Attachment 1 - Specifications - Revised.pdf

PDF 317 KB Posted

Attached to
Purchase Two Scientific Laboratory Steam Sterilizers Federal contract opportunity
Solicitation number
1232SA25Q0620
Issued by
Department of Agriculture Agricultural Research Service

About this file

This Specifications Document details sterilizer system requirements for the U.S. Meat Animal Research Center (USMARC), a USDA facility developing technologies to prevent meat product contamination. The technical requirements specify two laboratory steam sterilizers with precise technical specifications: steam-jacketed units with a 20 x 20 x 38" chamber size, manual vertical sliding or single hinged doors, temperature ranges from 110°C to 135°C, programmable up to 120 minutes of steam exposure, and an integrated electric steam generator. The sterilizers will be installed in separate buildings, with one recessed through a wall and the other free-standing.

Key deliverables include the procurement and installation of two Laboratory Steam Sterilizers within 45 days of award, meeting specific technical requirements for sterilizing bacterial growth medium, experimental tools, small equipment, and inactivating lab waste microbials. The document notes that the requirement can be met with two specific sterilizer models: Consolidated Sterilizer Systems Model SSR-3A or PRIMUS Sterilizer Company Model A (single door). The government point of contact is Rong Wang at rong.wang@usda.gov, located at the USMARC facility in Clay Center, Nebraska.

View the file

Other files for this federal contract opportunity

Other files attached to Purchase Two Scientific Laboratory Steam Sterilizers, newest first.
File Type Posted
RFQ_1232SA25Q0620.pdf PDF
Attachment 1 - Specifications.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications

Specifications Document: Sterilizer Systems

1.0 Background and Purpose:

Scientists at the Roman L. Hurska U.S. Meat Animal Research Center (USMARC) are developing scientific information and new technologies to prevent meat product contamination by foodborne pathogens. As part of the effort, they require autoclaves to sterilize bacterial growth medium, experimental tools and small equipment, and to inactivate all microbials in lab waste.

CONTRACTOR REQUIREMENTS

2.0 Technical Requirements/Tasks:

Small Steam Sterilizer o Scientific Laboratory Steam Sterilizer o Steam-jacketed small sterilizers with a chamber size of 20 x 20 x 38” (508 x 508 x 965 mm).

o Manual operated vertical sliding door or single door hinged.

o o The selectable temperature range is from 110oC to 135oC and from 104oC to 135°C for liquid cycles with temperature control of liquid load.

o Programmable for times up to 120 minutes of steam exposure.

o The program settings must be accessed controlled and can be communicated with an on-board printer to document cycle information.

o Control panel must have an internal deflection routes steam vapor and moisture away from the electrical control.

o Typical applications include wrapped and unwrapped hard goods, liquids (media and reagents), and biohazardous waste.

o Single Door Cabinet with steam heating (minimum 75 psig steam required) and an integrated standard electric steam generator utilized on the unit fed with softened water (

– 45 kW).

o Maintenance service contract is optional.

o Standard Piping: Brass and Copper Threaded.

3.0 Deliverable/Schedule/

Key Deliverables

4.0 Performance Requirement Measures:

5.0 Government Point of Contact: