Inductively Coupled Plasma - Optical Emission Spectrometry

Open Solicitation Posted

Solicitation number
W911S227U0002
Agency
Responses due
Set-aside
Total Small Business

Opportunity facts

Place of performance
2744 Rodman Ave. Rock Island, Illinois 61299

Notice details come from SAM.gov. Updated .

About this opportunity

The Department of the Army's Mission and Installation Contracting Command Fort Drum (MICC Fort Drum) seeks a bench-top inductively coupled plasma optical emission spectrometer (ICP-OES) system, with proposals due November 16, 2026. The procurement is set aside for Total Small Business, and the notice carries a solicitation amendment.

Place of performance is 2744 Rodman Ave, Rock Island, IL 61299. The solicitation was posted October 1, 2026 under NAICS 334516, Analytical Laboratory Instrument Manufacturing, and PSC 6630, Chemical Analysis Instruments.

The mandatory minimum requirements state a preferred instrument comparable to the Agilent 5800, and vendors may offer any equivalent system that meets the specified requirements. Key requirements include a simultaneous-reading, bench-top, dual-view vertical plasma system; a solid-state echelle polychromator with 253 mm focal length and resolution below 0.0065 nm at 200 nm; a view of at least 96% of the 167–785 nm emission spectrum; and floor-mount designs are not acceptable. The detector must be solid-state, Peltier-cooled to -40°C or below, and hermetically sealed. The RF generator must be solid-state at 750–1500 W. All plasma gases must use mass flow controllers. The system must run aqueous and organic matrices, including oil, MIBK, kerosene, and hexanes. Power draw is 2.9 kVA or less, and exhaust capacity is at least 2.5 m³/min (88 ft³/min).

Notice text

2 versions

Update #2 · Latest ·

Solicitation Amendment for Inductively Coupled Plasma - Optical Emission Spectrometry (ICP-OES)

Update #1 ·

MANDATORY MINIMUM REQUIREMENTS: Preferred instrument: ICP- OES comparable to the Agilent 5800; vendors may offer any equivalent system that meets the required specifications.1. Spectrometer RequirementsSimultaneous reading ICP OES w/ solid state detector.Bench top design only. Floor mount not acceptable.Echelle polychromator, 253 mm focal length, single detector, single entrance slit. Resolution: less than 0.0065 nm @ 200 nm; less than 0.013 nm @ 327.4 nm.Purged + thermostated optical enclosure; purge 0.7 L/min w/ 3 L/min boost option (less than 189 nm).Pixel level anti blooming detector.Vertical plasma; single piece torch. Optional torches must be available.View equal to or greater than 96% emission spectrum from 167-785 nm.Simultaneous UV + Vis determinations in one measurement.2. ICP System RequirementsBench top system; dual view vertical plasma (synchronous axial/radial).Must run aqueous + organic matrices including oil, MIBK, kerosene, hexanes.Monitors gas pressures/flows, water flows, air pressure, exhaust, plasma stability; auto shutdown on interlock breach.Operates at 15-30 degrees C and 25-80% RH; tolerates 2 degrees C/hr change with no performance loss.Exhaust requirement: equal to or greater than 2.5 m^3/min (88 ft^3/min).Power: equal to or less than 2.9 kVA, 200-240 VAC, 50-60 Hz, equal to or less than 15 amps3. Detector RequirementsSingle solid state detector; simultaneous reading of all wavelengths.Peltier cooled to equal to or less than -40 degrees C; Adaptive Integration capability.Hermetically sealed; no gas purge required.4. RF Generator RequirementsSolid state generator, 750-1500 W, 10 W increments.equal to or greater than 75% power transfer efficiency; typical stability 0.1%.5. Gas Flow ControlsAll plasma gases (coolant, auxiliary, nebulizer, makeup) must use mass flow controllers.Must meet flow ranges + increments listed in Government technical criteria.Integrated, computer controlled oxygen addition for organics.6. Sample Introduction SystemCassette style vertical torch; no manual adjustments required after mounting.Double pass glass cyclonic spray chamber + glass concentric nebulizer.Five channel variable speed peristaltic pump; up to 80 rpm.7. Software Requirements64 bit software; Windows 11 compatible.Automated background + spectral interference correction; automatic IEC recalculation.QC system supporting blanks, spikes, duplicates, and corrective actions.Full raw data saving + complete post run reprocessing.8. Performance RequirementsMust meet all EPA ICP OES detection limits and CRDLs using concentric nebulizer + cyclonic spray chamber.9. Additional RequirementsNew instrument; refurbished/demo units not acceptable.One year warranty (parts, labor, travel).Separate, remotely placeable water chiller.Autosampler holding equal to or greater than 180 sample positions + equal to or greater than 11 standards10. Hydrofluoric Acid Compatibility RequirementSample introduction systems must be or include compatible hardware for hydrofluoric acid use. This shall include, at a minimum, an HF-resistant nebulizer and spray chamber11. Certified Metal Standards Accuracy RequirementThe instrument shall be capable of analyzing within +/- 10% of certified values of metal standards (either chip or atomized metal powder) for typical alloy steel grades, typical stainless steel grades, typical copper alloys, typical cast and wrought aluminum, and titanium alloys with proper preparation for metallurgical elements excluding elements atypical of interrogation by ICP-OES methods such as carbon, nitrogen, hydrogen, and oxygen.

Similar open opportunities

Opportunity Agency Set-aside Responses due
ICP-OES for Oil Analysis Yuma Proving Ground301 C Street Bldg 2410Yuma, Az 85365 85365 · NAICS 334516 MICC Fort Drum Total Small Business
Rugged Spectrometers 4223 2nd St.Dugway, UT 84022 84022 · NAICS 334516 MICC Fort Drum Total Small Business
Sources Sought for (ICP-OES) Spectrometer System with an Autosampler Annapolis, Maryland 21402, United States · NAICS 334516 Naval Supply Systems Command No set-aside
Minicams parts Wdtc Rcvg Dugway Proving Ground POC: Ben Hunt 4223 2nd St. Dugway, UT 84022 84022 · NAICS 334516 MICC Fort Drum Total Small Business
Test Equipment, Source Measure Units Attn: D/Eecs Aaron St. Leger, 646 Swift Road, West Point, NY 10996, 10996 · NAICS 334515 MICC Fort Drum Total Small Business

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity

GovTribe for Federal Contractors