Salient Characteristics for nerve monitoring system
- 1 to 4 channels that are individually and simultaneously selectable.
- Input sensitivity 5 – 10,000 μV peak-to-peak AC Coupled ± 5% at 10 mV to ± 30% at 100 mV peak-to peak AC Coupled
- Input Impedance > 10 Meg Ohm
- Patient Isolation Wired 1,000 Vrms 60Hz < 100 µA
- Measurement Range:
- Electrodes: 0K to 2K Ohm ± 500 Ohm >2K-175K Ohm ± 20% Ground and Stim Returns < 25K.
- Bipolar Electrocautery Rejection Continuous Monitoring During Bipolar Cautery < 40 watts
- Monopolar Electrocautery (ESU) Interference Automatic detection and muting in wired or wirefree modes
- Muting Console Input Sensitivity
- Muting (0.6 - 2.0 Volts Vrms) Non-Muting (<0.3 Volts Vrms)
- Muting Detector Input ESU Immunity
- ESU < 100 Watts Cut / Coag or (<3.0 Vrms 100-800 KHz Sq. Wave)
- Stimulus Output Accuracy ± .01 mA (or ± 10% of reading at 1 K load) over Stimulus Range
- Stimulus Measurement Accuracy ± .02 mA (or ± 10% of reading at 1 K load) over Stimulus Range
- Waveform Monophasic, Bi-phasic, Alternating Polarity, or Pulse Train of square pulses
- Interpulse Interval 50uS, 100uS, 500uS, 1-5mSec
- Interphase Delay 0, 50, 100, 500, uS