D

ES

IG

, I nc

8-

3-

So ut h ar va rd

, S ui te

, T ul sa kl ah om a

7-

En gi ne er in g es ig n, In c.

50 50

40 32

(BASED ON 17LPM FLOW PER WELL @ 91,440mm DEPTH)

MANIFOLD

IDENTIFICATION

NUMBER (TYP)

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

IDENTIFICATION

NUMBER (TYP)

DO NOT

INSULATE

TYPICAL

1-10 15-10

15-1

GEOTHERMAL WELL LOCATION

DIMENSIONS

FROM TO DIMENSION

A 1-1 59.88 M B 1-1 97.62 M A 1-10 92.07 M B 1-10 117.07 M A 15-1 40.73 M B 15-1 7.76 M A 15-10 83.53 M B 15-10 68.42 M

WHICH MAY BE INCORPORATED HEREIN AS A RESULT.

RESPONSIBLE FOR ANY ERRORS OR OMISSIONS

ACCURACY OF THIS INFORMATION AND SHALL NOT BE

THE ARCHITECT/ENGINEER HAS NOT VERIFIED THE

BASED ON INFORMATION PROVIDED BY OTHERS.

THESE RECORD DRAWINGS HAVE BEEN PREPARED

RECORD DRAWING

LOWER LEVEL

ORIGINAL SCALE: 1 = 100

0 1 m 3 m

SH2

FD

SP

SH1

DF

WC1

UR

WC2

MB

LV2

SK3

SH1 SH2

LV1

SK1

SK6WB

FS

SK4

SP

150 SS BELOW GRADE,

RE: CIVIL DWGS FOR CONTINUATION

250 SD BELOW GRADE,

100 DOM. WATER BELOW

GRADE, RE: CIVIL DWGS

FOR CONTINUATION

150 FIRE SPRINKLER LINE BELOW

GRADE W/ PRESSURE REDUCER

AND BACK-FLOW PREVENTOR,

RE: CIVIL DWGS FOR CONTINUATION31347SF

100 BELOW MAIN SLAB

WT, RE: B/M1.01

FOR DETAIL

150 BELOW GRADE TO

FIRE DEP'T POST

RE: CIVIL DRAWINGS

SH1

SK6

50 HW UP

CO

50 SD

50 SD DN

DN

CONTINUE TO CURB

150 W DN

CO CO

RESERVED FOR GEO PIPING

50 SD 50 SD 50 SD

250 SD BELOW LOWER SLAB

SD

75 TYP

ALL FD'S

150 W

75V TURN UP

75 MAIN VENT

IN LOWER

CLG SPACE

VENT UP

25 HWR DN

M1.01

PLUMBING PLAN

PAWNEE HEALTH CENTER, LOWER LEVEL

100 SD

ES/EW

P09

MARK DESCRIPTION

E-77-AA VINYL BUMPERGUARD, 1453-BB STRAINER. 75 W, 50 V, 15 CW, 15 HW.

832-AA HOSE AND HOSE BRACKET, MSG2424 WALL GUARD,

MOP SINK BASIN, MOLDED-STONE, 24"x24"x10". FIAT #MSB-2424. 830-AA FAUCET, GENERAL NOTES:

1. FURNISH AND INSTALL HYDRO-PNEUMATIC WATER HAMMER ARRESTORS, SIZE AS

REQUIRED. CONFORM TO A.S.S.E. STD. 101. INSTALL ON BOTH COLD AND HOT

WATER PIPING SYSTEMS IN ACCORDANCE WITH MINIMUM LOCAL AND STATE CODE

REGULATIONS AND REQUIREMENTS. ZURN #Z-1700 APPROVED. ADDITIONALLY,

INSTALL AIR CHAMBER ARRESTORS AT ALL NEW FIXTURES FOR BOTH HOT AND

COLD WATER PIPING.

2. EQUAL PRODUCTS MANUFACTURED BY ZURN, JOSAM, SYMMONS, CRANE, KOHLER, WADE,

DELANY, HAWS/SUNROC, MCGUIRE, MCGUIRE PROWRAP AND ELKAY ARE APPROVED.

STANDARD SAN. P-TRAP. 3"W, 1.5"V.

CO REFER TO SPEC. MANUAL.

SK2 SINGLE SINK, 2 HOLE, SELF RIM, 15"x17.5"x7"D. ELKAY #BLR-1517. 1 1/2" 17GA P-TRAP W/C.O., WB WASHING MACHINE BOX, RECESSED WALL MOUNT. SYMMONS MODEL W-602-X.

STANDARD SANITARY P-TRAP, MT TOP AT 1200 A.F.F. 50 W, 40 V, 15 CW, 15 HW.

WC1 WATER CLOSET, WALL MOUNT, FLUSH VALVE, ELONGATED BOWL, WHITE. KOHLER #K-4330, WC2 WATER CLOSET, WALL MOUNT, HANDICAP, FLUSH VALVE, ELONGATED BOWL, WHITE.

SLOAN OPTIMA 8111-MC (1.6 GAL) BATTERY-POWERED SENSOR-OPERATED FLUSH VALVE W/

FLUSH VALVE W/ OVER-RIDE BUTTON, OLSONITE 95 SEAT. MT TOP OF FLUSH VALVE 30"

KOHLER #K-4330. SLOAN OPTIMA 8111-MC (1.6 GAL) BATTERY-POWERED SENSOR-OPERATED

UR URINAL, WALL MOUNT, ADA, FLUSH VALVE, WHITE, CRANE #7-187E. SLOAN OPTIMA

ZR-1221 CARRIER. 50 W, 40 V, 20 CW.

8180-1.0-MC (1.0 GAL) BATTERY-POWERED SENSOR-OPERATED FLUSH VALVE, ZURN

W/ LAM. NOZZLE & DRAIN, MCGUIRE 155WC OFFSET DRAIN, ADA W/ RIM HEIGHT OF 34" A.F.F., WALL MOUNT, ROUNDED LAV., WALL MOUNT, 21"x22", WHITE. AMER. STANDARD MURRO #0954.000. ZURN ZR-1231 CARRIER, 1 1/4" 17GA P-TRAP W/C.O., DELTA 516-WFHDF FAUCET

3/8" FLEX RISERS & L.K. STOPS, INCLUDE OPTIONAL SHROUD. 40 W, 40 V, 15 CW, 15 HW.

TB BATHTUB, 1500 x 900, ADA, WHITE. LASCO 2603-NW LASCOAT GELCOAT FINISH, INTEGRAL

SOAP SHELF & DRAIN. DELTA #1412-LHP ANTI-SCALD PRESS. BALANCE TUB VALVE W/

A.F.F., CENTER OF WATER CLOSET 18" FROM SIDE WALL, MOUNT RIM @ 18" A.F.F., ZURN

100 W, 50 V, 25 CW.

HANGER, 1 1/4" 17GA P-TRAP W/ C.O., FLEX RISER & L.K. STOP, 4.5FLA AT 115V/1. MT ONE

@ STD HT & ONE @ ADA HT. 40 W, 40 V, 15 CW.

FD FLOOR DRAIN, 3" STANDARD DUTY, 6" DIA. POLISHED NICKEL STRAINER. ZURN #ZN-415-6B.

ELEC. WATER COOLER, WALL MOUNT, TWO LEVEL, ADA. SUNROC #ADA-8-BLN. WALL

LK-35 DRAIN, FLEX RISERS & L.K. STOPS. 50 W, 40 V, 15 CW, 15 HW.

OVER-RIDE BUTTON, OLSONITE 95 SEAT. ZURN 1203 OR 1204 CARRIER, MT RIM AT 15" A.F.F.

SK4 DOUBLE SINK, 2 HOLE, SELF RIM, 29"x18". ELKAY #LR-2918. 1 1/2" 17GA P-TRAP W/C.O., DELTA 711-WFHDF SINGLE HANDLE GOOSENECK FAUCET W/ LAM. FLOW NOZZLE, ADA, SK5

CHROME ADA KNOB, RIGHT HAND DRAIN. 50 W, 40 V, 15 CW, 15 HW.

SH1 SHOWER. DELTA 1325-WSHDF CHROME ANTI-SCALD PRESSURE BALANCE VALVE W/ VANDAL

RESISTANT HANDLE & SHOWER HEAD, VACUUM BREAKER. 15 CW, 15 HW.

SH2 SHOWER. DELTA 11T5143 CHROME ANTI-SCALD PRESSURE BALANCE VALVE W/ VANDAL

RESISTANT HANDLE, ADA, ADJUSTABLE SLIDE BRACKET & VACUUM BREAKER. 15 CW, 15 HW.

DOUBLE SINK, 2 HOLE, SELF RIM, 37"x22". ELKAY #LR-3722. 1 1/2" 17GA P-TRAP W/C.O., ELKAY LK-35 DRAIN, FLEX RISERS & L.K. STOP. 50 W, 40 V, 15 CW, 15 HW.

ELKAY LK-35 DRAIN, FLEX RISERS & L.K. STOPS. 50 W, 40 V, 15 CW, 15 HW.

SINGLE DEEP SINK, 2 HOLE, SELF RIM, 22"x22"x12". ELKAY #DLR-222212. 1 1/2" 17GA P-TRAP

W/C.O., DELTA 711-WFHDF SINGLE HANDLE GOOSENECK FAUCET W/ LAM. FLOW NOZZLE,

ADA, ELKAY LK-35 DRAIN, FLEX RISERS & L.K. STOPS. 50 W, 40 V, 15 CW, 15 HW.

DELTA 711-WFHDF SINGLE HANDLE GOOSENECK FAUCET W/ LAM. FLOW NOZZLE, ADA, ELKAY

A.R.E. INTERIOR, STANDARD SAN. P-TRAP. 75 W, 40 V.

FS FLOOR SINK, 12"x12", 8" DEPTH, POLISHED NICKEL GRATE, DOME STRAINER. ZURN #ZN-1901.

SINGLE SINK, 2 HOLE, SELF RIM, 17"x16". ELKAY #LR-1716. 1 1/2" 17GA P-TRAP W/C.O., ELKAY LK-35 DRAIN, FLEX RISERS & L.K. STOPS. 50 W, 40 V, 15 CW, 15 HW.

SK6 LAUNDRY SINK, MOLDED STONE, 23"x22". FIAT #FL-1, A-1 DECK MOUNTED FAUCET, STAINLESS STEEL LEGS & 75 STANDARD TRAP. 75 W, 50 V, 15 CW, 15 HW.

SK2-PT SAME AS 'SK2' ABOVE PLUS PLASTER TRAP.

#1203 OR #1204 CARRIER. 100 W, 50 V, 25 CW.

SAME AS 'WC2' BUT REPLACE AUTOM. FLUSH VALVE W/ MANUAL SLOAN REGAL 811-XL.WC3

SK7 CLINICAL SINK, FLOOR MOUNT, FLUSHING RIM, WHITE. KOHLER K-6676, ZURN Z-6000XL-BW

(3.5 GAL) FLUSH VALVE W/ VAC. BREAKER, K-8935 STAINLESS STEEL RIM GUARDS, MT ON

700x350x250 HIGH TERRAZZO BASE. 100W, 50V, 25CW. INCLUDE FAUCET - 15 HW, 15 CW.

HWRP RETURN HOT WATER CIRCULATING PUMP, 8 GPM @ 22 FT HD. GRUNDFOS MODEL UP26-96BF.

SPECIFICALLY MANUFACTURED FOR OPEN SYSTEM. 1/12 HP @ 120V/1 SERVICE.

CS OVAL CUP SINK, STRAINER, 4.5"x10.5", DURCON CS-12. 40 W, 25 V.

HB HOSE BIBB, EXPOSED, 3/4" FREEZE PROOF, WALL MOUNT. ZURN #Z-1310-3.

ANTI-SIPHON, AUTOMATIC DRAINING, 3/4"CW.

ES/EW EMERGENCY SHOWER/EYEWASH. BRADLEY MODEL S19-310 W/ PLASTIC SHOWER HEAD

AND EYE WASH BOWL, SAFETY COATED COMPONENTS. 40 CW, 15 HW.

MechCD-LowerFlr

100 CW TO MAIN

FLR CLG SPACE.

100 CW AT LOWER

FLOOR STRUCTURE

INSTALL BUILDING SHUTOFF VALVE

WITH BACKFLOW PREVENTOR

1.

GATE VALVE

CHECK VALVE

COLD WATER PIPE

HOT WATER PIPE

VENT PIPE

WASTE PIPE

RD ROOF DRAIN

HOT WATER RETURN PIPE

ALL AREAS WITHIN LIMITS OF CONSTRUCTION

TO BE FULLY SPRINKLED. CONTRACTOR TO VERIFY

SPECIFICATION IS A PERFORMANCE SPEC.

SIZE OF SERVICE SPRINKLER LINE TO SERVE AREA.

2. ALL THERAPEUTIC POOL PIPING, CONTROLS, PUMPS, WATER

HEATER AND CONNECTIONS WILL BE BY POOL CONTRACTOR.

PROVIDE CONNECTIONS AS INDICATED ON THIS PLAN.

CO WASTE LINE CLEAN OUT

INSTALL DOMESTIC WATER PIPING ABOVE CLG & BELOW INSULATION

SO THAT NONE OF THE PIPING IS EXPOSED TO UNINSULATED SPACE.

WH WATER HEATER, FLOOR MOUNT, GLASS LINED, INSULATION. A.O. SMITH #DVE-120. 119 GAL.

HOT WATER STORAGE TANK, VERTICAL, 200 GALLON CAPACITY, GLASS LINED, INSULATED

A.O. SMITH #TJV-200. 50mm INLET & OUTLET. PIPE T&P TO FLOOR DRAIN.

HB

FIRE SPRINKLER RISER

20 CW

T'STAT SET TO 110°F, 50 INLET & OUTLET, PIPE T&P TO FLOOR DRAIN. INCLUDE AMTROL

CAPACITY, 212 GPH @ 70°F RISE, TOTAL INPUT 36KW, 6 - 6KW ELEMENTS, @ 480V/3.

WT

WH

EW EMERGENCY EYE/FACE WASH. BRADLEY MODEL S19-220BF W/ WRAP AROUND SKIRT

(BARRIER FREE), STAINLESS STEEL. 40 W, 40 V, 15 CW.

PROVIDE A SOLENOID VALVE IN FIRE SPRINKLER LINE FEEDING ELEVATOR

SHAFT AND ELEVATOR EQUIPMENT ROOM. INSTALL HEAT DETECTOR AND

SMOKE DETECTOR IN ELEVATOR SHAFT AND ELEVATOR EQUIPMENT ROOM.

IS TO RESPOND TO NORMAL ELEVATOR CAPTURE FEATURE. ONLY UPON

ON GENERAL ALARM OR LOBBY SMOKE DETECTOR ALARM, THE ELEVATOR

IS ELEVATOR TO GO TO DESIGNATED FLOOR, OPEN DOORS, RELEASE

SMOKE OR HEAT DETECTION IN ELEVATOR SHAFT OR EQUIPMENT ROOM

SOLENOID VALVE AND CONNECT POWER VIA SHUNT TRIP BREAKER.

#AST-42 THERMAL EXPANSION TANK.

SP STAINLESS STEEL SUBMERSIBLE SUMP PUMP, GRUNDFOS BOSS 200, 1/3HP @ 115V/1.

SK3-PT SAME AS 'SK3' ABOVE PLUS PLASTER TRAP.

P01 ACCESORY CONSOLE - ADEC #5531, REF. TEMPLATE

# 86.014?.00, TECHNICAL PACKET 86.0142.00.

STUB AND CONNECT PER MANUFACTURE'S INSTRUCTION

THE FOLLOWING UTILITIES:

1/2" COLD WATER, MAN. SHUTOFF, 3/8" COMPR. FITTING

1/2" HOT WATER, MAN. SHUTOFF, 3/8" COMPR. FITTING

1/2" COMP AIR, MAN. SHUTOFF, 3/8" COMPR. FITTING

3/4" C. W/ 2#12, 1#12G, CONN. TO SUPPLIED QUAD RECEPT.

1-1/2" PVC DRAIN W/ AIR ADMITTANCE VALVE FOR VENT.

2" EMPTY PVC TO TREATMENT CONSOLE, SWEEP 90'S.

P02 CENTRAL CONSOLE - ADEC #5543, REF. TEMPLATE

# 86.0145.00, TECHNICAL PACKET 86.0142.00.

1/2" HOT WATER, MAN. SHUTOFF, 3/8" COMPR. FITTING

1/2" COMP AIR, MAN. SHUTOFF, 3/8" COMPR. FITTING

3/4" C. W/ 2#12, 1#12G, CONN. TO SUPPLIED QUAD RECEPT.

1-1/2" PVC DRAIN W/ AIR ADMITTANCE VALVE FOR VENT.

3/4" C. W/ 2#12, 1#12G, DED. CIRC, CONN. TO EXRAY.

P03 CASCADE CHAIR - ADEC #1040, REF. TEMPLATE

# 85.0498.00, TECHNICAL PACKET 86.0142.00.

1/2" COMP AIR, MAN. SHUTOFF, 3/8" COMPR. FITTING

3/4" C. W/ 2#12, 1#12G, CONN. TO SUPPLIED QUAD RECEPT.

P04 TREATMENT CONSOLE - ADEC #5580.42, REF. TEMPLATE

# 86.1046.00, TECHNICAL PACKET 86.0142.00.

1/2" COMP AIR, MAN. SHUTOFF, 3/8" COMPR. FITTING

3/4" C. W/ 2#12, 1#12G, CONN. TO SUPPLIED QUAD RECEPT.

RECEIVE 2" PVC'S FROM 5531, 5543 & 1040, SWEEP 90'S.

1/2" PVC VACUUM TO TRUNK LINE IN TRENCH.

P05 15 CW FOR COFFEE MAKER, MT 1200 A.F.F.

P06 RUN 15 CW & HW DN WALL & BELOW SLAB TO SINK

P.B.O. IN DENTAL EQUIPMENT. INSTALL ISOLATION

VALVE FOR H & CW ABOVE CLG.

P09 TAP COLD WATER LINE SERVING SINK WITH 1/4"

LINE TO SERVE COFFEE MAKER.

3. PROVIDE CONNECTIONS & CONNECT ALL EQUIPMENT PROVIDED

BY OWNERE PER MANUFACTURER'S INSTRUCTIONS.

75 MAIN VTR

100 W

75 V

15 HW30 CW

L.K. STOPS. 40 W, 40 V, 15 CW, 15 HW.

LAVATORY, OVAL, 20"x17", SELF RIM, WHITE. CRANE #1-280-V. 1 1/4" 17GA P-TRAP W/C.O., DELTA 516-WFHDF FAUCET & DRAIN, MCGUIRE 155WC OFFSET DRAIN. FLEX RISERS &

BRANCH OFF 75 CW WITH BACK-FLOW

PREVENTOR FOR LANDSCAPE IRRIGATION.

NO SCALE PLUM

BRONZE SWING TYPE

CHECK VALVE INSTALLED

HORIZONTALLY

32 COLD WATER INLET

LINE OF FINISHED CEILING

EMERGENCY SHOWER HEAD

SHOWER ACTIVATION VALVE

HAND PULL FOR EMERGENCY

FINISHED FLOOR

HAND OR FOOT OPERATED

EYEWASH ACTIVATION VALVE

EMERGENCY EYEWASH

EMERGENCY EYEWASH DRAIN

PEDAL ACTIVATION CHAIN

PEDAL FOR EMERGENCY

FLOOR DRAIN

FLOOR DRAIN TRAP ARM

TO BRANCH WASTE LINE

15 HOT WATER INLET

32 TEMPERED WATER SUPPLY

PIPE TO FIXTURE

FINISHED WALL

TEMPERED WATER SUPPLY

TO EMERGENCY SHOWER

TEMPERED WATER SUPPLY

TO EMERGENCY EYEWASH

FIXTURE SUPPORT PIPE

32 WASTE FROM EYEWASH

TOWARD FLOOR DRAIN

FIXTURE BASE SUPPORT FLANGE

EXTENDED BEYOND SUPPORT

FLANGE AND TURNED DOWN

1" AIR GAP

COMBINATION UNIT WITH TEMPERED WATER SUPPLY EMERGSH1

SH3 SHOWER MODULE, HANDICAP, CRANE MODEL ADA-3636S, ONE PIECE ACRYLIC, WHITE.

COMPLETE W/ S-5 FOLD-UP SEAT W/ PHENOLIC SLATS, INTEGRAL SOAP & SHAMPOO

SHELVES, STN. STL. GRAB BAR, INTEGRAL BACK & HEAD SUPPORT, ATP-1 DOME LIGHT,

ASSEMBLY TO BE DELTA 1305-WSHDF CHROME PRES. BALANCE VALVE W/ VANDAL

H-30 2" CHROME PLATED DRAIN W/ STRAINER GRATE, STN. STL. CURTAIN ROD. SHWR

BAR, IN-LINE VACUUM BREAKER. 50 W, 40 V, 15 CW, 15 HW.

RESISTANT HANDLE & W/ RPW1111HDF HAND SHWR W/ ADJ. SLIDE BRACKET, WALL

FINISHED FLOOR

FINISHED GRADE

V

IF

Y

NONE PLUM

SIZE OF ALL LINES.

FIRE PROTECTION SUB CONTRACTOR RESPONSIBLE FOR ALL

FIRE PROTECTION PIPING INSIDE BUILDING UP TO 5'-0" OUTSIDE

BUILDING LINE.

BEYOND 1500 OUTSIDE BUILDING LINE.

RE: CIVIL SITE UTILITIES DRAWING FOR CONTINUATION

FIRESP2

PRESSURE GAUGES

DETECTOR CHECK VALVE

OS&Y VALVE W/ TAMPER SWITCH

100 SWING CHECK VALVE

GENERAL NOTE:

IT IS THE RESPONSIBILITY OF

THE CONTRACTOR TO CONFIRM

TO FD

50 MAIN DRAIN

FIRE PROTECTION MAIN TO F/A PANEL

VANE TYPE

FLOW SWITCH

ALARM CHECK VALVE

WITH MAIN DRAIN

ZONE MAIN DRAIN

TO NEAREST MOP SINK

TO ZONE

MULTIPLE ZONES, MARK CEILING WITH RED DOT AT EACH ZONE VALVE.

ZONE (OS&Y) VALVE W/ TAMPER SWITCH

ZONE

PRESSURE GAUGE

FIRE DEPT.

CONN. POST

PIPE TO FD (TYP)

RWH

WT#1 WT#2

HWCP

NONE

HWRP

EXP TANK

50 HWS

25 HWR

50 CW SUPPLY

T THERMOMETERPUMP CONNECTIONS

UNION

BALL VALVE

T&P Relief Valve (typ) typ

METALLIC ELECT. CONDUIT

1/ 2" = 16mm 3/ 4" = 21mm 1" = 25mm 1-1/4" = 35mm 1-1/2" = 41mm 2" = 53mm 2-1/2" = 63mm 3" = 78mm 3-1/2" = 91mm 4" = 103mm 6" = 155mm

MECHANICAL PIPE & FITTINGS

ELECTRICAL PVC CONDUIT

1/ 2" = 15mm 3/ 4" = 20mm 1" = 25mm 1-1/2" = 40mm 2" = 50mm 2-1/2" = 65mm 3" = 75mm 4" = 100mm 6" = 150mm 8" = 200mm

HWCP STORAGE TANKS CIRCULATING PUMP, 14 GPM @ 7 FT HD. GRUNDFOS MODEL UP26-64SU&SF.

SPECIFICALLY MANUFACTURED FOR OPEN SYSTEM. 1/12 HP @ 120V/1 SERVICE.

20CW

ELECTRICAL PANELS

50 CW

TAP 100 CW LINE AND CONNECT SYSTEM MAKE UP WATER FOR:

1. HEAT PUMP CIRCULATION SYSTEM (TS&TR) - 20CW.

2. WATER TO WATER HEAT PUMPS FOR OA UNIT - 20CW.

3. HEAT RECOVERY WATER LOOP - 20CW.

4. BACK FLOW SOURCE FOR CIRCULATION LOOP FILTER - 40CW.

INCLUDE BACKFLOW PREVENTER AND PRESSURE REGULATOR

WITH EACH. INCLUDE DDC FLOW SENSOR FOR ALL EXCEPT FOR

FILTER BACKWASH SUPPLY.

ALL BACKFLOW PREVENTERS

WILL BE REDUCED PRESSURE

VALVE - DOUBLE CHECK VALVE

TYPE IS NOT APPROVED.

AI

LE

TR

IC

AL

IT

W/W HEAT PUMPS, TYP.

FILTER

/A

T

PU

PS

YP

OA AHU

EXP TNK

PUMP

Pump

BELOW LOWER SLAB

LANDSCAPE IRRIGATION 75 LINE,

200 SLEEVE FROM INSIDE TO OUTSIDE

PAST WALK, RE 1/L3.01.

75 CW

DRAIN VALVE

RELIEF VALVE (TYP)

UNION (TYP)

BALL VALVE (TYP)

REGULATOR (TYP)

FILTER (TYP)

PRESS. GAUGE

SOUND REDUCING

COVER (TYP)

20mm MAIN

15mm

FLEX

CONNECTOR

DRIP LEG &

EXHAUST, ROUTE TO

OUTDOORS PER NFPA 99

VACUUM

PRES SW

TO SYSTEM

LEAD-LAG

PARALLEL

CONTROLLER

RAMVAC

S1 ELECTROLS

#8-600-425

MSTR SLAVE

THRU FLAT ROOF

VAC. GAGE

ALL PIPING TO BE 32mm SCH 40 PVC (1-1/4" I.P.S.), EXCEPT NOTED.

REMOTE

SWITCH

LEAD-LAG

PARALLEL

CONTROLLER

REMOTE

SWITCH

AIR TECHNIQUES

53251, LABEL "AIR1 AIR2"

CONTRACTOR TO

MANUFACTURE CUSTOM

STEEL RACK FOR AIR

COMPRESSORS,

PAINT BLACK.

NONE

15mm

PROVIDE ALL EQUAL PIPE LENGTHS BETWEEN WT#1 AND WT#2 FOR EVEN FLOW.

sp ac e re se rv ed fo r e le ct ric al no p ip g, n o du ct s.

space reserved for electrical, no piping, no ducts.

WATER SOFTENER & BRINE TANK - "WS"

MARK DESCRIPTION

AC

CONTINUOUS, 20mm MANIFOLD IN TRENCH, 15mm TO CABINET. INSTALL PRES. REGULATOR AIR COMPRESSOR, AIR TECHNIQUES #AIRSTAR 50C, 3 HP @ 208V/1. 9.2 CFM PEAK, 5.2 CFM

SET TO 689kPa (100 PSI). CONTACT INFO - (516)433-7676, (800)247-8324-2 (MARK).

VAC

DISPLACEMENT, 114 LITER (30 GAL.) TANK, MOUNT 1-1/4" MANIFOLD ON TRENCH WALL WITH

UPGRADEABLE VACUUM AIR, RAMVAC BISON E-SERIES COMBO 3, 3HP @ 208V/1. 1.8 LITER,

1/100 SLOPE. CONTACT INFO - JOE OR BOB (800)572-6822.

INCLUDE OPTIONAL STAND, LEAD-LAG / PARALLEL CONTROL FOR TWO UNITS.

INCLUDE SOUND COVER, LEAD-LAG / PARALLEL CONTROL FOR TWO UNITS.

BYPASS VALVE

100 CONC PAD

WS WATER SOFTENER, COMMERCIAL GRADE, 360,000 MAX. GRAIN CAPACITY, 65 GPM SERVICE

FLOW RATE, 1200 LBS. SALT STORAGE, MacCLEAN MODEL KS12000. 2" SOLID BRASS

CONTROL VALVE, HEAVY DUTY FIBERGLASS MINERAL TANK, QUARTZ GRAVEL UNDER

DRAINAGE, HEAVY DUTY BRINE MAKER, PROGRAMMER WITH 6 CYCLE WITH TIMED BRINE

REFILL, CAPABLE OF BEING PROGRAMMED TO REGENERATE DURING LOW DEMAND

PERIODS. SYSTEM DESIGNED FOR 100° F, AND 100 PSI PRESSURE.

SK6

50mmPVC50mmPVC

RM G133

RM G133

FIRE SPRINKLER

WATER GONG, +2700

1 2 3

5 5.3 5.5 7

D.2

2.1

ALCOVE

B135

VESTIBULE

B143

EXERCISE

B140

LOUNGE

B141

STORAGE

B144

B139

WORKROOM

B124

OPEN

SPACE

B122

OFFICE

B123

STAIR 1

SB002

STAIR 2

SB001

MEN

SHOWERS

B133

B134

WOMEN

B136

B137

CORRIDOR

B131

B132

RR

ID

OR

B1

B121

B128

MEDICAL GAS

B130

SOILED LINEN

B142

HAZARDOUS

WASTE

B145

RECEIVING

B129

B127

B126

B119

HOUSEKEEPING

B102

ELEVATOR

EQUIPMENT

B103

CLINICAL ENG.

B118

B117

SHOP

B115

B116

B114

B120

MECHANICAL

ROOM

B101

B104

B110

B109 OFFICE

B108

B107

LABORATORY

B106

B105

B113

CONCESSIONS

B138

HALL

B112

SINK3

UP

LOW WALL LOW WALL

SAFE AREA

PLUMBING FIXTURE SCHEDULE

GENERAL PLUMBING NOTES

PLUMBING ABBREVIATIONS

FIRE SPRINKLER SYSTEM

PLUMBING PIPING SYMBOLS

EMERGENCY ELEVATOR OPERATION

PLUMBING KEYNOTES

C EMERGENCY SHOWER/EYEWASH D FIRE PROTECTION PIPING DETAILB WATER HEATER WITH STORAGE

E DENTAL VACUUM AND AIR COMPRESSOR EQUIPMENT DETAILS

DENTAL EQUIPMENT SCHEDULE

50 HW

40 CW25 HW

ACCURACY OF THIS INFORMATION AND SHALL NOT BE

THE ARCHITECT/ENGINEER HAS NOT VERIFIED THE

hgonzalez

Highlight

Inserted Text

Text Box

6"

4"

Line

MAIN LEVEL

16882SF +

14492SF =

31374SF

15333SF

50SD DN

14491SF

P01

P02

P03

P04 P01P02

P04

P01P02

P03P01

5d fu

9d df u

4dfu

6dfu

18dfu

30dfu

P06 P06P06 P06 P06

15dfu

16dfu

MechCD-MainFlr

Dental Section

Admin Section

Bottom Right Section

PAWNEE HEALTH CENTER, MAIN LEVEL

3/ 4" = 21mm 1" = 25mm 1-1/4" = 35mm 1-1/2" = 41mm 2" = 53mm 2-1/2" = 63mm 3" = 78mm 3-1/2" = 91mm

1" = 25mm 1-1/2" = 40mm 2" = 50mm 2-1/2" = 65mm 3" = 75mm 4" = 100mm

TURN UP THRU ROOF

5d

12dfu

7dfu

50 V dfu

50 V50 V

75 VTR75 V

51dfu

75 V75 V 88dfu

9dfu dfu

58dfu

73dfu

65 V

75 V 65 V

99dfu, 185ft

113dfu

G603

G619

HYDROTHERAPY

G618

CONFERENCE

G601

DRESSING

G617

TOILET

G615

JANITOR

G614

WAITING

G612

RECEPTION

G624

G625

G623

G622

PROCEDURE

G621

G620

G110

G109

G164

WOMEN SHOWER

G165

MEN SHOWER

G161

G160

G159

G137

G150

G148

G149

G144

G145

CEPH/PAN

G138

FILM PROC.

G139

G143

G136

G134

G140

G141

G128

G127

G126

G129

G133

G131

TESTING

G130

NURSE

STATION

G214

G213

STAFF

G219

SOILED

LINEN

G218

G216

G226

G227

G240

G239

G237

CLEAN

G235

G233

G238

G236

G234

EXAM

G232

G241

G244

G245

G246

G247

G220

G248

NURSE'S

G221

G249

G251

G250

G252

TRIAGE

G253

G255

VITAL

G257

G254

G258

TRIBAL

G262

G260

ISOLATION EXAM

G259

PHARMACY

G205

G224

G225

G228

HOUSE

KEEPING

G229

G231

G230

G201

G204

G203

G202

G125

G209

G208

PREPARATION

G210

G264

G206

G607

SHREDDER

G606

LIBRARY

G605

VENDING

G608

G609

G611

G4

CLOSET

G602

PHYSICAL

THERAPY

G613

G404

G405

G406 WORKROOM

G407

G410

G409

G408

G411MICRO

G415

G416

G420

G419

G417

LAB

G413

G402

G414 RECEPTION

G412

G401

MAIL AREA

G503

G531

G527

G529

G526

G525

G524

G523

G522

G521

G520

G519

G530

G516

G323

G544

G543

G542

G541

G540

G539

G537

ART

G538

G536

COPIER ALCOVE

G535

G533

G517

G502

G504

G506

G515

G513OFFICE

G512OFFICE

G511

G510

G509

G508

G507

SUPPLY/DUPLICATION

G320

G319

G318

G317

G316

G315

G314

STUDY

G322

INTERVIEW

G321

G313

G162

G166

G167

G123

G122

G121

G120

G104

G105

G112

G114 ALCOVE

G113

G117

G103

G108

G107

REGISTRATION

G102

G156

G157

G155

SUBWAITING

G223

G222

G312

G309

G311

G308

WORK

G304

G310

TECH

AREA

G306

VISION

FIELD

G305

G303

EYEWEAR

G302

G301

G151

UTILITY

G153

G604

G610

SG001

G403

HA

LL

G2

G217

G263

G242

G215

G256

G501

G545

OPEN OFFICE

G518

G528

SG002

G1

G546

G5

G114

G532

G307

G158

G124

G119

G146

G152

G135

OFFICE SPACE

G106

LOBBY/

COMMON WAITING

G101

G100

G111

G115

G116

G211

G147

SINK5

SI

NK

SINK

P4700

OTOSCOPE

RE

REF

CUBCURTAI N2CUBCURTAI N2

REFRIG3

UN

DE

RC

OU

NT

ER

FI

CA

BI

NE

FAX

2020CPR

UNDERCOUNTERFI LECABI NET

UNDERCOUNTERFILECABINET

PROJECTOR

3350E

1 2 3 5

6 7 8 9

D.8

1.2

1 2 3 4 6 7 8 9 10

5.3 5.52.1

6.1

D.7

CANOPY

DOWN

LOW WALL LOW WALLLOW WALL LOW WALL LOW WALL LOW WALL LOW WALL

LOW WALL LOW WALLLOW WALLLOW WALLLOW WALLLOW WALL

LOW WALL LOW WALL LOW WALL

LOW WALL LOW WALL LOW WALL LOW WALL LOW WALL LOW WALL

LO

L

LOW WALL

PAY TELEPHONE

WHEEL CHAIRS

HEALTH RECORDS

GENERAL PLUMBING NOTES

M1.02

100 BELOW MAIN SLAB

50 HW DN

SK2

EW

WC3

SK1, LOOKS LIKE SK2

WC3 LV1

WC3LV1 SK1

SK3-PT

SK3 SK1

SK5

WC3LV1

WC2WC2

WC1WC1

URWC1 LV

LV2MB

SK2-PT

SK2-PTSK2

3LV1

250 SD, 3/1000 SLOPE

BELOW SLAB

300 SD, 3/1000 SLOPE

AS HIGH AS POSSIBLE

BELOW MAIN SLAB

250 SD DN FROM

ATTIC SPACE

250 SD DN FROM

ATTIC SPACE250 SD DN

50SD DN

250 ESD DN 250 ESD, 3/1000 SLOPE

300 SD OUT FOR EMERGENCY

ROOF DRAINAGE, BELOW GRADE.

100 W BELOW

MAIN SLAB

75 W, 40 V, 15 CW & 15 HW ROUGH

IN FOR FUTURE SINK (TYP. IN OFF.)

CAP W, V, CW & HW IN WALL.

75 W, 40 V, 15 CW & 15 HW

ROUGH IN FOR FUTURE SINK

CAP W, V, CW & HW IN WALL.

ESD

ESDESDESD

SDSD

SDSD

100 W BELOW

MAIN SLAB

DN

50SD DN 50 SD DN

300 SD BELOW LOWER SLAB,

RE: A/M1.01 FOR CONTINUATION

50 V BELOW SLAB

SH

TB

LV

WC1WC1

SK4

WB

WC2WC2

SK2-PT

75W

SK7

25 HW

20 HW

300 SD

250 ESD,

3/1000 SLOPE

EMPTY 50 PVC

EMPTY 50 PVC

150 W DN

75 W

CS

15 HW

100 CW UP

20 HW & 25 CW

& CW

40 CW

25 CW

15 HW & 25 CW

20 HW25 CW

15 CW

20 CW 15 CW

25CW

15CW

15HW

WCO

75 W75 W

HUB DRAIN

20HW

15CW40 CW

15 HW &

VAC-2

1. INSTALL DOMESTIC WATER PIPING ABOVE CLG & BELOW

INSULATION SO THAT NONE OF THE PIPING IS EXPOSED

2. PROVIDE CONNECTIONS & CONNECT ALL EQUIPMENT PROVIDED

BY OWNER PER MANUFACTURER'S INSTRUCTIONS.

3. REFER TO SHEET M1.01 FOR PLUMBING KEYNOTES.

75 V FROM

75 VTR

HB20CW

20CW

TO UNINSULATED SPACE.

VAC-1 AC-1,2

STACKED

32 VACUUM

LINE IN TRENCH.

ATTACH TO SIDE

TO FACILATATE

SLOPE.

20 AIR COMPRESSOR

MANIFOLD LINE IN

TRENCH.

v

TWO (2) 50 VACUUM

DENTAL DISCHARGE

LINES, MT AS HIGH

AS PRACTICAL.

1/4" TAP

FOR MODEL

TRIMMER,

+1100

75 MAIN VENT UP

FROM LOWER LEVEL

ROUTE TO

BATHROOM

BELOW

VENT PLBG FIX. W/

AIR ADMIT. VAVLE

CS

VENT PLBG FIX. W/

AIR ADMIT. VAVLE

25 HWR

ACCURACY OF THIS INFORMATION AND SHALL NOT BE

THE ARCHITECT/ENGINEER HAS NOT VERIFIED THE

ATTIC LEVEL

PAWNEE HEALTH CENTER, ATTIC LEVEL

1 m 3 mPLUMBING PLAN

150 SD VERTICAL

65 SD VERTICAL

75 SD

VERTICAL

50 SD VERTICAL

75 SD, 3/1000 SLOPE

ABOVE ATTIC DECK

100 DN

75 SD 3/1000 SLOPE

ABOVE ATTIC DECK

895SF 1549SF

100 SD AS HIGH

AS POSSIBLE

8759SF

200 SD 3/1000 SLOPE

15333SF

250 SD 3/1000 SLOPE

15295SF

250 SD 3/1000 SLOPE AS

HIGH AS POSSIBLE8123SF

200 SD 3/1000 SLOPE

16882SF

250 DN TO BELOW

MAIN FLR SLAB

1549SF

8721SF

200 SD, 3/1000 SLOPE

8721SF7320SF

660SF

654SF

6574SF 7172SF

1587SF

235SF

M1.03

MechCD-Attic

50 DN 50 DN 50 DN

50 DN 50 DN 50 DN

14491SF

100 SD 3/1000 SLOPE

75 SD VERTICAL

EM

150 SD

EMEM

100 DN

150 SD EM

7320SF

7171SF

AS HIGH AS POSSIBLE200 SD, 3/1000 SLOPE

250 SD

3/1000 SLOPE

PLATFORM

A105 PLATFORM

A104

A106

A108

A109

STAIR

SA002

A101

A1

1936SF

915SF 915SF 75 SD 465SF

1065SF 100 SD

810SF

1315SF

2125SF

100 SD 3/1000 SLOPE

150 SD 3/1000 SLOPE

3440SF

4755SF

537SF840SF915SF915SF493SF 75 SD 75 SD 75 SD 75 SD75 SD

705SF 75 SD 1305SF 100 SD 1265SF 100 SD 1265SF 100 SD 781SF

3/ 4" = 21mm 1" = 25mm 1-1/4" = 35mm 1-1/2" = 41mm 2" = 53mm 2-1/2" = 63mm 3" = 78mm 3-1/2" = 91mm

1" = 25mm 1-1/2" = 40mm 2" = 50mm 2-1/2" = 65mm 3" = 75mm 4" = 100mm

FD, PLASTIC FLOOR DRAIN

WITH NO STRAINER

FD FD

FDFD

FD FD

TWO (2) 50 VACUUM

DENTAL DISCHARGE

LINES, TERMINATE

ABOVE ROOF 180°

FITTING W/ SCREEN

& BACKDRAFT DAMPER/

VALVE.

PER NFPA 99

MAINTAIN 600 AIR

GAP ABOVE ROOF.

UP THRU ROOF FROM BELOW

PVC ROOF DRAIN

HUB ONLY.

50 AIR GAP

GUTTER DOWN LEAD

FLAT ROOF

METAL ROOF

GUTTER

ESCUTCHEON

SEAL PER DETAIL 7/A3.21a

INSULATE

RE: PLANS

FOR SIZE

SUPPORT STRAP

CLEAN AND SAND

ALL EXPOSED PVC,

PAINT WITH ACRYLIC

BLACK SPRAY PAINT

TO PREVENT

ULTRAVIOLET

DEGRADATION

OF PIPE.

ORIGINAL SCALE: NONE

ROOF DRAIN DETAIL

SECTION AT METAL ROOF TO FLAT ROOF

DISPLAY

WALL BELOW

IDO

COMMUNICATIONS

A103

SA001

A107

A105

5.3 5.5

ROOF HATCH

SK

YL

IGH

PVC COND. PIPING

W/ INSULATION FOR

SOUND DEADENING.

ACCURACY OF THIS INFORMATION AND SHALL NOT BE

THE ARCHITECT/ENGINEER HAS NOT VERIFIED THE

550 x 350, UP

SPACE ALLOCATED FOR

FIRE SPRINKLER RISER

G3

C1

C2

OUTSIDE AIR

HANDLING UNIT

EXPANSION TANK

800x800 UP TO

ATTIC SPACE

SUPPLY AIR

SUPPLY & RETURN PIPING

FROM GEOTHERMAL FIELD

RE: B/M0.01 FOR CONNECTIONS

& B/M3.02 FOR DETAILS.

DOMESTIC WATER

HEATERS

SY

ST

YP

AS

BLADDER

AIR VENT, AMTROL #706

EXP TANK

TEMP SENSOR (TYP)

TWO WAY 2 POSITION

VALVES, INTERLOCK W/

RESPECTIVE PUMP.

TYPICAL FOR EACH

GEOTHERMAL FIELD.

GEOTHRMGEOTHRM

GEOTHRM

EO

TH

EL

PUMP - 'G1'

PUMP - 'G2' PUMP - 'G3'

AIR PURGER, AMTROL #462 (150mm) - MOUNT AT HIGHEST POINT

WATER

FILTER

BACKFLUSH

TO DRAIN

TS

HEAT PUMPS

SYSTEM

CIRC

PUMPS

100 100 100

100 100 100

150150

SA180

SA200, 200

SV60

SA190, 200

SA190, 200

SA220SA230

SA

0, SA140,

SA130, 150

SA130,

SA130, 150 SA130, 150

SD200SA220

SA190

M2.01

HVAC PLAN

SA190,

SC90

SC140

SA40 SA180 SA180

RC270

RA500SA100

RA100

RA360

RA340

RA500,

RA180

SC260

SC260

SC250

EB100

SC100

SC100

SD60

PAWNEE HEALTH CENTER, LOWER LEVEL

SA250

A2

RA400RB690

SA240SA240

SA230

EB50

SA140

SA200

RA130

SA100

RA500

RA440

SA160

SC60

EB300

EB250,

350EXH 1400EXH

850EXH

EB250

SC220

2150EXH

RA600

RA160

SC130

RA210

RA200, 200

DAMPER

THERMOSTAT

SQUARE TO ROUND FITTING

RETURN AIR GRILLE

EXHAUST GRILLE

SPIN-IN DUCT TAKEOFF

ROUND OR RECTANGULAR DUCT

CEILING SUPPLY DIFFUSER

45° DUCT TAKEOFF

PAINT ALL ROOF & WALL MTD EXH. FANS & PENETRATIONS THRU BLDG

TO MATCH BUILDING EXTERIOR.

RELIEF AIR DAMPER, TYPICAL ALL AHU-5, RE: D/M2.01M01

M02

M01

500x400 I.D.

600x400 I.D.

SD30

200EXH

RA780

RA170

RA340,

SC150

SA140 SA140 SA140

300CFMEXH

300EXH

EF120, MTD ON WALL ABOVE CANOPY

EB120

e d

4" FILTER

RAOA

VOLUME DAMPER

POSITION DAMPER

MOTORIZED TWO

500* CFM

600x400

8" DIAMETER RELIEF

AIR WITH BACK-DRAFT

cws&r d

EDGE OF

MECH SPACE

TOP

500x400

ABOVE

INSTALL 1" ACCOUSTICAL

DUCT LINER FROM UNIT

SUPPLY AND RETURN DUCTS.

TYPICAL FOR BOTH

TO MIN. OF 3600 LENGTH.

* INITIAL SETTING

UNLESS NOTED ON

FLEX

CONNECTION

RA

RAILING

HP UNIT

TYPICAL HP

VERT. UNIT

PLAN.

6" STAND,

BRAMEC 4008

6x6 ANTI-VIBRATION

PAD, BRAMEC #6136

400x600

400x500

1" COND. TO FLOOR

DRAIN, TYPICAL

FLEX CONNECTION

12'x4'

1700x900

900x800, HUG CLG.

STRUCTURE

800x800

TWO (2) 5 TON WATER TO WATER UNITS STACKED*

THREE (3) 5 TON WATER TO WATER UNITS STACKED*

THREE (3) 5 TON WATER TO WATER UNITS STACKED*

* CONTRACTOR TO FABRICATE SUITABLE

STEEL RACK TO SUPPORT UNITS AND

TO PROVIDE MFR'S CLEARANCES.

DESCRIPTIONMARK

UNIT

SIZE

(CFM)

S.P.

GPM GPMGPM

100% O.A.

ER COIL

MARK

OA

PRIMARY COIL

HP

FAN

NOTES

CONN CONN CONN

25 2.5" 2.5" 2.5"48

UNITS

1, 2, 3, 412000 48

SECONDRY COIL

MODULAR AIR HANDLER, VARIABLE VOLUME, MULTIPLE WATER COILS, TRANE CLIMATE CHANGER. INCLUDE

THE FOLLOWING MODULES IN ORDER:

VERTICAL INTAKE MODULE (40" L), "G" FLAT FILTER MODULE WITH 4" 30% EFF PLEATED FILTERS, "R" MEDIUM MODULE WITH 6-ROW UNIT (UW) ENERGY RECOVERY COIL, "K" SMALL INSPECTION MODULE WITH ACCESS DOOR, "R" MEDIUM COIL MODULE WITH 8-ROW UNIT (UW) PRIMARY COIL, "K" SMALL INSPECTION MODULE WITH ACCESS DOOR, "R" MEDIUM COIL MODULE WITH 8-ROW UNIT (UW) SECONDARY COIL, "AA" VERTICAL DISCHARGE FC FAN, TYPE "B" WITH VARIABLE FREQUENCY DRIVE AND ISOLATION TRANSFORMER. ALL COILS TO HAVE LESS THAN 5 FT WG PRESSURE DROP AT GIVEN GPM. DO NOT EXCEED 500 FPM VELOCITY AT FACE OF

COILS, SUPPLY LARGEST FACE AREA COIL AVAILABLE FOR MODULE.

IN-LINE

TYP

5 TON

WTR-WTR

PR

IM

AR

EC

AR

IL

DDC ANALOG INPUT

UNION P/T PLUG

TEMPERATURE SENSOR

AIR COIL

BALL VALVEUNION

THERMOMETER

MANUAL AIR VENT

COIL # HPS

PRIMARY

SECONDARY

PIPE SIZE

GPM

48 65

1. UNIT TO INCLUDE 2" SPRING ISOLATORS FOR FAN & MOTOR ASSEMBLY.

2. UNIT TO INCLUDE BASE RAIL.

EF

ED

PE

YS

FO

. O

PROVIDE ANTIFREEZE

SOLUTION TO PREVENT

FREEZING AT -5°F OA

ALL OUTSIDE AIR UNIT COILS.

EXP TANK,

AMTROL SX-90V

3. FAN MOTOR TO BE RATED FOR VARIABLE SPEED DRIVE USE, 460/3 VOLT.

PE

AT

SE

SO

FAN MODULE OA

BLANK MODULE W/ACCESS DOOR

FLEX CONNECTION (TYP)

BASE RAIL

PROVIDE NEOPRENE ISOLATION UNDER ALL BASE RAILS

COIL MODULE* W/ 6-ROW ENERGY RECOVERY COIL

CONTINUOUS - MASON TYPE "W" WAFFLE PAD

600x600

ACCESS

DOOR

FILTER MODULE W/ 4"-30% EFF. PLEATED FILTERS

COIL MODULE* W/ 8-ROW PRIMARY COIL

COIL MODULE* W/ 8-ROW SECONDARY COIL

4" DRAINABLE ALUMINUM

LOUVER WITH INSECT SCREEN

BY ARCHITECT

OF BUILDING OR AS DIRECTED

PAINT TO MATCH EXTERIOR

2 - POSITION

INSULATED,

WEATHER-

TIGHT BLADES.

4. CARRIER, TEMTROL AND YORK MANUFACTURERS ARE APPROVED PROVIDED PROPOSED EQUIPMENT WILL MEET OR

EXCEED SPECIFIED QUALITY AND PERFORMANCE.

CONDENSING WATER

FILTER ASSEMBLY

TO GEOTHERMAL

FIELD, RE: M0.01

15 - 75mm LINES.

FROM GEOTHERMAL

FIELD, RE: M0.01

15 - 75mm LINES.

INTAKE OUTSIDE AIR

LOUVER "LV", MOUNT

AS HIGH AS PRACTICAL.

(3600x1200)

65 65

SW SANITARY WASTE

AW ACID WASTE

RD ROOF DRAIN

TS CONDENSING WATER SUPPLY

TR CONDENSING WATER RETURN

CA CONTROL OR COMPRESSED AIR

TW DOMESTIC TEMPERED WATER

VA NON-MEDICAL VACUUM AIR

20CW MAKE-UP

7 GAL. (26 L) BLADDER

TYPE EXPANSION TANK

EXHAUSTEXHAUST

AUTOMATIC

AIR VENT

12,000 CFM

5500 CFM 3800 CFM

ENERGY RECOVERY COIL

(TYPICAL OF 2)

25 GPM (95LPM)

ENERGY RECOVERY COIL

50 GPM (189LPM)

LOCATED IN MECH RM

MOUNTED IN OA UNIT

BASE MT PUMP

50 GPM (189 LPM),

30 FEET HEAD (90kPa)

3/4" (20mm) WITH HOSE NIPPLES

CHEM FEED

50% ANTI-FREEZE

MechCD-LowerFlr

200 200

150 200

200 EB300,

35048 0c fm

800cfm

1650cfm 1880cfm

430cfm

850cfm

1650EXH 1850EXH

450 500 x 350 550 x 350

1040cfm 1300cfm

280cfm 600cfm

RA605

940cfm 1300cfm

250 1540cfm

500cfm 700cfm

760cfm

1010cfm

1260cfm

1780cfm

300 300 300

200300

500x400 I.D.

EB100,

3.5" 15

COIL CONNECTIONS

OUTSIDE AIR SYSTEM

FLOW DIAGRAM

M-1

M-2

M-3

M-4

M-5

M-6

DDC ANALOG INPUT

FLEX HOSE

P/T PLUG

VALVE WITH P/T TAPS

MANUAL FLOW BALANCETEMP. SENSOR

HEAT

RETURN (40)

SUPPLY (40)

HP CONNECTIONS

2 WAY

2 POSITION

VALVE

+100 CONC PAD

+100 CONC PAD (TYP)

GATE VALVE

BUTTERFLY VALVE

STRAINER

BALANCE VALVE

FLOW DIRECTION

PRESSURE REDUCTION VALVE

MOTOR OPERATED VALVE

L/S, liters/second (CFM) AIR FLOWMARK

AIR DISTRIBUTION SYSTEM

AIR TERMINALS:

2ND ENTRY 3RD ENTRY1ST ENTRY

EXHAUST

VAV

TYPE

AIR VOLUME QTY

ETC.

EXAMPLE:

SA100, 150 SUPPLY DIFFUSER, TYPE A, 50 L/S, 150 DUCT

SA50 25 (50)

SA100 50 (100)

SA250 125 (250)

SA400 200 (400)

RA1000 500 (1000)

EA300 150 (300)

1080cfm

1460cfm

150x150

1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1

1 1 1 1 1 1 1 1 1 1 11

1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1

1 1 1 1 1 1 1 1 1 1 1 1 1 1 1

1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1

200250

900cfm fm300

1460cfm400

320cfm

360cfm

1100cfm 1360cfm

EA200200

1540cfm

1360cfm

BRANCH OFF SA DUCT & CONNECT TO SA DIFFUSER WITH 200

DUCT AND DAMPER.

180cfm

370cfm

250 200

910cfm 1780cfm1780cfm

200 250

30056

M03 BRANCH OFF RA DUCT & CONNECT TO RA GRILLE WITH 200 DUCT,

SPIN-IN AND DAMPER.

M03

450cfm

410cfm 970cfm

ELECTRICAL PANELS

AI

3/ 4" = 21mm 1" = 25mm 1-1/4" = 35mm 1-1/2" = 41mm 2" = 53mm 2-1/2" = 63mm 3" = 78mm 3-1/2" = 91mm

1" = 25mm 1-1/2" = 40mm 2" = 50mm 2-1/2" = 65mm 3" = 75mm 4" = 100mm

DUCT MOUNTED

SMOKE DETECTOR

DDC FLOW SENSOR

20CW MAKE-UP

DDC FLOW SENSOR

AIR COIL

COIL CONNECTIONS

DUCT MOUNTED

SMOKE DETECTOR

E:

/M

0.

LANDSCAPE IRRIGATION 75 LINE,

200 SLEEVE FROM INSIDE TO OUTSIDE

PAST WALK, RE 1/L3.01.

500X400

600X400

600X400 500X400

350x300

W610x113 BEYOND W410x39

CEILING

600x400 WITH 37.5 INSULATION - RETURN DUCT

MAX HT OF LTG FIX

300 DIA WITH 37.5 INSULATION

CENTER BEAM

CROSS BEAM

LOOKING NORTHWEST

4TH ENTRY

DUCT SIZE

1. INDICATED EXAMPLES ONLY.

WATER SOFTENER & BRINE TANK

'C1' & 'C2'

M-2 M-5

M-1

M-6

M-3

M-4

1 1 1 1 1 1 1 1 1 1 1

"R"

100 DIA

DRYER

VENT

ALUM VENT

WALL CAP,

BACKDRAFT

DAMPER, NO

SCREEN

HVACNO SCALE HTPMP

HVACNO SCALE ERLOOP

HVAC SYSTEM

HVAC SYSTEM HVACSCALE: 1 = 25 SECTIONHVAC SYSTEM

HVACNO SCALE ERLOOP

HVAC SYSTEM

600x500

CLEAR

DUCT

O.A.

HVACNO SCALE HTPMPHVAC SYSTEM

EF

3.

FO

:2

SC

PL

AN

VA

IP

RE: A/M3.02

FOR FLOW PLAN

1 2 3

5 5.3 5.5 7

B135

B143

EXERCISE

B140

B141

B144

B139

B124

B122

B123

SB002

SB001

B133

B134

B136

B137

B131

B132

B121

B128

MEDICAL GAS

B130

SOILED LINEN

B142

HAZARDOUS

WASTE

B145

RECEIVING

B129

B127

B126

B119

HOUSEKEEPING

B102

EQUIPMENT

B103

CLINICAL ENG.

B118

B117

SHOP

B115

B116

B114

B120

MECHANICAL

B101

B104

B110

B109 OFFICE

B108

B107

B106

B105

B113

CONCESSIONS

B138

B112

SAFE AREA

HVACNO SCALE HTPMPHVAC SYSTEM

8FT

DUCTER RETURN

TR TS

ER SUPPLY

4'6" MAXIMUM

BETWEEN SUPPORT

TRAPEZE SUPPORTED

AT EACH JOIST

DISTRIBUTE WEIGHT

EQUALLY ABOUT

CENTER LINE

CENTERLINE OF

TRAPEZE & BEAM

TWO WAY 2 POSITION VALVE

FOR CHEM TREATMENT

ONE PUMP PER SYSTEM

HOSE BIBBS CONN.

IN LINE PUMP CONNECTION

MOTORIZED TWO POSITION DAMPER,

ENERGIZE WITH 120V POWER FROM

NEAREST REC CIRCUIT, OPERATE WITH

FAN AND CO2 SENSOR. OPEN - CO2 OVER

SETPOINT, CLOSED BELOW AND CLOSED

WITH FAN OFF.

200FIRE DMPR

100° FUSE

9x4 DUCT UP TO

EF120 MOUNTED

TO SIDEWALL

PLAN VIEW SIDE VIEW

ABBREVIATIONS

FLOW DIAGRAM SYMBOLS

AIR TERMINAL NOTATION

OUTSIDE AIR HANDLER EQUIPMENT SCHEDULE

SYSTEM SCHEDULE

100% O.A. HANDLER DETAIL

GENERAL MECHANICAL NOTES

MECHANICAL KEYNOTES

MECHANICAL SYMBOLS

D TYPICAL GEOTHERMAL HEAT PUMP UNIT DETAIL

E ENERGY RECOVERY LOOP DIAGRAM F SECTION AT DUCT CROSSING

B CONDENSING WATER FLOW DIAGRAM

C ATTIC CROSS SECTION @ HEAT PUMPSG NEW SECTION UTILITY SUPPORT

ACCURACY OF THIS INFORMATION AND SHALL NOT BE

THE ARCHITECT/ENGINEER HAS NOT VERIFIED THE

M2.02

BRANCH OFF SA DUCT & CONNECT TO SA DIFFUSER WITH 200

M04

M05

& ENERGY RECOVERY PIPING THRU MAIN FLOOR STRUCTURE.

UP TO

M-13

UP TO M-12

M-10

UP TO M-11

UP TO M-7

UP TO M-8

TO M-9

-2

-1

TO M-18

UP TO M-26

TO M-27

TO M-34

M-23

TO M-22

TO M-21

TO M-20

550x500

EF3800

SA220

SA300, 250

EA250

SA100 SA295 SA295

EB200,

SA280

SA200SA200SA200SA200SA180

SA230,

SA230, 200

RB1000

SB50

SA160 SA160

SA160, 200

RA300

RA800

EB80

EA50

SA160SA260

SA320

SA220,

SA140 SA110 SA200

SA140 SA110 SA110SA110

EA160

SA240 SA250

SA330

SA240

SA140, 150

SA200, 200SA190, 200

SA170

SB110

SA150, 150

SA130, 150

SA130

SB100

SA90

SA150

SA280,

SB100 SB100

SA120

SB90

SA90 SA290

SA270

SA310

SC80

EB

SC

EA160

SA160, 200

SA150SA150

EA100

SA110

SA90

SA280 SA280

SA300 SA300

EB150

SA190 SA200, 200

SA250SA250

SA240, 200

EB70

0, SA180SA140,

200 EB100

SA190 SA190, 200

SB40

SB130

SA260

EA100

SA200 SA250 SA250 SA250

EA130

SA240 SA240 SA240 SA240SA240

SA190 SA190 SA190 SA190,200

SA180SA240,200

SA190 SA190 SA190SA220SA160SA280

SA210

SA210

SA210 SA140

SA350

SA320 SA320 SA320

SA80

SA200 SA220 SA230

SA100 SA210

EB100 SA130 SA130

SB70

SA180, 200

SA190,

SB210

SC160,200

SA50

SE140

SC160

SA350SA350

SA350SA350

SA150SA110

SA80

SA310

EA250, 200

EA150

SA170, 200

SA230SA360

SA150,

SB80

650x300, FROM BELOW

SD80

RB1350

RB1800

RA120

RA820

RA110

RA200

RA700

RA280

SD250

RA180,

RA170,

RA130,150

RA150

RA130,

RA180,

THERMOSTAT

SQUARE TO ROUND FITTING

RETURN AIR GRILLE

EXHAUST GRILLE

SPIN-IN DUCT TAKEOFF

ROUND OR RECTANGULAR DUCT

CEILING SUPPLY DIFFUSER

45°…

This is the start of the file's text. The full file is on GovTribe.

Other files

File Type Posted
SF 24.pdf PDF
http //www.wdol.gov/wdol/scafiles/davisbacon/OK13.dvb v 13 —
PawneeCTHVACAmendment0001.docx.pdf PDF
52.213-4.pdf PDF
Additional Provisions-Terms and Conditions-Non 8(a).docx DOCX document
Insurance Requirements.docx DOCX document
Pawnee-Div-15-Mech.pdf PDF
Pawnee X-Ray Room Modification —
PawneeCTMechanicalSF18.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

File details come from the government source that posted it. Updated .