Questions and Answers Amendment 0002.docx

DOCX document 132 KB Posted

Attached to
Calibration of Pipettes and Scientific Equipment at NBAF in Manhattan, KS
Solicitation number
12805B24Q0301
Issued by

About this file

This document contains Questions and Answers related to Solicitation 12805B24Q0301 for the Calibration of Pipettes and Scientific Equipment at the National Bio and Agro-Defense Facility (NBAF) in Manhattan, KS.

The key details include:

  • The government is open to mail-in service for pipettes, but on-site service is required for other equipment like centrifuges.
  • The government will not provide the model numbers for the Rainin pipettes, but has listed the size ranges.
  • The government will not share the budget or government estimate.
  • The government confirmed that past performance for non-government projects will not impact the proposal evaluation.
  • The government provided the on-site address at the NBAF facility.
  • The government requires a Level 4 calibration for the pipettes, which includes 4 As-Found and 10 As-Left data points.
  • The government provided the required test point details for the various pieces of scientific equipment, including centrifuges, pH meters, balances, and other lab equipment.
  • The government confirmed that ISO 17025 accreditation is only required for the pipette calibration services, not the other equipment.

View the file

Other files

File Type Posted
12805B24Q0301 0002.pdf PDF
Attachment 1 PWS Amended.docx DOCX document
12805B24Q0301 0001.pdf PDF
12805B24Q0301 Combined Synopsis Solicitation.pdf PDF
Attachment 2 Schedule of Services Pricing.xlsx XLSX spreadsheet
Attachment 1 PWS.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Questions and Answers Solicitation 12805B24Q0301

1) We were wondering if your client would consider mail-in service instead of on-site service?

a. We would be open to mail-in service for pipettes, but other equipment would not be able to be mailed (centrifuges), so onsite visitation for calibration of those items would still be required.

2) Could you please provide the model numbers for the pipettes, highlighted below?

Description

Rainin Pipettes Single Channel 17025 Standard / please provide model numbers

Rainin Pipettes Multi Channel (8) 17025 Standard / please provide model numbers

Rainin Pipettes Multi Channel (12) 17025 Standard / please provide model numbers

Make
Model
Size
Rainin (Single Channel)
L-2 XLS
0.1µL - 2µL
Rainin (Single Channel)
L-20 XLS
2µL - 20µL
Rainin (Single Channel)
L-200 XLS
20µL - 200µL
Rainin (Single Channel)
L-1000 XLS
100µL - 1000µL
Rainin (8-Channel)
L-200 XLS
20µL - 200µL
Rainin (8-Channel)
L-1200 XLS
100µL - 1200µL
Rainin (12-Channel)
L-200 XLS
20µL - 200µL
Rainin (12-Channel)
L-1200 XLS
100µL - 1200µL

a.

3) Are you able to share a budget?

a. The budget/government estimate will not be shared.

4) Per the PWS under Qualifications, it states:

The contractor must provide written or electronic documentation showing the specialists performing work under this PWS are individuals that have certified training and/or documented experience in calibration of the pipettes and other identified equipment. The contractor must provide proof of familiarity that qualifies the specialist to perform the calibration procedures on pipettes to the ISO 17025 standard or related standards/expertise for other identified equipment.

Should this condition be met, would the evaluation of the proposal be impacted if:

Past Performance documentation- is NOT necessarily for a government project?

a. The evaluation of the proposal will not be impacted by past performance documentation listing services being provided to non-government institutions/industries/businesses/etc.

5) The onsite address for the job?

a. National Bio and Agro-Defense Facility 1980 Denison Ave. Manhattan, KS 66502 (Amended PWS page 3, identified in red font)

6) I see the requested ISO-17025 accredited but please confirm if they want a level 3 which is 4 As-found and As-left as data points or a level 4, which is 4 As-found and 10 As-left data points.

The level of calibration you would like quoted:

a. Level 4

7) Please provide the test point requirements for the Corning 6770.

a. Temperature, speed, timer, imbalance detection, brake function, safety lid lock test, and other tests identified by manufacturer in the user manual or contractor determines necessary/beneficial/equivalent for successful/safe general laboratory functionality.

8) Please provide the test point requirements for the Ultra Beckman Coulter Optima XE-100.

a. Temperature, speed, timer, imbalance detection, vacuum system check, rotor ID system check, brake function, safety lid lock test, and other tests identified by manufacturer in the user manual or contractor determines necessary/beneficial/equivalent for successful/safe general laboratory functionality.

9) Please provide the test point requirements for the Refrigerated Eppendorf 5910ri.

a. Temperature, speed, timer, imbalance detection, brake function, safety lid lock test, temperature uniformity, and other tests identified by manufacturer in the user manual or contractor determines necessary/beneficial/equivalent for successful/safe general laboratory functionality.

10) Please provide the test point requirements for the Thermofisher Sorvall X4RF Pro-MD.

a. Temperature, speed, timer, imbalance detection, brake function, safety lid lock test, temperature uniformity, digital display, rotor ID system, applicable run profiles, and other tests identified by manufacturer in the user manual or contractor determines necessary/beneficial/equivalent for successful/safe general laboratory functionality.

11) Please provide the test point requirements for the Thermo Scientific TSGP15D.

a. Temperature, temperature uniformity, temperature stability, timer, over-temperature safety test, heating/cooling rate, digital display, water level sensor, and other tests identified by manufacturer in the user manual or contractor determines necessary/beneficial/equivalent for successful/safe general laboratory functionality.

12) Please provide the test point requirements for the Mettler Toledo ME104TAOD.

a. Repeatability, linearity, accuracy, sensitivity, tare function, stability, digital display, provide feedback on ideal scale location, and other tests identified by manufacturer in the user manual or contractor determines necessary/beneficial/equivalent for successful/safe general laboratory functionality.

13) Please provide the test point requirements for the Atron Systems Inc BX-600+.

a. Repeatability, linearity, accuracy, sensitivity, tare function, stability, digital display, provide feedback on ideal scale location, and other tests identified by manufacturer in the user manual or contractor determines necessary/beneficial/equivalent for successful/safe general laboratory functionality.

b. Due to the weight requirements to calibrate this scale we understand adaptability might be required. Liquid might be an allowable equivalent to large weights, and the scale will primarily be used to weigh liquid. Other identified methods by contractor can be discussed and accepted.

14) Please provide the test point requirements for the Mettler Toledo SevenCompact S200.

a. Standard buffer calibration, accuracy, repeatability, temperature compensation test, stability, response time, electrode check, slope and offset, user settings verification, and other tests identified by manufacturer in the user manual or contractor determines necessary/beneficial/equivalent for successful/safe general laboratory functionality.

15) Please provide the test point requirements for the Drummond Pipet-Aid XP 2.

a. Volume accuracy, volume precision, filter/nozzle check, leakage, and other tests identified by manufacturer in the user manual or contractor determines necessary/beneficial/equivalent for successful/safe general laboratory functionality.

16) Please provide the test point requirements for the pipettes (2 volumes or 3 volumes).

a. 3 volumes is preferred.

17) For the other non-pipette equipment, do you require ISO 17025 Accredited Services?

a. No for other non-pipette equipment we do not require ISO 17025 Accredited Services, however we would still want adequate calibration/or equivalent documentation provided upon completion of services.

image1.png