1155089-Specification- Fuji X-ray CR imager.pdf

PDF 139 KB Posted

Attached to
Brand Name or Equal to Fujifilm FCR Carbon XL with FDX Console Package (FCR XL Image Reader
Solicitation number
1232SA25Q0636
Issued by

About this file

This specification outlines requirements for a Computed Radiography (CR) system, specifically the Fujifilm FCR Carbon XL or an equivalent. The system must utilize reusable imaging plates, support 24 x 30 cm plates, and have a throughput of at least 92 images per hour. Key technical requirements include high-resolution imaging (50 microns), automated plate processing, a console for image acquisition and management, and advanced image processing software with capabilities like magnification, masking, and annotation.

The CR system must integrate with an existing Faxitron MX-20 X-ray unit and Evo-View Diagnostic Workstation PACS, support multi-user network access, and be compatible with a 2TB Western Digital external hard drive. Additional requirements include a compact, portable design suitable for laboratory use, comprehensive installation and training, and a manufacturer's warranty covering parts and labor. The specification emphasizes the need for a system that can efficiently process images, particularly for examining bee nests and potential parasites, with robust data integrity and image management features.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specification

Fujifilm FCR Carbon XL - Brand Name or Equal Specification

This s pecification outlines the requirements for a Computed Radiography (CR) s ys tem. While a s pecific brand and model (Fujifilm FCR Carbon XL with FDX Cons ole Package) were mentioned, this document details the key functionalities and performance parameters that any equivalent s ys tem mus t meet.

A. CR Sys tem - Core Functionality & Performance:

• Imaging Type: Computed Radiography (CR) s ys tem utilizing reus able imaging plates .

• Res olution: Capable of acquiring high-res olution images , with a minimum res olution equivalent to 50 microns (20 lp/ mm) for optimal detail in examining bee nes ts and potentia l paras ites .

• Plate Size: Mus t s upport 24 x 30 cm imaging plates .

• Throughput: Mus t have a throughput of a t leas t 92 images per hour.

• Cas s ette-Bas ed Sys tem: The s ys tem s hall utilize cas s ettes to hous e and protect the imaging plates during acquis ition and proces s ing.

• Automated Plate Proces s ing: The s ys tem mus t automatically proces s the plates without requiring manual removal by the us er, protecting the plates from damage.

• Cons ole Functionality: Mus t include a cons ole for image acquis ition, pla te identification (e.g., barcode s canning), image proces s ing, and data management.

• Data Integrity: The s ys tem mus t ens ure data integrity by uniquely identifying and linking images with res earch data (e.g., through barcode s canning of pla tes / cas s ettes ).

B. Software Functionality:

• Image Proces s ing: The s ys tem's s oftware s hould provide advanced image proces s ing techniques to maximize res olution and image quality (e.g., Multi-objective Frequency Proces s ing (MFP), dynamic range control, flexible nois e control, and grid pattern removal).

• Image Manipulation: The s oftware s hould enable manipulation of images to enhance vis ualization, including magnification, mas king, cropping, and meas urement tools .

• Labeling & Annotation: The s oftware mus t provide functionality for labeling and annotating images with res earch data .

• Archiving & Storage: The s ys tem s hould integrate with exis ting PACS for image archiving and s torage.

C. Sys tem Integration & Compatibility:

• Exis ting X-ray Unit Compatibility: The CR s ys tem s hould be compatible with the exis ting Faxitron MX-20 X-ray unit.

• PACS Integration: The s ys tem mus t integrate s eamles s ly with the exis ting PACS (Evo-View Diagnos tic Works tation) and allow for the viewing and manipulation of images by multiple us ers a t their des ktop computers .

• Network Capabilities : The s ys tem s hould be networkable for multi-us er acces s and concurrent viewing of images by s everal s cientis ts .

• External Hard Drive: The s ys tem requires compatibility with the exis ting 2TB USB external Wes tern Digita l Hard Drive for additional s torage.

D. Phys ical & Environmental Requirements :

• Space Saving: The CR proces s or unit s hould be compact and des igned for us e in labs with limited s pace (e.g., s mall footprint, option for a cart for eas y relocation).

• Portability: The s ys tem s hould be movable on a cart to maximize res earch space utilization.

E. Ins ta lla tion, Training & Support:

• Ins ta lla tion & Training: Ins ta lla tion and comprehens ive tra ining on s ys tem operation for a ll relevant pers onnel are required.

• Warranty: Each component of the s ys tem mus t come with a manufacturer's warranty covering parts and labor.

File details come from the government source that posted it. Updated .