Specular Microscope

Closed Solicitation Posted

Solicitation number
FA8601
Agency
AFLCMC Wright Patterson AFB Lifecycle Management Center, Department of Defense
Responses due
Set-aside
No set-aside

Opportunity facts

NAICS code
333314 Optical Instrument and Lens Manufacturing
PSC
6515 Medical And Surgical Instruments, Equipment, And Supplies
Place of performance
Wright Patterson AFB, Ohio 45433, United States

Notice details come from SAM.gov. Updated .

About this opportunity

The Department of the Air Force Materiel Command Lifecycle Management Center at Wright-Patterson Air Force Base is soliciting proposals for a specular microscope through solicitation number FA8601, with proposals due by April 21, 2022. The microscope must be able to non-contactly photograph the cornea over a 0.25mm by 0.54mm range, include auto and manual measurement modes, capture measurements at the center and 12 peripheral points of the cornea, and measure and analyze corneal thickness, cell density, area, and other parameters. It must also come with a standard one-year warranty.

There is no set-aside designated for this requirement. The NAICS code is 334510 and PSC code is J070. The sole source justification identifies the Tomey EM 4000 as the only device meeting requirements. Delivery of one microscope is required within 14 days to the Wright-Patterson Medical Center in Ohio to allow for accurate pre-operative and post-operative endothelial cell measurements for patients undergoing eye surgeries at that location.

Notice text

The Wright-Patterson AFB medical center is in need of a Specular Microscope. It is required to measure endothelial cells. Due to the mechanical stress of the surgery or UV exposure, endothelial cells will be lost. This device is used to get an accurate pre/post op measurement of how much effect the procedure had and continues to have on the posterior cornea. 

Requirements:

  • o Photographing method – non-contact
  • o Photographing range – 0.25mm x 0.54mm
  • o Measurement mode – auto/manual/auto alignment manual shot
  • o Capturing position – center + 12 peripheral points
  • o Cornea thickness measurement accuracy - +/- 10pm
  • o Analysis method – Core method
  • o Analysis values – Number (the number of analyzed cells)

                                                CD (cell density)

                                                AVG (average cell area)

                                                 SD (standard deviation of cell area)

                                                CV (coefficient of variation of cell area)

                                                Max (maximum cell area)

                                                Min (minimum cell area)

  • o Histogram – Area (Polymegathism: Distribution by areas)

                     Apex (Plemorphism: Distribution by polygonal shapes types)

  • Warranty: Standard 1 Year.

Attachments

Files attached to this notice, newest first
File Type Posted
Statement of Work-EM4000 Specular Microscope.docx DOCX document
Single Source Justification-Tomey EM 4000 Redacted.pdf PDF

On GovTribe

Work this opportunity on GovTribe

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity