Statement of Work Project: Differential Scanning Calorimeter Thermal Analyzer
GENERAL INFORMATION
1.0 Scope of Work: Acquisition of a new thermogravimetric analyzer with differential scanning calorimeter to replace an obsolete DSC822e.
2.0 Background: The DSC822e differential scanning calorimeter currently in the Agroecosystems Management Research Unit has been a productive instrument used for thermal analysis. The DSC822e is an obsolete instrument, and it has been determined the research unit should replace the instrument with a more advanced thermal analyzer with thermogravimetric capabilities, to be able to process the volume of samples we will perform analysis on.
CONTRACTOR REQUIREMENTS
3.0 Technical Requirements/Tasks:
Minimum requirements:
Furnace temperature range: 0°C to 1000°C Furnace temperature accuracy: +/- 0.3K Furnace temperature precision: +/- 0.2K Furnace temperature resolution: 0.001 K Furnace heating time: 10 min Furnace heating rate: 150 K/min Cooling system: mini-chiller Able to perform dual analysis of thermogravimetry and differential scanning calorimetry Microbalance range: 0-1g Microbalance resolution: 1.0 ug Must be compatible with existing STARe evaluation software package Qty: 1 system
4.0 Government Furnished: USDA escort to lab for minimal installation duties as detailed above during normal business hours excluding State and Federal Holidays
5.0 Deliverables/Schedule Equipment shall be delivered onsite within 60 days of date of award.
6.0 Evaluation Factors Technical capability, deliverability and price with technical capability and deliverability more important than price
7.0 Technical Point of Contact: N/A
8.0 Place of Delivery: USDA-ARS National Laboratory for Agriculture and the Environment; 1015 N. University Blvd., Ames, IA 50010