1155089-Specification- Fuji X-ray CR imager.pdf
PDF 139 KB Posted
- Attached to
- Brand Name or Equal to Fujifilm FCR Carbon XL with FDX Console Package (FCR XL Image Reader
- Solicitation number
- 1232SA25Q0636
About this file
This specification outlines requirements for a Computed Radiography (CR) system, specifically the Fujifilm FCR Carbon XL or an equivalent. The system must utilize reusable imaging plates, support 24 x 30 cm plates, and have a throughput of at least 92 images per hour. Key technical requirements include high-resolution imaging (50 microns), automated plate processing, a console for image acquisition and management, and advanced image processing software with capabilities like magnification, masking, and annotation.
The CR system must integrate with an existing Faxitron MX-20 X-ray unit and Evo-View Diagnostic Workstation PACS, support multi-user network access, and be compatible with a 2TB Western Digital external hard drive. Additional requirements include a compact, portable design suitable for laboratory use, comprehensive installation and training, and a manufacturer's warranty covering parts and labor. The specification emphasizes the need for a system that can efficiently process images, particularly for examining bee nests and potential parasites, with robust data integrity and image management features.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 1232SA25Q0636 - LPTA - RFQ - TCs -.pdf | ||
| 1232SA25Q0636 SF 1449 (11_2021) SOL.pdf | ||
| 1155089 1232SA25Q0636- Combined Syn-SOL -.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specification
Fujifilm FCR Carbon XL - Brand Name or Equal Specification
This s pecification outlines the requirements for a Computed Radiography (CR) s ys tem. While a s pecific brand and model (Fujifilm FCR Carbon XL with FDX Cons ole Package) were mentioned, this document details the key functionalities and performance parameters that any equivalent s ys tem mus t meet.
A. CR Sys tem - Core Functionality & Performance:
• Imaging Type: Computed Radiography (CR) s ys tem utilizing reus able imaging plates .
• Res olution: Capable of acquiring high-res olution images , with a minimum res olution equivalent to 50 microns (20 lp/ mm) for optimal detail in examining bee nes ts and potentia l paras ites .
• Plate Size: Mus t s upport 24 x 30 cm imaging plates .
• Throughput: Mus t have a throughput of a t leas t 92 images per hour.
• Cas s ette-Bas ed Sys tem: The s ys tem s hall utilize cas s ettes to hous e and protect the imaging plates during acquis ition and proces s ing.
• Automated Plate Proces s ing: The s ys tem mus t automatically proces s the plates without requiring manual removal by the us er, protecting the plates from damage.
• Cons ole Functionality: Mus t include a cons ole for image acquis ition, pla te identification (e.g., barcode s canning), image proces s ing, and data management.
• Data Integrity: The s ys tem mus t ens ure data integrity by uniquely identifying and linking images with res earch data (e.g., through barcode s canning of pla tes / cas s ettes ).
B. Software Functionality:
• Image Proces s ing: The s ys tem's s oftware s hould provide advanced image proces s ing techniques to maximize res olution and image quality (e.g., Multi-objective Frequency Proces s ing (MFP), dynamic range control, flexible nois e control, and grid pattern removal).
• Image Manipulation: The s oftware s hould enable manipulation of images to enhance vis ualization, including magnification, mas king, cropping, and meas urement tools .
• Labeling & Annotation: The s oftware mus t provide functionality for labeling and annotating images with res earch data .
• Archiving & Storage: The s ys tem s hould integrate with exis ting PACS for image archiving and s torage.
C. Sys tem Integration & Compatibility:
• Exis ting X-ray Unit Compatibility: The CR s ys tem s hould be compatible with the exis ting Faxitron MX-20 X-ray unit.
• PACS Integration: The s ys tem mus t integrate s eamles s ly with the exis ting PACS (Evo-View Diagnos tic Works tation) and allow for the viewing and manipulation of images by multiple us ers a t their des ktop computers .
• Network Capabilities : The s ys tem s hould be networkable for multi-us er acces s and concurrent viewing of images by s everal s cientis ts .
• External Hard Drive: The s ys tem requires compatibility with the exis ting 2TB USB external Wes tern Digita l Hard Drive for additional s torage.
D. Phys ical & Environmental Requirements :
• Space Saving: The CR proces s or unit s hould be compact and des igned for us e in labs with limited s pace (e.g., s mall footprint, option for a cart for eas y relocation).
• Portability: The s ys tem s hould be movable on a cart to maximize res earch space utilization.
E. Ins ta lla tion, Training & Support:
• Ins ta lla tion & Training: Ins ta lla tion and comprehens ive tra ining on s ys tem operation for a ll relevant pers onnel are required.
• Warranty: Each component of the s ys tem mus t come with a manufacturer's warranty covering parts and labor.