The National Oceanic and Atmospheric Administration (NOAA) is seeking qualified U.S. businesses to provide an LNI Swissgas Sonimix 4001 Ozone Generator. This stand-alone ozone primary transfer standard is required to calibrate NOAA's ozone measuring research instruments, which are used to measure ozone from the surface to the stratosphere. The key requirements include the ability to generate ozone in a range from 10 ppb to 35 ppm, with high accuracy and precision, as well as features like pressure regulation, sonic nozzles, and UV photometer measurement. Interested firms must provide a brief narrative and supporting documentation demonstrating their capability and experience by August 7, 2024.

This is a sources sought notice, not a pre-solicitation announcement, and no set-aside designation is specified. The NAICS code is 334516, Analytical Laboratory Instrument Manufacturing, with a size standard of 1,000 employees. No potential incumbents or award values are provided, and the notice states that this request for information does not commit the government to any future procurement action.

Notice text

Notice Number: 145036240036SM

Notice Type: Sources Sought Synopsis

Title: LNI Swissgas Sonimix 4001 Ozone Generator

Response Date: August 7, 2024

This is a SOURCES SOUGHT SYNOPSIS and is not a pre-solicitation announcement. The purpose of this Market Research and Sources Sought Notice is to gain knowledge of potential qualified U.S. Domestic businesses, which could include Small Business, HUBZone, 8(a), Service Disabled Veteran Owned, Women Owned Small Business firms and even Large Businesses. Responses to this Sources Sought Synopsis will be used by the Government in making appropriate acquisition decisions.

The National Oceanic and Atmospheric Administration (NOAA), Oceanic and Atmospheric Research (OAR), Earth Systems Research Laboratories (ESRL), Chemical Sciences Laboratory (CSL) has multiple ozone measuring research instruments and conducts projects that require measurement of ozone from the surface to the stratosphere. Ozone is an important trace gas species in Earth’s atmosphere. The ozone layer in the stratosphere is critical for protecting Earth’s surface from harmful UV radiation, while at the surface ozone is a criteria air pollutant which negatively impacts both human health and the environment (e.g. crop production). Due to the strong gradient in ozone mixing ratios between the upper troposphere (UT) and lower stratosphere (LS), ozone is also a critical measurement for characterizing the atmospheric state in UTLS studies. A primary standard (PS) ozone generator is necessary to regularly calibrate these instruments to ensure research data quality.

INTENDED USE/PURPOSE:

The LNI Swissgas Sonimix 4001 is a stand-alone ozone primary transfer standard able to generate ozone together with dry or wet air allowing it to calibrate the ozone monitors in several points. The instrument uses a pressure regulator, sonic nozzles, and an ozone generator including thermostatisation and regulation by UV light photometer measurement with atmospheric pressure and temperature corrections. It is built with 1 reference UV light detector that measures the UV intensity in alternance without any ozone, and through the generated ozone. This is the only known instrument that can meet the Government’s needs, and this notice’s intent is to identify any other products that meet the full requirements listed below.

SPECIAL REQUIREMENTS/ESSENTIAL SALIENT FEATURES:

General Specifications:

Range: From 10ppb to 10 ppm @ 3 L/min + 10ppm to 35 ppm @ 0.9 L/min

Dilution principle: Sonic nozzles and high precision pressure regulators

Dilution air: Dry air from internal zero air (option) or external dry air source

Generated gas flow: 900ml/min for range 2

Accuracy: 10 ppb to 100 ppb: +/- 2 ppb; >100 ppb: 1.5% relative

Ozone repeatability: Better than 1% relative

Warm up time: < 45 minutes

Stabilization time: < 15 minutes

O3 regulation: Measure of the UV light absorption of the generated ozone by UV reference photometer. (US EPA method and Beer Lambert law).Including atmospheric pressure and room temperature corrections

Dimensions: 19cm x 36cm, 50cm deep

Net weight: 15 Kg

UV Section:

UV lamp life time: About 10’000 hours (with a mean concentration of 1 ppm)

Zero Air Section:

Life time: Annual preventive maintenance

Quality: O3, NO and NO2 < 1 ppb – H2O < -40°C dew point

Communication:

Outlet port: Teflon 1/4‘’ Swagelok fitting

Electrical Data:

Power supply voltage: 230V/50-60Hz – 117V/60Hz – 100V/60Hz

DELIVERY SCHEDULE:

Deliver by 60 days after date of award.

The North American Industry Classification Code (NAICS) is 334516 with a size standard of 1,000 employees.

Interested firms shall provide, in writing, a brief narrative (no more than 3 pages in length) and documentation to support your firm’s capability and experience meeting the requirements herein and include your firm’s name, address, point of contact, telephone number, UEID Number, CCR CAGE Number, business type and size (i.e., small, large, non-profit, educational, etc.). Any response that does not include this information will not be considered. The Government will review your response to determine if your firm has the experience, knowledge and capability to provide the items as specified above.

This request for information does not constitute a commitment, implied or otherwise, that a procurement action will be issued. Nor does this request for information constitute a commitment, implied or otherwise, that any procurement action may be issued for the Government’s requirements.

No entitlement to payment of direct or indirect costs or charges by the Government will arise because of the submission of contractor’s information. The Government shall not be liable for or suffer any consequential damages for proprietary information not properly identified. Proprietary information will be safeguarded in accordance with the applicable Government regulations.

All contractors doing business with the Government are required to be registered and active with the System for Award Management (SAM) website.

NOTE: THIS SYNOPSIS IS NOT CONSIDERED A REQUEST FOR PROPOSAL BUT A REQUEST FOR INFORMATION ONLY. No solicitation document is available. No contract will be awarded based on responses received to this notice.

Notice history

Notice Type Posted
LNI Swissgas Sonimix 4011 Ozone Generator Latest special notice Special Notice
This notice Original Pre-Solicitation

Similar open opportunities

Opportunity Agency Set-aside Responses due
Instruments Upgrade - Nitrogen & Hydrogen Generators Indian Head, Maryland, United States · NAICS 334519 Naval Sea Systems Command No set-aside
Sources Sought Notice for Continuous Wave Parametric Oscillator (CW OPO) Laser System Gaithersburg, Maryland 20899, United States · NAICS 334516 National Institute of Standards and Technology No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity

GovTribe for Federal Contractors