Performance Work Statement for OHD Respirator Unit Calibration and Maintenance
This Performance Work Statement (PWS) describes maintenance, calibration, and repair services for twenty-six (26) OHD Quantifit2 Respirator Test Fit machine units at Pearl Harbor Naval Shipyard & Intermediate Maintenance Facility (PHNSY & IMF). The contract period runs from Base Year 2026–2027 with four optional years extending through 2030–2031. The contractor must provide annual preventive maintenance, remedial repairs, calibration to National Institute of Standards and Technology (NIST) standards, and comprehensive reporting. Preventive maintenance includes disassembly and cleaning of units, inspection and replacement of components (scattering chambers, sampling systems, electrical connections, pumps, pistons, tubing, ports, and gaskets), firmware updates, and leak checks. All parts replaced during preventive maintenance are included at no additional cost to the government and must be new, not refurbished. The contractor must return serviced units within four (4) days of receipt and provide service reports detailing work performed and parts replaced, along with NIST calibration certificates for each unit.
Remedial repairs identified during preventive maintenance require prior written cost estimates (hours, labor, materials, and total cost) submitted to the Contracting Officer Representative (COR) for approval by the Contracting Officer (KO) before work commences. Equipment that falls out of calibration during the certificate period must be re-calibrated and re-certified at no additional government cost. The contractor is responsible for all shipping costs to return re-calibrated equipment. The contractor must be an authorized OEM-trained, certified technician (not a sub-contractor) restricted to U.S. citizens and nationals per Department of Defense requirements. The contractor must establish a project manager with full authority to represent them, maintain a single point of contact with the government COR, develop and maintain a Quality Control Plan, and ensure all services comply with the PWS requirements and OEM guidelines. Service delivery performance requires 100% compliance for preventive maintenance and calibration certificates and 90% compliance for timely return of serviced equipment.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PERFORMANCE WORK STATEMENT (PWS)
OHD Respirator Unit Calibration and Maintenance for Pearl Harbor Naval Shipyard
& Intermediate Maintenance Facility (PHNSY & IMF)
1.0 BACKGROUND
1.1 The primary mission of PHNSY & IMF is to deliver regional maintenance, at the depot and intermediate levels, to keep the surface ships and submarines of our nation's navy "Fit to Fight."
As Hawaii’s regional maintenance center, capabilities include excellence in overhauling, repairing, converting, alteration, refurbishing, and decommissioning of Navy vessels.
2.0 OBJECTIVE & SCOPE
2.1 PHNSY & IMF seeks a contractor that is authorized and qualified to perform maintenance repairs and calibration services of the twenty six (26) OHD Quantifit2 Respirator Test Fit machine units. The contractor shall provide annual preventive maintenance, calibration, and remedial repair. In addition, the contractor shall provide an inspection report, service report and calibration certificate for each unit serviced.
3.0 PERFORMANCE REQUIREMENTS
3.1 The contractor shall provide annual preventive maintenance, repair, calibration and remedial service of the twenty six (26) OHD Quantifit2 Respirator Test Fit machine units in accordance with current
Original Equipment Manufacturer (OEM) guidelines. In addition, the contractor shall provide an inspection and service report for each unit serviced. The contractor shall be responsible for insuring all services under this contract are in accordance with the terms and conditions outlined herein and conform to the specifications within this Performance Work Statement (PWS). Required turnaround time is four (4) days after receipt of equipment.
PERIOD OF PERFORMANCE
Base Year - 2026 – 2027
Option 1 - 2027 – 2028
Option 2 - 2028 – 2029 Option 3 - 2029 – 2030
Option 4 - 2030 – 2031
3.2 Preventive Maintenance
The contractor shall provide the following preventive maintenance services:
• Reviewing each unit and the documentation of problem(s) experienced by the government.
• Inspect the operating function of each unit to determine the current condition.
• Disassemble and clean the unit’s exterior and interior housing.
• Removing, inspecting, cleaning, replacing, repairing the following as needed.
o Scattering Chambers o Sampling System o Testing Electrical Connections o Performing Voltage Measurements o Tightening Hardware o Aligning Optics o Pump o Piston o Internal Flow Tubing and Internal Pressure Tubing o All Ports (Exhaust, Valve and Internal Connection)
DRAFT
o All Port O-Rings/Gaskets
• Clean and checking functions of all accessories sent with each unit as applicable. Repair and replace as needed.
• Reassembling and conducting leak checks
• Updating internal program chips to latest version (if applicable)
• Updating manual and firmware to latest version (if applicable)
3.3 Remedial Repairs
Remedial repair (repairs required outside scope of annual preventive maintenance) identified during preventive maintenance shall be accomplished if completion of the repair is possible within the period of performance. The contractor is required to provide a written estimate of the cost (time, labor & materials) of the additional work to the Contracting Officer Representative (COR) prior to the commencement of work. The quote shall include the following as a minimum: number of hours required, parts and materials cost, labor cost, any additional cost and total cost. The COR shall evaluate the written estimate and provide the recommendation to the Contracting Officer (KO). The
KO is the approval authority for any remedial repairs and when approved an in-scope contract modification will be issued.
3.4 Parts
a) Preventive Maintenance Parts
Parts that are replaced during preventive maintenance service are included at no additional cost to the government. Materials used shall be new and not refurbished or reconditioned materials. All parts shall meet manufacturer specifications.
b) Remedial Repair Parts
Parts that are replaced as part of the remedial repair is to be issued on a separate purchase order, and approved by the government. Materials used shall be new and not refurbished or reconditioned materials. All parts shall meet manufacturer specifications.
3.5 Calibration
a) The contractor shall calibrate the equipment to National Institute of Standards and Technology
(NIST) standards.
b) Equipment which falls out of calibration specifications during the period of time for which the calibration certificate provides, shall be re-calibrated and re-certified by the contractor at no additional cost to the government. The contractor shall be responsible for any cost associated in returning the re-calibrated equipment to the government.
3.6 Reports
a) Service Reports
The contractor shall provide the government a service report no later than one week after completion of service. The report shall detail the nature of the work performed, problems identified during preventive maintenance, remedial repair actions, and parts or materials replaced. Service report shall include part numbers and prices, name of technician, date of performed service and unit identification.
b) Calibration Certificate
The contractor shall provide the government a National Institute of Standards and Technology (NIST)
Calibration Certificate for each unit when returning serviced units.
3.7 Shipping To/From
OHD Quantifit2 Respirator Test Fit machine unit(s) will be shipped to the contractor packaged in
Testing machines.
7.2 The contractor shall establish clear organizational lines of authority and responsibility to ensure effective management. The contractor must maintain continuity between Code 900T TPOC and the
COR. The contractor shall provide a project manager who shall be responsible for the performance and quality of meeting the requirements of this PWS. The project manager or alternate shall have full authority to act on behalf of the contractor on all contract matters relating to operations of this contract.
7.3 The contractor shall have a single point of contact between themselves and the government COR to support and meet the requirements of the PWS.
7.4 The contractor shall be aware that the absence of any contract requirement from the PWS does not detract from its enforceability nor limit the rights or remedies of the government under any other provision of the contract. The contractor shall ensure the PWS requirements are accomplished.