Nerve Monitoring System, Salt Lake City VA Health Care System

Closed Solicitation Posted

Solicitation number
36C25923Q0969
Agency
Responses due
Set-aside
No set-aside

Opportunity facts

NAICS code
334510 Electromedical and Electrotherapeutic Apparatus Manufacturing
PSC
6515 Medical And Surgical Instruments, Equipment, And Supplies
Place of performance
Salt Lake City, Utah 84148, United States

Notice details come from SAM.gov. Updated .

Notice text

Salient Characteristics for nerve monitoring system

  • 1 to 4 channels that are individually and simultaneously selectable.
  • Input sensitivity 5 – 10,000 μV peak-to-peak AC Coupled ± 5% at 10 mV to ± 30% at 100 mV peak-to peak AC Coupled
  • Input Impedance > 10 Meg Ohm
  • Patient Isolation Wired 1,000 Vrms 60Hz < 100 µA
  • Measurement Range:
    • Electrodes: 0K to 2K Ohm ± 500 Ohm >2K-175K Ohm ± 20% Ground and Stim Returns < 25K.   
  • Bipolar Electrocautery Rejection Continuous Monitoring During Bipolar Cautery < 40 watts
  • Monopolar Electrocautery (ESU) Interference Automatic detection and muting in wired or wirefree modes
  • Muting Console Input Sensitivity
    • Muting (0.6 - 2.0 Volts Vrms) Non-Muting (<0.3 Volts Vrms)
  • Muting Detector Input ESU Immunity
    • ESU < 100 Watts Cut / Coag or (<3.0 Vrms 100-800 KHz Sq. Wave)
  • Stimulus Output Accuracy ± .01 mA (or ± 10% of reading at 1 K load) over Stimulus Range
  • Stimulus Measurement Accuracy ± .02 mA (or ± 10% of reading at 1 K load) over Stimulus Range
  • Waveform Monophasic, Bi-phasic, Alternating Polarity, or Pulse Train of square pulses
  • Interpulse Interval 50uS, 100uS, 500uS, 1-5mSec
  • Interphase Delay 0, 50, 100, 500, uS