This document is a salient characteristics specification for a Thermal Gravimetric Analyzer (TGA) system that the USGS Geology, Energy & Minerals Science Center intends to procure and deliver to Reston, VA within 56 days of order receipt. The TGA must be a stand-alone, fully integrated system with robotic autosampler capability for high-throughput analysis, installed dimensions not exceeding 21" W x 25" D x 25" H, and weight limited to 100 lbs. The system must include a built-in gas controller supporting multiple gases (air, argon, carbon dioxide, nitrogen, oxygen, helium) at 1–3 bar pressure with 3 ml/min flow rate tolerance, a water bath cooling system, and expansion options for peripheral gas analysis devices. The vendor must provide available annual service plans and initial delivery, installation, and calibration services.
The furnace must heat to at least 1100°C with 0.5°C temperature precision, accommodate maximum heating rates of at least 250°C/min, include a cooling jacket with built-in thermocouple sensor, and demonstrate capability to analyze common geologic samples such as coals, carbonates, and clay samples. The solicitation (RFQ 140G0126Q0089) issued approximately April 7, 2026, requires written quotes by April 15, 2026, with an anticipated award date of April 17, 2026. The NAICS code is 334516 with a 1,000-employee size standard. Vendors must be registered at sam.gov to be considered, and quotes should be submitted to rlanderson@usgs.gov. This solicitation is issued by the Department of the Interior, US Geological Survey Office of Acquisitions and Grants.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Sol_140G0126Q0089_Amd_0001.pdf | ||
| D01_Attachment_-_52_212-5.pdf | ||
| Sol_140G0126Q0089.pdf | ||
| B08_Attachment_-_Salient_Characteristic.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics:
OVERVIEW TGA SYSTEM
· Must be a stand-alone and totally integrated thermal gravimetric analyzer (TGA) with software support
· Must have a gas controller capable of supporting multiple input gases
· Must come with a robotic autosampler for high throughput analyses
· Must have expansion options to include supporting peripheral devices for gas analysis
· Must have installed dimensions no greater than 21” W, 25”D, 25” H
· Must weigh no more than 100 lbs (44kg) installed
· Must have factory set power supply capable of operation at 115V.
· Must have an available annual service plan option
· Must have available initial delivery, installation, and calibration service
· Must have a demonstrated capability of analyzing common geologic samples such as coals, carbonates, clay samples, etc.
GAS CONTROLLER AND GAS TYPE
· The TGA must have a built-in gas controller, capable of operation at 1-3bar pressure
· Must be tunable/adjustable with instrument software
· The following gases must be approved for use with the TGA and included gas controller: air, argon, carbon dioxide, nitrogen, oxygen, helium
· The controller must have a flow rate tolerance of ±3ml/min
WATER BATH COOLING SYSTEM
· The TGA must come with a water bath cooling system or equivalent to cool the furnace and microbalance during operation to prevent overheating
FURNACE
· The TGA system must come with a furnace of sufficient size to support a variety of material volumes and crucible types
· The furnace must contain a cooling jacket
· Must contain a built-in sensor with a cell thermocouple
· Must be capable of heating to at least 1100°C
· Must have temperature precision of ±0.5°C (K)
· Must have a maximum heating rate of at least 250°C/min
· Must have a cooling time <25 mins
BALANCE
· The TGA system must come with a built-in microbalance
· Must have a measurement precision of ±1 mg
· Must be temperature controlled for improved precision via connected cooling system
· Must be purgeable with non-reactive gas to prevent damage to the microbalance
SAMPLING ROBOT
· The TGA system must come with a sampling robot automate loading of the sample carousel
· The robot must place sample crucibles on a turntable prior to analysis and remove them after analysis is complete
SOFTWARE
· The TGA system must come with software with support
· The software must allow for network connectivity
· The software must allow for automated back-ups of raw instrument data
SUPPORTED PERIPHERAL DEVICES
· The TGA system must be able to be expanded to support peripheral devices that analyze the gas composition including:
· Mass Spectrometer (TGA-MS)
· Fourier Transform Infrared Detector (TGA-FTIR)
· Fourier Transform Infrared Mass Spectrometer (TGA-FTIR/MS)