19WHAR18Q0004 ATTACHMENT 1
STATEMENT OF WORK (SOW)
INL BOGOTA – MEDELLIN HUMAN ID LABORATORY EQUIPMENT & SUPPLIES
DATE: June 26, 2018
1. BACKGROUND
The Office of International Narcotics and Law Enforcement (INL) Affairs at the U.S. Embassy in Bogotá, Colombia, needs to purchase Human Genetics Identification laboratory equipment for the Colombian Attorney General Office’s Technical Investigative Group (CTI) in Medellín, Antioquia. CTI has confirmed that the purchase and installation of this equipment complies with Colombia’s National Agency for Accreditation (ONAC) ISO 17025: 2017 (the primary international standard used to by Colombia to accredit laboratory tests and calibration).
The purchase includes: installation, training, warranty and guarantee of the equipment to be donated to the Government of Colombia (GOC) for law enforcement purposes. The objective of the purchase is to assist the GOC expedite the processing the hundreds of unresolved cases involving human remains, the majority of which are thought to be linked to the armed conflict. Currently the GOC’s ability to process these human remains is limited because there are only two labs in the country, in Bogotá and Barranquilla, with the appropriate equipment to undertake the necessary forensics tests. Therefore, the purchase of Human Genetics Identification laboratory equipment is expected to significantly increase the capacity of the Colombian Government, more specifically CTI to identify human remains. As a part of its commitment to this project CTI- Fiscalia made the necessary changes and renovations to adapt the building, as needed, where the equipment will be installed and donated.
2. TECHNICAL SPECIFICATIONS
The following equipment is being acquired on a brand name or equal basis in accordance with FAR 52.211-6. To be considered for award “or equal” products must:
a. Meet or exceed all the technical specifications;
b. Not exceed the ranges of requested dimensions (otherwise the equipment cannot be installed in the space that is currently available in the Human Identification Laboratory); and
c. In the case of items 5 (3500 Genetic Analyzer for Human Identification), 11 (Real-Time PCR System for Human Identification), 16
(Thermal cycler 3 x 32-well), and 39 (Qiacube. Reference: Cat No./ID: 9001882), the proposed “or equal”.
2.1. Specifications:
No. ITEM Quantit y Width
(cm) Depth
(cm) Height
(cm) Voltage
(AMP or VOLT)
Class II Type A2 Biological Safety Cabinet Microprocessor controller
Protection type: product and personnel.
Conformance: NSF 49, UL.
Range exhaust volume: 200-300 CFM.
Lighting: fluorescent, ultraviolet.
Includes:
∗ Base stand with casters.
∗ Work surface and 4 units HEPA filter or ULPA.
2 100 - 110 70 - 80 130 - 160 110 - 120 volts, 60 Hz
Polymerase Chain Reaction Cabinet Microprocessor controller
Lighting: fluorescent, ultraviolet.
Supply HEPA filter, 99.99% efficient on particles 0.3 micron.
Includes:
∗ Base stand with casters.
1 90 - 110 65 - 80
85 – 95 (150-170 with base stand)
110 - 120 volts, 60 Hz
∗ Work surface.
∗ HEPA filter.
Ductless Fume Hood Microprocessor controller
Lighting: fluorescent.
Exhaust volume: 100 fpm (220 CFM) o higher with sash opening of 9.4" o higher.
Includes:
∗ Base stand with casters.
∗ Work surface.
∗ 2 units HEPA filter.
∗ 2 units Carbon filters.
1 100 - 110 65 - 80
90 - 120 (150-190 with base stand)
110 - 120 volts, 60 Hz
Ductless Fume Hood Microprocessor controller
Lighting: fluorescent.
Exhaust volume: 100 fpm (220 CFM) or higher, with sash opening of 9.4" or higher.
∗ Base stand with casters.
1 100 - 110 65 - 80
90 - 120 (150-190 with base stand)
110 - 120
∗ Work surface.
∗ 4 units Carbon filters.
Please refer to pictures enclosed.
3500 Genetic Analyzer for Human Identification
Format: 8-well strips, 96-well plate.
Run time: 30 to 140 min. (varies depending on run module).
Dimensions: electrophoresis unit: 122 cm (W) (open door) x 61 cm
(D) x 72 cm (H), Electrophoresis unit width (closed door): 61 cm.
Throughput: ≥ 312 samples/day (HID36_POP4), ≥ 88 samples per day (standard sequencing with POP-7™ and 50 cm capillaries).
Requirement electrical: 120 V.
Range: 80 to 368 sequencing samples per day (varies depending on run module).
Read length: > 850 bp using a 50 cm capillary and POP-7™ polymer, ≥ 420 bp using a 36 cm capillary and POP-4™.
Current limit: 15 amps.
Voltage limit: electrophoresis voltage - up to 20 kV.
Laser features: Long-life single-line 505 nm solid-state laser.
Capillary length: 36 cm, 50 cm.
Operating system: Windows 7 (Service Pack 1).
Analysis software: GeneMapper® ID-X software.
External computer: Memory: 16 GB (2x 8GB) 1600 MHz DDR3 Non-ECC. 4th Gen Intel Core I7 Processor- 3.1 GHz Turbo Processor, Operating System: Windows 7 SP1, Hard Drive: 2x 500GB SATA 3.0 Gb/s and 8 MB Data Burst Cache.
1 61 61 72 110 - 120
Compatible polymers: POP-4™, POP-7™, POP-6™.
Includes:
∗ Capillary Electrophoresis instrument.
∗ 3500 (8-capillary) array and POP™ polymer.
∗ Human identification reagents and other consumables for system qualification.
∗ Computer workstation with flat-screen monitor.
∗ Integrated software for instrument control, data collection, quality control, and basecalling or sizing of samples.
Digital Dry Baths/Block Heaters
Temperature range (metric): ambient +5° to 130°C or superior (ambient at 25°C).
Time to max. temperature: ≤ 20 min. 30° to 130°C.
Timer range: 0 to 99:59 min or continuous.
Temperature uniformity: ≤ ± 1°C.
Temperature accuracy: ≤ ± 0.5°C.
∗ 2 units blocks for digital dry baths/block heaters.
Product type: 1.5 ml tube.
∗ 2 units blocks for digital dry baths/block heaters.
Product type: 2.0 ml tube.
1 20 - 30 25 - 35 10 - 15 110 - 120
Humidity and Temperature Datalogger
Temperature range: -40 to 158°F / -40 to 70°C.
Resolution: 0.1°F/°C.
Humidity range: 0 to 100% RH. Resolution: 0.1% RH.
Datalogging interval: 1 second or 2 seconds to 24 hours.
Memory: Temperature: 16,000 points. Relative humidity: 16,000 points.
Includes:
∗ 3.6 V lithium batteries, 2-pack.
∗ 1 - mounting bracket.
∗ 1 - Software.
6 4 - 5 3 - 5 9 - 10 N.A.
Portable Ice Maker & Dispenser
Type: portable ice maker.
Daily ice making capacity: 25 - 30 pound.
Ice shape: bullet.
Installation: freestanding.
Display type: LED.
Full ice basket indicator: yes.
Add water indicator: yes.
Number of ice sizes: 2.
Includes ice scoop: yes.
Automatic overflow protection: yes.
1 25 - 35 25 - 40 35 - 45 110 - 120
Ice storage: yes.
Harris Micro Punch
Includes: Replacement tips and cutting mats.
∗ Polished-steel tips are precision tooled, heat treated, rust resistant, and case hardened.
∗ Self-healing inert cutting mat ensures clean sample cuts and extends the life of the punch.
∗ 2.0 mm size is recommended.
2 N.A. N.A. N.A. N.A.
1.2 mm Electric Punch Kit
Includes:
∗ One punch with 1.2 mm tip and cover cap.
∗ Two additional 1.2 mm micro-punch tips, power supply with multi-plug adapters, 6.6 x 8 inch cutting mat, carry case, optional cord wrist strap.
∗ User guide.
1 N.A. N.A. N.A. N.A.
Real-Time PCR System for Human Identification
Sample capacity: 96 well; 0.2 ml block.
Dynamic range: 9 or more logarithmic units.
Dye compatibility: minimum FAM™/SYBR green, VIC™/ JOE™, NED™/ TAMRA™/Cy3™, ROX™/ Texas red® and Cy5™.
1 25 - 35 45 - 55 40 - 49 110 - 120
Peak block ramp rate: minimum 2.5°C⁄sec.
Sensitivity: 1 copy.
Temperature uniformity: 0.4 - 0.5°C.
Temperature accuracy: 0.25°C.
Heating/cooling method: Peltier.
Optical system: CCD camera with halogen lamp excitation or Complementary Metal-Oxide Semiconductor (CMOS) camera with bright white LED.
Validated for HID quantification assays.
The package includes one real-time PCR instrument, one tower computer, software.
Centrifuge w/rotor
Max. RCF: 24,000 - 30,130 × g.
Max. Speed: 16,000 - 17,500 rpm.
Max. Capacity: 24 × 1.5/2.0 mL, 6 × 50 mL.
Timer: 99 h : 59 min, with continuous run function.
Noise level: < 61 dB(A) with rotor F-45-30-11.
Control system: microprocessor.
Drive system: direct, brushless induction drive.
Acceleration / Deceleration rates: 2 (standard and soft).
Centrifugation chamber: brushed stainless steel.
1 33 - 37 40 - 48 25 - 35 110 - 120
∗ 1 units conical tube fixed angle rotor with lid: 6 x 15/50mL and 15 ml conical adapters.
∗ 1 units conical tube fixed angle rotor with lid: 24 x 1.5/2.0 mL and 1.5 & 2 ml conical adapters, pack of 24.
∗ 0.5/0.6 mL microliter tube adapters, pack of 24 or more.
Centrifuge w/rotor
Max. RCF: 24,000 - 30,130 × g.
Max. Speed: 16,000 - 17,500 rpm.
Max. Capacity: 2 o 4 × microplate.
Timer: 99h:59min, with continuous run function.
Noise level: < 61 dB(A) with rotor F-45-30-11.
Control system: microprocessor.
Drive system: direct, brushless induction drive.
Centrifugation chamber: brushed stainless steel.
Includes:
∗ Microplate swinging bucket rotor with 2 buckets incl. wind shield and rotor lid (2 o 4 microplates).
1 33 - 37 40 - 48 25 - 35 110 - 120 volts, 60 Hz
Hotplate Stirrer
Top plate diameter range: 140 – 190 mm.
Speed: 30 to 2,000 rpm.
Temperature range: 50° to 350°c.
1 16 - 21 25 - 30 10 - 20 110 - 120
Body material: aluminum.
Top plate material: ceramic-coated aluminum.
Includes:
∗ 2 units stir bars. Accessories hotplate stirrer: diameter: 26 mm; length: 50 mm; type: magnetic stir bar.
∗ 2 units stir bars. Accessories hotplate stirrer: diameter: 17 mm; length: 30 mm; type: magnetic stir bar.
Liquid Nitrogen Transfer Vessels
Capacity range: 10 - 20 L.
Diameter range: 25 - 40 cm.
Diameter inside neck: 5.0 - 6.0 cm.
Height: 50 - 70 cm.
Type: LN2 Transfer Vessel.
Includes:
∗ 2 units withdrawal device, 8L/min.
Description: LN2 withdrawal device, 8L/min Type: LN2 Withdrawal Device; For Use with (Equipment):
LN2 Transfer Vessels.
2 N.A. N.A. N.A. N.A.
Thermal Cycler 3 x 32-well
Block format: interchangeable.
Capacity: 3 x 32-well x 0.2 ml tubes.
Maximum dimensions: 33 cm (13 in) (W) x 27 cm (11 in) (H) x 57 cm (22 in) (D).
Display interface: touchscreen (8.4 in. TFT LCD).
Format: 0.2 ml tubes, 12-strip wells.
Instrument memory: USB and on-board.
Peak block ramp rate: 6.0°C⁄sec.
Program features: auto re-start (after power outages), program overwrite protection.
Reaction speed: fast, standard.
Reaction volume range: 10 - 80 µl.
Sample ramp rate: ± 4.4°C⁄sec.
Temperature accuracy: ± 0.25°C (35°C to 99.9°C).
Temperature range (metric): 0 to 100.0°C.
Temperature uniformity: < 0.5°C (20 sec after reaching 95°C).
2 33 57 27 volts, 60 Hz
Maxwell FSC Instrument.
Processing time: 30 min.
Number of samples: Up to 16.
Weight: 11Kg.
1 30 - 51 30 - 56 25 - 58
Include:
* barcode reader integrated.
* 1. Tablet PC preloaded with the FSC Software application.
* 1 USB cable for connection of the FSC to the Tablet PC.
* 1 Power cable for the FSC instrument.
* 1 Power cable for tablet PC.
* 1 sample concentrator cover tray.
* 1 UV lamp (installed).
* 1 Tablet PC support.
* 1 Quick Start Guide.
* 1 User's manual.
* 1 Package x 50 CW spin baskets.
* 1 Pack x 50 CW microfuge tubes 1.5ml.
* 1 Kit FSC DNA IQ Casework.
Reference: AS1550.
* 48 FSC Cartridges.
* 50 FSC Plungers.
* 50 Elution Tubes, 0.5ml.
* 20ml Elution Buffer.
* 32ml Lysis Buffer.
* 1 Casework Kit Extraction Kit. Reference: DC6745.
* 50ml Casework Extraction Buffer.
* 2 × 10mg Proteinase K.
* 900μl 1-Thioglycerol.
* 1.25ml Nuclease-Free Water.
* Installation and operation instruction.
Oscillating Saw
∗ Battery oscillating saw: universal use for bone osteotomy.
∗ Offers variable speed from 0 to 15,000 rpm.
∗ Depending on the adapter.
∗ Material in Titanium.
∗ Maximum force 250 W.
∗ Weight 1.6 kg with built-in rechargeable batteries, adjustment of saw blades by quick coupling without keys.
∗ It is possible to rotate the head of the saw 360 degrees, steam sterilizable 134 degrees.
∗ 2 bars for 25 minutes.
∗ 2 units rechargeable battery.
NiMH battery with charging capacity of 1950, without electrodes, weight 400 g, voltage 9.6 V, material aluminum and plastic, non-sterilizable.
∗ 1 unit battery charger.
Electronic controlled charger, for 4 battery stations, robust
1 N.A. N.A. N.A.
110 - 120 design in plastic material, weight 3.5 kg, voltage range from 110 to 240 vac, side plug.
∗ 1 unit cable to connect to the network.
∗ 20 units oscillating saw blade, AESCULAP connection. 50 mm long. Reference: GC219R. Dimension 50 mm, 30 mm,
0.7 mm, 0.9 mm.
∗ 4 units oscillating saw blade.
Reference GC681R. Sterile, reusable, 75 mm wide, 0.7 mm thick, working length 35 mm, in stainless steel. Dimension 35 mm, 75 mm, 0.7 mm, 0.9 mm.
Single-channel Pipette, variable, (0.1- 2.5 µL o 0.1 - 2 µL)
Features:
∗ Single-channel plunger pipette variable volume from 0.1 to
2.5 μl.
∗ With volume selector and independent tip ejector, where the volume selector arrangement is such that it prevents the accidental change of the selected volume or has a protection system against volume change.
∗ Plunger stroke of 13 mm or less, ergonomic design.
∗ 4-digit volume indication.
∗ Maximum allowed error that complies with ISO 8655-2:
2002.
∗ Complete pipette sterilizable in autoclave and using conventional tips not exclusive of the brand.
6 N.A. N.A. N.A. N.A.
∗ Guarantee of supply of spare parts and consumables (seals, pistons, etc.), repairable, provision and supply of spare parts easily achieved in the market.
Single-channel Pipette, variable, (0.5 – 10 µL)
Features:
∗ Single-channel plunger pipette variable volume from 0.5 to 10 μl.
∗ With volume selector and independent tip ejector, where the volume selector arrangement is such that it prevents the accidental change of the selected volume or has a protection system against volume change.
∗ Plunger stroke of 13 mm or less, ergonomic design.
∗ 4-digit volume indication.
∗ Maximum allowed error that complies with ISO 8655-2:
2002.
∗ Complete pipette sterilizable in autoclave and using
Single-channel Pipette, variable, (2 – 20 µL)
Features:
∗ Single-channel plunger pipette variable volume from 2 to 20 μl.
∗ With volume selector and independent tip ejector, where the volume selector arrangement is such that it prevents the accidental change of the selected volume or has a protection system against volume change.
∗ Plunger stroke of 13 mm or less, ergonomic design.
∗ 4-digit volume indication.
∗ Maximum allowed error that complies with ISO 8655-2:
2002.
∗ Complete pipette sterilizable in autoclave and using conventional tips not exclusive of the brand.
Brand: Eppendor, Rainin or Pipetman (Gilson)
Single-channel Pipette, variable, (10 – 100 µL)
Features:
∗ Single-channel plunger pipette variable volume from 10 to 100 μl.
∗ With volume selector and independent tip ejector, where the volume selector arrangement is such that it prevents the accidental change of the selected volume or has a protection system against volume change.
∗ Plunger stroke of 13 mm or less, ergonomic design.
∗ 4-digit volume indication.
∗ Maximum allowed error that complies with ISO 8655-2:
2002.
∗ Complete pipette sterilizable in autoclave and using
Single-channel Pipette, variable, (20 – 200 µL)
Features:
∗ Single-channel plunger pipette variable volume from 20 to 200 μl.
∗ With volume selector and independent tip ejector, where the volume selector arrangement is such that it prevents the accidental change of the selected volume or has a protection system against volume change.
10 N.A. N.A. N.A. N.A.
∗ Plunger stroke of 13 mm or less, ergonomic design.
∗ 4-digit volume indication.
∗ Maximum allowed error that complies with ISO 8655-2:
2002.
∗ Complete pipette sterilizable in autoclave and using conventional tips not exclusive of the brand.
Brand: Eppendor, Rainin or Pipetman (Gilson)
Single-channel Pipette, variable, (100 – 1000 µL)
Features:
∗ Single-channel plunger pipette variable volume from 100 to 1000 μl.
∗ With volume selector and independent tip ejector, where the volume selector arrangement is such that it prevents the accidental change of the selected volume or has a protection system against volume change.
∗ Plunger stroke of 13 mm or less, ergonomic design.
∗ 4-digit volume indication.
∗ Maximum allowed error that complies with ISO 8655-2:
2002.
∗ Complete pipette sterilizable in autoclave and using conventional tips not exclusive of the brand.
Carousel Adapter of Pipettes or Pipette Stand
Features:
∗ Pipette carousel for 6 pipettes and/or other manual liquid handling instruments (minimum).
∗ Adapter suitable for repeater.
Compatible with items 19 to 24.
6 N.A. N.A. N.A. N.A.
Combination Lab Refrigerator/Freezer
Capacity: 2 x 10 cu. ft. (2 x 283.1L).
Temperature range (refrigerator): 1° to 12°C.
Temperature range (freezer): -20° to 0°C.
Chamber material: aluminum.
Defrost: automatic.
Cabinet: high-quality steel.
Shelves: four epoxy-coated (two per chamber).
Doors: double, solid stainless-steel half doors.
Performance:
∗ Dual refrigeration system.
∗ Digital temperature controller for each chamber with intuitive interface.
∗ Temperature stability of ± 3°C.
∗ High-density, fluorine-free insulation.
2 65-75 65-90 180-210 110-120
∗ Automatic defrost.
Security:
∗ Keyed on/off switch.
∗ High/low temperature; audio/visual alarms.
∗ Easily accessible contacts for backup monitoring.
∗ Door ajar alarms.
∗ Lockable controller settings.
Quality:
∗ High-quality steel exterior front and aluminum interior.
∗ Casters for easy mobility.
∗ Quiet refrigeration system.
∗ cUL listed.
Voltage Regulator 3 KVA
Output capacity (FP = 0.7): 3,000 VA.
Output capacity (actual): 2,100 VA.
Regulation: ± 5% typical, ± 8% maximum.
Number of regulation steps: four (4).
Waveform: 100% sinusoidal (according to network).
Harmonic distortion: < 1% (according to the voltage THD of the electrical network).
Regulation with variable load: < 3.5% (0 to 100% nominal load).
2 20-25 20-30 45-50 110-120
Input connection: terminal.
Connection of outputs: terminal board and 1 tripolar socket.
Efficiency: > 96% at full load.
Operating temperature: 0°C to 45°C.
Quality assurance: certificate No. HK05 / 01579.
pH, ORP, ISE, Temperature Laboratory Bench Meter
Range: pH -2.00 to 16.00 pH; mV ±699.9 mV / ±2000 mV; ISE
0.001 to 19999 ppm.
Temp: -5.0 to 105.0°C.
Resolution: pH 0.01 pH; mV 0.1 mV / 1 mV; ISE 0.001 (0.001 to 9.999) ppm; 0.01 (10.00 to 99.99) ppm; 0.1 (100.0 to 999.9) ppm; 1 (1000 to 19999) ppm.
Temp: 0.1°C.
Accuracy: pH ±0.01 pH; mV ±0.2 mV / ±1 mV; ISE ±0.5% Full Scale; Temp ±0.4°C / ±0.8°F. Rel mV offset: ±2000 mV.
pH calibration: 1 or 2 point-calibration, with 7 memorized buffers.
ISE calibration: 1 or 2 point calibration, 6 standard solutions available.
Temperature compensation: automatic, from -5.0 to 105.0°C or manual, without temperature probe.
Input impedance: 1012 Ohm.
Slope calibration: from 80 to 108%.
Includes:
∗ pH Electrode Double Junction Refillable pH Electrode.
1 23-25 16-18 9,5 - 11 110-120
Temperature probe power supply: 12 VDC power adapter.
Electrode Holder
∗ 4.01 pH buffer solution.
∗ 7.01 pH buffer solution.
∗ 10.01 pH buffer solution.
∗ Electrode cleaning solution.
∗ Application software.
∗ RS232 connection cable with 2 M cable.
Microwave Oven
Capacity: 2.00 cu. ft.
Cavity finish: powder coat.
Cooking power: 1000 w.
Turntable diameter size: 12.0 in or higher details.
CFM motor class: 400.0.
Clock: yes.
Night light: yes.
Number of power levels: 10.
Timer: yes.
Features:
∗ CleanRelease® non-stick interior.
1 75,8 - 80,8 40 - 45 43,5 - 48,5 110-120
∗ 4-Speed, 400 CFM vent fan.
∗ Hidden vent.
LP Vortex Mixer
Speed: 0 to 3,000 rpm.
Load bearing capacity: 0.5 kg (1.1 lb.).
Shipping weight: 6.8 lb.
Shipping weight (metric): 3.1 kg.
Operating modes: continuous / touch.
Plug type: USA.
Includes:
∗ 6 units accessories for the LP vortex mixer.
∗ Description: rubber platform.
∗ Type: rubber platform.
∗ For use with (equipment): vortex mixer.
6 154 - 159 210 - 215 83 - 85 110-120 60 Hz
Acid Base Cabinet
Standing surface material: polyethylene.
Collection volume [l]: 18 (4.76 - gal).
Door design: wing door.
Further TÜV testing (1): DIN EN 14727.
Declaration of conformity: CE.
Material outer carcass: decor panels, with melamine resin coating.
1 56 - 60 66 - 70 63,5 - 70 N.A.
Outer carcass color: light grey RAL 7035.
Inner carcass color: light grey RAL 7035.
Door color: light grey RAL 7035.
Base color: slate grey RAL 7015.
Standing surface material: polyethylene.
Load-bearing capacity per standing surface with uniformly distributed load [kg]: 40 (88.18 lbs).
Number of storage shelves: 2.
Exhaust air connection included: NW 75.
Door technology: classic.
Door hinge: right.
Cabinet Flammable Liquids
Product group: safety storage cabinet.
Fire resistance (FR): type 90.
Model size: M-5.
Interior: 1 pull-out tray, automatically extending 1 perforated sheet insert.
Standing surface material: sheet steel.
Powder-coated, collection volume [lt]: 13.9 (3.43 gal).
Door design: wing door.
Door technology: classical.
Adjustable feet for level equalization: no.
Type-tested by recognized test institutes according to: DIN EN 14470-1.
1 88,8 - 90 49,3 - 50 63,1 - 65 N.A.
TÜV testing according to: DIN EN 14470-1, EK5/AK4 09-10.
Further TÜV testing (1): DIN EN 14727.
Further TÜV testing (2): TRGS 510 annex 3.
Mark of approval by TÜV: GS mark.
Added value mark of approval by TÜV: High Quality.
Declaration of conformity: CE.
Material outer carcass: sheet steel, powder-coated.
Material inner carcass: decor panels.
Outer carcass color: light grey RAL 7035.
Inner carcass color: light grey RAL 7035.
Door color: zinc yellow RAL 1018.
Base color: slate grey RAL 7015.
Earth connection according to TRGS 727 available: rear of the cabinet.
Number of earth connections: 2.
Standing surface material: sheet steel.
Powder-coated, collection volume (lt): 13 (3.43 gal).
Max. volume of the largest individual container [lt]: 12 (3.17 gal).
Load-bearing capacity per standing surface with uniformly distributed load [kg]: 40 (88.18 lbs.).
Adjustment height 1 [mm]: 340 (13.39 inch).
Base version: classical, with slots and floor extraction.
Storage shelf interior: automatically extending.
Number of storage shelves: 1.
Perforated sheet interior: perforated sheet insert.
Number of perforated sheet inserts: 1.
Exhaust air connection included: NW 75, doors can be opened with minimum effort.
Design: leaf doors.
Door technology: classic.
Plated Diamond Cutting Disc, 2". Reference: A-DD9
Trade Mark: Foredom Electric Co.
Description: 2" diameter plated diamond cutting disc recommended for use with A-M14 mandrel. 20,000 RPM max speed.
Units: 25.
25 N.A. N.A. N.A. N.A.
Mandrel, Screw type, 1/8" Shank. Reference: A-M14
Trade Mark: Foredom Electric Co.
Units: 10.
10 N.A. N.A. N.A. N.A.
Metal Mesh Gloves
Specifications:
∗ The gloves are constructed from individually-welded 4.2mm rings of AISI 304L stainless steel.
∗ The gloves to be comply with European specification
EN1082.
∗ Shaped, ergonomic design for comfort and fit.
∗ Stand up to cleaning chemicals and steam cleaning.
∗ Have excellent bacteria resistance.
∗ Have excellent resistance against oils, fat, blood, diluted leaches and acids.
∗ Stay flexible in very cold and very hot temperatures (up to
110° Celsius/ 230° Fahrenheit).
∗ Are recyclable.
Total units = 10 ∗ 2 units size S.
∗ 6 units size M.
∗ 2 units size L.
Curved Jaw Locking Pliers with Wire Cutter
Description:
∗ Jaw capacity: 15/16" / 24 mm.
∗ Size: 4" / 100 mm.
Features:
∗ Curved jaw with built-in wire cutter good for a variety of applications and material shapes.
∗ Classic trigger release for maximum locking force.
∗ Wire cutter included.
∗ Constructed of high-grade heat-treated alloy steel.
∗ Teeth are designed to grip from any angle.
2 N.A. N.A. N.A. N.A.
∗ Hex key adjusting screw tightens to adjust pressure and draw materials together and provides a controlled release.
Refrigerated Centrifuge
Temperature range: -10° to +40°C.
Capacity: 4 x 1000 mL.
Max. RCF: 25,314 x G.
Max. Speed: 15,200 rpm.
Refrigerated: yes.
Control: microprocessor.
Display: digital.
Memory: 99 programs (5 with direct access key).
Motor type: direct, brushless induction, low-profile.
Noise level: < 61dBA. Timer: 99 hr : 59min plus HOLD.
Includes:
∗ TX-750 4 + 750 mL swinging bucket rotor. Description:
length (metric) tube: 139 mm; radius (metric): 195 mm (max); angle: 90°; capacity: 4 x 750 mL; diameter (metric) tube: 98 mm.
∗ TX-750 4 + 750 mL swinging bucket rotor. Description: 750 mL round buckets for TX-750 rotor (set of 4).
∗ TX-750 4 + 750 mL swinging bucket rotor. Description:
clickseal biocontainment lids for round buckets (set of 4).
∗ TX-750 round bucket adapters. Description: adapters for 15 mL conical tube (set of 4).
1 40 - 57 65 - 70 35 - 85 110-120
∗ TX-750 round bucket adapters. Description: adapters for 50 mL conical tube (set of 4).
∗ Installation.
Double Rotisserie Incubator without Internal Electrical Outlet
Specifications:
∗ Double chambers ∗ Temperature range: ambient +5 to 80°C.
∗ Temperature accuracy: ± 0.5°C.
∗ Speed range: 0 to 40 rpm.
∗ 4 units. 16 vessel (50 ml) rotisserie, 50 mL conical tubes, vertical mount.
1 40 - 50 40 - 50 70 - 110 110-120 60 Hz
Qiacube. Reference: Cat No./ID: 9001882
Trade mark: QIAGEN.
Samples per run; throughput: 1–12 samples per run.
Protocols/main application on this instrument: DNA purification, RNA purification, protein purification.
Kits designed for this instrument: most QIAGEN manual kits (e.g., QIAamp, QIAquick, QIAprep).
Technology: depends on kit used.
Instrument dimensions: 650 x 620 x 570 mm.
1 65 62 57 110-120
Technical data of the instrument: 70.5 kg.
Processing: automated.
Centrifuge: 12,000 x g max, swing-out rotor with 12 positions, max 45°.
Shaker: speed 100-2000 rpm, amplitude 2 mm, heating range of room temperature to 70°C, ramp-up time of < 5 min from room temperature to 55°C ± 3°C.
Pipetting system: syringe size 1 ml, pipetting range 5-900 μl.
Touchscreen: transmissive TFT, 640 x 860 x 65 mm, white LED backlight, high brightness.
Software: QIAGEN protocols are preinstalled on the QIAcube or can be downloaded at www.qiagen.com/MyQIAcube.
Power: 100-120 V AC, 50-60 Hz, 650 VA.
Operating temperature: 18 - 28°C.
Humidity: Relative humidity of 10 - 75% (noncondensing).
Altitude: Up to 2000 m.
Place of operation: For indoor use only.
Storage temperature: 15°C to 30°C in manufacturer's package.
Transportation temperature: -25°C to 50°C in manufacturer's package.
Pollution level: 2.
Overvoltage category: II.
∗ Starter pack, QIAcube Ref: 990395.
http://www.qiagen.com/MyQIAcube
REAGENTS AND SUPPLIES:
No. ITEMS Quantity
N-Lauroylsarcosine Sodium SALT: reagent powder.
∗ Bioultra molecular biology grade ≥ 99% (HPLC).
∗ Case: 137-16-6. [] 99.00%. Minimum expiration time of 4 years from the delivery in the laboratories of the CTI, bottle per 25 grams.
41 EDTA, Disodium Salt (Dihydrate), molecular biology grade x 500 g. 6
42 Sodium Hydroxide. 2
43 FTA purification reagent (500mL). 1
44 DNA IQ ™ Casework Pro Kit for Maxwell FSC Instrument (minimum 48 samples). 2
45 Proteinase K (Lyophilized) 100 mg. 100
PCR kit for coamplification in a single reaction of minimum twenty-two (22).
∗ STR's systems, an indel and amelogenin, for application in human identification.
Including the markers D3S1358, vWA, D16S539, CSF1P0, TPOX, D8S1179, D21S11, D18S51 , DYS391, D2S441, D19S433, TH01, FGA, D22S1045, D5S818, D13S317, D7S820, SE33, D10S1248, D1S1656, D12S391, D2S1338, and INDEL.
∗ Amelogenin, which have been implemented for use in highly degraded DNA samples and / or contain inhibitors, labeled with fluorochromes for use in Genetic Analyzers 3130 XL and 3500 XL that include the reaction mixture containing taq.
∗ Primer mixture, DNA control, allelic ladder.
∗ Kit for 200 reactions.
∗ Reference: Globalfiler.
47 LIZ 600 v 2. 4
48 DS-36 Matrix Standard (Dye Set J6). 1
49 Powerplex Y 23. 1
Human DNA quantification kit by real-time PCR, compatible with Real-Time PCR System.
∗ Kit for 400 reactions.
∗ Internal control in each reaction to identify samples that may contain PCR inhibitors.
∗ Reference: Quantifiler Human.
51 GA 3500s, Performance Optimized Polymer POP4; (960 iny); Applied Biosystems. 4
52 GA 3500s, Cathode Buffer Container - CBC; (4 pk); Applied Biosystems. 2
53 GA 3500s, Anode Buffer Container - ABC; (4 pk); Applied Biosystems. 2
PCR kit for coamplification of twenty three (23) loci STR's and amelogenin, for application in human identification.
∗ Including: D1S1656, D10S1248, D22S1045, D12S391, D2S441, D18S51, D21S11, TH01, D3S1358, D16S539, D2S1338, FGA, D8S1179 , vWA, D19S433, D13S317, Penta E, Penta D, CSF1PO, D7S820, D5S818, TPOX, DYS391 and amelogenin, which have been implemented for use in highly degraded DNA samples and / or containing inhibitors.
∗ Labeled with fluorochromes for use in ABI PRISM 3130, 3130 XL and 3500 XL.
∗ Containing: reaction mixture with included Taq DNA polymerase, control DNA, allelic ladder and size standard.
∗ That includes buffer that allows the direct amplification of samples in FTA card without previous purification of the card.
∗ Reference PowerPlex® Fusion kit for 800 reactions, DC2408
PCR kit for seventeen (17) STR's and amelogenin systems for human identification applications.
∗ Including the systems D1S1656, D10S1248, D22S1045, D12S391 and D2S441 with amplicon length less than
200 nucleotides and additionally the systems D18S51, D21S11 , TH01, D3S1358, D16S539, D2S1338, FGA, D8S1179, vWA, SE33, D19S433, and amelogenin, which have been implemented for use in highly degraded DNA samples and / or containing.
∗ Inhibitors, labeled with fluorochromes for use in ABI PRISM 3130 And 3130 XL.
∗ Containing: reaction mixture with Taq DNA polymerase included, DNA control, allelic ladder and size standard.
∗ That includes buffer that allows the direct amplification of samples in FTA card without previous purification of the card.
∗ Reference PowerPlex® ESX17 Kit for 400 reactions, DC6720.
PCR kit for coamplification of seven (7) STR's loci, for application in human identification.
∗ Including: LPL, F13B, FESFPS, F13A01, Penta D, Penta C and Penta E, marked with fluorochromes for use in
ABI PRISM 3130 , 3130 XL and 3500 XL.
∗ Containing: reaction mixture with included Taq DNA polymerase, control DNA, allelic ladder and size standard.
∗ Kit for 100 reactions.
57 Matrix 4 colors DG 4800. 1
58 Matrix 5 colors DG 4850. 1
59 QIAquick PCR Purification Kit for 250 reactions. Catalog qiagen 28106. 6
60 Absolute ethanol box of 4 bottles of 4 liters. 1
61 Buffer PB catalog Qiagen 19066. 6
62 Buffer PE catalog Qiagen 19065. 6
63 Magnetic separation rack 0.5 ml (for twelve tubes) 2
For 3500 Genetic Analyzer and Real-Time PCR System for Human Identification – Includes service, installation, and set-up of the system in its place of operation and start kit of reagents and sufficient supplies for installation and training.
OPERATIONAL ASPECTS:
The automated genetic analyzer equipment should be offered with:
a. All system modules, equipment and software must correspond to the most recent version available in the market.
b. It must be delivered accompanied with their respective catalogs and operating manuals in English and Spanish
c. .
d. The training or demonstration offered must be given in Spanish to at least 4 genetic experts identified by INL. This must include the handling of the equipment and each of the analysis software, maintenance procedures and all those parameters that are needed for its proper use.
e. The duration of manufacturer or vendor support in terms of labor, parts and supplies must be at least five (5) years from the date of start-up of the equipment at the CTI Centro de Idenficacion Humana in Medellin, Colombia.
f. The equipment must be new, not remanufactured, original, with all the elements and accessories; duly nationalized and guaranteed by the manufacturer, must have a minimum warranty of one (1) year for defects in operation and manufacturing from the date of delivery to satisfaction. During this period, preventive maintenance service will be provided at least once a year during the warranty period, additionally repair or replace free of charge the defective parts of the equipment, service limited by any of the following limitations or exclusions:
Products that have suffered from misuse, abuse, negligence, accident or which have undergone modifications in serial number, name, function or application.
Defects or damages caused directly or indirectly by installation or maintenance by personnel not authorized by the supplier and / or by use of unauthorized parts.
Connection to electrical networks without an appropriate voltage regulator.
Connection to electrical networks without pole connection to ground.
Connection to electrical networks without an appropriate voltage regulator.
During the guarantee period, and subsequently, if corrective maintenance is required, the response time must not exceed 72 working hours from the receipt of the written request by the user for the review and diagnosis of the damage reported and the resolution, repair and / or replacement of spare parts, if applicable, will be 20 business days from the date of diagnosis.
2.2. Installation:
The Vendor must estimate and include the costs of transportation and installation of the equipment in Colombia within this purchase, at following address: Centro de Identificación Humana – CIH, Subdirección Seccional de Policía Judicial CTI Medellín - Attorney General Office Calle 80 No. 65 – 154 Piso 3 Barrio Córdoba, Medellín, Antioquia (Colombia). Installation includes setup of the equipments, technical support and transportation of equipment.
2.3. Training:
The Vendor must estimate and include the costs of proper training for the Genetics technical staff of the CTI Centro de Identificación Humana – CIH in Medellin, Colombia within this purchase, in following address: Centro de Identificación Humana – CIH, Subdirección Seccional de Policía Judicial CTI Medellín - Fiscalía General de la Nación Calle 80 No. 65 – 154 Piso 3 Barrio Córdoba, Medellín, Antioquia (Colombia). Training includes: technical support by a fluent Spanish speaker; the agenda and schedule for training; air and ground transportation, lodging and meals for the technical expert(s) from company awarded to the place of the training (likely Medellín).
The training on the use of the equipment should be provided to the designated, three to six, Genetics technical staff members of the CTI Human Identification Laboratory in Medellín located at Calle 80 No. 65 - 154 Piso 3, Barrio Córdoba, Medellín, Colombia. The training provided for each equipment must be for a minimum of 8 hours and is necessary for the equipment that contains a special operating software, including:
3500 Genetic Analyzer for Human Identification or similar equipment selected.
Real-Time PCR System for Human Identification.
Thermal cycler 3 x 32-well.
Qiacube.
This training should be coordinated in advance with the Contracting Officer Representative (COR).
2.4. Maintenance:
2.4.1. Maintenance for equipment: Vendor must estimate and include the costs of preventive maintenance for two (2) years in this purchase including supplies and technical material required during each visit. A minimum of (1) one visit per year is required as well a tentative schedule for those visits. Technical support by (a) fluent Spanish speaker(s) is required; air and ground transportation, lodging and meals for the technical expert(s) support from company awarded to the place of the maintenance needs to be included.
2.4.2. Corrective Maintenance for equipment: Vendor must estimate and include the costs of corrective maintenance for one year in this purchase, including supplies, spares and technical material required on each visit. Technical by a fluent Spanish speaker is required, air and ground transportation, lodging and meals for the technical expert support from company awarded in the place of the maintenance needs to be included.
Item No. Description Quantity Installation
Included? Start-up? Training Included?
Required In-Country Warranty
Period
Manufacturer- Recommended
Preventive Maintenance
Included?
1 Class II Type A2 Biological Safety Cabinet. 2 Yes Yes No At least two
(2) years.
Yes; for at least two (2) years after delivery.
2 Polymerase Chain Reaction Cabinet. 1 Yes Yes No At least two
(2) years.
Yes; for at least two (2) years after delivery.
3 Ductless Fume Hood. 2 units Carbon filters. 1 Yes Yes No At least two
(2) years.
Yes; for at least two (2) years after delivery.
4 Ductless Fume Hood. 4 units Carbon filters. 1 Yes Yes No At least two
(2) years.
Yes; for at least two (2) years after delivery.
5 3500 Genetic Analyzer for Human Identification. 1 Yes Yes Yes; for 3-6 officials.
At least two
(2) years.
Yes; for at least two (2) years after delivery.
6 Digital Dry Baths/Block Heaters. 1 Yes Yes No At least two
(2) years.
Yes; for at least two (2) years after delivery.
7 Humidity and Temperature Datalogger. 6 No No No At least one
(1) year.
Yes; for at least one (1) year after delivery.
8 Portable Ice Maker & Dispenser. 1 Yes Yes No At least one
(1) year.
Yes; for at least one (1) year after delivery.
9 Harris Micro Punch. 2 No No No No N.A.
10 1.2 mm Electric Punch Kit. 1 No No No At least one
(1) year. No
11 Real-Time PCR System for Human Identification. 1 Yes Yes Yes; for 3-6 officials.
At least two
(2) years.
Yes; for at least two (2) years after delivery.
12 Centrifuge w/rotor. 1 Yes Yes No At least two
(2) years.
Yes; for at least two (2) years after delivery.
13 Centrifuge w/rotor. 1 Yes Yes No At least two
(2) years.
Yes; for at least two (2) years after delivery.
14 Hotplate Stirrer. 1 Yes Yes No At least one
(1) year.
Yes; for at least one (1) year after delivery.
15 Liquid Nitrogen Transfer Vessels. 2 No No No At least one
(1) year.
Yes; for at least one (1) year after delivery.
16 Thermal Cycler 3 x 32-well. 2 Yes Yes Yes; for 3-6 officials.
At least two
(2) years.
Yes; for at least two (2) years after delivery.
17 Maxwell FSC Instrument. 1 Yes Yes No At least two
(2) years.
Yes; for at least two (2) years after delivery.
18 Oscillating Saw. 1 No No No At least one
(1) year.
Yes; for at least one (1) year after delivery.
19 Single-channel Pipette, variable, (0.1 –
2.5 µL o 0.1 – 2 µL). 6 No No No At least one
(1) year.
Yes; for at least one (1) year after delivery.
20 Single-channel Pipette, variable, (0.5 – 10 µL). 6 No No No At least one
(1) year.
Yes; for at least one (1) year after delivery.
21 Single-channel Pipette, variable, (2 – 20 µL). 6 No No No At least one
(1) year.
Yes; for at least one (1) year after delivery.
22 Single-channel Pipette, variable, (10 – 100 µL). 6 No No No At least one
(1) year.
Yes; for at least one (1) year after delivery.
23 Single-channel Pipette, variable, (20 – 200 µL). 10 No No No At least one
(1) year.
Yes; for at least one (1) year after delivery.
24 Single-channel Pipette, variable, (100 – 1000 µL). 10 No No No At least one
(1) year.
Yes; for at least one (1) year after delivery.
25 Carousel Adapter of Pipettes or Pipette Stand. 6 No No No No No
26 Combination Lab Refrigerator/Freezer. 2 Yes Yes No At least two
(2) years.
27 Voltage Regulator 3 KVA. 2 No No No At least one
(1) year. No
28 pH, ORP, ISE, Temperature Laboratory Bench Meter. 1 Yes Yes No At least one
(1) year.
Yes; for at least one (1) year after delivery.
29 Microwave Oven. 1 No No No At least one
(1) year.
Yes; for at least one (1) year after delivery.
30 LP Vortex Mixer. 6 No No No At least one
(1) year.
Yes; for at least one (1) year after delivery.
31 Acid Base Cabinet. 1 Yes No No At least one
(1) year. No
32 Cabinet Flammable Liquids. 1 Yes No No At least one
(1) year. No
33 Plated Diamond Cutting Disc, 2".
Reference: A-Dd9. 25 No No No No No
34 Mandrel, Screw Type, 1/8" Shank.
Reference: A-M14. 10 No No No No No
35 Metal Mesh Gloves. 10 No No No No No
36 Curved Jaw Locking Pliers with Wire Cutter. 2 No No No No No
37 Refrigerated Centrifuge. 1 Yes Yes No At least two
(2) years.
Yes; for at least two (2) years after delivery.
38 Double Rotisserie Incubator without Internal Electrical Outlet. 1 Yes Yes No At least two
(2) years.
39 Qiacube. Reference: Cat No./Id: 9001882. 1 Yes Yes Yes; for 3-6 officials.
At least two
(2) years.
3. DELIVERY (SHIPPING AND HANDLING INSTRUCTIONS)
3.1. REQUIRED DELIVERY DATE: Selected vendor should have no more than 60 days after the award of the contract to deliver the full order.
3.2. DELIVERY PLACE: Final delivery place is the Centro de Identificación Humana – CIH, Subdirección Seccional de Policía Judicial CTI Medellín - Fiscalía General de la Nación Calle 80 No. 65 – 154 Piso 3 Barrio Córdoba, Medellín, Antioquia (Colombia), all charges for delivery and transportation to the final destination is the responsibility of the Contractor.
The Contractor shall ensure that all items are prepared for safe shipment and include all necessary export approvals, licenses, and paperwork. Price must include delivery to the above-mentioned address and compliance with all applicable export licensing requirements. If an export license is required, the items shall not be shipped until necessary export license is received. Vendor shall replace any defective material or equipment including airline transportation at no cost to the USG for warranty by one year.
Partial deliveries may be considered and approved by the Government.
4. WARRANTY
The Vendor must provide Quality warranty for two years for the equipment. The maintenance of the supplied equipment must be carried out on final destination at Centro de Identificación Humana – CIH, Subdirección Seccional de Policía Judicial CTI Medellín - Fiscalía General de la Nación Calle 80 No. 65 – 154 Piso 3 Barrio Córdoba, Medellín, Antioquia (Colombia). In the maintenance needed requires that the equipment to be removed or exchanged, that equipment must be delivered to the Human Identification Center of Medellín. A quality policy provided by Insurance company is required (Please provides a concise description of the warranty provided).
5. CUSTOMS CLEARANCE AND TAX EXEMPTIONS
The U.S. Government (USG) will assist the Contractor in obtaining customs clearance and tax exemption certificates from the Government of Colombia.
6. INSPECTION AND ACCEPTANCE
FINAL ACCEPTANCE: The USG will perform final acceptance upon delivery at Centro de Identificación Humana – CIH, Subdirección Seccional de Policía Judicial CTI Medellín - Fiscalía General de la Nación Calle 80 No. 65 – 154 Piso 3 Barrio Córdoba, Medellín, Antioquia (Colombia). The USG representative employee will receive and inspect the equipment being procured of this document upon final destination and will monitor the deliverables with appropriate USG Name Tags if required for End Use Monitoring during the period of performance established: 60 days upon final acceptance, the vendor may invoice for the number of items delivered.
INL Point of Contact:
Milena Sanchez de Boado Justice Advisor SanchezdeBoadoM@state.gov
7. PERIOD OF PERFORMANCE (POP)
FOR COMMODITIES: 60 days after the award of the contract, all items must be received and accepted by the Government.
mailto:SanchezdeBoadoM@state.gov
File details come from the government source that posted it.